25 October 2016 Researches on stability of microfocus electron-impact x-ray source
Author Affiliations +
Abstract
A microfocus electron-impact X-ray source with micro-beam was introduced in this paper. The tungsten cathode electrogun is used as emitting system, and the focusing system is consists of two magnetic solenoid lenses, it is effective, light and handy. The matching problems between emitting system focusing system are studied on the microfocus X-ray source. The current of the first focusing lens and the second focusing lens is 0.8A and 1.64A at the voltage of 90kV respectively, and the filament current is 2.5A. Under the condition, the micro-beam spot X-ray is gained. The test results of stability showed that the X-ray source have a excellent stability, X-ray intensity of which is 110.6±0.03μSr/hr, target current of which is 185.5±1.5μA, and the target temperature of which is 96.5±0.5°C.The resolution of micro-focus X-ray source is about 4μm by the analysis of JIMA, which meet the application requirement of microfocus X-ray source.
© (2016) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Yu-tian Ma, Jun-Biao Liu M.D., Wei-xia Zhao, Geng Niu, and Li Han "Researches on stability of microfocus electron-impact x-ray source", Proc. SPIE 9685, 8th International Symposium on Advanced Optical Manufacturing and Testing Technologies: Design, Manufacturing, and Testing of Micro- and Nano-Optical Devices and Systems; and Smart Structures and Materials, 96850J (25 October 2016); https://doi.org/10.1117/12.2243125
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
X-ray sources

X-rays

Electron beams

Magnetism

Spatial resolution

Lenses

Tungsten

RELATED CONTENT

The Imaging X-ray Polarimetry Explorer (IXPE)
Proceedings of SPIE (July 11 2016)
The use of Kumakhov lenses in diagnostics of osteoporosis
Proceedings of SPIE (December 06 2006)
New type of x ray source for lensless laboratory nano...
Proceedings of SPIE (September 02 2010)
Measurements of micro focus spot size in x ray tubes...
Proceedings of SPIE (December 06 2006)
CdZnTe arrays for astrophysics applications
Proceedings of SPIE (July 07 1997)