Modifications made to the Long Trace Profiler (LTP II) system at the Advanced Photon Source at Argonne National Laboratory have significantly improved the accuracy and repeatability of the instrument. The use of a Dove prism in the reference beam path corrects for phasing problems between mechanical errors and thermally-induced system errors. A singe reference correction now completely removes both error signals from the measured surface profile. The addition of a precision air conditioner keeps the temperature in the metrology enclosure constant to within +/- 0.1 degree(s)C over a 24 hour period and has significantly improved the stability and repeatability of the system. We illustrate the performance improvements with several sets of measurements. The improved environmental control has reduced thermal drift error to about 0.75 microradian RMS over a 7.5 hour time period. Measurements made in the forward scan direction and the reverse scan direction differ by only about 0.5 microradian RMS over a 500 mm trace length. We are now able to put 1-sigma error bar of 0.3 microradian on an average of 10 slope profile measurements over a 500 mm long trace length, and we are now able to put a 0.2 microradian error bar on an average of 10 measurements over a 200 mm trace length. The corresponding 1-sigma height error bar for this measurement is 1.1 nm.
PERSONAL Sign in with your SPIE account to access your personal subscriptions or to use specific features such as save to my library, sign up for alerts, save searches, etc.
The alert did not successfully save. Please try again later.
Peter Z. Takacs, Cynthia J. Bresloff, "Significant improvements in Long Trace Profiler measurement performance," Proc. SPIE 2856, Optics for High-Brightness Synchrotron Radiation Beamlines II, (22 November 1996); https://doi.org/10.1117/12.259878