20 April 2005 Evaluation of a photon-counting breast tomosynthesis imaging system
Andrew Maidment, Michael Albert, Stefan Thunberg, Leif Adelow, Ola Blom, Johan Egerstrom, Mathias Eklund, Tom Francke, Ulf Jordung, Tomas Kristoffersson, Karin Lindman, Lars Lindqvist, Daniel Marchal, Hans Olla, Erik Penton, Juha Rantanen, Skiff Solokov, Christer Ullberg, Niclas Weber
Author Affiliations +
Abstract
Digital breast tomosynthesis promises solutions to many of the problems currently associated with projection mammography, including elimination of artifactual densities due to the superposition of normal tissues and increasing the conspicuity of true lesions that would otherwise be masked by superimposed normal tissue. We have investigated tomosynthesis using 45 photon counting, orientation sensitive, linear detectors which are precisely aligned with the focal spot of the x ray source. The x-ray source and the digital detectors are scanned in a continuous motion across the object (patient); each linear detector collecting an image at a distinct angle. Simulations of the imaging system were performed to evaluate the effect of: (1) the range of angles over which projection images are acquired; and (2) the number of projection images acquired used in the tomosynthetic reconstruction. Two different simulations were evaluated; the first was a numerical simulation of a tungsten wire; the second consisted of tomosynthetic reconstructions of a cadaveric rabbit, in which the number and/or range of projection angles was varied. We have shown, analytically and through these simulations, that both the use of more projection angles and the use of a larger range of projection angles improve the image quality of tomosynthetic image reconstructions. The use of a photon-counting x-ray detector system allows us to consider image acquisition geometries with a large number of projection angles, as there is no additive detector noise to degrade the projection or reconstructed images. The maximum number of projection angles and the range of projections angles do have upper practical limits; the range of projection angles is determined predominantly by the detector element size.
© (2005) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Andrew Maidment, Michael Albert, Stefan Thunberg, Leif Adelow, Ola Blom, Johan Egerstrom, Mathias Eklund, Tom Francke, Ulf Jordung, Tomas Kristoffersson, Karin Lindman, Lars Lindqvist, Daniel Marchal, Hans Olla, Erik Penton, Juha Rantanen, Skiff Solokov, Christer Ullberg, and Niclas Weber "Evaluation of a photon-counting breast tomosynthesis imaging system", Proc. SPIE 5745, Medical Imaging 2005: Physics of Medical Imaging, (20 April 2005); https://doi.org/10.1117/12.597317
Lens.org Logo
CITATIONS
Cited by 39 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Sensors

Modulation transfer functions

Imaging systems

X-rays

Computer simulations

Image quality

Image restoration

Back to Top