Paper
20 January 2004 Evaluation of CZT detectors with capacitive Frisch grid structure
Author Affiliations +
Abstract
Recent results have shown that capacitive Frisch grid structures significantly improve spectroscopic performance of planar CZT detectors especially at higher energies. This paper presents results obtained with larger detectors than those previously reported on. Devices with various aspect ratios and grid length-to-device thickness ratios were fabricated and evaluated. A FWHM energy resolution of approximately 2% at 662 keV was obtained for a device with dimensions of 5 mm x 5 mm x 9.2 mm.
© (2004) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Leonard J. Cirignano, Hadong Kim, Kanai S. Shah, Misha Klugerman, Philip Wong, Michael R. Squillante, and Longxia Li "Evaluation of CZT detectors with capacitive Frisch grid structure", Proc. SPIE 5198, Hard X-Ray and Gamma-Ray Detector Physics V, (20 January 2004); https://doi.org/10.1117/12.507891