Paper
22 March 2016 Ultrasound waveform tomography with the second-order total-generalized-variation regularization
Author Affiliations +
Abstract
Ultrasound waveform tomography with the total-variation regularization could improve reconstructions of tumor margins, but the reconstructions usually contain unwanted blocky artifacts. We develop a new ultrasound waveform tomography method with a second-order total-generalized-variation regularization scheme to improve tomographic reconstructions of breast tumors and remove blocky artifacts in reconstruction results. We validate our new method using numerical phantom data and real phantom data acquired using our synthetic-aperture breast ultrasound tomography system with two parallel transducer arrays. Compared to reconstructions of ultrasound waveform tomography with modified total-variation regularization, our new ultrasound waveform tomography yields accurate sound-speed reconstruction results with significantly reduced artifacts.
© (2016) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Youzuo Lin and Lianjie Huang "Ultrasound waveform tomography with the second-order total-generalized-variation regularization", Proc. SPIE 9783, Medical Imaging 2016: Physics of Medical Imaging, 978356 (22 March 2016); https://doi.org/10.1117/12.2216422