1 January 2002 Three-dimensional diffuse optical tomography of bones and joints
Yong Xu, Nicusor Iftimia, Huabei Jiang, L. Lyndon Key, Marcy B. Bolster
Author Affiliations +

1.

Introduction

Near-infrared (NIR) diffuse optical tomography (DOT) has found its applications in imaging thick tissues such as breast and brain.1 2 3 4 5 6 7 The idea of DOT is to extract the spatial maps of intrinsic tissue absorption and scattering coefficients from the measured optical data through model-based reconstruction methods. The spectra of these optical properties will then allow one to obtain tissue functional images such as hemoglobin, water content, and cellular structure. In DOT, an effective reconstruction algorithm is crucial. To date various linear and nonlinear reconstruction algorithms have been successfully tested and evaluated using considerable laboratory phantom measurements and initial clinical data from different hardware systems including continuous wave (cw), frequency, and time domain.8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27

DOT has been recently applied to image other tissues such as human joints and associated bones. This particular application of DOT was conceived theoretically by Klose et al.;28 and demonstrated experimentally by Xu et al.;29 very recently. Diseases related to bones and joints such as osteoporosis and arthritis are a major cause of morbidity in the population over 50, affecting more than 68 million Americans. Conventional imaging modalities including x-ray radiography, computed tomography, magnetic resonance imaging, and ultrasonography usually can provide only tissue structural information which limits their use especially for early detection of diseases.

In the work of Xu et al.;,29 we have shown that two-dimensional (2D) DOT images of in vitro and in vivo bones and joints can be quantitatively reconstructed using NIR tomographic measurements. Encouraged by this successful experience, in a separate study, we have applied DOT to image a patient with osteoarthritis (OA).30 The findings from this study suggest that the diseased synovial membrane/fluid does show significant different optical properties compared to the joints from two healthy volunteers. While we have evidenced the potential high sensitivity of DOT for detecting OA (based on the quantitative nature of optical images at the joints), we have also seen that some of the important small joint structures/compositions such as cartilage, blood vessels, and synovial membrane could not be clearly differentiated in these 2D images presented. In part we believe that it would be difficult for 2D imaging methods to capture these small components distributed between the two bones with a separation distance less than 1 mm. The strong three-dimensional (3D) scattering nature of the bones would affect the 2D joint imaging significantly. This necessitates our consideration of optical image reconstruction in a 3D framework.

While 3D optical image reconstructions have been primarily limited to simulations,31 32 33 Jiang et al.;34 and Arridge et al.;35 have recently demonstrated full 3D reconstructions using breast or brain phantom model from cw and time-domain data, respectively. Assisted with ultrasound localization, Holboke et al.;36 have attempted to obtain a 3D optical characterization of the breast in a human subject with frequency-domain measurements. Most recently, Jiang et al.;37 have presented a full 3D reconstruction of absorption images of the human breast from cw measurements. In this paper we demonstrate for the first time that 3D volumetric absorption images of in vitro and in vivo bones and joints can be reconstructed using our DOT approach. The tomographic data were obtained with our automated multichannel DOT system, and the 3D images were recovered with our finite-element based nonlinear algorithm. The 3D images obtained show details of the joints/bones that are not available from the 2D images we achieved previously.

2.

Materials and Methods

Our multichannel frequency-domain DOT imager has been described in detail.26 29 34 Briefly, light from a 785 nm 50 mW diode laser (Thorlabs) is sequentially sent to the sample by sixteen 3 mm fiber optic bundles. For each source position, the diffused light is received at 16 detector positions along the surface of the sample and sequentially delivered to a photomultiplier tube (R928, Hamamatsu). The multiplexing of the source/detector fibers is accomplished by two automatic moving stages (17NST101, Melles-Griot). Data acquisition is realized by Labview routines (National Instruments). Both source and detector fiber bundles are firmly held by a metal ring structure with a diameter of 5 cm. Sixteen detector fiber bundles are arranged equally spaced around the annulus with another 16 source fiber bundles interspersed between the detector bundles. Tomographic data collection at multiple measurement planes are needed for 3D reconstruction. In the current configuration data collection time is about 10 min per measurement plane.

For the in vitro experiment, the sample consisted of two chicken bones embedded in a cylindrical solid phantom with a diameter of 5 cm. The solid phantom has contained a scattering medium composed of Intralipid (a fat emulsion suspension) with India ink as an absorber, which has had μa=0.005 mm −1 (absorption coefficient) and μs =1.0 mm −1 (reduced scattering coefficient). Agar powders have been used to solidify the Intralipid/ink suspensions. A schematic of this phantom is given in Figure 1(a). The tomographic measurements have been conducted at three different planes. For the in vivo experiment, the sample was the index finger of a healthy volunteer inserted into the same cylindrical solid phantom (the protocol was approved by the institutional review board and written informed consent was obtained from the participant). Figure 2 displays a schematic of the finger measurement configuration. A translation stage coupled with a high precision digital scale was used to vertically position the ring holding the source and detector fibers. The in vivo tomographic measurements have been conducted also at three different planes.

Figure 1

(a) Schematic of the in vitro test configuration including the solid phantom, chicken bones, and measurement volume. (b), (c), and (d) Reconstructed absorption images at three cut-planes from phantom/bones measurements (the dark region corresponds to the bones; the gray background corresponds to the solid phantom).

009201j.1.jpg

Figure 2

Schematic of the finger measurement configuration.

009201j.2.jpg

The absorption image of the sample was recovered using our 3D nonlinear, finite element based reconstruction algorithm. This algorithm uses a regularized Newton’s method to update an initial optical property distribution iteratively in order to minimize an object function composed of a weighted sum of the squared difference between computed and measured optical data at the sample surface. The computed optical data (i.e., photon density) is obtained by solving the photon diffusion equation with the finite element method. The mathematical details of our 3D algorithm and its evaluation with phantom and in vivo breast tissues have been described elsewhere.34 37 Briefly, the core procedure in our reconstruction algorithm is to iteratively solve the regularized matrix equation

Eq. (1)

(T+λI)δχ=T[Φ(m)Φ(c)],
where ℜJ is the Jacobian matrix that should be formed by ∂ϕ/∂D and ∂ϕ/∂μa at the boundary measurement sites; ϕ is the photon density; μa is the absorption coefficient; D is the diffusion coefficient which can be written as D=1/3μs , where μs is the reduced scattering coefficient. I is the identity matrix and λ can be a scalar or a diagonal matrix. Δχ=(ΔD1,ΔD2, … ,ΔDN,Δμa,1,Δμn,2, … ,Δμa,N)T is the update vector for the optical property profiles, where N is the total number of nodes in the finite element mesh used; ϕ(m)=[ϕ1 (m)2 (m), … ,ϕM (m)] and ϕ(c)=[ϕ1 (c)2 (c), … ,ϕM (c)] where ϕi (m) and ϕi (c), respectively, are measured and calculated data for i=1,2,…,M boundary locations.

In DOT, the goal is to update the μa and D or μs distributions through the iterative solution of Eq. (1) so that a weighted sum of the squared difference between computed and measured data can be minimized. To generate the 3D images within 3.0 cm height from in vitro measurements and 2.0 cm height from in vivo measurements, we have used a cylindrical finite element mesh with 1267 nodes and 5328 tetrahedral elements for the 3D reconstruction. Uniform distributions of optical properties that were 30–50 off the exact values of the background phantom medium were used as the initial guesses for the iterative procedure. The images reported were the results of five iterations when the initial sum of squared errors between measured and computed boundary data reached the steady state with about 1.5 h per iteration in a 600 MHz Pentium III PC.

3.

Results and Discussion

Figures 1(b), 1(c), and 1(d) show the reconstructed 3D absorption images at three different cut-planes from the in vitro measurements [see Figure 1(a) for the reference coordinate system]. We note that the absorption images show a significant absorption increase in the bone regions relative to the surrounding phantom background. As can be seen, the bone shapes at the joint as well as small structures such as cartilage are imaged.

Figures 3 4 display the optical images at the joint and off-joint planes along X, Y, and Z directions from the in vivo measurements. Figure 3 shows the reconstructed absorption images as a series of transverse slices parallel to the XOY plane (see Figure 2 for the reference coordinate system). As can be seen, the bones and several important joint structures can be clearly delineated. The bones and bone/cartilage are imaged, shown in Figures 3(a), 3(b), 3(e), and 3(f). Small structures such as blood vessels and extensor tendons are differentiated in Figures 3(b) and 3(c). The synovium is also identified [Figure 3(d)]. To view the bones and joint at different directions, we have provided series of images in Figure 4, at cut-planes parallel to XOZ and YOZ, respectively. Due to the strong absorption of the bones some joint structures cannot be seen in these longitudinal images, but the appearances of the bones, cartilage, synovium, and blood vessel are evident.

Figure 3

Transverse slices of reconstructed absorption images from in vivo measurements obtained around the proximal interphalangeal (PIP) joint of the index finger (the dark region corresponds to the finger; the gray background corresponds to the solid phantom). The axes (left and bottom) illustrate the spatial scale, in millimeters, whereas the gray scale (right) records the absorption coefficient, in inverse millimeters. B: bone; B/C: bone/cartilage; BV: blood vessel; T: tendon; S: synovium.

009201j.3.jpg

Figure 4

Longitudinal slices of reconstructed absorption images from in vivo measurements obtained around the PIP joint of the index finger (the dark region corresponds to the finger; the gray background corresponds to the solid phantom). The axes (left and bottom) illustrate the spatial scale, in millimeters, whereas the gray scale (right) records the absorption coefficient, in inverse millimeters. B/C: bone/cartilage; S: synovium.

009201j.4.jpg

It is interesting yet important to examine these absorption images shown in Figures 3 4 in terms of the absorption coefficient of each structure reconstructed, because the clinical potential of DOT would mostly lie on the difference of the optical properties between diseased and healthy bone/joint tissues. For example, an osteoporosis patient may show changed optical properties in the bone, while a rheumatoid arthritis patient may have increased optical properties in the synovium.38 39 40 41 The average absorption coefficients of bone and cartilage from our images are calculated to be 0.18 and 0.04/mm, respectively, which are in excellent agreement with the reported values of 0.15–0.18/mm for human bone41 and 0.033/mm for rabbit cartilage.42 The average absorption coefficient of synovium is found to be 0.094/mm, which is also in excellent agreement with that reported in the literature.39 40 The average absorption coefficients of blood vessels and tendons are approximately 0.056 and 0.044/mm, respectively. We note that the recovered absorption coefficient of the surrounding phantom background is 0.0054/mm which is about 8 off its exact value.

It is also important to recognize some of the advantages that 3D DOT has over 2D DOT. In our previous 2D images,29 critical joint structures such as synovium, cartilage, blood vessel, and tendon cannot be delineated. In addition, the 2D images have significantly underestimated the absorption coefficient of bones. In terms of the imaging resolution, objects as small as 1–2 mm in diameter such as the blood vessels are clearly identified from the 3D images presented which is typical for imaging a 5 cm diameter phantom/finger sample using DOT. We believe that if smaller core diameter fibers were used so that the source/detector fiber probe could directly contact the finger joints without using a large volume of phantom materials, the imaging resolution would be improved. We plan to perform such studies in the near future. In Figure 4, we have also noticed some boundary effects at the top and bottom edges in the bone regions. This was caused most probably by the strong absorption and scattering of the bones within the limited volume measured. We can increase the measuring volume to minimize this boundary impact on the region of interest.

In summary, we have demonstrated in vivo that bone and joint tissues can be imaged by 3D DOT. To the best of our knowledge, this study presents the first report of quantitative absorption images of in vitro and in vivo bones/joints that were obtained by using three-dimensional diffuse optical tomography with model-based image reconstruction methods.

Acknowledgement

This research was supported in part by a grant from the National Institutes of Health (NIH) (AR 46427).

REFERENCES

1. 

A. Yodh and B. Chance , “Spectroscopy and imaging with diffusing light,” Phys. Today , 48 34 –40 (1995). Google Scholar

2. 

S. Colak , M. van der Mark , G. tHooft , J. Hoogenraad , E. van der Linden , and F. Kuijpers , “Clinical optical tomography and NIR spectroscopy for breast cancer detection,” IEEE J. Sel. Top. Quantum Electron. , 5 1143 –1158 (1999). Google Scholar

3. 

V. Ntziachristos , A. Yodh , M. Schnall , and B. Chance , “Concurrent MRI and diffuse optical tomography of breast after indocyanine green enhancement,” Proc. Natl. Acad. Sci. U.S.A. , 97 2767 –2772 (2000). Google Scholar

4. 

B. W. Pogue , S. P. Poplack , T. O. McBride , W. A. Wells , K. S. Osterman , U. L. Osterberg , and K. D. Paulsen , “Quantitative hemoglobin tomography with diffuse near-infrared spectroscopy: pilot results in the breast,” Radiology , 218 261 –266 (2001). Google Scholar

5. 

H. Jiang, N. Iftimia, J. Eggert, L. Baron, K. Klove, Y. Xu, Y. Yang, and T. Pope, “Near-infrared optical imaging of breast,” Acad. Radiol. (in press).

6. 

D. A. Benaron , P. R. Contag , and C. H. Contag , “Imaging brain structure and function, infection and gene expression in the body using light,” Philos. Trans. R. Soc. London, Ser. B , 352 755 –761 (1997). Google Scholar

7. 

B. Chance , E. Anday , S. Nioka , S. Zhou , L. Hong , K. Worden , C. Li , T. Murray , Y. Ovetsky , D. Pidikiti , and R. Thomas , “A novel method for fast imaging of brain function, non-invasively, with light,” Opt. Express , 2 411 –423 (1998). Google Scholar

8. 

J. R. Singer , F. A. Grunbaum , P. Kohn , and J. P. Zubelli , “Image reconstruction of the interior of bodies that diffuse radiation,” Science , 248 990 –993 (1990). Google Scholar

9. 

R. L. Barbour, H. Graber, R. Aronson, and J. Lubowsky, “Model for 3D optical imaging of tissue,” in Proc. 10th Annual Int. Geoscience Remote Sensing Symp., Vol. II, pp. 1395–1399 (1990).

10. 

S. R. Arridge , M. Schweiger , and D. T. Delpy , “Iterative reconstruction of near infra-red absorption images,” Proc. SPIE , 1767 372 –383 (1992). Google Scholar

11. 

M. A. O’Leary , D. A. Boas , B. Chance , and A. G. Yodh , “Experimental images of heterogeneous turbid media by frequency-domain diffusing-photon tomography,” Opt. Lett. , 20 426 –428 (1995). Google Scholar

12. 

H. Jiang , K. D. Paulsen , U. L. Osterberg , B. W. Pogue , and M. S. Patterson , “Simultaneous reconstruction of absorption and scattering maps in turbid media from near-infrared frequency-domain data,” Opt. Lett. , 20 2128 –2130 (1995). Google Scholar

13. 

S. B. Colak , D. Papaioannou , G. tHooft , M. van der Mark , H. Schomberg , J. Paasschens , J. Melissen , and N. van Asten , “Tomographic image reconstruction from optical projections in light diffusing media,” Appl. Opt. , 36 180 –213 (1997). Google Scholar

14. 

S. A. Walker , S. Fantini , and E. Gratton , “Image reconstruction by backprojection from frequency-domain optical measurements in highly scattering media,” Appl. Opt. , 36 170 –179 (1997). Google Scholar

15. 

R. L. Barbour, H. Graber, Y. Wang, J. Chang, and R. Aronson, “Perturbation approach for optical diffusion tomography using continuous-wave and time-resolved data,” Medical Optical Tomography: Functional Imaging and Monitoring, G. Muller, Ed., pp. 87–120, SPIE, Bellingham, Washington (1993).

16. 

J. Schotland , J. Haselgrove , and J. Leigh , “Photon hitting density,” Appl. Opt. , 32 448 –453 (1993). Google Scholar

17. 

X. Li , T. Durduran , A. Yodh , B. Chance , and D. Pattanyak , “Diffraction tomography for biochemical imaging with diffuse-photon density waves,” Opt. Lett. , 22 573 –575 (1997). Google Scholar

18. 

C. Matson and H. Liu , “Analysis of the forward problem with diffuse photon density waves in turbid media by use of a diffraction tomography model,” J. Opt. Soc. Am. A , 16 455 –466 (1999). Google Scholar

19. 

X. Chen and D. Boas , “Diffuse optical reflection tomography with continuous-wave illumination,” Opt. Express , 3 118 –123 (1998). Google Scholar

20. 

S. R. Arridge and M. Schweiger , “Image reconstruction in optical tomography,” Philos. Trans. R. Soc. London, Ser. B , 352 717 –728 (1997). Google Scholar

21. 

A. Hielscher , A. Klose , and K. Hanson , “Gradient-based iterative image reconstruction scheme for time-resolved optical tomography,” IEEE Trans. Med. Imaging , 18 262 –271 (1999). Google Scholar

22. 

R. Roy and E. Sevick-Muraca , “Truncated Newton’s optimization scheme for absorption and fluorescence optical tomography: Part I Theory and formulation,” Opt. Express , 4 353 –371 (1999). Google Scholar

23. 

H. Jiang , K. D. Paulsen , U. L. Osterberg , B. W. Pogue , and M. S. Patterson , “Optical image reconstruction using frequency-domain data: simulations and experiments,” J. Opt. Soc. Am. A , 13 253 –266 (1996). Google Scholar

24. 

H. Jiang , K. D. Paulsen , and U. L. Osterberg , “Optical image reconstruction using DC data: simulations and experiments,” Phys. Med. Biol. , 41 1483 –1498 (1996). Google Scholar

25. 

H. Jiang , K. D. Paulsen , U. L. Osterberg , and M. S. Patterson , “Frequency-domain optical image reconstruction in heterogeneous media: an experimental study of single-target detectability,” Appl. Opt. , 36 52 –63 (1997). Google Scholar

26. 

N. Iftimia and H. Jiang , “Quantitative optical image reconstruction of turbid media using dc measurements,” Appl. Opt. , 39 5256 –5261 (2000). Google Scholar

27. 

B. W. Pogue , T. McBride , J. Prewitt , U. Osterberg , and K. Paulsen , “Spatially variant regularization improves diffuse optical tomography,” Appl. Opt. , 38 2950 –2961 (1999). Google Scholar

28. 

A. Klose , A. Hielscher , K. Hanson , and J. Beuthan , “Two- and three-dimensional optical tomography of finger joints for diagnostics of rheumatoid arthritis,” Proc. SPIE , 3566 151 –159 (1998). Google Scholar

29. 

Y. Xu , N. Iftimia , H. Jiang , L. Key , and M. Bolster , “Imaging of in vitro and in vivo bones and joints with continuous-wave diffuse optical tomography,” Opt. Express , 8 447 –451 (2001). Google Scholar

30. 

H. Jiang, N. Iftimia, Y. Xu, L. Key, and M. Bolster, “In vivo detection of osteoarthritis with diffuse optical tomography,” IEEE J. Sel. Top. Quantum Electron. (submitted).

31. 

H. Jiang, “Three-dimensional optical image reconstruction: Finite element approach,” in Proc. Advances in Optical Imaging and Photon Migration, Optical Society of America, pp. 156–158 (1998).

32. 

M. Schweiger and S. Arridge , “Comparison of two- and three-dimensional reconstruction methods in optical tomography,” Appl. Opt. , 37 7419 –7428 (1998). Google Scholar

33. 

M. Eppstein , D. Dougherty , D. Hawrysz , and E. Sevick-Muraca , “Three-dimensional optical tomography,” Proc. SPIE , 3597 97 –105 (1999). Google Scholar

34. 

H. Jiang , Y. Xu , and N. Iftimia , “Experimental three-dimensional optical image reconstruction of heterogeneous turbid media,” Opt. Express , 7 204 –209 (2000). Google Scholar

35. 

S. Arridge , “A method for three-dimensional time-resolved optical tomography,” Int. J. Imaging Syst. Technol. , 11 2 –11 (2000). Google Scholar

36. 

M. J. Holboke , B. Tromberg , X. Li , N. Shah , J. Fishkin , D. Kidney , J. Butler , B. Chance , and A. Yodh , “Three-dimensional diffuse optical mammography with ultrasound localization in a human subject,” J. Biomed. Opt. , 5 237 –247 (2000). Google Scholar

37. 

H. Jiang, Y. Xu, N. Iftimia, L. Baron, and J. Eggert, “Three-dimensional optical tomographic imaging of breast in a human subject,” IEEE Trans. Med. Imaging (in press).

38. 

V. Prapavat , R. Schuetz , W. Runge , J. Beuthan , and G. Muller , “In vivo investigations on the detection of chronical polyarthritis using a cw transillumination method at interphalangeal joints,” Proc. SPIE , 2626 360 –366 (1995). Google Scholar

39. 

J. Beuthan , V. Prapavat , R. Naber , O. Minet , and G. Muller , “Diagnostic of inflammatory rheumatic diseases with photon density waves,” Proc. SPIE , 2676 43 –53 (1996). Google Scholar

40. 

V. Prapavat , W. Runge , J. Mans , A. Krause , J. Beuthan , and G. Muller , “The development of a finger joint phantom for the optical simulation of early inflammatory rheumatic changes,” Biomedizinische Technik , 42 319 –326 (1997). Google Scholar

41. 

J. Beuthan, O. Minet, G. Muller, and V. Prapavat, “IR-diaphanoscopy in medicine,” in Medical Optical Tomography: Functional Imaging and Monitoring, G. Muller, B. Chance, R. Alfano et al.;, Eds., pp. 263–282, SPIE Press, Bellingham, Washington (1993).

42. 

W. F. Cheong , S. Prahl , and A. Welch , “A review of the optical properties of biological tissues,” IEEE J. Quantum Electron. , 26 2166 –2185 (1990). Google Scholar
©(2002) Society of Photo-Optical Instrumentation Engineers (SPIE)
Yong Xu, Nicusor Iftimia, Huabei Jiang, L. Lyndon Key, and Marcy B. Bolster "Three-dimensional diffuse optical tomography of bones and joints," Journal of Biomedical Optics 7(1), (1 January 2002). https://doi.org/10.1117/1.1427336
Published: 1 January 2002
Lens.org Logo
CITATIONS
Cited by 69 scholarly publications and 10 patents.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Bone

Absorption

3D image processing

In vivo imaging

Diffuse optical tomography

In vitro testing

3D metrology

Back to Top