11 April 2014 Simultaneously measuring the refractive index and thickness of an optical sample by using improved fiber-based optical coherence tomography
Chih-Ta Yen, Jen-Fa Huang, Ming-Jia Wu, Yu-Fan Lee, Chia-Tsai Huang, Shu-Fan Huang, Hsu-Chih Cheng
Author Affiliations +
Abstract
An improved structure for the optical coherence tomography (OCT) scheme based on a 4×4 fiber coupler for simultaneously measuring the refractive index and thickness of optical samples is presented. The proposed structure incorporates the optical path length difference of interference light and is used to calculate the refractive index and thickness of an unknown sample without any prior knowledge of the sample parameters. Two methods (time-domain and Fourier-domain OCT) of obtaining information about an unknown sample are proposed, and a high-speed high-resolution OCT system was developed.

1.

Introduction

Optical coherence tomography (OCT) is a novel, noninvasive high-resolution imaging technique that can be used to analyze the inner structure of a detected sample in detail. The researchers frequently use OCT to inspect elements in the biomedical and industrial fields. Generally, OCT can be divided into two categories: time-domain OCT (TD-OCT) and Fourier-domain OCT (FD-OCT). TD-OCT and FD-OCT can be used to scan complex samples. These noninvasive approaches have become vital techniques for medical diagnosis because they have the advantages of revealing the surface of an organism in high resolution. The OCT technique continues to develop with the expansion of the diagnosis region and improvements in high-quality analysis. The most common application of OCT is in ophthalmology1 for retinal imaging to detect quantitatively the thickness of nerve fibers and produce high-resolution retinal images. As the OCT technique has been developed, it has been used to measure the inner structures of biological samples to provide high sampling rates and sensitivity, and to facilitate three-dimensional imaging analysis.25

OCT is a new biomedical detection technique in which a Michelson interferometer, Mach-Zehnder interferometer, or a common-path interferometer is often used as the system structure. The earliest OCT application, proposed a few decades ago, involved the combined structure of a white-light interferometer and Michelson interferometer, which was used as a one-dimensional optical length detection technique.6 Recent studies have presented numerous designs and OCT methods for measuring the optical parameters of a substance. Several studies have adopted ellipsometry techniques to determine the refractive index and thickness of a subject by using the Fresnel equation. 7,8 Jin et al. proposed a method for measuring the geometrical thickness and the refractive index of a silicon wafer by using spectral domain analysis. The geometrical thickness was determined without the use of any moving parts.9 Swillo et al. described an optical fiber coherence tomography system characterized by a Fizeau-based configuration with doubled pulses. This method was used to calculate the transmission coefficient of a sample by comparing suitable amplitudes of harmonic components in the detected signals.10 Chen et al. presented a simple fiber-optic sensor based on the Fabry-Perot interference for measuring the optical glass refractive index.11 Verma et al. demonstrated the use of common-path spectral-domain OCT for measuring the refractive index of a biomimetic material and a single biological cell.12 Tsai et al. demonstrated a new method for evaluating the properties of indium tin oxide conduction glass and identifying its defects. This approach can be used to estimate several parameters simultaneously, including thickness, refractive index, reflection coefficient, and transmission coefficient.13 Several previous studies have incorporated a wavelength-tuning interferometer for measuring the absolute optical thickness of a mask glass14 or the thickness and refractive index of a transparent plate simultaneously.15,16

Based on the previous discussion of simultaneously measuring thickness and refractive index,17 and the subsequent improvement achieved using a fiber loop mirror to reduce the experimental time for convenience and speed,18 FD-OCT was used in this study to expedite the overall experimental measurements. Compared with previously used free-space OCT,17 this construction offers the advantages of compactness, portability, low cost, and low complexity, and reduces the alignment time. Moreover, the same optical design can be used to measure the refractive index and thickness of an optical sample by performing FD-OCT. FD-OCT can be used to acquire information on a sample immediately by analyzing the spectrum of the interference signal, such as the direct current (DC) term, autocorrelation term, and conjugate terms, and, therefore, requires less time for detection than TD-OCT does. In addition, the position of the reference mirror in the reference arm does not need to be changed when using the scanning device to acquire axial depth information on the sample.

2.

Theory and Analysis Methods

Figure 1 shows a sketch of the proposed structure. All optical components are connected by a 4×4 fiber coupler (25% splitting ratio). This structure consists of a low-coherence light source, an optical spectrum analyzer (OSA), a photodetector, a data acquisition (DAQ) card, a 4×4 fiber coupler, two collimators, two mirrors, the sample (S), and a fiber loop mirror. In addition, the mirror M1 is placed on a movable stage with a speed that is controlled by a computer. In this optical path design, the arm with mirror M1 is the reference arm, and the arm with M0 is the detection arm. The fiber loop mirror is the other reference arm in this design. The optical direction indicates that the light reflected from the reference and detection arms are coupled by the 4×4 fiber coupler, and the coupled light is transmitted to the photodetector containing the DAQ card and OSA for analysis. Finally, information on the unknown sample can be acquired by analyzing the interference signals of the reference and detection arms. By using the photodetector containing the DAQ card and OSA, conducting two methods of analysis, TD-OCT and FD-OCT, is possible. Both methods involve using the optical length difference corresponding to the interference between M1 and the fiber loop mirror, between M1 and the left side of the sample surface, between M1 and the right side of the sample surface, and between M1 and the mirror M0.

Fig. 1

Schematic drawing of the improved structure. This structure consists of a low-coherence light source, an optical spectrum analyzer (OSA), a photodetector, a data acquisition (DAQ) card, a 4×4 fiber coupler, two collimators, two mirrors (M0 and M1), the sample (S), and a fiber loop mirror.

OE_53_4_044108_f001.png

Figure 2 shows that P1, P2, P3, and P4 are the optical path lengths of reflected light from the two samples’ surfaces, M0, and the fiber loop mirror, respectively. Therefore, before measuring the unknown sample, adjusting the mirror M0 to obtain the same optical path length as that of the fiber loop mirror is necessary. At this step, the arm of mirror M1 is inserted in the shutter for preventing unnecessary interference. According to Fig. 2, the relationships among P1, P2, P3, and P4 can be described as follows:

Eq. (1)

d1=P2P1=ns×L,

Eq. (2)

d2=P3P4=(nsna)×L,
where d1 and d2 are the optical path length difference between P1 and P2, and P3 and P4, respectively; ns and na are the refractive indices of the unknown sample and air, respectively; and L is the thickness of the sample. The refractive index of air was na=1 in this experiment. However, this formula is based on a homogeneous medium. By solving Eqs. (1) and (2), deriving Eq. (3) for measuring the refractive index of an unknown sample, and substituting the refractive index ns back into Eq. (1), determination of thickness L is possible.

Eq. (3)

ns=nad1d2d1.

Fig. 2

Schematic diagram of optical path in detection arm.

OE_53_4_044108_f002.png

This analysis indicates that knowing any parameters of the unknown sample in advance is not required. It is only necessary to obtain the optical path length difference between each path to calculate the refractive index and thickness. Two methods can be used to obtain the optical path length difference: TD-OCT and FD-OCT analysis. These methods involve the use of only a few mathematical operations to obtain the optical path lengths P1, P2, P3, and P4.

To perform measurements by using TD-OCT, before the unknown sample is inserted, adjusting the optical path length between the arm with mirror M0 and the fiber loop mirror is necessary. This can be achieved by modifying the position of mirror M0. In this process, a light shutter in the reference arm M1 is used to prevent unnecessary interference signal reflections caused by mirror M1.When the optical path length of the arm with mirror M0 is the same as that of the fiber loop mirror, the unknown sample can be inserted into the position between the collimator and M0, and then the measuring process can begin. Thereafter, mirror M1 is placed in a suitable position for the scanning process. The unknown sample, mirror M0, and the fiber loop mirror remain stationary when the scanning operation is performed by using the translational stage to move the reference arm with M1 at a constant velocity. When the reference arm with M1 changes the optical path length, it generates interference signals. Finally, the optical path length can be acquired by multiplying the time intervals between the interference signals by the constant velocity of the translation stage. The refractive index and the thickness of the unknown sample can be obtained using Eqs. (1), (2), and (3).

For performing measurements by using FD-OCT, first, the optical path difference between the arm with mirror M0 and the fiber loop mirror is adjusted to zero before beginning the measurement process. The unknown sample is then inserted, and the reference arm with mirror M1 is adjusted to an appropriate position and fixed for the following measurement. When beams are backscattered and reflected from the reference arm and sample arm at the coupler, the interference signal is detected by the OSA. By using Fourier transform on the spectrum of the interference signal, calculating the interference in the z axis, which contains the information on the depth of the sample arm, is possible. Thus, the interference signals can be obtained, and the optical path difference between each signal can be calculated. Finally, the information on the unknown sample, including the refractive index and thickness, can be obtained using Eqs. (1), (2), and (3) during TD-OCT analysis. Figure 3 shows the flow chart for TD-OCT and FD-OCT.

Fig. 3

Flow chart of postprocessing steps of the measurement.

OE_53_4_044108_f003.png

3.

Experimental Results and Discussion

The experimental setup includes an ASLD155-200 light source with a power of 17.5 mW, a central wavelength of 1550 nm, and a spectral bandwidth of 61.92 nm. In the proposed OCT system, the axial resolution Δz is determined by the coherence length lc of the light source. The axial resolution can be calculated as follows if the spectrum of the light source has a Gaussian distribution:

Eq. (4)

Δz=lc2=2ln2πλ02Δλ0.44·λ02Δλ,
where λ0 is the central wavelength of the light source and Δλ is the bandwidth. Equation (4) shows that the resolution of the proposed system is 17.072μm. Therefore, under this restriction, the thickness of the tested sample cannot be smaller than the system resolution, or the proposed method cannot be performed successfully. The performance of the proposed system in measuring the thickness and the refractive index of the glass sample was evaluated using an optical sample with a thickness of 1.090 mm and refractive index of 1.516. The parameters of the optical sample were measured using TD-OCT and FD-OCT.

In TD-OCT analysis, the two-step calibration procedure described in Sec. 2 was performed prior to the measurement process. TD-OCT was conducted using a Labview software package, which interacted directly with the translation stage and the photodetector. The interference signal was received by the photodetector containing the DAQ card in the time domain from 0 to 10 s. The translation stage was acquired from Sigma Koki Co., Ltd. (Model SGSP 20-35, Japan). During the scanning experiments, the stage was moved with a constant velocity of 0.5mm/s while performing TD-OCT. The translation stage was controlled using a GSC-0 stage controller (Sigma Koki Co., Ltd.) through the RS-232C interface of a personal computer.

As Fig. 4 shows, the four interference signals (A, B, C, and D) between the reference arms and the detection arm can be identified after scanning. These four interference signals correspond to the left (A) and right (B) surfaces of the sample and the surfaces of mirror M0 (C) and the fiber loop mirror (D). The optical path distance shown in Fig. 4 can be obtained by multiplying the time intervals between the interference signals by the constant velocity of the translation stage in the reference arm. Therefore, Eqs. (1) and (2) were used to obtain the values of d1 and d2, respectively. From Eq. (3), the refractive index of sample ns was determined to be ns=1.544. After obtaining the value of the refractive index, ns was substituted back into Eq. (1) to determine the thickness of the optical sample, which was L=1.069mm.

Fig. 4

The interference signals are obtained by Labview in time-domain analysis.

OE_53_4_044108_f004.png

Table 1 shows five repeated measurement results obtained from the same glass sample (ns=1.516, L=1.09mm). The mean values of the refractive index and the thickness were calculated to be 1.542 and 1.072 mm, respectively. The standard deviations of the refractive index and the thickness obtained from the five repeated measurements were 0.0046 and 0.0047 mm, respectively. According to Table 1, the maximal error and average error of the refractive index between the measured results and the real parameters of the glass sample at 1550 nm were 2.11 and 1.74%, respectively. Furthermore, the maximal error and average error of the thickness between the measured results and the real parameters of the glass sample were 2.20 and 1.65%, respectively. Comparisons between the measured and real parameters of the glass sample indicated good agreement. Using the proposed TD-OCT scheme has the advantage of simultaneously measuring the refractive index and the thickness of the sample.

Table 1

Five repeated measurement results of a glass sample with time-domain optical coherence tomography.

A (s)B (s)C (s)D (s)d1=BA (s)d2=CD (s)ns=d1/(d1d2)L=(d1×0.5)/ns (mm)
12.9456.2478.9337.7703.3021.1631.5441.069
21.5724.8737.8696.7013.3011.1681.5481.066
32.4005.7117.9966.8403.3111.1561.5361.078
42.4145.7248.0066.8453.3101.1611.5401.075
54.1457.4559.0057.8393.3101.1661.5441.072
Mean1.5421.072
Standard deviation (σ)0.00460.0047
Mean±σ1.542±0.00461.072±0.0047
Real parameters of glass sample1.5161.090
Maximum error (%)2.112.20
Average error (%)1.741.65

In FD-OCT analysis, the two-step calibration procedure described in Sec. 2 was performed prior to the measurement process. The optical spectrum of the interferogram in FD-OCT was measured using an OSA (Anritsu MS9710C) with a resolution of 0.1 nm (from 1500 to 1600 nm, 1001 sampling points). The depth range of FD-OCT is determined by the maximal number of fringes within the spectral width recorded using the OSA. The maximal depth position zmax is limited by the OSA wavelength resolution δλ and is given by

Eq. (5)

zmax=λ24δλ.

Therefore, in this experimental setup, zmax was 6 mm. To increase the depth range, increasing the spectral resolution or the number of sampling points is necessary.

Figure 5 shows the signals produced using FD-OCT, and the inverse Fourier transform (IFT) of the interference spectrum detected by the OSA. Three signals were obtained using the IFT calculation: the DC term, autocorrelation term, and the interference signals produced by sample and mirror surfaces (A, B, C, and D) and its conjugate terms. In Fig. 5, the four interference signals correspond to the left (A) and right (B) surfaces of the sample and the surfaces of mirror M0 (C) and the fiber loop mirror (D). The signals produced by the surfaces include A, B, C, and D, which are 1.170, 2.820, 4.554, and 3.996 mm on the z axis as shown in Fig. 5, respectively. The autocorrelation term signal may affect the determination of the correct interference signals, but the sample signals and autocorrelation term signal can be distinguished. Lengthening the optical path length difference between reference arm M1 and the left surface of the sample in the sample arm to increase the distance between the sample signals and the autocorrelation term signal is possible. Therefore, d1 (i.e., d1=BA) and d2 (i.e., d2=CD) in Eqs. (1) and (2) can be obtained by calculating each interval in Fig. 5. Finally, the refractive index of the optical sample was ns=na×1.650/(1.650×0.558)=1.511. The thickness of the optical sample was acquired by substituting the value of ns back into Eq. (1) as follows: L=d1/ns=1.092mm.

Fig. 5

The demodulated signal of the interferogram along the z axis.

OE_53_4_044108_f005.png

The five repeated measurement results obtained using FD-OCT analyses are shown in Table 2. The mean value of refractive index and thickness were calculated to be 1.516 and 1.087 mm, respectively. The standard deviation of the refractive index and thickness derived from the five repeated measurements were 0.0104 and 0.0089 mm, respectively. According to Table 2, the maximal error and average error of the refractive index between the measured results and the real parameters of the glass sample at 1550 nm were 0.99 and 0.03%, respectively. Furthermore, the maximal error and average error of the thickness between the measured results and the real parameters of the glass sample were 1.47 and 0.26%, respectively. As shown in Tables 1 and 2, the measurement results were similar to the real parameters of the glass sample.

Table 2

Five repeated measurement results of a glass sample with Fourier-domain optical coherence tomography.

A (mm)B (mm)C (mm)D (mm)d1=BA (mm)d2=CD (mm)ns=d1/(d1d2)L=d1/ns (mm)
11.1702.8204.5543.9961.6500.5581.5111.092
22.9884.6385.4364.8841.6500.5521.5031.098
31.2242.8685.3044.7341.6440.5701.5311.074
41.0802.7304.6324.0681.6500.5641.5191.086
52.5204.1644.9684.4101.6440.5581.5141.086
Mean1.5161.087
Standard deviation (σ)0.01040.0089
Mean±σ1.516±0.01041.087±0.0089
Real parameters of glass sample1.5161.090
Maximum error (%)0.991.47
Average error (%)0.030.26

4.

Conclusion

In this study, a fiber-based OCT structure was developed for simultaneously measuring the refractive index and thickness of a sample without any previous information. This structure involves the use of two methods for measuring the sample: TD-OCT and FD-OCT. The experimental results indicate only a minimal difference between the experimental values obtained using these methods. The two methods have advantages and disadvantages. In general, TD-OCT requires an extensive period for performing the scan; consequently, the measurement speed is low. However, TD-OCT has several advantages: it is low cost because the structure comprises a photodetector containing a DAQ, and it does not require complex calculations. Conversely, FD-OCT can be used to acquire information on a sample immediately by analyzing the spectrum of the interference signal, such as the DC term, autocorrelation term, and conjugate terms and, therefore, requires less time for detection than the TD-OCT does. In addition, using the scanning device is not required to change the position of the reference mirror in the reference arm while acquiring the axial depth information of the sample. However, although FD-OCT involves the use of a low-cost spectrometer, which is constructed by combining a diffraction grating and a charge-coupled device detector, the price is higher than that of using TD-OCT, and the analysis and calculations are more complex. In the experiments conducted in this study, the dispersion effect of the sample was not considered, which may have affected the experimental results and the system resolution.

Acknowledgments

The authors gratefully acknowledge the financial support provided for this study by the National Science Council under Grant Nos. NSC 100-2221-E-150-010 and NSC 102-2221-E-150-002.

References

1. 

A. F. FercherK. MengedohtW. Werner, “Eye-length measurement by interferometry with partially coherent light,” Opt. Lett., 13 (3), 1867 –1869 (1988). http://dx.doi.org/10.1364/OL.13.000186 OPLEDP 0146-9592 Google Scholar

2. 

W. Clivazet al., “High-resolution reflectometry in biological tissue,” Opt. Lett., 17 (1), 4 –6 (1992). http://dx.doi.org/10.1364/OL.17.000004 OPLEDP 0146-9592 Google Scholar

3. 

D. Huanget al., “Optical coherence tomography,” Science, 254 (5035), 1178 –1181 (1991). http://dx.doi.org/10.1126/science.1957169 SCIEAS 0036-8075 Google Scholar

4. 

R. LeitgebC. K. HitzenbergerA. F. Fercher, “Performance of Fourier domain vs. time domain optical coherence tomography,” Opt. Express, 11 (8), 889 –894 (2003). http://dx.doi.org/10.1364/OE.11.000889 OPEXFF 1094-4087 Google Scholar

5. 

L. VabreA. DuboisA. C. Boccara, “Thermal-light full-field optical coherence tomography,” Opt. Lett., 27 (7), 530 –532 (2002). http://dx.doi.org/10.1364/OL.27.000530 OPLEDP 0146-9592 Google Scholar

6. 

R. C. YoungquistS. CarrD. E. N. Davies, “Optical coherence domain reflectometry: a new optical evaluation technique,” Opt. Lett., 12 (3), 158 –160 (1987). http://dx.doi.org/10.1364/OL.12.000158 OPLEDP 0146-9592 Google Scholar

7. 

M. H. ChiuJ. Y. LeeD. C. Su, “Complex refractive-index measurement based on Fresnel’s equations and the uses of heterodyne interferometry,” Appl. Opt., 38 (19), 4047 –4052 (1999). http://dx.doi.org/10.1364/AO.38.004047 APOPAI 0003-6935 Google Scholar

8. 

C. C. HsuJ. Y. LeeD. C. Su, “Thickness and optical constants measurement of thin film growth with circular heterodyne interferometry,” Thin Solid Films, 491 (1–2), 91 –95 (2005). http://dx.doi.org/10.1016/j.tsf.2005.05.037 THSFAP 0040-6090 Google Scholar

9. 

J. Jinet al., “Thickness and refractive index measurement of a silicon wafer based on an optical comb,” Opt. Express, 18 (17), 18339 –18346 (2010). http://dx.doi.org/10.1364/OE.18.018339 OPEXFF 1094-4087 Google Scholar

10. 

R. SwilloL. R. JaroszewiczB. Pura, “New optical fiber coherence tomography—compact and stabile system for medical application,” IEEE Sens. J., 8 (7), 1138 –1144 (2008). http://dx.doi.org/10.1109/JSEN.2008.926180 ISJEAZ 1530-437X Google Scholar

11. 

J. H. Chenet al., “Extrinsic fiber-optic Fabry–Perot interferometer sensor for refractive index measurement of optical glass,” Appl. Opt., 49 (29), 5592 –5596 (2010). http://dx.doi.org/10.1364/AO.49.005592 APOPAI 0003-6935 Google Scholar

12. 

Y. Vermaet al., “Use of common path phase sensitive spectral domain optical coherence tomography for refractive index measurements,” Appl. Opt., 50 (25), E7 –E12 (2011). http://dx.doi.org/10.1364/AO.50.0000E7 APOPAI 0003-6935 Google Scholar

13. 

M. T. Tsaiet al., “Defect detection and property evaluation of indium tin oxide conducting glass using optical coherence tomography,” Opt. Exp., 19 (8), 7556 –7566 (2011). http://dx.doi.org/10.1364/OE.19.007559 OPEXFF 1094-4087 Google Scholar

14. 

Y. Kimet al., “Optical thickness measurement of mask blank glass plate by the excess fraction method using a wavelength-tuning interferometer,” Opt. Laser Eng., 51 (10), 1173 –1178 (2013). http://dx.doi.org/10.1016/j.optlaseng.2013.04.003 OLENDN 0143-8166 Google Scholar

15. 

T. FukanoI. Yamaguchi, “Separation of measurement of the refractive index and the geometrical thickness by use of a wavelength-scanning interferometer with a confocal microscope,” Appl. Opt., 38 (19), 4065 –4073 (1999). http://dx.doi.org/10.1364/AO.38.004065 APOPAI 0003-6935 Google Scholar

16. 

H. J. Choiet al., “Measurement of refractive index and thickness of transparent plate by dual-wavelength interference,” Opt. Express, 18 (9), 9429 –9434 (2010). http://dx.doi.org/10.1364/OE.18.009429 OPEXFF 1094-4087 Google Scholar

17. 

H. C. ChengY. C. Liu, “Simultaneous measurement of group refractive index and thickness of optical samples using optical coherence tomography,” Appl. Opt., 49 (5), 790 –797 (2010). http://dx.doi.org/10.1364/AO.49.000790 APOPAI 0003-6935 Google Scholar

18. 

S. F. Huanget al., “Simultaneous measurement of group refractive index and sample thickness by optical coherence tomography,” in 16th Opto-Electronics and Communications Conference, 277 –278 (2011). Google Scholar

Biography

Chih-Ta Yen received his BS from the Department of Electrical Engineering at Tamkang University, Taiwan, in 1996, his MS from the Department of Electrical Engineering, National Taiwan Ocean University, Taiwan, in 2002, and his PhD from the Department of Electrical Engineering at National Cheng Kung University, Taiwan, in 2008. He is currently an associate professor in the area of communication technology at the Department of Electrical Engineering, National Formosa University, Yunlin, Taiwan.

Jen-Fa Huang received his MASc and PhD degrees from the Department of Electrical Engineering at the University of Ottawa, ON, Canada, in 1981 and 1985, respectively. Since 1991, he has been a member of the Department of Electrical Engineering, National Cheng Kung University, Taiwan, where he is currently an adjunct professor in the Institute of Computer and Communication Engineering and Institute of Electro-Optic Science and Engineering, Department of Photonics.

Ming-Jia Wu received the MS degree from the Department of Electro-Optical Engineering, National Formosa University, Taiwan, in 2013.

Yu-Fan Lee received the MS degree from the Department of Electro-Optical Engineering, National Formosa University, Taiwan, in 2012.

Chia-Tsai Huang is currently working toward the MS degree in the area of fiber-optic communications and sensor with the Institute of Electro-Optic Science and Engineering, Department of Photonics, National Cheng Kung University, Taiwan.

Shu-Fan Huang received MS degree from the Institute of Electro-Optic Science and Engineering, Department of Photonics, National Cheng Kung University, Taiwan, in 2012.

Hsu-Chih Cheng received his BS degree from the Electronics Department of National Taiwan University of Science and Technology in 2000. His MS and PhD degrees in electrical engineering were received from the National Cheng Kung University, Tainan, Taiwan, in 2002 and 2006, respectively. He is currently an associate professor at the Department of Electro-Optical Engineering, National Formosa University, Yunlin County, Taiwan. His major interests lie in optical communications, optical system design, and optics fiber sensor.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 Unported License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Chih-Ta Yen, Jen-Fa Huang, Ming-Jia Wu, Yu-Fan Lee, Chia-Tsai Huang, Shu-Fan Huang, and Hsu-Chih Cheng "Simultaneously measuring the refractive index and thickness of an optical sample by using improved fiber-based optical coherence tomography," Optical Engineering 53(4), 044108 (11 April 2014). https://doi.org/10.1117/1.OE.53.4.044108
Published: 11 April 2014