Open Access
18 January 2018 Estimation of cerebral metabolic rate of oxygen consumption using combined multiwavelength photoacoustic microscopy and Doppler microultrasound
Author Affiliations +
Abstract
The metabolic rate of oxygen consumption is an important metric of tissue oxygen metabolism and is especially critical in the brain, yet few methods are available for measuring it. We use a custom combined photoacoustic-microultrasound system and demonstrate cerebral oxygen consumption estimation in vivo. In particular, the cerebral metabolic rate of oxygen consumption was estimated in a murine model during variation of inhaled oxygen from hypoxia to hyperoxia. The hypothesis of brain autoregulation was confirmed with our method even though oxygen saturation and flow in vessels changed.

1.

Introduction

The brain is one of the most energy-demanding and metabolically active organs of the body. Although it comprises only 2% of body weight, it receives about 15% of cardiac output and uses 20% of total body oxygen and 25% of total body glucose.1 Under normal physiological conditions, oxidative metabolism of glucose is the primary way to produce adenosine triphosphate, the most important energy source for the brain. Oxygen needs to be delivered to the cerebral tissue at a rate that is biochemically appropriate to neurons’ metabolic needs.1 The cerebral metabolic rate of oxygen (CMRO2) depends on the density of neurons and on their state of functional activation.2 Many of the common disorders of the brain, such as Alzheimer’s, Parkinson’s, Huntington’s, and others, have been found to be associated with alterations in the cerebral oxygen metabolism.35 Therefore, the measure of CMRO2 would not only enable a better understanding of the normal physiology during rest, sleep, anesthesia, aging, or functional brain tasks6,7 but also be important for research and clinical applications related to brain disorders, stroke, and help to improve the management of patients who are at the risk of developing brain hypoxia-ischemia.8,9

Since the extraction of oxygen from cerebral tissue is closely matched to the brain’s metabolic needs, CMRO2 should be equal to the total amount of oxygen delivered to the cerebral tissue per unit time minus the amount leaving in the venous circulation per unit time.2 As shown in Eq. (1), the cerebral blood flow (CBF) and the arteriovenous oxygen difference (AVDO2) are the key parameters to estimate CMRO2 based on Fick’s principle.10 Under steady-state physiologic conditions, CMRO2 remains relatively constant over a range of arterial partial pressure of oxygen (PaO2), typically 23 to 100 mm Hg, due to the cerebral autoregulation mechanism.2 During mild hypoxia (PaO2<90  mm Hg), the oxygen supply may be inadequate; however, metabolic demand may still be met by elevated CBF or reduced venous blood saturation, which means the higher extraction of oxygen from the blood.11 Severe hypoxia (PaO2<30mm Hg) may cause significant decrease in CMRO2 due to insufficient blood flow or oxygen extraction compensation.12 The metabolic autoregulation is a negative feedback system that seeks to balance the oxygen supply to its demand in a certain range13

Eq. (1)

CMRO2=CBF×AVDO2.

Currently, techniques to obtain CMRO2 are limited. The important criteria for such techniques include noninvasiveness, adequate spatial and temporal resolution, low-radiation exposure, good safety profile, and widespread clinical availability.14 Unfortunately, no in vivo method can fulfill all of these requirements. In early studies, the jugular vein oximetry involving catheterization has been used for a surrogate measure of CMRO2 but is invasive.15 Positron emission tomography (PET) with O15-labled radiotracers provides the most direct measurement of CMRO2.16 Yet, the spatial resolution is poor (1  cm for clinical system), and the high ionizing radiation dose prevents its repeated use on the same patient. Magnetic resonance imaging (MRI) is also applied to image CMRO2 by utilizing the blood oxygenation-level-dependent (BOLD) effect. BOLD MRI is noninvasive and can be performed with high spatial resolution, but it is only sensitive to deoxy-hemoglobin and has trouble distinguishing between changes in oxygen saturation and blood flow.17 In recent studies, phase-contrast MRI has been proposed to quantify the blood flow in major inflow vessels; however, for the oxygen saturation estimation, only a large-sized vein (superior sagittal sinus) can be targeted.18 Moreover, the complex setup and high expenses preclude PET and MRI for many bedside clinical applications.

Imaging techniques based on optical contrast, which are sensitive to blood functional parameters, have been widely employed for the assessment of tissue oxygen metabolism. Near-infrared spectroscopy offers a means to estimate CMRO2 noninvasively with the blood flow information provided by diffuse correlation spectroscopy.19,20 The main drawback is that the spatial resolution is relatively low due to strong optical scattering. Photoacoustic imaging overcomes this limitation by ultrasonically imaging optical contrast through the photoacoustic effect.21 Optical-resolution photoacoustic microscopy (OR-PAM) has been used to estimate oxygen consumption in superficial tissues.22 This technique utilizes fine optical focusing to achieve high lateral resolution (5  μm); however, the imaging depth is shallow (<1  mm) and the detectable flow speed (<12  mm/s) is limited due to low laser repetition rate.23 While recent work suggests hope for photoacoustic flow estimation in deep tissues (>1  mm) using acoustic-resolution photoacoustic imaging,24 it is less established than Doppler ultrasound methods we selected in this paper.

Here, we propose CMRO2 estimation using photoacoustic and Doppler ultrasound methods. The combination of local blood oxygenation estimation and flow estimation in deep vessels is accomplished with a custom photoacoustic and microultrasound scanning system. We previously demonstrated oxygen flux estimation in a phantom study using such a system.25 In this paper, the CMRO2 is estimated in vivo by combining the measurements of arterial and venous oxygen saturation and flow rate of the internal jugular vein.

2.

Materials and Methods

2.1.

Experiment Setup

Our custom photoacoustic and microultrasound system was constructed by a combined light delivery and ultrasound probe, a tunable laser, a computer-controlled scanning, and triggering and data acquisition system. As the irradiation source, a tunable optical parametric oscillator (Surelite OPO Plus, Continuum, California) was pumped by a Q-switch Nd:YAG nanosecond-pulsed laser (Surelite III, Continuum, California) at a repetition rate of 10 Hz. The tuning range is from 410 to 710 nm. The laser output was coupled into a bifurcated fiber-bundle light guide with a core diameter of 185  μm (900 fibers, CeramOptec Industries, Germany). An acrylic probe was designed to position the bifurcations of the light guide around a 25-MHz single-element-focused ultrasound transducer (V324-SM, Olympus Panametrics-NDT, Massachusetts). The transducer was placed such that the focus of the transducer and the center of the illumination spot (1  cm2) were aligned. The illumination fluence was measured to be about 10  mJ/cm2. This probe was mounted on a three-axis motion system for B-scan and C-scan imaging. A peripheral component interconnect (PCI) motion card (NI7350, National Instruments Inc., Texas) was used to control a high-precision three-axis motion stage with three integrated stepper motors (23Y002D-LW8, Anaheim Automation, California). Fast scanning was accomplished with a voice-coil actuator as previously described25,26 while vertical and elevational translation were accomplished using the stepper motors.

Signals from the transducer were amplified by a variable-gain preamplifier then further amplified by an ultrasound pulser/receiver (5073PR, Panametrics-NDT, Massachusetts). The variable-gain stage (AD603ARZ evaluation board, Analog Devices, Massachusetts) was used to implement time-gain compensation (TGC). The TGC curves in a range of 40 dB for both ultrasound and photoacoustic imaging were controlled via a digital input–output (DIO) card (NI PCI-6542, National Instruments Inc., Texas). Signals were digitized using a multichannel data acquisition card (CS8289 Gage Cobra, Gage Applied Systems Inc., Illinois) with 12-bit dynamic range and sampling rates as high as 125  MSamples/s and postprocessed using MATLAB® (R2012a, MathWorks, Massachusetts).

Our system provided combined photoacoustic and Doppler ultrasound mode. The details are referred in our previous work.26 In brief, a pulse sequence was generated by the DIO card and sent to interleave laser and ultrasound triggers. Photoacoustic and Doppler ultrasound images were coregistered after signal processing. A schematic of the constructed system is shown in Fig. 1, which also shows a photograph of the probe with bifurcated fiber-bundle light delivery.

Fig. 1

A schematic of light delivery and data acquisition setup and photograph of the fiber bundle delivery surrounding the transducer. DIO, digital input–output; US, ultrasound; P/R, pulser/receiver; and DAC, data acquisition card.

JBO_23_1_016009_f001.png

2.2.

Animal Preparations

Sprague Dawley rats (Charles River Breeding Laboratories, Massachusetts), female, weighing about 50 g, were employed in this study. Each rat was first anesthetized in an induction chamber with 1.25% to 1.5% (volume to volume) vaporized isoflurane (IsoFlo, Zoetis Inc., New Jersey). After the rat was anesthetized, it was immobilized on a customized animal holder and fitted with a nose breather. The inhaled gas was controlled by an animal anesthesia system (EZ-SA800 Single Animal System, E-Z Anesthesia, Philadelphia) mixing 1% vaporized isoflurane with certain ratios of oxygen and nitrogen. Body temperature was maintained at 37°C with a temperature—controlled heating pad, and the blood oxygenation levels and heart rates were monitored by a pulse oximeter (Nonin 8600V, Nonin Medical Inc., Minnesota) clamping on the hind leg of the animal. All in vivo procedures and protocols were reviewed and approved by the University of Alberta Animal Care and Use Committees.

The physiological conditions (hyperoxia, normoxia, and hypoxia) of the animal can be achieved by varying the fraction of inspired oxygen (FiO2), which alters the arterial oxygen saturation acutely.12 In this study, mild hypoxia and mild hyperoxia (PaO2 within the range of 30 to 100 mm Hg) were introduced to avoid significant changes in cerebral oxygen consumption. According to Kelman’s oxygen hemoglobin dissociation curve,27 the arterial oxygen saturation should be within the range of 60% to 100%. The animal was exposed to different FiO2 during gas anesthesia by mixing the output of regulated oxygen and nitrogen tanks with precise flow measurements read from the respective flow meters on the tanks. The global arterial oxygen saturation was recorded at each FiO2.

2.3.

Blood Flow Estimation using Doppler Ultrasound

Phantom imaging studies of flow estimation using double-dye mixtures and sheep blood were previously reported.25 Pulsed color Doppler ultrasound was used for flow velocity estimation. A sequence of modulated ultrasound pulses was transmitted along a certain Doppler angle, and the flow velocities were calculated from detected phase shifts in received echoes.28 However, when the Doppler angle approaches 90 deg, the phase shift almost vanishes and the transverse flow velocity cannot be identified. In phantom studies,25 the appropriate Doppler angle was possible owing to oblique vessel orientation; however, the in vivo measurements were to be done along vessels almost perpendicular to the ultrasound beam axis. Considering the difficulty of creating such a Doppler angle with our wide-footprint probe, we adopted a different approach termed Doppler bandwidth broadening to transverse flow estimation, similar to that used in previous ultrasound and OR-PAM techniques.23,29 In brief, assuming negligible velocity variation within the sample volume and sufficient velocity resolution, the bandwidth of the Doppler spectral broadening is given as

Eq. (2)

Bw=2f0c0DFvsinθ,
where Bw is the Doppler bandwidth, f0 is the center frequency, c0 is the speed of sound, D is the transducer aperture, F is the focal length, v is the flow velocity, and θ is the Doppler angle of the flow relative to the ultrasound beam axis. Since the square of Doppler bandwidth Bw can be approximated by the variance of the Doppler spectrum,29 the transverse flow (θ=90  deg) can be estimated as

Eq. (3)

vT=c02f0FDσ2,
where σ2 is the variance of Doppler spectrum, which is calculated routinely in correlation-based color Doppler ultrasound algorithm.28

Our system also provided power Doppler ultrasound to image blood volume fractions. Different from flow velocity estimation, the intensity of power Doppler image is proportional to the number of scatterers moving within a vascular volume rather than the frequency shifts they generate,30 which makes it more sensitive to blood flow detection and independent of Doppler angles.31 Since tissues surrounding blood vessels often generate much stronger scattered signals than moving blood, clutter filtering is necessary before velocity estimation and power calculations. Here, an adaptive eigenfilter was designed and implemented based on the statistics of the clutter and tissue movement. The correlation matrix of the received signals was decomposed into eigenvectors and eigenvalues by the discrete Karhunen–Loeve transform (DKLT), and then eigenvalues were sorted in decreasing amplitude. Since the eigenvalue spectrum represents the power distribution of the frequency components32 and the power of clutter signals arising from tissues is much higher than other components, the clutter space and flow space can be partitioned in the spectrum. Appropriate filter order was determined experimentally to give the best trade-off between clutter rejection and preservation of flow.

2.4.

Oxygen Saturation Estimation Using Photoacoustic Methods

Blood oxygen saturation (sO2) estimation was validated in phantom studies25 by relating photoacoustic estimates of sO2 to sO2 levels measured from knowledge of measured pO2 and pH and given known oxygen dissociation curves. As this is significantly more challenging in vivo, in this paper, we chose to implement two independent photoacoustic methods to cross-validate sO2 estimates. The first and most traditional approach was to solve for unknown concentration of oxy-hemoglobin [CHbO2(x)] and concentration of deoxy-hemoglobin [CHb(x)], given measured photoacoustic signals p(x,λk) and known molar extinction coefficients ϵHb(λk) and ϵHbO2(λk)

Eq. (4)

[p(x,λ1)p(x,λn)]=Γ(x)[ϕ(x,λ1)ϵHb(λ1)ϕ(x,λ1)ϵHbO2(λ1)ϕ(x,λn)ϵHb(λn)ϕ(x,λn)ϵHbO2(λn)][CHb(x)CHbO2(x)].

This approach suffers from unknown wavelength-dependent fluence ϕ(x,λk), which is often assumed a constant or requires wavelength-dependent calibration. The relative Gruneisen parameter and relative sensitivity are lumped into the wavelength-independent factor Γ(x) and will not impact the estimate of oxygen saturation of hemoglobin

Eq. (5)

sO2(x)=CHbO2(x)CHb(x)+CHbO2(x)×100%.

A second approach, proposed in Ref. 33, is to use multiwavelength photoacoustic measurements at different oxygen saturation levels and solve a system of equations to estimate the different sO2 levels. The absorption coefficients given two sO2 levels can be written as

Eq. (6)

μa1(x,λi)={ϵHb(λi)[1sO2(1)(x)]+ϵHbO2(λi)sO2(1)(x)}CHbT(x),

Eq. (7)

μa2(x,λi)={ϵHb(λi)[1sO2(2)(x)]+ϵHbO2(λi)sO2(2)(x)}CHbT(x),
where CHbT(x) is the total concentration of hemoglobin. The sO2 at the respective time points can thus be estimated from photoacoustic measurements as

Eq. (8)

sO2(1)=bp2(λ1)ap2(λ2)p1(λ2)p2(λ1)p1(λ1)p2(λ2),

Eq. (9)

sO2(2)=bp1(λ2)ap1(λ1)p1(λ2)p2(λ1)p1(λ1)p2(λ2),
with

Eq. (10)

a=[p1(λ1)p2(λ1)]ϵHb(λ1)ϵHbO2(λ1)ϵHb(λ1),

Eq. (11)

b=[p1(λ2)p2(λ2)]ϵHb(λ2)ϵHbO2(λ2)ϵHb(λ2).

This approach requires no wavelength-dependent fluence calibration factor but assumes that the optical properties of tissue between the vessels of investigation and the tissue surface do not change their optical properties. Thus, neither approach is devoid of approximations, yet consistency in estimates would provide cross substantiation.

2.5.

Cerebral Metabolic Rate of Oxygen Consumption Estimation

According to Eq. (1), assuming that CBF is the only way of oxygen delivery to the brain, the global CMRO2 can be estimated by combining the measurements of CBF, arterial oxygen saturation (sO2,a), and venous oxygen saturation (sO2,v)

Eq. (12)

CMRO2=CBF×CHbT×τ×(sO2,asO2,v),
where CHbT is the total concentration of hemoglobin with a unit of g/ml and τ is the oxygen-carrying capacity of 1.39-ml oxygen per gram of hemoglobin on the basis of a typical CHbT level (0.147  g/ml at a hematocrit of 0.42).34 Based on anatomic and hemodynamic studies,1,2 oxygen is supplied to the brain by four feeding vessels: the left and right internal carotid arteries and the left and right vertebral arteries. Although the contribution of each artery to the total cerebral perfusion is different, the oxygen saturation levels in these vessels are comparable to each other. For the blood draining from the brain, the pair of internal jugular veins provides the majority (80%) of the venous drainage via deep veins and dual sinuses. In our in vivo study, the combined imaging system scanned across the neck of the subject and the sO2,a and sO2,v were estimated from the carotid artery and jugular vein in the field of view. For CBF estimation, we used flow rates of internal jugular veins by measuring the cross-sectional area (A¯IJV) and the mean flow velocity (v¯IJV) of the vessel. Thus, the representative CMRO2 can be expressed as

Eq. (13)

CMRO2=A¯IJV×v¯IJV×CHbT×τ×(sO2,asO2,v).

The unit for CMRO2 is ml/100  g/min of the tissue.

3.

Results

First, the bandwidth broadening Doppler ultrasound approach was validated in phantoms with embedded tubing by comparing estimated flow velocity against preset flow speeds, as shown in Fig. 2. The construction of the cornstarch-gelatin phantom can be referred to our previous work.31 Under the assumption of fully developed parabolic flow, we observed excellent consistency and sensitivity to flow velocities of several mm/s, sufficient for in vivo flow estimation in murine carotid artery and jugular veins. The correlation of determination R2 of linear least-square fitting is 0.99 with a slope of 0.94 and a zero-offset of 4.2  mm/s.

Fig. 2

Comparison between the preset and the measured flow velocity of phantom using Doppler bandwidth broadening

JBO_23_1_016009_f002.png

Next, we estimated CMRO2 in vivo using Doppler ultrasound for flow estimation and multiwavelength photoacoustic methods for sO2 estimation. The animal was prepared as described in Sec. 2.2 and scanned across the neck at the speed of 5  mm/s under different physiological conditions. The global arterial sO2 was recorded by the pulse oximeter during the modulation of FiO2. Figure 3 shows representative in vivo coregistered photoacoustic (acquired at 580 nm), Doppler bandwidth broadening, and power Doppler images of the carotid artery and jugular vein when global arterial sO2 was measured at 85%. The artery was located >2  mm in depth, and the vein was located 1  mm under the skin. As in Fig. 3(b), the mean flow speed was quantified as 146  mm/s. For the task of cross-sectional area estimation, a circular region was selected so that all the color pixels in the power Doppler image [Fig. 3(c)] were fitted into the circle.

Fig. 3

Representative in vivo coregistered images of carotid artery (labeled as “A”) and jugular vein (labeled as “V”): (a) photoacoustic image acquired at 580 nm, (b) Doppler bandwidth broadening image, the color bar shows the absolute value of flow speed and direction, and (c) power Doppler image.

JBO_23_1_016009_f003.png

Figure 4 shows the measured venous mean flow velocities as a function of global arterial sO2 measured by the pulse oximeter. The appropriate ultrasound pulse-repetition-rate was chosen to ensure that the flow speeds were within the detectable range.35 The mean flow velocity of the internal jugular vein begins to drop above a global arterial sO2 of ˜85%.

Fig. 4

Venous mean flow velocities measured from hyperoxia (global arterial sO2 100%) to hypoxia (global arterial sO2 65%) conditions.

JBO_23_1_016009_f004.png

Figure 5 shows photoacoustic sO2 estimates in both arteries and veins as a function of measured global arterial sO2 from a pulse oximeter. As described in Sec. 2.4, we used two photoacoustic methods to cross validate our sO2 estimation. Method 1 is the traditional approach, which solves for unknown concentrations of oxy- and deoxy-hemoglobin using measured photoacoustic signals and known molar extinction coefficients, whereas method 2 uses multiwavelength photoacoustic measurements at different sO2 levels to solve a system of equations to estimate the different sO2 levels. Figure 5(a) shows the comparison of arterial sO2 estimations between method 1 and method 2. The estimates are consistent with p<0.05. It can also be observed that the sO2 estimates from arteries are close to the pulse oximeter values. The consistency ensures the validation of sO2 estimation in arteries and offers confidence in sO2 estimation in veins.

Fig. 5

Comparison of photoacoustic sO2 estimations between method 1 and method 2 in (a) carotid artery and (b) jugular vein. The arterial sO2 error bars are larger than venous sO2 error bars, likely due to flow pulsatility in arteries.

JBO_23_1_016009_f005.png

The representative CMRO2 was quantified according to Eq. (9) in Sec. 2.5. Here, we used a typical CHbT level of 0.147  g/ml at a hematocrit of 0.42.34 The results are shown in Fig. 6 as a function of global arterial sO2 as measured using pulse oximetry. As anticipated, the CMRO2 remains relatively constant during mild hypoxia and hyperoxia (global arterial sO2 of 65% to 100%), consistent with the hypothesis of brain metabolic autoregulation.

Fig. 6

The estimated representative CMRO2 (ml/100  g/min) as a function of varying global arterial sO2 measured using pulse oximetry during modulation of FiO2 under anesthesia. Error bars are standard deviations estimated according to the theory of propagation of errors.

JBO_23_1_016009_f006.png

4.

Discussion

Given the major significance of oxygen metabolism in health and disease, our work represents an approach to measure this metric using photoacoustic and ultrasound technologies. Current efforts use high-frequency ultrasound and visible light for animal models owing to limited penetration requirements. It should be noted, however, that depths investigated represent unreachable penetrations for previous oxygen metabolism progress using OR-PAM.

In this study, the total CBF is quantified by measuring outflow instead of inflow,36 as internal jugular veins provide the majority (80%) of the venous drainage from the brain. Since not all the blood vessels are counted for the measurements, the CMRO2 we measure is only a representative measure of oxygen consumption. Given the bifurcation of the common carotid artery into internal and external, our estimate of CMRO2 may be influenced by including facial oxygen flux. Encouragingly, the representative CMRO2 remains relatively constant during the modulation of the FiO2, which agrees with the hypothesis of cerebral metabolic autoregulation in the range of 23 to 100 mm Hg of PaO2. During mild hypoxia, the metabolic demand may be met by elevated CBF or higher extraction of oxygen from the blood.11 We have observed that the mean flow speed of internal jugular vein was dropping at global arterial sO2 of 85%. This may be the consequence of vasodilation, which helps to increase the cerebral blood volume under the hypoxia condition.2

Uncertainty due to measurement variability is quantified by measuring sO2 and flow rates at least three times and estimating standard deviations. Future work should aim to reduce uncertainties by optimization of light fluence, using higher-sensitivity detectors, and by a potential combination of motion compensation and judicious averaging. Array transducers may improve frame rates to real time if sufficient wavelength-switching capabilities are available, whereas current ultrasound B-scan speeds are limited to 15  fps while photoacoustic B-scans are limited by the 10-Hz repetition rate of the laser to acquire a single A-scan.

Future work should investigate conditions where brain autoregulation is impacted, such as stroke, trauma, acute sepsis, brain damage, and potentially neurodegenerative conditions. Preclinical investigation with animal models may offer invaluable insight prior to human translation. Human translation may be possible with infrared light and diagnostic-frequency transducers, especially in infants and children, considering the superficial depths of internal carotid and jugular vessels in these subjects.

5.

Conclusion

We have demonstrated in vivo estimation of CMRO2 metabolism in murine models using a custom combined photoacoustic and microultrasound system. Estimates of CMRO2 are based on arterial and venous oxygen saturation measurements using photoacoustic methods and flow measurements based on Doppler ultrasound. Photoacoustic estimates of oxygen saturation in arteries are validated with pulse oximetry while venous sO2 is cross validated using two independent photoacoustic methods. The estimated CMRO2 remains relatively constant from mild hypoxia to mild hyperoxia conditions, which agrees with the hypothesis of cerebral metabolic autoregulation. The demonstrated method for estimating oxygen consumption may have significant applicability in other tissues and has potential for clinical translation owing to its label free nature.

Disclosures

Roger Zemp is the cofounder and chief scientific officer of illumiSonics Inc., which, however, did not support this work. The authors have no other disclosures or conflicts of interest to report.

Acknowledgments

We gratefully acknowledge funding from the Canadian Institutes of Health Research (No.CPG 134739), the Natural Sciences and Engineering Research Council of Canada (Nos. NSERC 355544-2008, 375340-2009, STPGP 396444, EQPEQ 440290, EQPEQ 423474, EQPEQ 423197, RGPIN 355544, and STPGP 494293-16), the Canadian Cancer Society (Nos. CCS 2011-700718 and CCS 702032), and the Prostate Cancer Canada (No. PCC MVBRDG D2013-40).

References

1. 

R. J. Person and R. Thies, Physiology, 2nd ed.Springer-Verlag Inc., New York (1989). Google Scholar

2. 

R. C. Tasker, “Brain vascular and hydrodynamic physiology,” Semin. Pediatr. Surg., 23 (4), 168 –173 (2013). http://dx.doi.org/10.1053/j.sempedsurg.2013.10.003 SPSUEH 1055-8586 Google Scholar

3. 

K. Ishii et al., “Decreased medical temporal oxygen metabolism in Alzheimer’s disease shown by PET,” J. Nucl. Med., 37 (7), 1159 –1165 (1996). Google Scholar

4. 

R. F. Peppard et al., “Cerebral glucose metabolism in Parkinson’s disease with and without dementia,” Arch. Neurol., 49 (12), 1262 –1268 (1992). http://dx.doi.org/10.1001/archneur.1992.00530360060019 Google Scholar

5. 

P. Santens et al., “Cerebral oxygen metabolism in patients with progressive supranuclear palsy: a positron emission tomography study,” Eur. Neurol., 37 18 –22 (1997). http://dx.doi.org/10.1159/000117398 Google Scholar

6. 

V. Bonhomme et al., “Influence of anesthesia on cerebral blood flow, cerebral metabolic rate, and brain functional connectivity,” Curr. Opin. Anesthesiol., 24 (5), 474 –479 (2011). http://dx.doi.org/10.1097/ACO.0b013e32834a12a1 Google Scholar

7. 

A. A. Scand, “Regional cerebral blood flow and glucose metabolism during propofol anesthesia in healthy subjects studied with positron emission tomography,” Acta Anaesthesiol. Scand., 56 (2), 248 –255 (2012). http://dx.doi.org/10.1111/aas.2012.56.issue-2 Google Scholar

8. 

J. B. De Vis et al., “Calibrated MRI to evaluate cerebral hemodynamics in patients with an internal carotid artery occlusion,” J. Cereb. Blood Flow Metab., 35 (6), 1015 –1023 (2015). http://dx.doi.org/10.1038/jcbfm.2015.14 Google Scholar

9. 

M. N. Diringer et al., “Effect of hyperoxia on cerebral metabolic rate for oxygen measured using positron emission tomography in patients with acute severe head injury,” J. Neurosurg., 106 (4), 526 –529 (2007). http://dx.doi.org/10.3171/jns.2007.106.4.526 JONSAC 0022-3085 Google Scholar

10. 

S. S. Kety and C. F. Schmidt, “The effect of altered arterial tensions of carbon dioxide and oxygen on cerebral blood flow and cerebral oxygen consumption of normal young men,” J. Clin. Invest., 27 (4), 484 –492 (1948). http://dx.doi.org/10.1172/JCI101995 JCINAO 0021-9738 Google Scholar

11. 

S. P. Chong et al., “Cerebral metabolic rate of oxygen (CMRO2) assessed by combined Doppler and spectroscopic OCT,” Biomed. Opt. Express, 6 (10), 3941 –3951 (2015). http://dx.doi.org/10.1364/BOE.6.003941 BOEICL 2156-7085 Google Scholar

12. 

F. Xu et al., “Effect of hypoxia and hyperoxia on cerebral blood flow, blood oxygenation, and oxidative metabolism,” J. Cereb. Blood Flow Metab., 32 (10), 1909 –1918 (2012). http://dx.doi.org/10.1038/jcbfm.2012.93 Google Scholar

13. 

M. J. Cipolla, The Cerebral Circulation, Morgan and Claypool Life Sciences, San Rafael, California (2009). Google Scholar

14. 

T. Christen, D. S. Bolar and G. Zaharchuk, “Imaging brain oxygenation with MRI using oxygenation approaches: methods, validation, and clinical applications,” Am. J. Neuroradiol., 34 (6), 1113 –1123 (2013). http://dx.doi.org/10.3174/ajnr.A3070 Google Scholar

15. 

T. S. Mayberg and A. M. Lam, “Jugular bulb oximetry for the monitoring of cerebral blood flow and metabolism,” Neurosurg. Clin. N. Am., 7 (4), 755 –765 (1996). Google Scholar

16. 

Y. Yamamoto et al., “Noninvasive quantification of regional myocardial metabolic rate of oxygen by O152 inhalation and positron emission tomography,” Circulation, 94 808 –816 (1996). http://dx.doi.org/10.1161/01.CIR.94.4.808 CIRCAZ 0009-7322 Google Scholar

17. 

E. M. Haacke et al., “In vivo measurement of blood oxygen saturation using magnetic resonance imaging: a direct validation of the blood oxygen level-dependent concept in functional brain imaging,” Hum. Brain Mapp., 5 341 –346 (1997). http://dx.doi.org/10.1002/(SICI)1097-0193(1997)5:5<>1.0.CO;2-S Google Scholar

18. 

V. Jain, M. C. Langham and F. W. Wehrli, “MRI estimation of global brain oxygen consumption rate,” J. Cereb. Blood Flow Metab., 30 (9), 1598 –1607 (2010). http://dx.doi.org/10.1038/jcbfm.2010.49 Google Scholar

19. 

G. Yu et al., “Time-dependent blood flow and oxygenation in human skeletal muscles measured with noninvasive near-infrared diffuse optical spectroscopies,” J. Biomed. Opt., 10 (2), 024027 (2005). http://dx.doi.org/10.1117/1.1884603 JBOPFO 1083-3668 Google Scholar

20. 

L. A. Steiner et al., “Near-infrared spectroscopy can monitor dynamic cerebral autoregulation in adults,” Neurocrit. Care, 10 (1), 122 –128 (2009). http://dx.doi.org/10.1007/s12028-008-9140-5 Google Scholar

21. 

L. V. Wang, “Prospects of photoacoustic tomography,” Med. Phys., 35 (12), 5758 –5767 (2008). http://dx.doi.org/10.1118/1.3013698 MPHYA6 0094-2405 Google Scholar

22. 

J. Yao et al., “Label-free oxygen-metabolic photoacoustic microscopy in vivo,” J. Biomed. Opt., 16 (7), 076003 (2011). http://dx.doi.org/10.1117/1.3594786 JBOPFO 1083-3668 Google Scholar

23. 

J. Yao and L. V. Wang, “Transverse flow imaging based on photoacoustic Doppler bandwidth broadening,” J. Biomed. Opt., 15 (2), 021304 (2010). http://dx.doi.org/10.1117/1.3339953 JBOPFO 1083-3668 Google Scholar

24. 

J. Brunker and P. Beard, “Velocity measurements in whole blood using acoustic resolution photoacoustic Doppler,” Biomed. Opt. Express, 7 (7), 2789 –2806 (2016). http://dx.doi.org/10.1364/BOE.7.002789 BOEICL 2156-7085 Google Scholar

25. 

Y. Jiang et al., “Blood oxygen flux estimation with a combined photoacoustic and high-frequency ultrasound microscopy system: a phantom study,” J. Biomed. Opt., 17 (3), 036012 (2012). http://dx.doi.org/10.1117/1.JBO.17.3.036012 JBOPFO 1083-3668 Google Scholar

26. 

T. Harrison et al., “Combined photoacoustic and ultrasound biomicroscopy,” Opt. Express, 17 22041 –22046 (2009). http://dx.doi.org/10.1364/OE.17.022041 OPEXFF 1094-4087 Google Scholar

27. 

G. R. Kelman, “Digital computer subroutine of the conversion of oxygen tension into saturation,” J. Appl. Physiol., 21 (4), 1375 –1376 (1966). Google Scholar

28. 

C. Kasai et al., “Real-time two-dimensional blood flow imaging using an autocorrelation technique,” IEEE Trans. Sonics Ultrason., 32 458 –464 (1985). http://dx.doi.org/10.1109/T-SU.1985.31615 IESUAU 0018-9537 Google Scholar

29. 

B. Lee et al., “Doppler angle and flow velocity estimations using the classic and transverse Doppler effects,” IEEE Trans. Ultrason. Ferroelectr. Freq. Control, 46 (1), 252 –256 (1999). http://dx.doi.org/10.1109/58.741534 ITUCER 0885-3010 Google Scholar

30. 

J. M. Rubin et al., “Power Doppler US: a potentially useful alternative to mean frequency-based color Doppler US,” Radiology, 190 853 –856 (1994). http://dx.doi.org/10.1148/radiology.190.3.8115639 RADLAX 0033-8419 Google Scholar

31. 

Y. Jiang et al., “Photoacoustic and high-frequency power Doppler ultrasound biomicroscopy: a comparative study,” J. Biomed. Opt., 15 (5), 056008 (2010). http://dx.doi.org/10.1117/1.3491126 JBOPFO 1083-3668 Google Scholar

32. 

D. E. Kruse and K. W. Ferrara, “A new high resolution color flow system using an eigendecomposition-based adaptive filter for clutter rejection,” IEEE Trans. Ultrason. Ferroelectr. Freq. Control, 49 (12), 1739 –1754 (2002). http://dx.doi.org/10.1109/TUFFC.2002.1159852 ITUCER 0885-3010 Google Scholar

33. 

J. Xia et al., “Calibration-free quantification of absolute oxygen saturation based on the dynamics of photoacoustic signals,” Opt. Lett., 38 (15), 2800 –2803 (2013). http://dx.doi.org/10.1364/OL.38.002800 OPLEDP 0146-9592 Google Scholar

34. 

J. B. West, Pulmonary Physiology and Pathophysiology: An Integrated, Case-Based Approach, 2nd ed.Lippincott Williams & Wilkins, Philadelphia (2007). Google Scholar

35. 

R. S. C. Cobbold, Foundations of Biomedical Ultrasound, Oxford University Press, New York (2007). Google Scholar

36. 

H. Ohsumi et al., “Internal jugular bulb blood velocity as a continuous indicator of cerebral blood flow during open heart surgery,” Anesthesiology, 81 (2), 325 –332 (1994). http://dx.doi.org/10.1097/00000542-199408000-00009 ANESAV 0003-3022 Google Scholar

Biographies for the authors are not available.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 Unported License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Yan Jiang and Roger Zemp "Estimation of cerebral metabolic rate of oxygen consumption using combined multiwavelength photoacoustic microscopy and Doppler microultrasound," Journal of Biomedical Optics 23(1), 016009 (18 January 2018). https://doi.org/10.1117/1.JBO.23.1.016009
Received: 13 June 2017; Accepted: 16 November 2017; Published: 18 January 2018
Lens.org Logo
CITATIONS
Cited by 6 scholarly publications.
Advertisement
Advertisement
KEYWORDS
Oxygen

Doppler effect

Photoacoustic spectroscopy

Ultrasonography

Brain

Veins

Arteries

Back to Top