Paper
22 December 2015 Sub-bandage sensing system for remote monitoring of chronic wounds in healthcare
Alex Hariz, Nasir Mehmood, Nico Voelcker
Author Affiliations +
Proceedings Volume 9668, Micro+Nano Materials, Devices, and Systems; 966853 (2015) https://doi.org/10.1117/12.2221917
Event: SPIE Micro+Nano Materials, Devices, and Applications, 2015, Sydney, New South Wales, Australia
Abstract
Chronic wounds, such as venous leg ulcers, can be monitored non-invasively by using modern sensing devices and wireless technologies. The development of such wireless diagnostic tools may improve chronic wound management by providing evidence on efficacy of treatments being provided. In this paper we present a low-power portable telemetric system for wound condition sensing and monitoring. The system aims at measuring and transmitting real-time information of wound-site temperature, sub-bandage pressure and moisture level from within the wound dressing.
The system comprises commercially available non-invasive temperature, moisture, and pressure sensors, which are interfaced with a telemetry device on a flexible 0.15 mm thick printed circuit material, making up a lightweight biocompatible sensing device. The real-time data obtained is transmitted wirelessly to a portable receiver which displays the measured values. The performance of the whole telemetric sensing system is validated on a mannequin leg using commercial compression bandages and dressings. A number of trials on a healthy human volunteer are performed where treatment conditions were emulated using various compression bandage configurations.
A reliable and repeatable performance of the system is achieved under compression bandage and with minimal discomfort to the volunteer. The system is capable of reporting instantaneous changes in bandage pressure, moisture level and local temperature at wound site with average measurement resolutions of 0.5 mmHg, 3.0 %RH, and 0.2 °C respectively. Effective range of data transmission is 4-5 m in an open environment.
© (2015) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Alex Hariz, Nasir Mehmood, and Nico Voelcker "Sub-bandage sensing system for remote monitoring of chronic wounds in healthcare", Proc. SPIE 9668, Micro+Nano Materials, Devices, and Systems, 966853 (22 December 2015); https://doi.org/10.1117/12.2221917
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Sensors

Sensing systems

Temperature metrology

Wound healing

Skin

LCDs

Medicine

Back to Top