17 March 2015 Transducer elements position calibration in a ring array USCT system
Satoshi Tamano, Takashi Azuma, Haruka Imoto, Shu Takagi, Shin-ichiro Umemura, Yoichiro Matsumoto
Author Affiliations +
Abstract
Development of Ultrasound Computed Tomography (USCT) transducer elements position calibration method was described. USCT experimental system having 1024 ring array transducer elements, diameter of 104mm, electronic High Voltage multiplexer circuits and commercially available ultrasound research platform was used. The result of applying the transducer elements position calibration to the received data, -6dB Tx / Rx beam width was 0.68mm for uncorrected, 0.20mm for corrected. Corrected value was almost same as simulated value. As a result, when reconstructing crosssectional images in the present experimental system, expected a precise image, not only reflection black-and-white images but also sound-of-speed transmission image would be reconstructed.
© (2015) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Satoshi Tamano, Takashi Azuma, Haruka Imoto, Shu Takagi, Shin-ichiro Umemura, and Yoichiro Matsumoto "Transducer elements position calibration in a ring array USCT system", Proc. SPIE 9419, Medical Imaging 2015: Ultrasonic Imaging and Tomography, 94190P (17 March 2015); https://doi.org/10.1117/12.2082323