Paper
1 September 2015 Phase change cells and the verification of gallium as a thermal calibration reference in space
Harri Latvikoski, Gail E. Bingham, T. Shane Topham, Igor Podolski
Author Affiliations +
Abstract
The validation of models of global climate change and accurate measurement of the atmosphere and surface temperatures require that orbital sensors have low drift rates, and are monitored or regularly recalibrated by accepted standards. Phase change materials (PCM), such as those that make up the ITS-90 standard, are the basis for international commerce and have been suggested for monitoring and recalibration of orbital temperature sensors. Space Dynamics Laboratory (SDL) and its partners have been developing miniaturized phase change reference technologies that could be deployed on an orbital blackbody for nearly a decade. A significant part of this effort has been the exploration of the behavior of gallium (Ga) and its eutectics, gallium-tin (GaSn) and gallium-indium (GaIn) in conditions expected to be encountered in this application. In this paper, these behaviors are detailed and an example of a hardware design that could be used as an infrared blackbody calibration monitor is presented. To determine if and how microgravity will affect the behavior of Ga, the authors conducted an experiment on the International Space Station (ISS) and compared the observed phase change temperature with earth-based measurements. This paper also provides a brief description of the experiment hardware, microgravity considerations, and the pre-flight, flight and post-flight data analysis.
© (2015) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Harri Latvikoski, Gail E. Bingham, T. Shane Topham, and Igor Podolski "Phase change cells and the verification of gallium as a thermal calibration reference in space", Proc. SPIE 9608, Infrared Remote Sensing and Instrumentation XXIII, 96080Q (1 September 2015); https://doi.org/10.1117/12.2189932
Lens.org Logo
CITATIONS
Cited by 4 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Gallium

Black bodies

Calibration

Temperature metrology

Sensors

Temperature sensors

Sensor calibration

RELATED CONTENT

Checking AIRS nonlinearity in flight
Proceedings of SPIE (September 09 2019)
Accurate blackbodies
Proceedings of SPIE (July 19 2010)
Uncertainties In Calibration
Proceedings of SPIE (November 10 1975)

Back to Top