1 August 1992 Liquid-nitrogen-cooled servo-stabilized Fabry-Pérot interferometer for the infrared
Newrick K. Reay, Krzysztof A.R.B. Pietraszewski
Author Affiliations +
Abstract
A cooled Fabry-Pérot etalon has been constructed for which the parallelism and mirror spacing can be servo-controlled and tuned using an existing commercially available etalon control system. Capacitance sensors monitor the mirror spacing and parallelism, and error signals produced as a consequence of changes in these values are used to stabilize the cavity with piezoelectric actuators. The etalon operates at liquid nitrogen temperature, has a free aperture of 50 mm and a nominal mirror spacing of 60 μm, and is coated for the 3- to 5-μm spectral region. Under servo-control at operating temperature, the response time of the etalon is 30 ms and the cavity tuning range is 6 μm, corresponding to ~ 3 orders of interference at the midrange wavelength of 4 μm.
Newrick K. Reay and Krzysztof A.R.B. Pietraszewski "Liquid-nitrogen-cooled servo-stabilized Fabry-Pérot interferometer for the infrared," Optical Engineering 31(8), (1 August 1992). https://doi.org/10.1117/12.58705
Published: 1 August 1992
Lens.org Logo
CITATIONS
Cited by 6 scholarly publications and 1 patent.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Fabry–Perot interferometers

Mirrors

Actuators

Cryogenics

Capacitance

Ferroelectric materials

Capacitors

Back to Top