1 August 1973 Calibration of a Field Spectroradiometer
Ross W. Leamer, Anthony J. La Rocca
Author Affiliations +
Abstract
The processes used to calibrate a field spectroradiometer are described. The spectro-radiometer uses four CVF's and four detectors with two of each in two separate instruments. Wavelength calibration procedure was to illuminate the aperture of the spectroradiometer with energy containing spectral lines at known wavelengths, and to note the linearity of the wavelength signal over the range of the spectral lines. Spectral radiance calibration consisted of comparing analog signals from the detectors to calibrated external references under known conditions.
Ross W. Leamer and Anthony J. La Rocca "Calibration of a Field Spectroradiometer," Optical Engineering 12(4), 124124 (1 August 1973). https://doi.org/10.1117/12.7971644
Published: 1 August 1973
Lens.org Logo
CITATIONS
Cited by 13 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Calibration

Spectral calibration

Sensors

Signal detection

Analog electronics

Back to Top