This paper describes the strategy and planned implementation of the ground calibration of the X-IFU microcalorimeter spectrometer on-board Athena, the X-ray observatory from ESA to be launched after the mid-2030s. X-IFU is the second generation of space microcalorimeter instrument, and its calibration strategy benefits from the experience acquired on the present Japanese-US built Resolve instrument on-board XRISM. This calibration plan takes into account the reformulation of the Athena mission and its instruments that took place in 2022-23. The X-IFU calibration strategy is presented along with the set of X-ray sources needed for the ground calibration and the the definition of the sequences for using them.
The Athena mission underwent a redefinition phase since July 2022, driven by the imperative to reduce the mission cost for the European Space Agency while maintaining its flagship character. This most notably lead to a modification of the X-ray Integral Field Unit (X-IFU) cryogenic architecture. Passive cooling via radiative panels at spacecraft level (V-grooves) now provides a 50K thermal environment to the X-IFU cryostat. 4K cooling is achieved through a remote active cooler, with a multi-stage Adiabatic Demagnetization Refrigerator delivering the final 50 mK temperature required by the detectors.
Amidst these changes, the core concept of the readout chain remains robust, utilizing Transition Edge Sensor microcalorimeters and Time-Division Multiplexing. Noteworthy is the introduction of a slower pixel, enabling an increase in the multiplexing factor (from 34 to 48) without compromising the energy resolution. This mitigates the reduction of the field of view (from 5’ to 4’) while reducing the number of channels and thus the resource demands on the system.
In this contribution, we will give an overview of this new instrument architecture, addressing its main functional chains.
The X-ray Integral Field Unit (X-IFU) instrument is the high-resolution X-ray spectrometer of the ESA Athena X-ray Observatory. X-IFU will deliver spectra from 0.2 to 12 keV with a spectral resolution requirement of 4 eV (3 eV design goal) up to 7 keV from 5" pixels, with a hexagonal field of view of 4' equivalent diameter. The main sensor array and its associated detection chain is one of the major functional chains of the X-IFU instrument, and is the main contributor to XIFU performance. CNES (Centre National d’Études Spatiales) is the prime contractor for the X-IFU and leads the project development and procurement aspects within the X-IFU Consortium; additional major partners of the main detection chain are NASA-GFSC, SRON, VTT, APC, NIST, IRAP, and IAP. The detection chain design for X-IFU has evolved in the past few years in order to secure the performances and development costs, in the frame of the New Athena mission. New TES pixels are implemented with slower time constant and a reduced sensitivity to magnetic field. The slower time constant directly allows an increase of the MUX factor and a reduction of the number of channels, together with the decrease of the number of proximity electronics boxes, or warm front end electronics (WFEE). The cryostat outer vessel temperature is now a 50 K thermal interface, cooled passively thanks to L-shaped thermal shield (L-grooves). This has a direct impact of the cryo-harness between the 4 K core interface and the WFEE interface. In the past years, we have performed early demonstration on the critical components in order to secure the detection chain design and performances. This paper presents the progress done on early demonstrations (warm electronics, cryo-harness breadboarding,...), while providing an update to the detection-chain design description.
The x-ray integral field unit (X-IFU) instrument is the high-resolution x-ray spectrometer of the ESA Athena x-ray observatory. X-IFU will deliver spectra from 0.2 to 12 keV with a spectral resolution of 2.5 eV up to 7 keV from 5" pixels, with a hexagonal field of view of 5' equivalent diameter. The main sensor array and its associated detection chain is one of the major sub-systems of the X-IFU instrument, and is the main contributor to X-IFU’s performance. CNES (the French Space Agency) is leading the development of X-IFU; additional major partners are NASA-GFSC, SRON, VTT, APC, NIST, and IRAP. This paper updates the B-phase definition of the X-IFU detection chain. The readout is based on time-division multiplexing (TDM). The different sub-components of the detection chain (the main sensor array, the cold electronics stages, and the warm electronics) require global design optimization in order to achieve the best performance. The detection chain’s sensitivity to the EMI/EMC environment requires detailed analysis and implementation of dedicated design solutions. This paper focuses on these aspects while providing an update to the detection-chain design description.
The X-IFU (x-ray integral field unit) onboard the large ESA mission Athena (advanced telescope for high energy astrophysics), planned to be launched in the mid 2030s, will be a cryogenic x-ray imaging spectrometer operating at 55 mK. It will provide unprecedented spatially resolved high-resolution spectroscopy (2.5 eV FWHM up to 7 keV) in the 0.2-12 keV energy range thanks to its array of TES (transition edge sensors) microcalorimeters of more than 2k pixel. The detection chain of the instrument is developed by an international collaboration: the detector array by NASA/GSFC, the cold electronics by NIST, the cold amplifier by VTT, the WFEE (warm front-end electronics) by APC, the DRE (digital readout electronics) by IRAP and a focal plane assembly by SRON. To assess the operation of the complete readout chain of the X-IFU, a 50 mK test bench based on a kilo-pixel array of microcalorimeters from NASA/GSFC has been developed at IRAP in collaboration with CNES. Validation of the test bench has been performed with an intermediate detection chain entirely from NIST and Goddard. Next planned activities include the integration of DRE and WFEE prototypes in order to perform an end-to-end demonstration of a complete X-IFU detection chain.
The X-ray Integral Field Unit (X-IFU) instrument to be on board the future ESA mission Athena X-ray Observatory is a cryogenic micro-calorimeter array of Transition Edge Sensor (TES) detectors aimed at providing spatially resolved high-resolution spectroscopy. As a part of the on-board Event Processor (EP), the reconstruction software will provide the energy, spatial location and arrival time of the incoming X-ray photons hitting the detector and inducing current pulses on it. Being the standard optimal filtering technique the chosen baseline reconstruction algorithm, a particular modification of this technique based on a truncation of the filter in the Time Domain (equivalent to 0-padding the pulse signal) was previously studied on simulated data, proving a better energy resolution results at a lower computational cost. However, the 0-padding technique also showed a larger sensitivity to instrumental conditions, making essential the analysis of its behaviour over real laboratory data. A comparative analysis of X-IFU-like TES laboratory data from NASA and NIST (at different instrumental conditions) with both the 0-padding filter and the standard optimal filters reconstruction shows that, once the corrections for the baseline drift and the jitter (phase introduced by the sampling rate) have been performed, the resolution values obtained using 0-padding are systematically lower than those of the standard filter of the same length and comparable or even lower to the values provided by the full-length filter. The shorter length of the 0-padding filter would be an additional benefit, reducing the computational cost of the reconstruction process.
We present a test platform for the Athena X-IFU detection chain, shared between IRAP and CNES. This test bench, housed in a commercial two-stage ADR cryostat provided by Entropy GmbH, will serve as the first demonstration of the representative end-to-end readout chain for the X-IFU, using prototypes of the future flight electronics and currently available subsystems. The focal plane array (FPA), placed at the 50 mK cold stage of the ADR, includes a 1024-pixel array of transition-edge sensor (TES) microcalorimeter spectrometers provided by NASA/GSFC, superconducting amplifiers (SQUIDs) from VTT, as well as superconducting readout electronics for frequency domain multiplexing (FDM), provided by SRON. The detection chain then continues with the prototype room temperature electronics for the X-IFU: the Warm Front-End Electronics (WFEE, provided by APC) and the Digital Readout Electronics (DRE, provided by IRAP). The test bench yields critical feedback on current subsystem designs and electronic interfaces, and in the future will also be used for refining the X-IFU calibration plan as well as laboratory astrophysics experiments relevant to future X-IFU science. In this presentation, we describe the characterization of the cryostat, various design trades for the FPA and readout chain, and recent results from our current setup.
The X-IFU instrument to be on board the X-ray Athena Observatory is a cryogenic microcalorimeter array of TES detectors aimed at providing spatially resolved high-resolution spectroscopy. The reconstruction software will provide energy, position and arrival time of the incoming X-ray photons that produce current pulses in the detector. Different modifications of the standard optimal filtering algorithm have been compared to process pulses shorter than those considered of high resolution (those where the full length is not available due to a close pulse after them) to select the best algorithm based on energy resolution and computing performance results. We can conclude that the best approach is that of the 0-padding filtering. However further checks about its sensibility to instrumental changes (baseline, bias voltage, etc.) are required.
The X-ray Integral Field Unit (X-IFU) will be the cryogenic X-ray spectro-imager of the ESA Athena space observatory. It will implement a matrix of 3168 superconducting Transition Edge Sensors at 90 mK to provide a very high spectral resolution of 2.5 eV up to the energy of 7 keV, across a 5’ large field of view. The X-IFU successfully passed its Preliminary Requirements Review in mid-2019. After providing an overview of the driving science cases, the key requirements, I will present a status on the instrument development.
The X-ray Integral Field Unit (X-IFU) of the ATHENA space observatory is based on a 3 kilo-pixels array of superconducting micro-calorimeters, that is read out with the Warm Front-End Electronics (WFEE) and then followed by the Digital Readout Electronics (DRE) outside the cryostat. We have developed low-noise/lowdrift amplifiers for this readout as well as performed simulations and measured the noise, bandwidth, linearity, harmonic distortion and thermal drift. Here we are presenting a realistic model of these parameters. These characterizations and models of the WFEE are necessary in order to get a full ATHENA X-IFU readout chain simulator. Impact of the WFEE low-frequency noise and its limited bandwidth on the entire X-IFU detection chain in Time Domain Multiplexing will be introduced.
The X-ray Integral Field Unit (X-IFU) is the high resolution X-ray spectrometer of the ESA Athena X-ray observatory. Over a field of view of 5’ equivalent diameter, it will deliver X-ray spectra from 0.2 to 12 keV with a spectral resolution of 2.5 eV up to 7 keV on ∼ 5” pixels. The X-IFU is based on a large format array of super-conducting molybdenum-gold Transition Edge Sensors cooled at ∼ 90 mK, each coupled with an absorber made of gold and bismuth with a pitch of 249 μm. A cryogenic anti-coincidence detector located underneath the prime TES array enables the non X-ray background to be reduced. A bath temperature of ∼ 50 mK is obtained by a series of mechanical coolers combining 15K Pulse Tubes, 4K and 2K Joule-Thomson coolers which pre-cool a sub Kelvin cooler made of a 3He sorption cooler coupled with an Adiabatic Demagnetization Refrigerator. Frequency domain multiplexing enables to read out 40 pixels in one single channel. A photon interacting with an absorber leads to a current pulse, amplified by the readout electronics and whose shape is reconstructed on board to recover its energy with high accuracy. The defocusing capability offered by the Athena movable mirror assembly enables the X-IFU to observe the brightest X-ray sources of the sky (up to Crab-like intensities) by spreading the telescope point spread function over hundreds of pixels. Thus the X-IFU delivers low pile-up, high throughput (< 50%), and typically 10 eV spectral resolution at 1 Crab intensities, i.e. a factor of 10 or more better than Silicon based X-ray detectors. In this paper, the current X-IFU baseline is presented, together with an assessment of its anticipated performance in terms of spectral resolution, background, and count rate capability. The X-IFU baseline configuration will be subject to a preliminary requirement review that is scheduled at the end of 2018.
The X-ray Integral Field Unit (X-IFU) is an imaging microcalorimeter being developed for ESA's Athena X-ray observatory to providing high spectral resolution imaging between 0.2-12 keV, with moderate count-rate capability and a large field-of-view. The X-IFU focal plane assembly (FPA) will contain the instrument's large-format transition edge sensor (TES) microcalorimeter array and its superconducting readout electronics, plus a second TES detector, located behind the main sensor array, is used to detect high-energy cosmic rays and secondary particles passing through the sensor array and enable the rejection of false events that they generate in the sensor array's event list. A Kevlar thermal suspension is used to isolate the detectors at 55 mK from the 2 K environment of the X-IFU instrument cryostat's cold stage, while three layers of shielding are used to allow the detector's to achieve their low-noise performance in the expected on-ground and in-flight electromagnetic and microvibration environment. This paper will describe the preliminary design concept of the X-IFU focal plane assembly and its critical technology building blocks.
The X-ray Integral Field Unit (X-IFU) is the cryogenic imaging spectrometer onboard the ESA L2 mission Athena. With its array of almost 3840 superconducting Transition Edge Sensors micro-calorimeters, the X-IFU will provide spatially resolved (5" over the field of view) high-resolution spectroscopy (2.5 eV FWHM up to 7 keV) in the 0.2-12 keV energy band. These transformational capabilities will allow the X-IFU to probe the Hot and Energetic Universe, and notably measure the physical properties of large-scale structures with unprecedented accuracy. Starting from numerically-simulated massive (1014M) galaxy clusters at different steps of their evolution, we investigate the capabilities of the X-IFU in recovering chemical abundances, redshift and gas temperature spatial distributions across time, making use of full field-of-view End-To-End simulations of X-IFU observations. This work serve as feasibility study for the Chemical Enrichment of the Universe science objective. We show that using 100 ks observations, the X-IFU will provide an unprecedented spatially-accurate knowledge of the physics of the ICM (abundances, temperature, bulk-motion). We also demonstrate that challenges related to the data analysis of extended sources with very high-resolution spectrometers (e.g. binning, line of sight mixing, particle background) need to be thoroughly addressed to maximise the science of the instrument.
The X-ray Integral Field Unit (X-IFU) is the cryogenic imaging spectrometer on board the future X-ray observatory Athena. With a hexagonal array of 3840 AC-biased Transition Edge Sensors (TES), it will provide narrow-field observations (5’ equivalent diameter) with unprecedented high spectral resolution (2.5 eV up to 7 keV) over the 0.2 – 12 keV bandpass. Throughout its observations, the X-IFU will face various sources of X-ray background. Specifically, the so-called Non-X-ray Background (NXB) caused by the interaction of high-energy cosmic rays with the instrument, may lead to a degradation of its sensitivity in the observation of faint extended sources (e.g. galaxy clusters outskirts). To limit this effect, a cryogenic anti-coincidence detector (CryoAC) will be placed below the detector plane to lower the NXB level down to the required level of 5⊗10−3 cts/s/cm2/keV over 2 - 10 keV. In this contribution, we investigate ways to accurately monitor the NXB and ensure the highest reproducibility in-flight. Using the limiting science case of the background-dominated observation of galaxy clusters outskirts, we demonstrate that a reproducibility of 2% on the absolute knowledge of the background is required to perform driving science objectives, such as measuring abundances and turbulence in the outskirts. Monitoring of the NXB in-flight through closed observations, the detector’s CryoAC or the companion instrument (Wide Field Imager) will be used to meet this requirement.
The Athena X-Ray Integral Field Unit (X-IFU) will provide spatially resolved high-resolution spectroscopy (2.5 eV FWHM up to 7 keV) over the 0.2 to 12 keV energy band. It will comprise an array of 3840 superconducting Transition Edge Sensors (TESs) operated at 90 mK, with an absolute energy scale accuracy of 0.4 eV. Slight changes in the TES operating environment can cause significant variations in its energy response function, which may result in degradation of the detector’s energy resolution, and eventually in systematic errors in the absolute energy scale if not properly corrected. These changes will be monitored via an onboard Modulated X-ray Source (MXS) and the energy scale will be corrected accordingly using a multi-parameter interpolation of gain curves obtained during ground calibration. Assuming realistic MXS configurations and using the instrument End-To-End simulator SIXTE, we investigate here both statistical and systematic effects on the X-IFU energy scale, occurring either during ground measurements or in-flight. The corresponding impacts on the energy resolution and means of accounting for these errors are also addressed. We notably demonstrate that a multiparameter gain correction, using both the pulse-height estimate and the baseline of a pulse, can accurately recover systematic effects on the gain due to realistic changes in TES operating conditions within 0.4 eV. Optimisations of this technique with respect to the MXS line configuration and correction time, as well as to the energy scale parametrization are also show promising results to improve the accuracy of the correction.
With its array of 3840 Transition Edge Sensors (TESs) operated at 90 mK, the X-Ray Integral Field Unit (XIFU) on board the ESA L2 mission Athena will provide spatially resolved high-resolution spectroscopy (2.5 eV FWHM up to 7 keV) over the 0.2 to 12 keV bandpass. The in-flight performance of the X-IFU will be strongly affected by the calibration of the instrument. Uncertainties in the knowledge of the overall system, from the filter transmission to the energy scale, may introduce systematic errors in the data, which could potentially compromise science objectives – notably those involving line characterisation e.g. turbulence velocity measurements – if not properly accounted for. Defining and validating calibration requirements is therefore of paramount importance. In this paper, we put forward a simulation tool based on the most up-to-date configurations of the various subsystems (e.g. filters, detector absorbers) which allows us to estimate systematic errors related to uncertainties in the instrumental response. Notably, the effect of uncertainties in the energy resolution and of the instrumental quantum efficiency on X-IFU observations is assessed, by taking as a test case the measurements of the iron K complex in the hot gas surrounding clusters of galaxies. In-flight and ground calibration of the energy resolution and the quantum efficiency is also addressed. We demonstrate that provided an accurate calibration of the instrument, such effects should be low in both cases with respect to statistics during observations.
The 96 read-out chains which are foreseen in the X-ray Integral Field Unit (XIFU) on ESA's L2 mission Athena to receive the signals from the 3840 X-ray microcalorimeter transition-edge sensors (TES), are based on the principle of Frequency Domain Multiplexing (FDM) with closed-loop baseband feedback (BBFB) to match the dynamic range of the read-out to that of the detectors. The XIFU instrument concept currently undergoes a Phase-A assessment. The Digital-to-Analogue Converters (DACs) which generate the carrier signals of the FDM and the signals of the BBFB loops were identified as critical elements. In this presentation we formulate the dynamic range requirements for the DACs and assess to what extend a current state-of-the-art system, based on Analog Devices AD 9726, meets these requirements. In this context, the need to place resonance frequencies on an exact grid, possibly with the assistance of frequency tuning PID loops, or increased accuracy of the lithographic production of the LC bandpass filters used in FDM, is discussed. Finally, the impact of pulse shape, in particular electrical bandwidth, on DAC performance is assessed.
The X-ray Integral Field Unit (X-IFU) is one of the two instruments of the Athena astrophysics space mission approved by ESA in the Cosmic Vision 2015-2025 Science Programme. The X-IFU consists of a large array of transition edge sensor micro-calorimeters that will operate at ~100 mK inside a sophisticated cryostat. A set of thin filters, highly transparent to X-rays, will be mounted on the opening windows of the cryostat thermal shields in order to attenuate the IR radiative load, to attenuate radio frequency electromagnetic interferences, and to protect the detector from contamination. Thermal filters are critical items in the proper operation of the X-IFU detector in space. They need to be strong enough to survive the launch stresses but very thin to be highly transparent to X-rays. They essentially define the detector quantum efficiency at low energies and are fundamental to make the photon shot noise a negligible contribution to the energy resolution budget. In this paper, we review the main results of modeling and characterization tests of the thermal filters performed during the phase A study to identify the suitable materials, optimize the design, and demonstrate that the chosen technology can reach the proper readiness before mission adoption.
The X-ray Integral Field Unit (X-IFU) is a next generation microcalorimeter planned for launch onboard the Athena observatory. Operating a matrix of 3840 superconducting Transition Edge Sensors at 90 mK, it will provide unprecedented spectro-imaging capabilities (2.5 eV resolution, for a field of view of 5’) in the soft X-ray band (0.2 up to 12 keV), enabling breakthrough science. The definition of the instrument evolved along the phase A study and we present here an overview of its predicted performances and their modeling, illustrating how the design of the X-IFU meets its top-level scientific requirements. This article notably covers the energy resolution, count-rate capability, quantum efficiency and non X-ray background levels, highlighting their main drivers.
In the framework of the ESA Athena mission, the X-ray Integral Field Unit (X-IFU) micro-calorimeter will provide unprecedented spatially resolved high-resolution X-ray spectroscopy. For this purpose, the on-board Event Processor (EP) must initially trigger the current pulses induced by the X-ray photons hitting the detector to proceed with a reconstruction which provides the arrival time, spatial location and energy of each event. The current event triggering design is implemented in two stages: one initial trigger of the low-pass filtered derivative of the raw data to extract records containing pulses and a second stage performing a fine detection to look for all the pulses in the record. In order to establish the current baseline detection technique of the EP in the X-IFU instrument, an assessment of the capabilities of different triggering algorithms is required, both in terms of performance (detection efficiency) and computational load, as processing will take place on-board. We present a comparison of two detection algorithms, the Simplest Threshold Crossing (STC) and the model-dependent Adjusted Derivative (AD). The analysis also evaluates the (possible) negative effect of different instrumental scenarios as a reduced sampling rate. The evaluations point out that the simplest algorithm STC shows worse performance than AD for the smallest pulses separations and the lowest secondary energies. Nevertheless, checking the expected number of such pulses combinations in a typical bright source observation, we conclude that it does not have impact in the science. Moreover, the savings in the computational resources and calibration needs make STC a valuable option.
The X-ray Integral Field Unit (X-IFU) on board the Advanced Telescope for High-ENergy Astrophysics (Athena) will provide spatially resolved high-resolution X-ray spectroscopy from 0.2 to 12 keV, with ~ 5" pixels over a field of view of 5 arc minute equivalent diameter and a spectral resolution of 2.5 eV up to 7 keV. In this paper, we first review the core scientific objectives of Athena, driving the main performance parameters of the X-IFU, namely the spectral resolution, the field of view, the effective area, the count rate capabilities, the instrumental background. We also illustrate the breakthrough potential of the X-IFU for some observatory science goals. Then we brie y describe the X-IFU design as defined at the time of the mission consolidation review concluded in May 2016, and report on its predicted performance. Finally, we discuss some options to improve the instrument performance while not increasing its complexity and resource demands (e.g. count rate capability, spectral resolution).
In this paper we present a first assessment of the impact of various forms of instrumental crosstalk on the science performance of the X-ray Integral Field Unit (X-IFU) on the Athena X-ray mission. This assessment is made using the SIXTE end-to-end simulator in the context of one of the more technically challenging science cases for the XIFU instrument. Crosstalk considerations may influence or drive various aspects of the design of the array of high-countrate Transition Edge Sensor (TES) detectors and its Frequency Domain Multiplexed (FDM) readout architecture.
The focal plane of the X-ray integral field unit (X-IFU) for ESA’s Athena X-ray observatory will consist of ~ 4000 transition edge sensor (TES) x-ray microcalorimeters optimized for the energy range of 0.2 to 12 keV. The instrument will provide unprecedented spectral resolution of ~ 2.5 eV at energies of up to 7 keV and will accommodate photon fluxes of 1 mCrab (90 cps) for point source observations. The baseline configuration is a uniform large pixel array (LPA) of 4.28” pixels that is read out using frequency domain multiplexing (FDM). However, an alternative configuration under study incorporates an 18 × 18 small pixel array (SPA) of 2” pixels in the central ~ 36” region. This hybrid array configuration could be designed to accommodate higher fluxes of up to 10 mCrab (900 cps) or alternately for improved spectral performance (< 1.5 eV) at low count-rates. In this paper we report on the TES pixel designs that are being optimized to meet these proposed LPA and SPA configurations. In particular we describe details of how important TES parameters are chosen to meet the specific mission criteria such as energy resolution, count-rate and quantum efficiency, and highlight performance trade-offs between designs. The basis of the pixel parameter selection is discussed in the context of existing TES arrays that are being developed for solar and x-ray astronomy applications. We describe the latest results on DC biased diagnostic arrays as well as large format kilo-pixel arrays and discuss the technical challenges associated with integrating different array types on to a single detector die.
We present the design of tessim, a simulator for the physics of transition edge sensors developed in the framework of the Athena end to end simulation effort. Designed to represent the general behavior of transition edge sensors and to provide input for engineering and science studies for Athena, tessim implements a numerical solution of the linearized equations describing these devices. The simulation includes a model for the relevant noise sources and several implementations of possible trigger algorithms. Input and output of the software are standard FITS- files which can be visualized and processed using standard X-ray astronomical tool packages. Tessim is freely available as part of the SIXTE package (http://www.sternwarte.uni-erlangen.de/research/sixte/).
We present simulations of the detection probability for absorption lines from ions in the warm and hot ionized medium (WHIM) with Athena in the spectra of Gamma-ray burst afterglows. The simulations are based on Swift XRT lightcurves of these afterglows and are performed using the end-to-end simulation framework SIXTE. We simulate both the case of single and multiple absorption lines, as well as results for line searches in absorption structures from a more complex medium. We show that the Athena X-IFU can detect WHIM lines with strong Ovii lines (equivalent widths larger than 0.14 eV) in spectra containing 3 x 106 counts.
The X-ray Integral Field Unit (X-IFU) microcalorimeter, on-board Athena, with its focal plane comprising 3840 Transition Edge Sensors (TESs) operating at 90 mK, will provide unprecedented spectral-imaging capability in the 0.2-12 keV energy range. It will rely on the on-board digital processing of current pulses induced by the heat deposited in the TES absorber, as to recover the energy of each individual events. Assessing the capabilities of the pulse reconstruction is required to understand the overall scientific performance of the X-IFU, notably in terms of energy resolution degradation with both increasing energies and count rates. Using synthetic data streams generated by the X-IFU End-to-End simulator, we present here a comprehensive benchmark of various pulse reconstruction techniques, ranging from standard optimal filtering to more advanced algorithms based on noise covariance matrices. Beside deriving the spectral resolution achieved by the different algorithms, a first assessment of the computing power and ground calibration needs is presented. Overall, all methods show similar performances, with the reconstruction based on noise covariance matrices showing the best improvement with respect to the standard optimal filtering technique. Due to prohibitive calibration needs, this method might however not be applicable to the X-IFU and the best compromise currently appears to be the so-called resistance space analysis which also features very promising high count rate capabilities.
Athena is designed to implement the Hot and Energetic Universe science theme selected by the European Space Agency for the second large mission of its Cosmic Vision program. The Athena science payload consists of a large aperture high angular resolution X-ray optics (2 m2 at 1 keV) and twelve meters away, two interchangeable focal plane instruments: the X-ray Integral Field Unit (X-IFU) and the Wide Field Imager. The X-IFU is a cryogenic X-ray spectrometer, based on a large array of Transition Edge Sensors (TES), offering 2:5 eV spectral resolution, with ~5" pixels, over a field of view of 50 in diameter. In this paper, we present the X-IFU detector and readout electronics principles, some elements of the current design for the focal plane assembly and the cooling chain. We describe the current performance estimates, in terms of spectral resolution, effective area, particle background rejection and count rate capability. Finally, we emphasize on the technology developments necessary to meet the demanding requirements of the X-IFU, both for the sensor, readout electronics and cooling chain.
We present an overview of the development of the end-to-end simulation programs for the instruments on the future European X-ray astronomy mission Athena. The overview includes the design considerations behind the simulation software and the current status and planned developments of the simulators for the X-ray Integral Field Unit and the Wide Field Imager.
Access to the requested content is limited to institutions that have purchased or subscribe to SPIE eBooks.
You are receiving this notice because your organization may not have SPIE eBooks access.*
*Shibboleth/Open Athens users─please
sign in
to access your institution's subscriptions.
To obtain this item, you may purchase the complete book in print or electronic format on
SPIE.org.
INSTITUTIONAL Select your institution to access the SPIE Digital Library.
PERSONAL Sign in with your SPIE account to access your personal subscriptions or to use specific features such as save to my library, sign up for alerts, save searches, etc.