20 April 2005 Improving the imaging of calcifications in CT by histogram-based selective deblurring
Empar Rollano-Hijarrubia, Frits van der Meer, Add van der Lugt, Harrie Weinans, Henry Vrooman, Albert Vossepoel, Rik Stokking
Author Affiliations +
Abstract
Imaging of small high-density structures, such as calcifications, with computed tomography (CT) is limited by the spatial resolution of the system. Blur causes small calcifications to be imaged with lower contrast and overestimated volume, thereby hampering the analysis of vessels. The aim of this work is to reduce the blur of calcifications by applying three-dimensional (3D) deconvolution. Unfortunately, the high-frequency amplification of the deconvolution produces edge-related ring artifacts and enhances noise and original artifacts, which degrades the imaging of low-density structures. A method, referred to as Histogram-based Selective Deblurring (HiSD), was implemented to avoid these negative effects. HiSD uses the histogram information to generate a restored image in which the low-intensity voxel information of the observed image is combined with the high-intensity voxel information of the deconvolved image. To evaluate HiSD we scanned four in-vitro atherosclerotic plaques of carotid arteries with a multislice spiral CT and with a microfocus CT (μCT), used as reference. Restored images were generated from the observed images, and qualitatively and quantitatively compared with their corresponding μCT images. Transverse views and maximum-intensity projections of restored images show the decrease of blur of the calcifications in 3D. Measurements of the areas of 27 calcifications and total volumes of calcification of 4 plaques show that the overestimation of calcification was smaller for restored images (mean-error: 90% for area; 92% for volume) than for observed images (143%; 213%, respectively). The qualitative and quantitative analyses show that the imaging of calcifications in CT can be improved considerably by applying HiSD.
© (2005) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Empar Rollano-Hijarrubia, Frits van der Meer, Add van der Lugt, Harrie Weinans, Henry Vrooman, Albert Vossepoel, and Rik Stokking "Improving the imaging of calcifications in CT by histogram-based selective deblurring", Proc. SPIE 5745, Medical Imaging 2005: Physics of Medical Imaging, (20 April 2005); https://doi.org/10.1117/12.595491
Lens.org Logo
CITATIONS
Cited by 9 scholarly publications and 2 patents.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Image restoration

Computed tomography

Deconvolution

3D image processing

Point spread functions

In vitro testing

Tissues

Back to Top