27 November 1989 Thermal Detectors As X-Ray Calorimeters: Theoretical And Experimental Results And Data Processing Techniques
R. Kessel, D. M. Walton, J. L. Culhane, W. S. Holland, I. D. Hepburn, P. A. R. Ade, M. J. Griffin
Author Affiliations +
Abstract
We report measurements on the energy resolution of germanium X-ray calorimeters for the energy range 1-50 keV. The detector we use is a modified infrared bolometer which is currently operated at 3He temperatures (0.31K) and is capable of attaining a FWHM of 172 eV for 5.9 keV X-rays. We have identified and characterized two types of event which can contaminate the spectrum of the bare germanium calorimeter. The first type is due to photons absorbed in the indium soldered contacts of the detector. The second type is due to β-particles striking the detector. The theoretical performance of such a device has been modelled by considering the individual noise sources that limit the energy resolution. We also present a discussion concerning the post-processing of data which employs both the techniques of matched filtering and maximum entropy method deconvolution.
© (1989) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
R. Kessel, D. M. Walton, J. L. Culhane, W. S. Holland, I. D. Hepburn, P. A. R. Ade, and M. J. Griffin "Thermal Detectors As X-Ray Calorimeters: Theoretical And Experimental Results And Data Processing Techniques", Proc. SPIE 1140, X-Ray Instrumentation in Medicine and Biology, Plasma Physics, Astrophysics, and Synchrotron Radiation, (27 November 1989); https://doi.org/10.1117/12.961806
Lens.org Logo
CITATIONS
Cited by 4 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Sensors

X-rays

Germanium

X-ray detectors

Crystals

Electrons

Infrared sensors

RELATED CONTENT


Back to Top