22 October 2019 Small form factor, flexible, dual-modality handheld probe for smartphone-based, point-of-care oral and oropharyngeal cancer screening
Author Affiliations +
Abstract

Oral cancer is a growing health issue in low- and middle-income countries due to betel quid, tobacco, and alcohol use and in younger populations of middle- and high-income communities due to the prevalence of human papillomavirus. The described point-of-care, smartphone-based intraoral probe enables autofluorescence imaging and polarized white light imaging in a compact geometry through the use of a USB-connected camera module. The small size and flexible imaging head improves on previous intraoral probe designs and allows imaging the cheek pockets, tonsils, and base of tongue, the areas of greatest risk for both causes of oral cancer. Cloud-based remote specialist and convolutional neural network clinical diagnosis allow for both remote community and home use. The device is characterized and preliminary field-testing data are shared.

1.

Introduction

Oral cancer incidence and death rates are an ongoing concern in low- and middle-income countries (LMICs). LMICs suffer the majority of worldwide new cancer cases and deaths along with <50% five-year survival rates in some locations.1,2 The major risk factor for oral squamous cell carcinoma and oral submucous fibrosis in many LMICs is betel quid chewing, with or without the inclusion of tobacco.3,4 Furthermore, oral and oropharyngeal cancers are increasing in younger populations (especially men) in middle- to high-income communities (MHICs) due to the spread of human papillomavirus (HPV), mostly affecting the tonsils and base of the tongue (BOT).57

Conventional oral examinations (COEs) provide a high specificity8 for visible lesions. However, lesions can be at an advanced stage before they are visible. Additionally, with higher incidence rates deep in the oral cavity and in oropharyngeal regions, accessibility can be an issue for a COE. Autofluorescence imaging (AFI) has the potential to be an important oral and oropharyngeal cancer adjunctive screening tool to aid in discrimination of oral potentially malignant lesions not visible during a COE.1 Previously, we described an AFI and white light imaging (WLI) smartphone-based oral cancer screening device with the ability to capture both whole cavity and intraoral images.9 Following is an intraoral probe with improved form factor to increase clinician ergonomics and patient comfort while improving access to the BOT and oropharynx, significant cancer sites.7,10

2.

Materials

We have developed a small form-factor, intraoral AFI, and polarized white light imaging (PWLI) handheld probe utilizing a commercially available camera module (OV5648, Omnivision, Santa Clara, California), connected by USB to an LG G4 smartphone (LG, Seoul, South Korea) for low-cost, point-of-care oral, and oropharyngeal cancer screening (Fig. 1).

Fig. 1

Intraoral probe system view showing the handheld probe: (a) overall system size; (b), (c) the flexible head; (d) the imaging and LED illumination apertures, along with the (e) installed smartphone running the custom Android application and (f) installed electronics.

JBO_24_10_106003_f001.png

The 3D-printed probe housing of VeroBlackPlus (Stratasys, Eden Prairie, Minnesota) provides mounting for the camera, camera printed circuit board (PCB), illumination LEDs, filters, polarizers, and wiring while also integrating a 25-mm-diameter handle and flexible section near the imaging head. The 135.5-mm usable probe length allows access to the BOT and oropharynx while the 14.4  mm×8.4  mm probe cross-section maintains patient comfort during imaging tasks. Significantly, the 24-mm flexible silicone (Mold Star 20T and Black Silc Pig, Smooth-On, Macungie, Pennsylvania) section provides up to 470% elongation before break, bending up to 45 deg in either direction (Fig. 1) to better access all areas of the oral cavity. A 250-μm thick, 11-mm wide piece of aluminum sheet metal embedded in the flexible section is a ductile core to the silicone elastomer, maintaining the probe head angle after bending to the desired position. The aluminum sheet keeps the two sections of the probe attached with notches to capture the screws joining the two halves of the probe head clamshell while also serving as a thermal sink for the illumination LEDs. To ensure the elastomer stretches with the probe as it is bent and does not pull away from the seams of the 3D-printed plastic/silicone interface (keeping the electronics from being exposed), the silicone overlaps the 3D-printed plastic a minimum of 5 mm and maximum of 11 mm on each end, and 750-μm vertical capture features in the plastic embed into the silicone. The interface is further reinforced with Dow Corning 700 (DowDuPont, Midland, Michigan) silicone adhesive. Creating the silicone section is the last assembly step; the 3D-printed mold is clamped around the two probe halves and the silicone injected through the vents. Opposing vents in the mold reduce air pockets in the final cured silicone shape. The various features and components of the aluminum stiffener, flexible section, and silicone mold are shown in Fig. 2.

Fig. 2

Cross-section images of the system mechanical design highlighting features related to the flexible silicone section. (a) A horizontal section view up to the aluminum stiffener/heatsink (rendered in teal for visibility) demonstrating its shape and position in the probe head. The stiffener is clamped between the two halves of the probe head and, along with the screw capture notches, keeps the two sections of the probe rigidly connected. (b) A vertical section view of the final assembly step, where the green-colored mold for the silicone is clamped around the probe head. The camera PCB and illumination LED PCB flex cables span the flexible section. The LED flex PCB surrounds the camera and is attached to the heatsink with electrically insulating epoxy. (c) A vertical section view of the device after the silicone has been injected into the mold and cured. Here, the silicone and plastic overlap region along with the capture features is visible. The capture features and overlap increase robustness in the joint. After curing, the mold is removed and the silicone/3D-printed plastic interface is reinforced with silicone adhesive.

JBO_24_10_106003_f002.png

Illumination for PWLI and AFI is accomplished with two white (4000 K) and four violet (400 nm) LEDs (Lumileds, Amsterdam, Netherlands), respectively. The LEDs are mounted on an annulus of flexible, polyimide PCB, surrounding the camera module in a plane-symmetric pattern to maximize illumination uniformity without the use of additional optical components. The PCB is filled with large copper planes to increase heat transfer while the backside copper is exposed and attached to the aluminum stiffener with an electrically insulating, thermally conductive epoxy (DP240, 3M, St. Paul, Minnesota). To enable AFI, a 425-nm shortpass excitation filter (Asahi Spectra, Tokyo, Japan) is installed in front of the violet LEDs and a complementary 470-nm longpass emission filter (Asahi Spectra) is installed in front of the camera module.

The fixed emission filter in the imaging channel distorts the color space during WLI and is corrected by applying the transformation matrix calculated using a standard 24-color color checker (X-Rite, Grand Rapids, Michigan).

Eq. (1)

A=|0.61980.05680.09570.08720.93910.10730.06920.36180.0699|.

The stray light aperture prevents unfiltered scatter from the edges of the excitation filter to reach the tissue plane. Orthogonal linear polarizers (Edmund Optics, Barrington, New Jersey) in front of the white LEDs and the camera module enable PWLI. The thermally sensitive film polarizer is positioned at a distance from the LED emission surface by a 1-mm-thick PMMA window.

The USB-connected camera module integrates an image sensor (5 MP, 3673.6  μm×2738.4  μm), 4.1  mm focal length imaging lens, and control circuitry on a rigid-flex PCB with the image sensor and lens mounted on the end of the 30-mm-long flexible cable. The long flex PCB cable is embedded in the silicone section of the probe, with the rigid portion of the PCB containing driving and readout circuitry in the nonmoving section. The vendor-supplied lens focal length is calculated from the maximum chief ray angle (29 deg) and the sensor diagonal (4.6 mm) provided in the sensor module datasheet. The lens is installed in a threaded barrel and the position adjusted to provide a working distance of 20 mm, providing a cropped field of view (FOV) of 15  mm×20  mm.

The probe is compatible with the previously reported 3D-printed structure for mounting of the smartphone, LED driving electronics, and whole-cavity imaging module.9 Additionally, the system maintains the custom Android application for image capture, processing algorithms, cloud server data upload, convolutional neural network (CNN) classification, and web browser remote diagnosis features of the previously reported system.9,11

During patient testing, the probe is covered with a hygienic sleeve (TIDI Products, Neenah, Wisconsin).

3.

Methods

To characterize the imaging system performance, the FOV and cutoff frequency were measured by imaging a 1951 USAF resolution test chart, and a slanted edge test12,13 was used to compare the actual on-axis modulation transfer function (MTF) to the theoretical calculated from the image sensor pixel pitch and lens magnification.

Uniformity and leakage were evaluated to benchmark the illumination system. Illumination uniformity was quantified using the coefficient of variation

Eq. (2)

uniformity=1cv=1σx¯,
where σ is the standard deviation and x¯ is the mean of the grayscale pixel values when imaging a 50% neutral value matte card (Munsell, Grand Rapids, Michigan). Illumination leakage was measured by imaging a mirror under violet illumination and measuring the returned signal on each color channel relative to the maximum possible signal for the channel

Eq. (3)

leakage=i=0N1xi255N,
where xi is the pixel value and N is the number of pixel in the channel.

The maximum permissible exposure (MPE) of the AFI and PWLI illumination was calculated according to the International Electrotechnical Commission 60825-1:2014.14 Since the AFI LED bandwidth has power in two wavelength regions (315 to 400 nm and 400 to 700 nm), the most stringent limit is applied. With an exposure time of 0.8 s, MPE is calculated as

Eq. (4)

MPE=C1  Jm2=5.6×103t0.25  Jm2=5296  Jm2.

With the following assumptions (a) LED power run at 90% of maximum 675 mW, (b) emission area of 1  mm2, (c) emission duration of 800 ms, (d) Lambertian emission, (e) the closest tissues could be to the emission plane is 1.5 mm, and (f) skin aperture diameter is 3.5 mm. The flux transfer from an LED to tissue given the correct configuration factor is15

Eq. (5)

Φtissue=ΦLEDF=ΦLED12[X(X24R22R12)12]=0.0065  W,
where R1=r1/d, the radius of the emitter divided by the distance between the two planes; R2=r2/d, the radius of the tissue plane divided by the distance between the two planes; and X=1+(1+R22)/R12, corresponding to an exposure of 541  Jm2, safely below the MPE. The white LEDs output is also below the MPE as they have a similar exposure to the violet LEDs but are run at 10% of maximum power and have a higher limit due to the wavelength band.

A pilot human subjects study is currently being performed to demonstrate the capability of the hardware to reach oral and oropharyngeal sites and the feasibility of using CNN classification algorithms. This study is institutional review board approved by Mazumdar Shaw Cancer Centre (NNH/MEC-CL-2016-394), University of California, Irvine (HS#2002-2805), and Rowsell Park Comprehensive Cancer Center (I 49117). For this study, all subjects provided informed written and oral consent before testing. The workflow for testing the handheld probe was reported previously.9 The patients first have a conventional oral exam, and then the smartphone-based exam is administered. WLI and PWLI are collected with the whole cavity module9 and with the handheld probe. With the probe, both the lesion area and the normal contralateral area are imaged to provide additional data for the CNN. After testing, the images are uploaded either to an internal electronic health record system or to the cloud for remote specialist review and classification by the CNN.

To provide further visual information to the diagnosing specialist, additional images are created from the AFI. Fluorescence loss can be visualized from a luminance map where the green channel mean is subtracted from the green channel

Eq. (6)

f=IGI¯G,
and a red/green ratio map16,17

Eq. (7)

f=IRIG
can signal additional areas of concern. The CNN uses both the relative green luminance information and the red/green ratio in its analysis.

4.

Results

Figure 3 shows the FOV of the system and a zoomed view to show group 4–3 of the resolution target is resolvable, a spatial frequency of 20.16  lp/mm. This result matches the calculated MTF from the slanted edge test in Fig. 3, where the estimated cutoff frequency of 25  lp/mm results in a calculated image space numerical aperture of NA=0.007. The resulting depth of field is18

Eq. (8)

LofDfDLoBLofDfD+LoB12.5  mm,
where Lo is the nominal object plane (20  mm), f is the focal length of the lens (4.1 mm), D is the entrance pupil diameter (0.073 mm), and B is the blur diameter chosen to be three pixels (4.2  μm).

Fig. 3

Image of a 1951 USAF resolution test chart showing (a) the 15  mm×20  mm FOV and (b) contrast limit. The zoomed contrast limit image (b) shows group 4–3 is resolvable, a cutoff frequency of 20.16  lp/mm. (c) A comparison of the theoretical diffraction limit and measured on-axis MTF performance.

JBO_24_10_106003_f003.png

Illumination uniformity was measured to be 83.80% for white light illumination and 89.84% for violet light illumination. Leakage was measured as 14.01%, 13.00%, and 28.97% for the red, green, and blue channels, respectively.

Oral and oropharyngeal images demonstrating the ability of the handheld probe to capture AFI and PWLI at the BOT and oropharynx with the maps created from Eqs. (6) and (7) are shown in Figs. 4 and 5. Figure 4 shows AFI and PWLI along with the maps created from Eqs. (6) and (7) for normal and OSCC clinical diagnoses. Figure 5 demonstrates the ability to capture AFI and PWLI at the BOT and oropharynx, areas of increased cancer risk from HPV. For these sets of images, the upper limit of the dynamic range of the fluorescence loss images of column c in Figs. 4 and 5 has been deliberately limited for better visualization of low fluorescence (low luminance) regions. Regions of high luminance in the red-to-green ratio images of column d indicate possible cancerous conditions. The pixelated regions in the corners of some of the red-to-green ratio images are artifacts of small number division.

Fig. 4

Sample (column a) white light, (column b) autofluorescence, (column c) green luminance subtraction, and (column d) red/green ratio handheld intraoral images with clinical diagnoses of (q) normal/variation, location: floor of the mouth; (r) normal/variation, location: tongue; and (s) OSCC, location: right buccal mucosa.

JBO_24_10_106003_f004.png

Fig. 5

Sample (column a) white light, (column b) autofluorescence, (column c) green luminance subtraction, and (column d) red/green ratio handheld intraoral images demonstrating the ability to image the BOT.

JBO_24_10_106003_f005.png

5.

Discussion

AFI with excitation at 405  nm19,20 along with PWLI is a promising modality for low-cost, low-complexity devices to increase detection rates and provide earlier detection of oral and oropharyngeal cancer.21 Here, decreased fluorescence signal at 500  nm indicates increasing dysplasia while increased signal at 635  nm can serve as an additional indicator.1 The necessary excitation wavelengths now have readily available LED sources at increasingly lower cost and the resulting autofluorescence emission spectrum peaks are easily detectable by a smartphone camera.

This new intraoral probe provides marked improvements in geometry to increase access to the significant oral and oropharyngeal cancer sites compared to other AFI devices. Table 1 highlights the differences in form-factor, features, and capabilities between this intraoral probe, our previous device,9 and two commercially available AFI devices, the VELscope™22 (Apteryx Imaging, Akron, Ohio) and the Identifi®23 (StarDental, Lancaster, Pennsylvania).

Table 1

Comparison of AFI device features and specifications.

ParameterThis workUthoff 20189VELscope™Identifi®
Usable intraoral length135.5 mm116.6 mmNot intraoralSimilar to this work
Minimum intraoral cross-sectional area113.1  mm2369.5  mm2Not intraoralSimilar to this work
Maximum intraoral cross-sectional area116.5  mm2594.0  mm2Not intraoralSimilar to this work
Working distance20 mm10 mm80 mmNo camera
Depth of field12.5 mm0.6 mm>20  mmNo camera
FOV area300  mm2247  mm21257  mm2No camera
Flexible jointYesNoNot intraoralMirror
Accesses BOTYesNoNoMaybe
Accesses tonsils and posterior pharynxYesNoYesMaybe
Native image capture, processing, transmissionYesYesW/peripheralNo
AFIYesYesYesYes
PWLIYesWLINoWLI
Green-amber imagingNoNoNoYes

The new probe has increased its usable intraoral length by 20  mm over our previous device while decreasing the maximum cross-sectional area by over 80%. Removing the requirement to use the smartphone camera for imaging enables a smaller and lighter system by eliminating the design constraints, bulk, and cost of a long-track-length, well-aligned optical system. These geometry differences are visualized in Fig. 6. The drastic reduction in cross-sectional area along with the flexible joint enables better accommodation of the oral cavity’s and oropharynx’s contours, leading to the consistent ability to image the BOT and tonsils. The previous device’s length and size were insufficient to reach the BOT, and its inflexible imaging head was unable to orient for imaging the tonsils. While the VELscope™ is not an intraoral device, its geometry and long working distance does allow for imaging some oropharyngeal sites (tonsils, posterior pharynx) but cannot access the BOT. From images, the size of the Identifi® is estimated to be similar to this work’s probe.

Fig. 6

Comparison of mechanical outlines for (a) the probe proposed in this paper and (b) the previously published probe9 demonstrating its greatly reduced intraoral dimensions.

JBO_24_10_106003_f006.png

The illumination uniformity is decreased compared to the previous probe due to fewer LED sources but still maintains a high uniformity approaching 90%. Leakage data were only available for the current device measuring under 15% for both the red and green channels, the channels of interest (the blue channel is discarded), an area for improvement.

This probe’s depth of field is increased by 22× compared to the previous probe while maintaining sufficient resolution to resolve vasculature as seen in Fig. 4. The VELscope™ similarly benefits from a long depth of field whereas the previous probe’s short depth of field led to >50% of the images with too much blur to analyze in our first pilot study.9 Importantly, the FOV is increased, increasing the ability of a remote clinician to orient themselves in the image, valuable feedback from testing the previous probe. The VELscope™ and Identifi® both offer a larger FOV (due to not being intraoral and due to not implementing a camera, respectively) than this probe but the downsides are the VELscope™ not being able to access the BOT and the Identifi® lacking native image capture outweigh this smaller FOV.

The new probe integrates a full feature set of AFI, PWLI (decreasing the area of images saturated from specular reflection), and native image capture, processing, and transmission, lacking only the green-amber illumination of the Identifi®, giving it a high size-to-feature ratio compared to the other devices. Additionally, the new form-factor is derived from commercially successful intraoral imaging devices, such as the CS1500 (Carestream Dental, Atlanta, Georgia) and Claris i5HD (SOTA Imaging, Orange, California), suggesting improved ergonomics for both the clinician and patient.

Future improvements should include decreasing leakage from the AFI illumination. Additional improvements could include a custom designed lens system to increase the FOV and optimization of the working distance based on doctor use and patient comfort feedback. A sample optical design meeting these criteria is shown in Fig. 7 with the optical prescription provided in Table 2. Here, the angular full FOV has been in increased to 120 deg along the diagonal, resulting in a spatial FOV of 45.3  mm×33.7  mm with increased optical performance as measured by MTF compared to the USB camera lens. The increased FOV would allow remote specialists to more easily orient themselves when analyzing images outside of the clinical setting. The design utilizes a single achromat with an object space NA=0.0035, image space NA=0.023, providing a 9.4-mm depth of field where the focal length f=1.875  mm and the entrance pupil diameter D=0.165mm. This large depth of field is critical for reducing the number of out-of-focus images from hand movement during image capture.

Fig. 7

Optical design for a 120-deg full FOV, image space NA=0.023 lens with increased MTF performance compared to Fig 3. (a) The layout of the optical design with dimensions; (b) provides the theoretical MTF performance.

JBO_24_10_106003_f007.png

Table 2

Prescription for imaging lens with increased FOV and resolution.

SurfaceMaterialRadiusThicknessConicα2
ObjectAir15.000
StopOKP4HT−2.5200.3−0.129
3E48R0.8330.6
4Air−0.7392.6840.4300.157
Image

The sag of the aspheric surfaces is defined by24

Eq. (9)

z=cr21+1(1+k)c2r2+α2r4,
where r is the radial distance from the optical axis, c is the surface curvature (1/R), k is the conic constant, and αn defines the polynomial coefficients. Specifying plastic lenses (OKP4HT, Osaka Gas Chemicals, Osaka, Japan; E48R, Zeon, Tokyo, Japan) allows for molding technologies to achieve low-cost parts even with aspheric surfaces.

With the small camera modules, there is opportunity to include two camera modules in the head, one camera with a wide FOV, low NA lens and the other with a high NA lens and narrow FOV, providing wide FOV imaging and high-resolution imaging at the same time. For high-resolution imaging, a camera module with autofocus capability would be preferred. Alternatively, the same lens could be used for both cameras to enable 3-D spectroscopic imaging.

Additionally, the ability to 3D-print the flexible area along with the plastic would greatly decrease assembly time. This was attempted with Stratasys Tango material along with VeroBlackPlus but the Tango’s elongation at break of only 200% was not sufficient as the interface tore after only a few bending cycles. New 3D-printing materials could enable this change in the future.

Lastly, with an exposure time of 250 ms, motion of the probe during image capture affects quality. Usability could be improved by the addition of a camera control button on the probe handle so the clinician could prompt image capture from the probe, as well as the smartphone. In addition, since the LEDs are run at 10% of the emission limit, an increase in LED power and a corresponding decrease in camera exposure time would further limit blur in the images.

6.

Conclusion

We have demonstrated a compact, dual-modality imaging device with a flexible head to enable good quality AFI and PWLI in oral (cheek pockets) and oropharyngeal (tonsils and BOT) areas with high cancer incidence rates from betel quid chewing and HPV but more difficult to access with conventional aided or unaided visual examinations, leading to earlier detection and diagnosis of precancerous or cancerous conditions. Advances in high-power, low-cost light sources, 3D-printing, and low-cost smartphones are instrumental to the creation of simple AFI and PWLI prototype devices for oral and oropharyngeal cancer screening in LMICs and MHICs. USB-connected cameras remove design constraints related to using the smartphone’s camera allowing better ergonomics for the clinician and increased comfort for the patient, a system better able to address the imaging need while keeping the desirable functions and portability of the smartphone.

7.

Supplementary Material

The design files for the LED driver have been released on GitHub under the GPL-3.0 license.25,26

Disclosures

The authors have declared that no competing interests exist.

Acknowledgments

We are grateful for our funding sources. This research was supported by the National Institute of Biomedical Imaging and Bioengineering of the U.S. National Institutes of Health (NIH) under Award No. UH2EB022623 and NIH Award No. 1S10OD018061-01A1. Uthoff received funding from the National Institutes of Health, National Institute of Biomedical Imaging and Bioengineering, Graduate Training in Biomedical Imaging and Spectroscopy Grant No. T32EB000809. We would like to thank Pier Morgan and the Center for Gamma-Ray Imaging (CGRI) for assistance with the rapid prototype printer, Ken Almonte for guidance on the LED driver design, and David Coombs for PCB assembly assistance.

References

1. 

X. Luo et al., “Accuracy of autofluorescence in diagnosing oral squamous cell carcinoma and oral potentially malignant disorders: a comparative study with aero-digestive lesions,” Sci. Rep., 6 29943 (2016). https://doi.org/10.1038/srep29943 SRCEC3 2045-2322 Google Scholar

2. 

J. Ferlay et al., “Cancer incidence and mortality worldwide: sources, methods and major patterns in GLOBOCAN 2012: Globocan 2012,” Int. J. Cancer, 136 (5), E359 –E386 (2015). https://doi.org/10.1002/ijc.29210 IJCNAW 1097-0215 Google Scholar

3. 

H. K. Amarasinghe et al., “Betel-quid chewing with or without tobacco is a major risk factor for oral potentially malignant disorders in Sri Lanka: a case-control study,” Oral Oncol, 46 (4), 297 –301 (2010). https://doi.org/10.1016/j.oraloncology.2010.01.017 EJCCER 1368-8375 Google Scholar

4. 

H. Song et al., “Betel quid chewing without tobacco: a meta-analysis of carcinogenic and precarcinogenic effects,” Asia Pac. J. Public Health, 27 (2), NP47 –NP57 (2015). https://doi.org/10.1177/1010539513486921 Google Scholar

5. 

D. Young et al., “Increase in head and neck cancer in younger patients due to human papillomavirus (HPV),” Oral Oncol., 51 (8), 727 –730 (2015). https://doi.org/10.1016/j.oraloncology.2015.03.015 EJCCER 1368-8375 Google Scholar

6. 

M. L. Gillison et al., “Epidemiology of human papillomavirus-positive head and neck squamous cell carcinoma,” J. Clin. Oncol., 33 (29), 3235 –3242 (2015). https://doi.org/10.1200/JCO.2015.61.6995 JCONDN 0732-183X Google Scholar

7. 

T. Ramqvist et al., “Human papillomavirus and tonsillar and base of tongue cancer,” Viruses, 7 (3), 1332 –1343 (2015). https://doi.org/10.3390/v7031332 Google Scholar

8. 

R. Sankaranarayanan et al., “Effect of screening on oral cancer mortality in Kerala, India: a cluster-randomised controlled trial,” Lancet, 365 (9475), 1927 –1933 (2005). https://doi.org/10.1016/S0140-6736(05)66658-5 LANCAO 0140-6736 Google Scholar

9. 

R. D. Uthoff et al., “Point-of-care, smartphone-based, dual-modality, dual-view, oral cancer screening device with neural network classification for low-resource communities,” PloS One, 13 (12), e0207493 (2018). https://doi.org/10.1371/journal.pone.0207493 POLNCL 1932-6203 Google Scholar

10. 

A. K. Agarwal et al., “Treatment delay in oral and oropharyngeal cancer in our population: the role of socio-economic factors and health-seeking behaviour,” Indian J. Otolaryngol. Head Neck Surg., 63 (2), 145 –150 (2011). https://doi.org/10.1007/s12070-011-0134-9 Google Scholar

11. 

B. Song et al., “Automatic classification of dual-modality, smartphone-based oral dysplasia and malignancy images using deep learning,” Biomed. Opt. Express, 9 (11), 5318 –5329 (2018). https://doi.org/10.1364/BOE.9.005318 BOEICL 2156-7085 Google Scholar

12. 

P. D. Burns, “Slanted-edge MTF for digital camera and scanner analysis,” in Proc. IS&T, 135 –138 (2000). Google Scholar

13. 

P. D. Burns, “Sfrmat3: SFR analysis for digital cameras and scanners,” LosBurns Imaging Software, (2009) http://www.losburns.com/imaging/software/SFRedge/index.htm Google Scholar

14. 

IEC 60825–1; International Standard: Safety of Laser Products—Part 1: Equipment Classification and Requirements, 3.0 ed.IEC, Geneva, Switzerland (2014). Google Scholar

15. 

A. Arecchi et al., Field Guide to Illumination, SPIE Press, Bellingham, Washington (2007). Google Scholar

16. 

R. J. Mallia et al., “Laser-induced autofluorescence spectral ratio reference standard for early discrimination of oral cancer,” Cancer, 112 (7), 1503 –1512 (2008). https://doi.org/10.1002/cncr.23324 CANCAR 0008-543X Google Scholar

17. 

A. Zuluaga et al., “Fluorescence excitation emission matrices of human tissue: a system for in vivo measurement and method of data analysis,” Appl. Spectrosc., 53 (3), 302 –311 (1999). https://doi.org/10.1366/0003702991946695 APSPA4 0003-7028 Google Scholar

18. 

J. E. Greivenkamp, Field Guide to Geometrical Optics, Number v. FG01 in SPIE Field Guides, SPIE Press, Bellingham, Washington (2004). Google Scholar

19. 

D. L. Heintzelman et al., “Optimal excitation wavelengths for in vivo detection of oral neoplasia using fluorescence spectroscopy,” J. Photochem. Photobiol., 72 (1), 103 –113 (2000). https://doi.org/10.1562/0031-8655(2000)072<0103:OEWFIV>2.0.CO;2 Google Scholar

20. 

D. Roblyer et al., “Objective detection and delineation of oral neoplasia using autofluorescence imaging,” Cancer Prev. Res., 2 (5), 423 –431 (2009). https://doi.org/10.1158/1940-6207.CAPR-08-0229 Google Scholar

21. 

A. Rashid et al., “The use of light-based (optical) detection systems as adjuncts in the detection of oral cancer and oral potentially malignant disorders: a systematic review,” J. Oral Pathol. Med., 44 (5), 307 –328 (2015). https://doi.org/10.1111/jop.2015.44.issue-5 JPMEEA 0904-2512 Google Scholar

22. 

Apteryx Imaging, “VELscope Vx system,” (2019) https://apteryx.com/product/velscope-vx-system/ Google Scholar

23. 

DentalEZ, “Identifi: oral mucosal visual enhancement system,” (2019) http://www.identafi.net/ Google Scholar

24. 

R. Kingslake et al., Lens Design Fundamentals, Academic Press, London (2009). Google Scholar

25. 

R. D. Uthoff, “Rossuthoff/multichannel_led_driver: KiCAD design files,” GitHub, (2018). https://doi.org/10.5281/zenodo.1239445 Google Scholar

26. 

R. D. Uthoff, “Rossuthoff/kicad_footprints: custom footprints for pcb designs,” GitHub, (2018). https://doi.org/10.5281/zenodo.1214890 Google Scholar

Biography

Ross D. Uthoff is a PhD candidate at the University of Arizona, James C. Wyant College of Optical Sciences. He received his BS degree in optical engineering from Rose-Hulman Institute of Technology. His research interest is optical system design, currently being applied to medical imaging using smartphones. He is a member of SPIE.

Biographies of the other authors are not available.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 Unported License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Ross D. Uthoff, Bofan Song, Sumsum Sunny, Sanjana Patrick, Amritha Suresh, Trupti Kolur, Keerthi Gurushanth, Kimberly Wooten, Vishal Gupta, Mary E. Platek, Anurag K. Singh, Petra Wilder-Smith, Moni Abraham Kuriakose, Praveen Birur, and Rongguang Liang "Small form factor, flexible, dual-modality handheld probe for smartphone-based, point-of-care oral and oropharyngeal cancer screening," Journal of Biomedical Optics 24(10), 106003 (22 October 2019). https://doi.org/10.1117/1.JBO.24.10.106003
Received: 17 July 2019; Accepted: 24 September 2019; Published: 22 October 2019