Paper
1 April 2003 Rectification of nonlinearity in pre-amplifier and A/D converter of biomedical instruments
Author Affiliations +
Abstract
Nonlinear distortions are always introduced to biomedical signals during the acquisition stage, which consequently fail the traditional linear independent component analysis (ICA) methods for further signal processing. This paper investigates the non-linearity system function in the pre-amplifier and A/D converter of the biomedical instruments. A polynomial general model structure with adjustable parameters to approximate the nonlinear relation is proposed for medical instruments. Model parameters are validated using a typical electrocardiograph (ECG) acquisition system with sinusoids of varying frequency and amplitude. Thus the inverse nonlinear transform is applied to acquired data to cancel the nonlinear distortions. The ICA method is then applied to the originally linear mixed data, non-rectified data and also rectified data and the results are favorably compared in the designed experiment using both clinical ECGs and the simulated data from cardiac simulators.
© (2003) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Jun Yang, Seong-Bin Yim, Ivica Kopriva, and Harold H. Szu "Rectification of nonlinearity in pre-amplifier and A/D converter of biomedical instruments", Proc. SPIE 5102, Independent Component Analyses, Wavelets, and Neural Networks, (1 April 2003); https://doi.org/10.1117/12.501740
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Independent component analysis

Electrocardiography

Complex systems

Amplifiers

Biomedical optics

Signal processing

Statistical analysis

Back to Top