This paper presents some aspects of the calibration plan that has been developed for NFIRAOS, the TMT first light MCAO system. The plan consolidates the best practices from current 8-meter class AO instruments, while also addressing the specificities of NFIRAOS. We present the calibration hardware that will be available in NFIRAOS, including artificial sources, diagnostic detectors and supporting software. We then discuss two key calibration procedures in more details: the measurement of the DM-WFS interaction matrix and the multi-conjugate AO correction of the NFIRAOS non-common path aberration.
© (2016) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Jean-Pierre Véran, Glen Herriot, David Andersen, and Lianqi Wang "Solving the NFIRAOS calibration puzzle", Proc. SPIE 9909, Adaptive Optics Systems V, 99090L (26 July 2016); https://doi.org/10.1117/12.2233669