15 March 2016 Pulsed photoacoustic flow imaging of whole blood with low frequency detection
Author Affiliations +
Abstract
Ultrasound flow imaging is widely used for quantification of blood flow in vivo, but estimating low flow velocities remains challenging. Pulsed photoacoustic flowmetry could be an alternative – but has not been shown capable of deep in vivo imaging so far. A new photoacoustic system is proposed, that has more potential for deep in-vivo applications. In this work the system is tested in vitro. In contrast to earlier research, 1064 nm NIR laser irradiation is used, that would allow deeper in-vivo light penetration. For detection, a 15 MHz transducer with lower in-vivo tissue ultrasound attenuation is used. Both changes are not trivial, as they reduce the overall visibility of the photoacoustic signals from red blood cells. This work shows the flow estimation performance of this system in vitro, and aims to serve as a point of reference when moving to an in-vivo application. Using this 15 MHz transducer, 1D flow profiles of flowing blood were extracted. The jitter (standard error) in the velocity estimation over the profiles was 10% when estimating flow of particles and 20% with whole blood; the bias in flow estimation was roughly 30%.
© (2016) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Pim J. van den Berg, Khalid Daoudi, and Wiendelt Steenbergen "Pulsed photoacoustic flow imaging of whole blood with low frequency detection", Proc. SPIE 9708, Photons Plus Ultrasound: Imaging and Sensing 2016, 97081C (15 March 2016); https://doi.org/10.1117/12.2212145
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Photoacoustic spectroscopy

Blood

Particles

Transducers

In vivo imaging

Ultrasonography

Error analysis

Back to Top