1.IntroductionIn x-ray microcalorimeter instruments, the purpose of dewar blocking filters is (1) to shield the detector array from thermal radiation from the instrument and optical and UV photons from the sky and (2) to shield the coldest stages of the instrument from the thermal radiation emanating from the warmer stages. Additional requirements on the filters are to allow x-ray transmission in the scientific band of interest—which has led to the use of thin films—and to be mechanically robust to allow handling, vacuum, and low-temperature operation as well as to withstand launch vibration. Because the x-ray transmission may be drastically reduced by even thin layers of molecular contamination on the filters, we also require that the filter and aperture assembly design minimize the probability of contamination and allow removal of contaminants via filter heaters that can be operated on-orbit. These requirements have led to the use of meshes for mechanical support and to aid in thermal conduction. Other attenuators in the optical path, which may combine to reduce the total effective area but can also aid in meeting some out-of-band attenuation requirements, include telescope thermal shields, filters on an external-to-the-dewar filter wheel, and the window on the dewar door, the latter of which is typically designed to withstand more than an atmosphere of differential pressure (beryllium is a typical material of choice). 1.1.HeritageThe Astro-H (Hitomi) soft x-ray spectrometer (SXS),1 launched in 2016, used a series of five thin-film radiation-blocking filters anchored to nested temperature stages that blocked long-wavelength radiation while minimizing x-ray attenuation. The SXS aperture assembly used a system of barriers, baffles, filter carriers, and filter mounts that supported the filters and inhibited their potential contamination. The two inner filters were freestanding with Al/polyimide thicknesses of 50/75 nm. The three outer (warmer) filters were equipped with thermometers and heaters for decontamination and used Al/polyimide thicknesses of 100/100 nm and micromachined silicon support meshes. Further details on the design, implementation, and calibration of the SXS filters and aperture assembly may be found in Refs. 2 and 3. Currently, we are rebuilding the SXS instrument, now named Resolve,4 for the X-Ray Imaging and Spectroscopy Mission (XRISM),5 the mission aimed at recovering the calorimeter science lost due to the early loss of the Hitomi satellite.6 The Resolve filters will be identical to those of SXS. As an example of out-of-band requirements for a microcalorimeter instrument, for Resolve, the required maximum out-of-band transmission of each of the thinner freestanding filters is in the visible and in the IR range (3 to ), of each of the thicker mesh-supported filters is in the visible and in the IR and of the combination of the filters and telescope thermal shield is at 40.8 eV (geo-coronal He II), at 21.2 eV, and at 10.2 eV. The SXS filters built on designs developed for the X-ray Quantum Calorimeter (XQC) sounding rocket experiment,7,8 which has flown six times and used even thinner thin-film layers than SXS, with nominal thicknesses of Al and 50-nm polyimide. Similar filters also flew on the Micro-X Sounding Rocket experiment in 2018.9 The X-ray Integral Field Unit (X-IFU)10,11 on the Athena x-ray observatory, scheduled for launch in 2031, will use transition-edge-sensor microcalorimeters. The X-IFU will also use a set of optical blocking filters, and these are currently under development. While the basic design will be similar to those developed for previous microcalorimeter missions, the X-IFU filters team is exploring several options, including meshes of different material than silicon and replacing the thin film to use SiN instead of polyimide.12,13 One interesting option currently under consideration for the X-IFU is to coat the microcalorimeter pixels with a thin layer of Au to increase the reflectivity in the optical and IR. The pixels under development for the X-IFU use Au/Bi absorbers, where the material facing the telescope is Bi, which has lower reflectivity in the optical and IR compared to the Au/Bi/Au stack under consideration. Increasing the reflectivity of the microcalorimeter x-ray absorbers means that the out-of-band transmission requirements of the filters will be relaxed. 1.2.Considerations (Differences) for Lynx X-ray MicrocalorimeterAlthough the Lynx filters will retain heritage based on the previous missions, there are a few important differences that will drive aspects of the design.
2.Lynx X-ray Microcalorimeter Optical Blocking Filter DesignFor the LXM baseline design, we have chosen to use six filters, four of which are “conventional,” with thin aluminized polyimide films supported by silicon meshes, and two of which are waveguide-cutoff filters that do not employ thin films. The driving factor behind this choice is the need for a higher transmission at low x-ray energies. Table 1 summarizes the baseline LXM filter stack in comparison to previous missions. The waveguide-cutoff filters reduce the x-ray transmission by a geometrical factor set by the mesh parameters, but because the holes are open, and not covered by a film, the transmission through the holes is unity across the x-ray waveband. For the conventional filters, we have chosen thinner films, again to increase the soft-band x-ray transmission. Each filter will be Al/20-nm polyimide. We expect the out-of-band long-wavelength attenuation of these thin filters will be limited by the few nm-thick oxide on each filter.15 Fabrication experts are working to study representative thin films and the resulting effects of the oxidation (see Sec. 3). The baseline films are thinner than those used for previous missions, and if we cannot meet the out-of-band rejection requirements or if they are too fragile (i.e., cannot survive handling or environmental tests with high reliability) we can employ thicker films, as described below. Table 1Summary of dewar blocking filters for heritage instruments and LXM. For both X-IFU and LXM, the size of the aperture requires all filters to have mechanical support meshes.
2.1.Need for Support MeshesFor large filters, and with the typical thin-film materials, support meshes are required even though they reduce the filter transmission. Figure 2 provides an example of the silicon support meshes used for Astro-H SXS. When filters are small enough, we may omit the mesh. But, unlike the case of SXS and Resolve, meshless filters are impractical for the Lynx microcalorimeter instrument for even the smallest filters. This statement holds for all filters in the current Lynx baseline design (even without the extended array). The size of the LXM detector array means that the dewar filters will be large (a few cm diameter), and thin, freestanding films with this diameter will be too fragile. For Astro-H, the largest freestanding filter was 12-mm diameter. The 18.5-mm-diameter filter and larger required a mesh for mechanical support. The baseline Lynx design has the smallest (coldest) Lynx filter at 26-mm diameter for the baseline array configuration. And, the proposed Lynx films are thinner (more fragile) than the Astro-H filters. We conclude that based on technology demonstrations to date, the baseline Lynx design should have meshes for all filters. 2.2.Rationale for Proposed Filter Composition and Resulting X-Ray TransmissionWe assume six filters, with support meshes on all filters. The total film thickness is 40-nm Al/80-nm polyimide, assumed to be split over 3 to 4 filters. These are quite thin films if split over four filters and moderately thin if on three filters. Waveguide-cutoff filters (no film) are assumed for 2 to 3 of the filters. See Sec. 3 for information about the waveguide-cutoff filters. The transmission of this set of filters is shown in Fig. 3, and comparisons to Astro-H SXS and Athena X-IFU are presented in Figs. 4 and 5. In this calculation, we assume the Athena support meshes are thick metal meshes with 2% blocking factor13 and Lynx support meshes are identical to the mesh of the Astro-H SXS dewar main shell filter, with 97.5% open-area coarse Si mesh ( Si) plus 97.5% open-area fine mesh ( Si). We assume that the waveguide-cutoff filters for Lynx have the same geometry as the Astro-H meshes, including a thick coarse mesh and a thinner fine mesh but are made of Au. For the “mesh Lynx” curves in Figs. 3–5, we assume two waveguide-cutoff filters, indicated by the larger number of meshes than films. A cross-sectional diagram of the filters in the LXM dewar is presented in Fig. 6 and details about the implementation, including the filter locations, temperatures, and sizes, are presented in Table 2. Table 2Left columns: dewar filter thermal interface, filter abbreviation, nominal operating temperature, clear aperture diameter, expected beam diameter from the telescope, and distance from the filter to the detector. The larger diameter of the clear aperture compared to the expected beam diameter provides alignment margin and allows greater illumination by the MXSs. Right columns: baseline thin-film component thicknesses.
The smallest LXM filter is similar to the size of the Astro-H SXS mesh filters, and the larger LXM filters are in-family with the size of the Athena X-IFU filters, which are currently under development.12,13 While we have baselined Si meshes similar to those of Astro-H SXS, the mechanical properties of such meshes at larger sizes will need to be studied in detail. This mechanical analysis has not yet been performed. A change in the mesh design may be required to change the resonance frequencies or quality factors, which could involve selectively thickening the bars. Studies of the mesh fabrication yield will also be performed, and, if the yield is not high enough for large Si meshes, then alternate mesh materials will be adopted. In addition to mechanical mesh failures, defects in the mesh or adhesive that bonds the mesh to the thin film can lead to filter film failures during environmental testing, in particular for the thin films baselined for LXM, so this will also be studied for Lynx and is currently under study for Athena X-IFU. We will incorporate any lessons learned from the X-IFU development program regarding mesh and filter design and fabrication. We do not anticipate a reduction in fabrication yield of the films themselves for the LXM filters compared to filters with smaller diameters. The final filter composition and transmission may change as follows:
3.Technology Development and Testing for Lynx X-ray Microcalorimeter Optical Blocking FiltersSeveral areas of technology development have been pursued to provide enhanced options for the LXM optical blocking filters. In this section, we present an overview of some of these studies, each of which will be explained in more detail in upcoming articles.17,18 The Luxel Corporation has undertaken several studies of thin-film filters for LXM. Similar to studies underway for Athena X-IFU filters, they have been exploring the possibility of using silicon nitride instead of polyimide to provide an increase in low-energy transmission. Figure 7 shows an example of the increase in transmission at energies below 100 eV offered by exchanging a 200-nm polyimide film for a SiN film of half the thickness. Luxel has also been studying the reflectance and absorption of thin aluminized polyimide films in the optical, NIR, and FIR. They measured samples of various aluminum thicknesses, into the regime where the aluminum becomes thinner than their standard products () and approaches the thicknesses specified for LXM. Figure 8 presents initial results of measurements on films with aluminum thicknesses ranging from 12 to 33 nm. The value of the initial reflective surface even in very thin coatings is apparent in Fig. 8. For example, two 12-nm Al films will have lower IR transmission than a single 24-nm Al film for wavelengths . While x-ray transmission will not be identical for these scenarios due to the native oxide layer on the Al surfaces, the data show that the optical blocking per unit thickness is improved with thinner Al films. The data do not follow this trend, indicating that the shorter wavelengths may be passing through discontinuities in the thin Al surface whereas longer wavelengths can span the discontinuities and see a more continuous surface. The data may suggest that thinner Al layers are particularly viable deeper in the Lynx filter stack to block infrared photons from warmer stages, whereas outer filters should maintain an optically continuous layer of to attenuate shorter wavelengths. Fabrication procedures that provide more continuous thin Al films are under investigation. In addition, to test whether extreme ultraviolet (EUV) radiation would increase the aluminum oxidation on-orbit and thus lessen the optical/IR attenuation, Luxel worked with partners at the National Institute for Standards and Technology (NIST), Gaithersburg, Maryland, and the Laboratory for Atmospheric and Space Physics to irradiate aluminized polyimide filters with EUV photons using the SURF-III facility at NIST. Some filters did exhibit increased oxidation; however, it was clear that while this effect must be considered for missions that will experience intense solar EUV radiation, this will not be a problem for the much lower intensity levels that Lynx will experience. Additional details will be available in a forthcoming paper.17 In addition to these studies of advanced thin-film filters, several groups (e.g., groups led by R. Fettig, D. McCammon, and E. Wollack) have been exploring the feasibility of using waveguide-cutoff filters to block long-wavelength radiation using a periodic array of hexagonal holes with thin walls. This configuration should result in a far-infrared cutoff frequency where the hole size is used to control the onset of the high-pass response and thickness determines the rejection. These filters have heritage in metal mesh filters that have been used for many years in the sub-mm and mm astronomy communities. Advances in experimental realizations of these filters will be described in Ref. 18, an upcoming paper from researchers in both the US and Germany and from both the Lynx and Origins Space Telescope study teams. The geometries that have been explored by various groups include meshes of different materials and different length-to-width ratios of the holes. The x-ray blocking fraction of recent designs is similar to that of the Si support meshes used for Astro-H SXS, although preliminary testing showed a slight increase in x-ray blocking fraction due to details of the filter fabrication. 4.Filter Wheel and Modulated X-Ray SourcesFigures 9 and 10 present views of the dewar model highlighting the current design of the LXM filters (external to the dewar) and MXSs. To avoid a large diameter wheel, the LXM filters are mounted on paddles that are stacked vertically on top of one another so that one (or more) filter can swing into the optical path. The filters are housed in the gray module shown in the bottom of the right panel of Fig. 9. The optical path is highlighted by the orange circle on the filter assembly. At the current level of development, we assume filters similar to those used for Astro-H SXS,19 including Be filters to reduce the low-energy flux and a neutral density filter to provide an energy-independent reduction in flux. For LXM, we plan to use an aluminized polyimide optical blocking filter instead of the polyimide-only contamination blocking filter on Astro-H, to add margin in the event that one of the dewar filters is damaged or degrades. This change will be implemented for XRISM Resolve as well. LXM will have a suite of eight MXSs for calibration of the detector gain, real-time monitoring of the gain drift, and measurements of the core line-spread function (energy resolution). The MXSs themselves are not shown in the model, but their mounting fixtures are apparent in the ring mounted between the dewar entrance and the external filter assembly. The MXSs are assumed to be similar to the ones used for Astro-H SXS19 and Athena X-IFU.20 For Astro-H SXS, there were two MXSs with “direct” illumination that could be used for gain drift monitoring and two “indirect” fluorescent sources that provided low-energy lines but were dim enough that they could only be used at high duty cycles (). In Fig. 9, the mounting spaces for the direct MXSs are shown as tubes pointing diagonally toward the detector. The indirect sources would be mounted parallel to the ring pointing at a fluorescent target. Other implementations may be possible as there are development efforts underway to build brighter MXSs that provide low-energy lines. In the case of Astro-H SXS, only one MXS was operated at a time and each MXS illuminated the entire detector array. This configuration meant that it was the sightline of the MXS that drove the requirements on the size of each filter clear aperture. For LXM, we plan to operate more than one MXS at a time to illuminate the entire detector array; with the current design, each MXS will only illuminate approximately half of the detector array. 5.ConclusionWe have presented the baseline design of the LXM optical blocking filters and compared the filters to those on recent and upcoming microcalorimeter instruments, such as the Astro-H SXS and Athena X-IFU. The LXM will aim to provide a higher filter transmission at low energies than the previous missions, which leads to the use of very thin Al/polyimide filters and waveguide-cutoff filters. The size of the LXM detector array, in the baseline configuration, means that the blocking filters will be of a spatial extent large enough that they will need mechanical support meshes (given technical demonstrations to-date) but are in-family with the size of filters required for the Athena X-IFU. If the extended array option is chosen, then significantly larger filters will be required and additional care will be needed to be taken with regard to mechanical properties and uniformity. We have briefly introduced a few of the filter testing and technology development projects underway to address the LXM filter baseline design and enhanced goals, including studies of filters with thin Al layers and of waveguide-cutoff filters using arrays of hexagonal holes. We find that fabricating films consistent with the LXM baseline design that meet the out-of-band blocking requirements will be challenging and need to be verified during the Lynx development program; the filters may need slightly thicker Al. We have also presented the basic configuration of the filters and MXSs mounted outside of the LXM dewar. AcknowledgmentsPart of this work was performed under the auspices of the U.S. Department of Energy by Lawrence Livermore National Laboratory under Contract No. DE-AC52-07NA27344. ReferencesR. L. Kelley et al.,
“The Astro-H high resolution soft x-ray spectrometer,”
Proc SPIE, 9905 99050V
(2016). https://doi.org/10.1117/12.2232509 Google Scholar
C. A. Kilbourne et al.,
“Design, implementation, and performance of the Astro-H soft x-ray spectrometer aperture assembly and blocking filters,”
J. Astron. Telesc. Instrum. Syst., 4 011215
(2018). https://doi.org/10.1117/1.JATIS.4.1.011215 Google Scholar
M. E. Eckart et al.,
“Ground calibration of the Astro-H (Hitomi) soft x-ray spectrometer,”
J. Astron. Telesc. Instrum. Syst., 4 021406
(2018). https://doi.org/10.1117/1.JATIS.4.2.021406 Google Scholar
Y. Ishisaki et al.,
“Resolve Instrument on X-ray Astronomy Recovery Mission (XARM),”
J. Low Temp. Phys., 193 991
–995
(2018). https://doi.org/10.1007/s10909-018-1913-4 JLTPAC 0022-2291 Google Scholar
M. Tashiro et al.,
“Concept of the X-ray Astronomy Recovery Mission,”
Proc SPIE, 10699 1069922
(2018). https://doi.org/10.1117/12.2309455 Google Scholar
T. Takahashi et al.,
“Hitomi (ASTRO-H) X-ray Astronomy Satellite,”
J. Astron. Telesc. Instrum. Syst., 4 021402
(2018). https://doi.org/10.1117/1.JATIS.4.2.021402 Google Scholar
D. McCammon et al.,
“A high spectral resolution observation of the soft x-ray diffuse background with thermal detectors,”
Astrophys. J., 576 188
–203
(2002). https://doi.org/10.1086/apj.2002.576.issue-1 Google Scholar
S. G. Crowder et al.,
“Observed limits on charge exchange contributions to the diffuse x-ray background,”
Astrophys. J., 758 143
(2012). https://doi.org/10.1088/0004-637X/758/2/143 Google Scholar
P. Wikus et al.,
“Progress on the Micro-X Sounding Rocket X-Ray Telescope: completion of flight hardware,”
Proc. SPIE, 7732 77321P
(2010). https://doi.org/10.1117/12.857034 PSISDG 0277-786X Google Scholar
D. Barret et al.,
“The ATHENA X-ray Integral Field Unit (X-IFU),”
Proc. SPIE, 10699 106991G
(2018). https://doi.org/10.1117/12.2312409 PSISDG 0277-786X Google Scholar
F. Pajot et al.,
“The Athena X-ray Integral Field Unit (X-IFU),”
J. Low Temp. Phys., 193 901
–907
(2018). https://doi.org/10.1007/s10909-018-1904-5 JLTPAC 0022-2291 Google Scholar
M. Barbera et al.,
“Baseline design of the thermal blocking filters for the X-IFU detector on board ATHENA,”
Proc. SPIE, 9144 91445U
(2014). https://doi.org/10.1117/12.2057403 PSISDG 0277-786X Google Scholar
M. Barbera et al.,
“ATHENA X-IFU thermal filters development status toward the end of the instrument phase-A,”
Proc. SPIE, 10699 106991R
(2018). https://doi.org/10.1117/12.2314450 PSISDG 0277-786X Google Scholar
S. R. Bandler et al.,
“The Lynx x-ray microcalorimeter—LXM.,”
J. Astron. Telesc. Instrum. Syst.,
(2019). Google Scholar
L. Sciortino et al.,
“Surface investigation and aluminum oxide estimation on test filters for the ATHENA X-IFU and WFI detectors,”
Proc SPIE, 9905 990566
(2016). https://doi.org/10.1117/12.2232376 Google Scholar
B. L. Henke, E. M. Gullikson and J. C. Davis,
“X-ray interactions: photoabsorption, scattering, transmission, and reflection at E = 50–30, 000 eV, Z = 1–92,”
At. Data Nucl. Data Tables, 54 181
–342
(1993). https://doi.org/10.1006/adnd.1993.1013 Google Scholar
B. Zeiger, H. Lopez and T. Ayers,
“Advances in optical blocking filters for the Lynx x-ray microcalorimeter,”
(2019). Google Scholar
E. Wollack et al.,
“Technology development of advanced filter design for the Lynx x-ray microcalorimeter,”
(2019). Google Scholar
C. P. de Vries et al.,
“Calibration sources and filters of the soft x-ray spectrometer instrument on the Hitomi spacecraft,”
J. Astron. Telesc. Instrum. Syst., 4 011204
(2018). https://doi.org/10.1117/1.JATIS.4.1.011204 Google Scholar
C. P. de Vries et al.,
“Simulating modulated x-ray calibration sources for future x-ray missions using GEANT4,”
Proc. SPIE, 10699 1069965
(2018). https://doi.org/10.1117/12.2305475 PSISDG 0277-786X Google Scholar
BiographyMegan E. Eckart is a research scientist in the Astrophysics and Advanced Diagnostics Group at the Lawrence Livermore National Laboratory. |