Presentation
6 March 2020 Ultra-sensitive photoacoustic microscopy based on high numerical aperture acoustic lens (Conference Presentation)
Author Affiliations +
Abstract
Due to limited ultrasound detection angle, photoacoustic microscopy may own a relatively low sensitivity. To break this limit, we develop an ultra-sensitive optical resolution photoacoustic microscopy based on a customized acoustic lens with high numerical aperture (0.74) (HNA-OR-PAM). The sensitivity of HNA-OR-PAM is improved to around 160% as the state-of-the-art OR-PAM. It has the capability to measure oxygen saturation of mice’s ear in vivo with ~10nJ pulse energy, reducing the nonlinear effect induced by high pulse energy. In addition, photoacoustic signal of tilted objects could be enhanced due to augmented ultrasound detection angle, which has been validated in our phantom study and the brain imaging experiment in vivo.
Conference Presentation
© (2020) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Mingsheng LI, Jiangbo Chen, and Lidai Wang "Ultra-sensitive photoacoustic microscopy based on high numerical aperture acoustic lens (Conference Presentation)", Proc. SPIE 11240, Photons Plus Ultrasound: Imaging and Sensing 2020, 1124020 (6 March 2020); https://doi.org/10.1117/12.2544750
Advertisement
Advertisement
KEYWORDS
Photoacoustic spectroscopy

Acoustics

Photoacoustic microscopy

Laser energy

Signal to noise ratio

Microscopes

Pulsed laser operation

RELATED CONTENT


Back to Top