Paper
1 July 1991 Sensitive fiber-optic immunoassay
Irene M. Walczak, Walter F. Love, Rudolf E. Slovacek
Author Affiliations +
Proceedings Volume 1420, Optical Fibers in Medicine VI; (1991) https://doi.org/10.1117/12.43859
Event: Optics, Electro-Optics, and Laser Applications in Science and Engineering, 1991, Los Angeles, CA, United States
Abstract
The principles of evanescent wave theory were applied to an immunological sensor for detecting the cardiac-specific isoenzyme creatine kinase-MB (CK-MB). The detection of the CK-MB isoenzyme is used in conjunction with the total CK measurement in the diagnosis of acute myocardial infarction. The clinical range for CK-MB is from 2-100 ng/ml. Previous work which utilized the fluorophor, Fluorescein isothiocyanate (FITC), was able to discriminate between 0 and 3 ng/ml CK-MB. Use of the fluorophor B-phycoerythrin (BPE) increased the assay sensitivity to 0.1 ng/ml CK-MB. The data was collected for 15 minutes using an optical launch and collection angle of 25 degree(s). This fiber optic based system is homogeneous and requires no subsequent washing, handling, or processing steps after exposure to the sample.
© (1991) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Irene M. Walczak, Walter F. Love, and Rudolf E. Slovacek "Sensitive fiber-optic immunoassay", Proc. SPIE 1420, Optical Fibers in Medicine VI, (1 July 1991); https://doi.org/10.1117/12.43859
Lens.org Logo
CITATIONS
Cited by 3 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Fiber optics sensors

Luminescence

Sensors

Fiber optics

Molecules

Interfaces

Medicine

Back to Top