Lateral flow devices (LFDs) are widely used point-of-care (POC) diagnostics. The basic LFD design remains largely unchanged since their first development and this limits their use in clinical applications due to lack of sensitivity. To enhance this, we report the use of laser-patterned geometric control barriers, in the form of a constriction, that leads to a slower flow rate and smaller test zone area. This high sensitivity LFD (HS-LFD) achieved 62% increase in test line colour intensity for the detection of procalcitonin (PCT) and reduced the LOD from 10 ng/ml to 1 ng/ml with contrived human samples
PERSONAL Sign in with your SPIE account to access your personal subscriptions or to use specific features such as save to my library, sign up for alerts, save searches, etc.
The alert did not successfully save. Please try again later.
Alice Iles, Ioannis Katis, Peijun He, Panagiotis Galanis, Collin Sones, Robert Eason, "Optimisation of flow path parameters for enhanced sensitivity lateral flow devices," Proc. SPIE PC11968, Optical Diagnostics and Sensing XXII: Toward Point-of-Care Diagnostics, PC1196804 (12 May 2022); https://doi.org/10.1117/12.2603880