3 May 2002 Approaching the theoretical x-ray sensitivity with Hgl2 direct detection image sensors
Robert A. Street, Steve E. Ready, Leonid Melekhov, Jackson Ho, Asaf Zuck, Barry N. Breen
Author Affiliations +
Abstract
The x-ray response of polycrystalline HgI2 for direct detection x-ray imagers, is studied using test arrays with 512 X 512 pixels of size 100 micron. We quantify the contributions to the x-ray sensitivity from electron and hole charge collection, x-ray absorption, effective fill factor and image lag, for x-ray energies from 25-100 kVp. The data analysis compares the measured sensitivity to the theoretical limit and identifies the contributions from various loss mechanisms. The sensitivity is explained by the ionization energy of approximately 5 eV, coupled with small corrections arising from incomplete x-ray absorption, incomplete charge collection, and image lag. Hence, imagers with HgI2 approach the theoretical maximum response for semiconductor detectors, with external array sensitivity demonstrated to within 50 percent of the limit.
© (2002) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Robert A. Street, Steve E. Ready, Leonid Melekhov, Jackson Ho, Asaf Zuck, and Barry N. Breen "Approaching the theoretical x-ray sensitivity with Hgl2 direct detection image sensors", Proc. SPIE 4682, Medical Imaging 2002: Physics of Medical Imaging, (3 May 2002); https://doi.org/10.1117/12.465584
Lens.org Logo
CITATIONS
Cited by 22 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
X-rays

Absorption

X-ray imaging

Ionization

Image sensors

Imaging systems

Signal attenuation

Back to Top