1 May 1990 Analysis of random noise, coil inhomogeneity, and ghost artifacts in spin echo imaging on a 1.5 T commercial magnetic resonance imager
Ramesh Chandra, Henry Rusinek
Author Affiliations +
Proceedings Volume 1245, Biomedical Image Processing; (1990) https://doi.org/10.1117/12.19547
Event: Electronic Imaging: Advanced Devices and Systems, 1990, Santa Clara, CA, United States
Abstract
Noise in spin echo (SE) images in MRI consists of random and structured components Structured noise arises mainly from the non-uniformity of the B1 field. We have studied these errors using a cylindrical phantom. Random noise analysis by subtraction of a 4,9, or 25 points smoothed image from the original image showed small but significant differences between the three methods, though it was uniform over the phantom and changed little with time over a span of 8 months. The rf coil nonuniformity ( coefficient of variation,CV) in two areas of interest (ROI). a central area (40 cm2) and a ring outside the central region (130 cm2) was measured to be 21% and 45% respectively for the saddle coil and 1 5% and 6.0% respectively for the bird cage coil Using the first day flood image for correction of the later images. CV in the two ROI's of the corrected images were 1.3% and 2.2% for the saddle coil respectively and 0.9% and 1.6% respectively.
© (1990) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Ramesh Chandra and Henry Rusinek "Analysis of random noise, coil inhomogeneity, and ghost artifacts in spin echo imaging on a 1.5 T commercial magnetic resonance imager", Proc. SPIE 1245, Biomedical Image Processing, (1 May 1990); https://doi.org/10.1117/12.19547
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Magnetic resonance imaging

Magnetism

Image processing

Biomedical optics

Imaging systems

Signal detection

Brain

Back to Top