You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
29 March 2016“Conventional” CT images from spectral measurements
Spectral imaging systems need to be able to produce "conventional" images, and it's been shown that systems with
energy discriminating detectors can achieve higher CNR than conventional systems by optimal weighting.
Combining measured data in energy bins (EBs) and also combining basis material images have previously been
proposed, but there are no studies systematically comparing the two methods. In this paper, we analytically
evaluate the two methods for systems with ideal photon counting detectors using CNR and beam hardening (BH)
artifact as metrics. For a 120-kVp polychromatic simulations of a water phantom with low contrast inserts, the
difference of the optimal CNR between the two methods for the studied phantom is within 2%. For a
polychromatic spectrum, beam-hardening artifacts are noticeable in EB weighted images (BH artifact of 3.8% for 8
EB and 6.9% for 2 EB), while weighted basis material images are free of such artifacts.
The alert did not successfully save. Please try again later.
Paurakh L. Rajbhandary, Norbert J. Pelc, "“Conventional” CT images from spectral measurements," Proc. SPIE 9783, Medical Imaging 2016: Physics of Medical Imaging, 97831Q (29 March 2016); https://doi.org/10.1117/12.2216988