6 July 1979 Performance Evaluation Of A Fourth-Generation Computed Tomography (CT) Scanner
Robert K. Cacak, William R. Hendee
Author Affiliations +
Proceedings Volume 0173, Application of Optical Instrumentation in Medicine VII; (1979) https://doi.org/10.1117/12.957143
Event: Application of Optical Instrumentation in Medicine VII, 1979, Toronto, Canada
Abstract
One of the first American Science and Engineering Co. (AS&E) CT scanners installed in a clinical environment is operational at the University of Colorado Medical Center. The performance characteristics of several aspects of this stationary detector array scanner have been measured. Included in these characteristics are the resolution capabilities of the scanner in terms of MTF for high contrast objects, and the perceptability capabilities for objects of lower contrast. Other characteristics include patient dose, beam profiles, CT number linearity, noise, algorithm dependences, and the performance of selected electromechanical features such as patient bed incrementing and light beam/radiation field coincidence. Equipment and techniques for measuring these quantities are described. Although data are presented solely for the CT scanner, the procedures for evaluation of performance and periodic quality control may be generalized to other CT scanners. A summary of quality assurance tests being performed and a brief description of these tests in tabular form are presented.
© (1979) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Robert K. Cacak and William R. Hendee "Performance Evaluation Of A Fourth-Generation Computed Tomography (CT) Scanner", Proc. SPIE 0173, Application of Optical Instrumentation in Medicine VII, (6 July 1979); https://doi.org/10.1117/12.957143
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Scanners

Computed tomography

X-ray computed tomography

Modulation transfer functions

Sensors

Medicine

Optical instrument design

RELATED CONTENT

Radiation Dosimetry In Computed Tomography (CT)
Proceedings of SPIE (December 27 1977)
Sensitometry In Computerized Tomography
Proceedings of SPIE (December 27 1977)
High resolution dynamic CT scanner based on a variable zoom...
Proceedings of SPIE (September 14 1993)
A Comparison Of Some Models Of X-Ray Beam Hardening
Proceedings of SPIE (December 23 1976)
Optimizing Computed Tomography (CT) Scanner Geometry
Proceedings of SPIE (July 06 1979)

Back to Top