1.IntroductionMonitoring and quantification of glucose level in blood is an important ongoing field of research in clinical analysis.1, 2 Specifically, monitoring blood glucose by near infrared spectroscopy (NIRS) has become one of the most active areas in biomedical optics and pharmaceutical research investigated over 20 years by many research groups.3, 4, 5, 6 More recently, interest in NIRS is on the rise for strict continuous control of blood glucose levels for patients in an intensive care unit (ICU) and operating room (OR) known as tight glycemic control (TGC).7 In TGC, blood glucose is frequently (four to six daily measurements) or continuously monitored in order to provide information needed to administer the proper amount of insulin or glucose in order to avoid hypoglycemia or hyperglycemia while maintaining normoglycemia (80 to 110 mg/dL).8 Recently, Tang report on the effect of 30 drugs used in hospitals on glucose measurement with handheld glucose meters and a portable glucose analyzer.9 To date, however, very little research is available describing the influence of drugs, found in many patients while treated in ICU, on NIRS measurements. The Food and Drug Administration of the United States also requires understanding regarding the effect of drugs on accuracy and precision for blood glucose assays using NIRS. In general, the measurement of blood glucose by any technique is inherently complex because of the wide range of potentially interfering components in blood and body tissues. Therefore, a study designed to screen the influence of interferent substances at clinically relevant concentrations is necessary. The clinical and laboratory standards institute (CLSI) refer to the term interference as a cause of clinically significant bias in the measured analyte concentration due to the effect of another component or property of the sample.10 By clinical significance we mean the importance of an error to its potential to alter a physician's diagnosis, treatment, or management of a patient. One of our goals in this study was to evaluate the effects of test interferent compounds on NIRS glucose prediction errors. The main test interferences for this evaluation are saccharide compounds, suspected as having the greatest potential to interfere with NIR glucose analysis. Saccharide compounds can be divided into three main groups: 1. monosaccharides (e.g., fructose, galactose, mannitol, and xylose), 2. disaccharides (e.g., maltose), and 3. polysaccharides (e.g., dextran and hetastarch). These saccharides are suspect since they have similar spectroscopic fingerprints to glucose and may be present physiologically at large relative concentrations. The selection of specific interferents and test concentration levels for interference testing are gathered from a variety of reference literature sources. Although not comprehensive in these literature, sources for interference information include the following: common handheld meter interferences, selected interferences from the EP7-A2 guidelines (high concentrations), and commonly used drugs in ICUs. To reduce or eliminate the effect of these saccharide compounds on glucose measurements, a preprocessing method such as orthogonal signal correction (OSC) can be used. OSC aims to remove strong structured variation from the spectral data, X, that is unrelated, or orthogonal, to the concentration Y.11 To achieve such orthogonality, X is decomposed into scores (t), which are required to be orthogonal to Y, and loading (p) vectors. Multivariate calibration by partial least squares (PLS) is accomplished with X calibrated against scores and new scores and loadings are recalculated.12 Next, new scores and loadings, which contain information not related to the concentration, are subtracted from X in order to correct the spectra. For new spectral data, X new, a new score vector is calculated and multiplied with the transpose of the loading vector previously found. Finally, the product of the two is subtracted from X new. This type of mathematical treatment has been applied here with one more element; during calibration and validation modeling the bias (intercept) and slope from the prediction data sets, that are not included in the calibration model, were determined. This process was repeated 100 times (∼10% of the model size across the entire set) in order to compute statistic power on the bias and slope parameter. For each new validation spectra, both the prediction vector, [TeX:] $\hat b$ , and the bias and slope from the calibration process were automatically applied yielding a standardized prediction. A flow chart describing the difference between OSC and slope and bias OSC (sbOSC) is depicted in Fig. 1. In this work, the ability to determine glucose concentration in the presence of saccharide interferences in NIRS measurements by using Fourier transform near-infrared spectroscopy (FT-NIRS) with a modification to OSC, for the first time to our knowledge, was investigated. Detailed results demonstrate the use of a single calibration for quantitative determination of glucose levels in whole blood samples from 14 healthy anonymous patients. These patient samples contained maximum physiological levels of six different saccharides. The measurement error effect of the individual interferents was determined based on the maximum interference effect (MIE) of each substance. Furthermore, the detailed use of FT-NIRS in transmission mode coupled with sbOSC and a single PLS calibration model for determination of glucose in whole blood will be described. 2.Material and Methods2.1.InstrumentationThe transmittance Fourier transform near-infrared (FT-NIR) spectrometer was designed and built in-house that has an operating wavelength range of 1.25–2.5 μm (4000–8000 cm−1) with a spectral resolution of 32 cm−1. It uses the same basic optical configuration of a Michelson interferometer platform.13 The spectrometer consists of a uniform, constant, blackbody light source (mini-igniter) with a color temperature of approximately 1035 °C, matched fused silica beam splitter and compensator plates, fixed and moveable (active) corner-cube retroreflectors, 1 mm diameter extended-wavelength InGaAs PIN photodiode, and a data acquisition system that digitizes the interferogram data and stores them on PC. The interferogram data was processed and analyzed off-line using MatLab. Light reaching the sample is high pass filtered by the use of a silicon collimator lens that attenuates wavelengths below 1.05 μm. A spring suspension system along with a voice coil actuator is used to move the active retroreflector in a strictly axial direction. The active mirror is continuously scanning at a constant velocity of 0.8 cm·s−1, 70% duty cycle, which results in approximately eight interferograms per second. The blood and background (Saline) samples were pumped through a borosilicate flow cell with a path length of 1 mm that was controlled at 34 °C. 2.2.Sample HandlingHuman blood was collected from healthy human donors in the morning between 7:30 to 9:00 am on each testing day. Using a venipuncture technique, blood is drawn into multiple 50 mL heparinized syringes. Blood is immediately transported to our lab by motor vehicle, which takes approximately 30 min. Once at the lab, blood syringes are transferred to 50 mL centrifuge tubes. A small and well-mixed sample is used for native hematocrit (Hct) and glucose readings. A VWR clinical 200 large centrifuge with Hct adapter and YSI 2300 glucose analyzers are respectively utilized to measure Hct and glucose values. Regardless of Hct volume percentage, all blood tubes are spun down at 3000 rpm for 5 min using the VWR centrifuge with 50 mL tube adapter and cell buffy coats are gently removed. As will be mentioned in Sec. 2.4, our model database consists two sample groups: (1) samples that are designed with different measurement conditions: flow rate, temperature, and flowcell pathlength with six different glucose levels and three Hct levels; and (2) saccharide containing samples. For the first group, red blood cells (RBC) are reconstituted with plasma to generate different Hct levels. A wide range of samples are created from three Hct levels and six glucose stock concentrations. Each blood sample is spiked with glucose concentration at least ten minutes before the sampling process and that the sample is gently mixed during that time. A small blood sample is also kept for Hct and glucose readings. For the second group, RBCs are then reconstituted with plasma to create a mid-range Hct level. High concentration glucose stock solution prepared with saline is used to spike the blood to three different glucose levels. Interference solutions are also prepared with saline and at ten times the tested concentration. Saccharide interferences are mixed at least 12 h before testing time to ensure the anomeric stability of solutions. These concentrated interferences are then added to blood at a 1:10 ratio to yield the final testing concentration and at least 10 min before the sampling process. Samples for the prediction set are prepared as follows: RBCs are reconstituted with plasma to create a mid-range Hct level. High concentration glucose stock solution prepared with saline is used to spike the blood. Saccharide interferences are mixed at least 12 h before testing time to ensure the stability of solutions. These concentrated interferences are then added to blood at a 1:10 ratio to yield the final testing concentration at least 10 min before the sampling process. 2.3.MeasurementsFT-NIRS measurements were performed on whole blood samples obtained from 14 anonymous patients. Spectra were collected in transmission mode in the spectral region from 4000 to 8000 cm−1 from two FT-NIR systems (Nos. 7 & 8). Symmetrical interferograms that contained 2048 point were measured at 32 cm−1 resolution at full throughput. Background spectra of Saline (NaCl 0.9%) were collected before and after blood sample measurement for 30 s to account for variation in instrument noise and drift profiles. On the other hand, blood measurements were obtained for 60 s. Each background and blood measurement is an average of ∼230 and ∼460 spectra, respectively. Reference measurements were made by two bench whole-blood glucose analyzers, YSI 2300, which has an accuracy of ±3 mg/dL; instrument A versus instrument B (1 sigma). Glucose samples were in the concentration range of 30–500 mg/dL with variable Hct levels ranging from 24 to 48 vol.%. For drug measurements to be included in the calibration and validation model the NIR interference is measured by spiking a control blood sample with an interferent and measuring the NIR glucose for the spiked sample versus an untreated control sample (the control sample has an identical volume of saline without the interferent). In this manner, interferent spiked and control samples are serially measured following CLSI guidelines (EP7-A2, Interference Testing). 2.4.Calibration ModelThe goal during NIR calibration is to increase analytical specificity or selectivity for glucose and decrease the effects of interferents. For that, a PLS1 model was used to develop a multivariate calibration model for predicting glucose concentration from the NIR spectra.14 Spectra here refer to the absorbance data given by the modified Beer–Lambert law15 Eq. 1[TeX:] \documentclass[12pt]{minimal}\begin{document}\begin{equation} \mu _a = \log _{10} \left({\frac{{I_{bck} }}{{I_{bld} }}} \right), \end{equation}\end{document}Calibrations for glucose must be designed such that they are sensitive to changes in glucose concentration but are insensitive to interferent chemicals. A method for making calibrations more selective to the analyte of interest and less selective toward an interferent includes the use of “interferent compensation.” This is accomplished by including low (nonzero), medium, and high concentration samples into the initial calibration and testing the sensitivity of the calibration to glucose and to the interferent following the incorporation of such samples into the initial calibration. A standard factorial design for including such samples is depicted in Fig. 2. This is implemented by taking the same basic blood sample with three different Hct levels (24, 36, and 48 vol.%) and spiking it to three levels of glucose (80, 140, and 200 mg/dL), and three levels of interferent (low, mid, and high expected physiological concentration). Table 1 summarizes these ranges of expected concentration for each inteferent. These intervals of values are near those commonly used in ICU and in OR. In addition, a combination of spectra from sample types with “real-world” effects (instrument type, sample temperature, flow rate, and flowcell pathlength) was added to the calibration model. Hence, our calibration data is formulated from a mixture of two groups of samples. The first group (N = 1007) is the real-world samples while the second (N = 361) contains the saccharide samples at constant: flow rate (0.5 ml/min), temperature (25ºC) and flowcell pathlength (1 mm) state. Table 1Three level of interferent concentration used in this work. To convert mg glucose/dL to mmol/L, divide by 18.015
Outlier detection based on both Mahalanobis distance (MD) and spectral F-ratio (SFR) metrics were applied to ensure the quality of the data used to construct the calibration model. A criterion of μ ± 3σ was used as an outlier screening for each metric. μ and σ represents the mean and the standard deviation receptively of the MD (or SFR). Samples that exhibit MD and SFR above this criterion were considered an outlier and removed from the calibration model. OSC was applied for preprocessing of the NIR spectra before being used in the PLS model. As stated in Sec. 1, the main goal of OSC is to determine and remove from the spectral data, X, the part of information that is not correlated with the concentration, Y, within a limited number of orthogonal scores (t) and loadings (p). Filtered data are obtained after iterative removal of the orthogonal components as follows: Eq. 2[TeX:] \documentclass[12pt]{minimal}\begin{document}\begin{equation} X_{OSC} = X - \sum\limits_{i = 1}^n {t_i \cdot p_i ^T }, \end{equation}\end{document}After outlier screening and OSC treatment, a PLS calibration matrix is formed. A leave-one-out method of cross-validation was performed on the calibration data set to determine the optimal number of factors (i.e., latent variables). The PLS model size was chosen as the number of factors resulting in the first minima of the standard error of cross-validation (SECV). Common calibration statistics such as SECV, mean relative error, coefficient of determination (R 2), etc. were computed to assess the fitting ability of the model. Once a model has been built, it can be used to predict the concentration of the unknown samples. 2.5.Validation ModelValidation data sets were composed of saccharide samples similar to the data included in the calibration model. These data sets differ from those in the calibration in their concentration level (Table 1, third column) and in the time when they were measured. The spectral data matrix, X val, was constructed by serially measuring new spectra of different interferences and control samples according to the following manner: Control, Interfernt1, Control, Interfernt2, Control, …, Interferent n, etc. such that each block-run consists of ten interferent and ten control samples. The interferent treated test samples are immediately measured following the control samples and include an identical dilution volume of saline only as an added volume. X val was then subtracted from the scores and loading vector obtained from the calibration model; Eq. 3[TeX:] \documentclass[12pt]{minimal}\begin{document}\begin{equation} \tilde X_{Val} = X_{Val} - t \cdot p'_{Cal}, \end{equation}\end{document}2.6.Computation and SoftwareAll computations were performed in Matlab software (Mathworks, Inc., R2008b) on an Intel Pentium (T4400) running at 2.2 GHz with 4GB memory. PLS program for calibration-prediction was obtained using PLS-Toolbox (Eigenvector Company, Ver. 2.1) and OSC m-file achieved from Ref. 17. Preprocessing algorithms were written in-house in Matlab language. During data processing the region of 4935–5300 cm−1, representing the water band, was cut off (i.e., trimmed) in order to eliminate noise caused by high water absorption. 3.Results and DiscussionThe aim of the first experiment was to investigate the effect of the saccharide interferences, on the prediction (validation) data sets. A PLS calibration was built with 1007 spectra from the real-world group as described in Sec. 2.4. The number of PLS factors used in the model and the cross-validation glucose prediction plot in the Clarke Error Grid Analysis (CEGA) format of these data sets are presented in Figs. 3a and 3b, respectively. Usually, CEGA is used to quantify clinical accuracy of blood glucose measurement techniques in comparison to a reference value.18 The grid consists of five zones A–E and provides an assessment of measurement errors and their respective clinical significance. For example, zone A is the region in which the NIR analytical values are within 20% of the reference values and indicates errors without serious clinical implications. As depicted, a linear correlation of R 2 = 0.99 between the NIR results and the YSI glucose concentrations, with SECV = 4.1 mg/dL is observed. These results, together with the fact that 100% of the prediction fall within zone A and zero% results on the other zones, confirm both the ability of PLS to calibrate models by using NIR spectra with real-world effects and the reliability of our FT-NIRS modality. Once the calibration is complete, the prediction data sets were build composed by six saccharide interferences: Dextran, Fructose, Galactose, Maltose, Mannitol, and Xylose. For interferent compensation, each interferent was spiked with different concentration levels and different glucose and Hct levels as per a factorial design depicted in Fig. 2; Table 2 shows an example for the Xylose experiment. The NIR predictions are reported under two basic conditions: 1. extrapolation testing without calibration interferent accommodation (compensation) of the test interferent compounds, where the calibration model predicts outside of its calibration space, and 2. interpolation with calibration interferent accommodation (compensation) of the test interferent compounds (i.e., including the extremes within the model), where the model predicts within its calibration space. Figure 4 together with Table 3 presents the results of the prediction in terms of maximum interference effect (MIE). MIE is expressed by Eq. 5[TeX:] \documentclass[12pt]{minimal}\begin{document}\begin{equation} MIE = \left| {NIR - YSI} \right| + 1.96 \times \left({\frac{{SEP}}{{\sqrt n }}} \right), \end{equation}\end{document}Eq. 6[TeX:] \documentclass[12pt]{minimal}\begin{document}\begin{equation} SEP=\sqrt{\frac{\sum({\hat y}-y)^2}{N_p},} \end{equation}\end{document}Table 2Example of Xylose sequence measurements. Total sample measurements are equal to 26. Discontinuity in Sample ID results from the measurement of other inteferent according to CLSI Guidelines.
Table 3Comparison of the prediction results of Maximum Interference Effect (MIE) for six saccharide interferences when exclude and include in the calibration model.
Before assessing the feasibility of sbOSC, a new experiment (experiment #2) was conducted in order to validate the efficacy of the standard OSC on interferent containing sample prediction. In this experiment a PLS calibration was constructed with 696 spectra obtained from a combination of a third set of real-world random samples (N = 335) and saccharide interferences (Ns = 361) having different concentrations. The performance of the calibration model before and after the use of OSC is shown in Figs. 5a and 5b, respectively. Inspection of the correlation plots in Fig. 5b in comparison with Fig. 5a reveals that with OSC processing the prediction results are considerably concentrated along the unity line with R 2 = 0.99 and a ∼70% decrease in SECV (19.1 mg/dL versus 5.8 mg/dL). In addition, Fig. 4b shows that a high percentage of the NIR predicted data points fall within the zones A and B of the Clarke grid; 99.9% and 0.1% of all NIR predictions fall in the A and B zones, respectively. On the contrary, without using OSC, 84.2%, 15.2%, and 0.6% of the predictions fall in the A, B, and D zones, respectively, resulting in a much higher error expressed as SECV. The next experiment's (experiment #3) purpose was to establish a statistical method for computing the calibration slope (m) and bias (i, intercept) used later during the sbOSC procedure. Eighty percent of the data set of random real-world samples (N = 807) were combined together with saccharide interferences (Ns = 361) in order to build an NIR calibration model. OSC was applied for preprocessing of the NIR spectra of these mixed samples (N T = 1168) before being used in the calibration modeling step. The validation data sets were composed of the same saccharide type molecules, however, as previously mentioned, with high physiological concentration levels (Table 1, third column). In addition, the spectra of these validation samples were collected on the same two FT-NIRS instruments 35 days after the samples used for calibration. A total of 100 calibration models were developed by selecting different random samples for calibration and validation in order to determine the distribution for the slope and bias. For each set of experiments m and i metrics were computed according to the following expressions: Eq. 7[TeX:] \documentclass[12pt]{minimal}\begin{document}\begin{eqnarray} m &=& \frac{{n \cdot \sum\limits_{i = 1}^n {(\hat y_i \cdot y_i) - \sum\limits_{i = 1}^n {\hat y_i \cdot \sum\limits_{i = 1}^n {y_i } } } }}{{n \cdot \sum\limits_{i = 1}^n {\left({\hat y_i^2 } \right) - \sum\limits_{i = 1}^n {\left({\hat y_i } \right)^2 } } }}, \nonumber\\ i &=& \frac{{\sum\limits_{i = 1}^n {\hat y_i^2 \cdot \sum\limits_{i = 1}^n {y_i - \sum\limits_{i = 1}^n {\hat y_i \cdot \sum\limits_{i = 1}^n {(\hat y_i \cdot y_i)} } } } }}{{n \cdot \sum\limits_{i = 1}^n {\left({\hat y_i^2 } \right) - \sum\limits_{i = 1}^n {(\hat y_i)^2 } } }}, \end{eqnarray}\end{document}Eq. 8[TeX:] \documentclass[12pt]{minimal}\begin{document}\begin{equation} \hat y_{new} = \frac{{\hat y_{old} - \bar i}}{{\overline m}}, \end{equation}\end{document}After identifying the slope and bias factors, a random data set of entire real-world samples (N = 200) were combined together with the saccharide interference samples (N = 361) to construct the PLS model for the calibration. Prediction data sets, not included in the calibration model, were employed as an independent test set. Therefore, our validation process is obtained on “blind” saccharide interferents samples. In addition the prediction equation from Eq. 4, was modified accordingly by Eq. 9[TeX:] \documentclass[12pt]{minimal}\begin{document}\begin{equation} \tilde Y_{Val} = \frac{{Y_{Val} - \bar i}}{{\overline m}}, \end{equation}\end{document}Eq. 10[TeX:] \documentclass[12pt]{minimal}\begin{document}\begin{equation} \tilde Y_{{Val} _{final}} = \tilde Y_{Val_{Intr}} - \tilde Y_{Val_{Cntrl}}. \end{equation}\end{document}Table 4Comparison of calibration and validation performance between sbOSC to several background correction methods. Validation results are present as the Maximum Interference Effect (MIE) in mg/dL. The validation data sets are saccharide interferences non-included in the calibration model.
Table 5Comparison of calibration and validation performance of sbOSC when Dextran excluded during modeling. Validation results are present as the Maximum Interference Effect (MIE) in mg/dL.
4.ConclusionsThere are medicines that can cause hypoglycemia (low blood sugar) or hyperglycemia (high blood sugar) in diabetic and nondiabetic patients. These drugs have a possible effect to interfere with NIR glucose measurements. In this paper, a method for determination of glucose concentrations in blood in presence of saccharide interferences using FT-NIRS and sbOSC as a pretreatment method was demonstrated. sbOSC is based on the traditional OSC platform but in addition uses a universal slope and bias correction on the prediction sets. The slope and bias were computed during calibration using 100 different calibration models and are used later for the validation prediction and routine analysis. A clear advantage of OSC and sbOSC presented in MIE value performance is observed in comparison to other basic preprocessing spectral correction techniques (Table 4). The novelty of this work is our ability to directly quantify glucose value in the presence of saccharide interferences in NIR region using FTS, and the ability of the sbOSC approach to further decrease the MIE values by an additional 10% for each interferent over OSC-PLS. To the best of our knowledge, no previous report of such a study and calibration development has been previously described. The overall results presented in this study clearly show the potential and versatility of our methodology, which could be applied for blood glucose assay in ICU and OR rooms. ReferencesM. B. Davidson, Davidson's Diabetes Mellitus: Diagnosis and Treatment, 5th ed.Saunders, Philadelpia
(2003). Google Scholar
O. S. Khalil, Diabetes. Technol. Ther., 6 660
(2004). https://doi.org/10.1089/dia.2004.6.issue-5 Google Scholar
K. J. Pezzaniti, T.-W. Jeng, L. McDowell, and
G. Oosta, Clin. Biochem., 34 239
(2001). https://doi.org/10.1016/S0009-9120(01)00198-9 Google Scholar
M. A. Arnold and
G. W. Small, Anal. Chem., 77 5429
(2005). https://doi.org/10.1021/ac050429e Google Scholar
J. T. Olsberg, M. A. Arnold, C. Mermelstein, J. Schmitz, and
J. Wagner, Appl. Spectr., 59 1480
(2005). https://doi.org/10.1366/000370205775142485 Google Scholar
J. T. Olesberg, L. Liu, V. V. Zee, and
M. A. Arnold, Anal. Chem., 78 215
(2006). https://doi.org/10.1021/ac051036i Google Scholar
K. Maruo, T. Oota, M. Tsurugi, T. Nakagawa, H. Arimoto, M. Tamura, Y. Ozaki, and
Y. Yamada, Appl. Spectr., 60 441
(2006). https://doi.org/10.1366/000370206779321508 Google Scholar
V. R. Kondepati and
H. M. Heise, Anal. Bioanal. Chem., 388 545
(2007). https://doi.org/10.1007/s00216-007-1229-8 Google Scholar
Z. Tang, X. Du, R. F. Louie, and
G. J. Kost, Am. J. Clin. Pathol., 113 75
(2000). Google Scholar
R. J. McEnroe, M. F. Burritt, D. M. Powers, D. W. Rheinheimer, and
B. H. Wallace, Interference Testing in Clinical Chemisrty; Approved Guidline – second ed., Pennsylvania,2005). Google Scholar
S. Wold, H. Antti, F. Lindgren, and
J. Ohman, Chem. Intell. Lab. Syst., 44 175
(1998). https://doi.org/10.1016/S0169-7439(98)00109-9 Google Scholar
H. Mark and
J. Workman, Chemometrics in Spectroscopy, Elsevier, Amsterdam
(2007). Google Scholar
A. P. Throne, Spectrophysics, 2ndChapman & Hall, London
(1988). Google Scholar
P. Geladi and
B. R. Kowalski,
“Partial Least-Squares regression; a tutorial,”
Anal. Chim. Acta, 185 1
(1986). https://doi.org/10.1016/0003-2670(86)80028-9 Google Scholar
V. V. Tuchin, Tissue Optics: Light Scattering Methods and Instruments for Medical Diagnosis, 2ndSPIE Press, Bellingham, Washington
(2007). Google Scholar
M. Blanco, J. Coello, I. Montoliu, and
M. A. Rometo, Anal. Chim. Acta., 434 125
(2001). https://doi.org/10.1016/S0003-2670(01)00820-0 Google Scholar
W. L. Clarke, D. Cox, L. A. Gonder-Frederick, W. Carter, and
S. L. Pohl, Diabetes Care, 10 622
(1987). https://doi.org/10.2337/diacare.10.5.622 Google Scholar
D. E. Honigs, G. M. Hieftje, H. L. Mark, and
T. B. Hirschfeld, Appl. Spectrosc., 57 2299
(1985). Google Scholar
J. W. Westerhuis, S. de-Jong, and
A. K. Smilde, Chemom. Intell. Lab. Syst., 56 13
(2001). https://doi.org/10.1016/S0169-7439(01)00102-2 Google Scholar
|