1.IntroductionBarrett’s esophagus is an acquired condition that predisposes patients to esophageal adenocarcinoma (EAC), an aggressive cancer with 5-year survival rates as low as 20%.1 Progression to cancer occurs through intermediate stages of low-grade dysplasia (LGD) and high-grade dysplasia (HGD).2 While the cancer risk in Barrett’s patients is typically less than 0.5% per year, the risk of progression may be up to 10% in the LGD that emerges within regions of Barrett’s.3,4 Early resection or radiofrequency ablation of these dysplastic regions can, however, be curative,5,6 which has motivated routine clinical surveillance of Barrett’s patients.7 Current clinical surveillance uses a combination of white light endoscopy (WLE) with targeted and random biopsies. Random biopsies are taken in each quadrant every 2-cm along the length of the Barrett’s segment according to the Seattle protocol.8 These random biopsies are time consuming, labor intensive, costly, and subject to high sampling error as a result of the patchy distribution of dysplastic regions.9 To reduce sampling error, targeted biopsies are guided by WLE, but poor contrast between dysplastic and Barrett’s tissues results in a sensitivity as low as 40% for dysplasia.10,11 To overcome this limitation, advanced wide-field surveillance techniques with improved detection of dysplastic regions and thus more accurate targeting of biopsies have been developed. These include: narrow band imaging;12–17 autofluorescence imaging;18–23 chromoendoscopy;24–27 and optical coherence tomography.28–30 Despite these recent advances, none has yet been recommended for use in routine clinical practice due to lack of robust evidence on the improved detection of dysplasia by their targeting of biopsies.7,9 Molecular imaging enables the visualization of complex biochemical processes involved in both normal physiology and disease states. The use of molecular imaging has the potential to achieve both high sensitivity and specificity for identification of dysplasia.31 Previous efforts in the development of molecular imaging contrast agents for this application have been limited by long clearance times of probes after intravenous administration, requiring at least 24 h between injection and endoscopy.32,33 Encouragingly, labels that can be topically applied during the endoscopic procedure have begun to emerge. A fluorescently labeled peptide that binds specifically to HGD and adenocarcinoma has recently been shown by Sturm et al.34 to delineate these pathologies in vivo in patients; however, the sensitivity to LGD has yet to be established. A follow-up study by Joshi et al.35 using a wide-field fluorescence imaging endoscope noted some important limitations, including: the challenges in visualizing the FITC fluorophore due to strong tissue autofluorescence; and the resulting low values of target-to-background ratio from diagnostic fluorescence features. We have previously demonstrated that cell-surface glycans are altered in the progression from Barrett’s esophagus to adenocarcinoma. The fluorescently labeled lectin wheat germ agglutinin (WGA) showed specific changes in binding patterns between HGD and LGD tissue and nondysplastic (ND) Barrett’s esophagus.36 The low-molecular weight fluorescent WGA showed potential for more accurate delineation of dysplasia than WLE and importantly for clinical translation, could be applied topically using a spray catheter and was easily displaced by washing with an excess of glucosamine.36 Progression of this and other promising molecular imaging agents into clinical trials, however, has encountered technological limitations. Existing endoscopy devices for detecting fluorescence emissions are optimized for detection of tissue autofluorescence in the visible wavelength range, meaning that the sensitivity, contrast, and dynamic range for detection of exogenously applied contrast agents is limited by the high endogenous background signal present within the images. To avoid the autofluorescent background, imaging can be performed using near-infrared (NIR) fluorophores in the range to 900 nm, where the tissue is relatively devoid of endogenous fluorescence.37 This has motivated several groups to develop flexible endoscopic devices for NIR fluorescence imaging in the GI tract, since none are commercially available at present.32,38–40 Pilot studies of such devices have been performed40,41 and several clinical trials are in progress (NCT01972373, NCT02113202, NCT02129933, and NCT02710838). Still, data supporting the use of flexible NIR fluorescence endoscopes in clinical trials for wide-field surveillance of Barrett’s esophagus are lacking. The overall goal of our work is to translate wide-field NIR fluorescence molecular imaging to improve the detection sensitivity and specificity for both low- and high-grade dysplastic lesions in Barrett’s esophagus. Toward this goal, we have labeled WGA with a NIR dye (IR800CW) to create WGA-IR800CW and developed a clinically translatable bimodal endoscope to detect both NIR fluorescence for molecular imaging as well as traditional white light reflectance. Our solution is based on an existing clinically approved accessory channel endoscope to ease the translational pathway. Here, we present the design and development of our bimodal endoscopic system, the technical characterization process undertaken along with preliminary biological validation and discuss the next steps required to achieve routine clinical application. 2.Materials and MethodsAll chemicals were purchased from Sigma-Aldrich unless otherwise stated. 2.1.Fluorescent Lectin SynthesisIRDye800CW NHS Ester was obtained from LI-COR Bioscience. Conjugation of the dye with wheat germ agglutinin (WGA, L9640) was performed according to the methods of Sato et al.42 to produce WGA-IR800CW. 2.2.Endoscope DesignA schematic of the endoscopic system is shown in Fig. 1. The system is based around an imaging fiber-optic bundle consisting of 10,000 individual fibers (PolyDiagnost). This is threaded inside a disposable catheter that protects the distal tip of the imaging fiber bundle from direct patient contact with a glass cover plane (PolyDiagnost). The catheter also contains a light guide for illumination and a 1.2 mm working channel. The combined imaging fiber bundle and catheter is known as the PolyScope flexible endoscope. The PolyScope may be used independently or introduced through the accessory channel of another endoscope. Designing our device around the PolyScope system offers several advantages. First, the lack of direct patient contact removes the need for sterilization of the imaging fiber between procedures, prolonging the lifetime of this expensive component of the system. Second, the system carries a CE mark for endoscopic retrograde cholangio-pancreatography, which will facilitate safety approval for use in the esophagus. Third, the catheter has a maximum diameter of 3.0 mm allowing it to be threaded into a 3.7 mm working channel of a commercial endoscope for direct comparison to gold standard white light reflectance imaging performed at the tip of the endoscope. Illumination is provided by a broadband halogen light source (OSL2B2, Thorlabs) clipped with a 750 nm short pass excitation filter (FESH0750, Thorlabs). This is directed onto the sample through the light guide embedded inside the PolyScope catheter. Light collected by the imaging fiber is relayed through an infinity corrected objective lens (421350-9970-000, Zeiss) and split into two channels with a long pass dichroic filter (DMLP650L, Thorlabs). Light from 400 to 633 nm is relayed by an achromatic doublet (ACA254-100-A, Thorlabs) onto a grayscale, complementary metal–oxide–semiconductor sensor (CMOS; Grasshopper3 GS3-U3-41C6M-C, PointGrey) while light is relayed by an achromatic doublet (ACA254-100-B, Thorlabs) onto an electron multiplying charge coupled device (EMCCD; ProEM+ eXcelon , Princeton) through an 800 nm long pass emission filter (FELH0800, Thorlabs), allowing simultaneous WL reflectance and NIR fluorescence imaging. The system is kept compact through the use of broadband mirrors (BB1-E02 and BB1-E03, Thorlabs) to bend the light path. The optics are securely housed inside a light-tight enclosure and mounted on an optical breadboard (MB4545/M, Thorlabs) that is fixed to a stainless steel trolley (FW2901-3, Freeway Medical) with a footprint of . This allows the system to be easily and safely transported around a crowded endoscopy suite. 2.3.Image Acquisition and CorrectionsA LabView (National Instruments) visual interface is used to control the cameras and acquire images. The white light images are saved as 8 bit tiff files and the NIR images are saved as 16 bit tiff files. Data analysis was carried out using MATLAB® (MathWorks). An overall schematic of the image corrections performed can be seen in Fig. 2. Since the imaging fiber bundle is composed of 10,000 individual fibers, a honeycomb structure appears in the images. To remove this, we first acquire a bright field calibration image from an externally illuminated diffusely reflecting surface (paper). Next, we identify the locations of each individual fiber within these bright field images using an algorithm adapted from Elter et al.43 The procedure is as follows:
Following identification of fiber centers, we reconstruct an image free of honeycomb artifact by calculating a bilinear interpolation of pixel values at fiber centers. Once this procedure is complete, we can define a lookup table that can be used to directly convert raw images into interpolated images without further calculation. This must be updated only if the system is realigned, which allows us to perform the correction in real time. The white light and NIR images can be coregistered using the fiber locations by selecting the four fibers located at the extreme and positions. The images are cropped at these locations and a similarity transformation is generated to successfully transform the white light image in and such that the extreme fibers are coregistered with the equivalent extreme fibers in the NIR image. 2.4.Optical Characterization2.4.1.Field of viewTo enable wide-field surveillance of the whole esophagus it is important that the endoscope has sufficient imaging performance across a wide field of view (FOV). In order to measure the FOV we captured 3 images of 1 mm graph paper at four different working distances (WD): 5.7, 8.7, 11.7, and 16.7 mm (). The resulting images will show a barrel distortion defined by where is the radial distance from the center of the ground truth image to a given vertex in mm, is the radial distance from the center of the distorted image to the same vertex in pixels, is a constant that describes the magnitude of the distortion, and is a constant used to convert between units of pixels and mm. For each of the 12 images acquired (four working distances × three replicates) we identified the position of the center of the fiber bundle and plotted the radial distance to each vertex in the graph paper, (in pixels), against the true distance, (in mm), which is known from the markings on the graph paper.2.4.2.PowerTo establish a maximum power for our system, we measured the maximum broadband power of a clinical endoscope (Olympus Evis Lucera CVL-260SL Xenon light source with an Olympus Gastroscope GIF-FQ260Z) and of our endoscope in normal operating conditions, at the same working distance, using a thermal power meter (A-02-D12-BBF-USB, LaserPoint). 2.4.3.ResolutionThe resolution of the system is fundamentally limited by the fiber bundle structure. To determine the limiting resolution, we took images of a 1951 USAF resolution test target (#53-714, Edmund Optics) at 4 working distances using external illumination from a broadband halogen light source (OSL2B2, Thorlabs). Background subtraction was performed using an averaged dark image of 10 frames acquired at the end of the experiment. White light images were analyzed since the test target is printed on white reflective photo paper intended for visible imaging. Since both imaging channels are able to resolve the individual fibers of the fiber bundle, the limiting resolution is determined by the fiber bundle properties rather than the properties of the cameras, and so the resolution measured using the white light images is applicable to both imaging channels. 2.4.4.SensitivityFor in vivo application it is important that we are able to detect WGA-IR800CW at low concentrations. The reasons for this are threefold. First, since binding of WGA-IR800CW decreases with the progression of the disease, we must be capable of detecting it at low concentrations to avoid false positives. Second, although WGA occurs naturally in food36 and IR800CW can be produced under cGMP conditions, minimizing the amount of any exogenous agent sprayed inside the esophagus is desirable for clinical translation. Third, we would like to minimize nonspecific binding effects that may become most problematic at high concentrations where they reduce our contrast. Therefore, determining the sensitivity of the system is a crucial aspect of the characterization process. In order to characterize sensitivity, we prepared a two-fold dilution series of WGA-IR800CW in phosphate buffered saline (PBS) and pipetted of each solution into a well plate (-Slide 18 Well—Flat, ibidi GmbH), which had been spray-painted matte black to avoid specular reflections. We then captured images of the dye at 5 different working distances (5, 10, 12, 16, and 20 mm) representative of the range that would be used in vivo. The NIR images were captured using an EM gain of 20 and an exposure time of 200 ms. The images were corrected and coregistered as described in Sec. 2.3. regions of interest (ROIs) were drawn on the white light images inside (signal) and outside (background) of the well. These ROIs were then applied to the coregistered NIR images for analysis in MATLAB®. The signal-to-noise ratio (SNR) was calculated using where is the mean of a Gaussian fitted to the pixel intensity distribution in the signal ROI, is the mean of a Gaussian fitted to the pixel intensity distribution in the background ROI, and is the standard deviation of a Gaussian fitted to the pixel intensity distribution in the background ROI.We also characterized the relationship between the SNR and electron multiplying (EM) gain of the NIR sensor by capturing images of of 3100 nM WGA-IR800CW with different levels of gain at working distances of 5, 9.5, and 12 mm. For an EMCCD we would expect with the EM gain of the sensor.442.5.Biological Validation Studies in Ex Vivo Tissue2.5.1.Imaging of mouse stomachsTo demonstrate the feasibility of imaging WGA-IR800CW on a tissue background we stained excised mouse stomachs with the fluorescently labeled lectin and acquired imaging data using our endoscope. All animal procedures were conducted in accordance with project and personal licenses issued under the United Kingdom Animals (Scientific Procedures) Act, 1986. Surplus mice from breeding were obtained post mortem (). Mouse stomachs were prepared by opening and pinning the excised stomachs on parafilm covered cork. The stomachs were washed with PBS to remove contents before the first round of imaging. This washing consisted of tilting the stomachs and pouring the PBS such that it ran from left to right over the stomachs to avoid cross contamination between the tissue types. Data acquisition was then performed using EM and . After the baseline imaging, of (780 nM) WGA-IR800CW was evenly pipetted onto the face of the stomachs and left to incubate for 10 min. The stomachs were then imaged four more times: immediately after incubation; after a wash with 5 mL of PBS; after a further wash of 15 mL of PBS; and after a final wash of 45 mL PBS. Each round of imaging captured three images: squamous tissue; gastric tissue; and parafilm covered cork to serve as a background control. For each image, an ROI was drawn in the center of the image and the mean signal calculated by fitting a Gaussian to the pixel intensity distribution. In order to confirm the differential binding of WGA to squamous and gastric tissue types rather than tissue geometry, we repeated the experiment using an additional wash with 33 ml of 1 mM -acetyl-D-glucosamine (Sigma-Aldrich) in PBS, which has been shown previously to compete with WGA for sialic acid binding.36 The washing was performed in the same manner as described above. Excess glucosamine that did not run off was left to incubate for 5 min before a final wash of 50 mL of PBS. 2.5.2.Imaging of human resection specimensIn order to validate our system in human tissue we performed a pilot study assessing five endoscopic mucosal resection (EMR) specimens from a patient undergoing endoscopic therapy for Barrett’s-related intramucosal EAC. The study was approved by the Cambridgeshire 2 Research Ethics Committee (09/H0308/118). Resections were first washed with PBS to remove superficial debris and excess mucus then stained with (268 nM) WGA-IR800CW using a small spray bottle to mimic the spray catheter application that would be used in an endoscopic practice. Tissues were left to incubate for 10 min at room temperature then washed with 15 mL of cold PBS to remove unbound fluorescent probe. The EMRs were imaged with a Fluobeam® 800 intraoperative wide-field fluorescence imaging system (Fluooptics, France), taken as the gold standard for imaging of NIR fluorescence, and then with our system. To fit the entire EMRs ( diameter) into a single field of view, we captured images at a working distance of using our endoscope, resulting in a decrease in illumination power and resulting SNR. To counter this, we increased our exposure time to 2 s, allowing us to achieve adequate SNR. In order to compare our endoscope images with the gold standard Fluobeam® images, we performed a simple coregistration of the images. We first manually drew ROIs around the EMR in the Fluobeam® image and the white light endoscope image (which is already coregistered with the NIR endoscope image as described in Sec. 2.3). Using these ROIs, we generated binary masks, which were then coregistered using the “imregtform” function in MATLAB®. This generated a similarity transformation that was applied to the raw Fluobeam® image to coregister it with the endoscope images. Finally, using both the endoscope mask and the coregistered Fluobeam® mask, the coregistered images were masked so that only image points inside both of the initial ROIs would be compared in the final coregistered images. This process is summarized in Fig. 3. After NIR imaging, EMR specimens were fixed in formalin and embedded in paraffin, according to standard histopathological procedures at Cambridge University Hospitals Human Research Tissue Bank. The EMR paraffin block was cut at intervals of 2 mm and sections mounted onto glass slides. Slides were H&E stained and scanned using an Axio Scan.Z1 and imported into Zen 2 lite software (both Carl Zeiss Microscopy, Jena). The EMR sections were scored by the study pathologist every 1 mm for pathological grade according to the Vienna classification, with the help of the ZEN software graphical grid. This allowed construction of a histology grid, which was superimposed manually onto FluoBeam® fluorescence images using methodology established previously.45 To facilitate histopathological correlation, the normal oesophageal squamous epithelium and oesophageal gastric/intestinal metaplasia (ND Barrett’s) were grouped together as “ND” and any grade of neoplasia, including indefinite for dysplasia (ID), LGD and HGD, and intramucosal cancer, were grouped together as “dysplastic.” To illustrate the relationship between the NIR fluorescence data and underlying pathology, we used EMR B (see Sec. 3.2.2) as there were similarly sized areas of ND and dysplastic tissue within this sample. 3.Results3.1.Optical Characterization3.1.1.Field of viewWe first assessed the FOV of the system. An example image displaying barrel distortion can be seen in Fig. 4(a). The distortion constant and the constant were determined by fitting Eq. 4 to the data shown in Fig. 4(b) (). The values of and were then used to determine the FOV () based on the diameter of the images in pixels (). Combining these data for the four working distances, we determined the angle of the FOV to be [Fig. 4(c)], which compares favorably to the manufacturer specified angle of 70 deg. 3.1.2.PowerTo establish a safe maximum power for our endoscope, we compared its maximum broadband power to that of a clinical endoscope. The maximum power from the clinical endoscope in white light reflectance mode was measured to be at a working distance of while the maximum power from our system was measured to be at a working distance of . 3.1.3.ResolutionResolution was determined by taking images of a USAF test target. At least four replicate data points for each line pair element were captured (centered within the FOV) for each of four working distances, 5, 10, 15, and 20 mm. The Michelson contrast was calculated for each element (before and after honeycomb artifact correction) and the results plotted against half the spatial frequency of the element, which is equal to the reciprocal of the width of a single line element (Fig. 5). Using a model of the modulation threshold,46 which has been shown to reproduce the results of several studies, we determined 1% to be the minimum contrast required for detection of the pattern by an observer viewing the images in normal operating conditions. By finding the intersect of this threshold with linear fits applied in the low frequency region, we determined the resolution of our system to be , , , and at working distances of 5, 10, 15, and 20 mm, respectively. These values are not significantly altered if we do not perform honeycomb correction, being , , , and , respectively. 3.1.4.SensitivityFinally, we established the system sensitivity by investigating the dependence of SNR on concentration; a SNR of 3 was used to determine the minimum detectable concentration at each working distance. Example images from the white light and NIR channels are shown in Fig. 6(a) while Fig. 6(b) shows the relationship between SNR and concentration at working distances of 5, 10, and 12 mm. At working distances beyond 16 mm, the dye was not detectable at any concentration . A linear fit is made to these curves for since below this we expect only noise in our data. For a working distance of 5 mm, the nonlinear region at high concentration, which is possibly due to saturation, was excluded from the linear fit. The minimum detectable concentrations assessed from Fig. 6(b) are and for working distances of 5 and 10 mm, respectively. Without the honeycomb correction outlined in Sec. 2.3 the minimum detectable concentrations are and for working distances of 5 and 10 mm, respectively, suggesting removal of the honeycomb artifact grants us an improvement in SNR. We also assessed the influence of working distance and EMCCD gain on these data. The relationship between sensitivity and working distance arises due to the dependence on the irradiance of the excitation light. This was investigated by capturing images of of 3100 nM WGA-IR800CW at eight working distances and plotting SNR against working distance [Fig. 6(c)]. The fit suggests that the SNR falls off as working distance to the power of , consistent with the inverse square law expected for an illumination cone. Figure 6(d) shows the fit of our data to Eq. (6) to assess the relationship between SNR and EMCCD gain. According to our data, using a gain of provides optimal SNR while preserving the lifetime of our EMCCD. 3.2.Biological Validation Studies in Ex Vivo Tissue3.2.1.Imaging of mouse stomachsTo demonstrate the feasibility of imaging WGA-IR800CW on a tissue background we stained excised mouse stomachs with the dye and acquired data using our endoscope. The upper nonglandular fore stomach, which has squamous tissue at the exposed surface, and the lower glandular stomach, which has simple columnar epithelium (gastric type) tissue at the exposed surface, provide us with a model of the corresponding tissue types found in the human esophagus in healthy (squamous) and diseased columnar-lined esophagus or Barrett’s, respectively. The different regions of tissue within the stomach are indicated in Figs. 7(a) and 7(b). Gastric tissue is clearly distinguishable (2 way ANOVA, ) from the squamous tissue using our system [Fig. 7(c)], even following extensive washing with PBS, suggesting that the WGA-IR800CW binds preferentially to this mouse tissue. The results shown in Fig. 7(d) show a significant reduction of the fluorescence to background level (2 way ANOVA, ) following the glucosamine wash, confirming our results are due to the differential binding of WGA to different tissue types. 3.2.2.Imaging of human resection specimensIn a first step towards clinical translation, we performed a pilot study on endoscopic mucosal resections (EMRs) collected from a patient with Barrett’s esophagus. The coregistered Fluobeam® and endoscope EMR images for all collected specimens are shown in Figs. 8(a) and 8(b), respectively. The high signal observed at the edge of the specimens is due to pooling of the dye between the edge of the tissue specimen and the underlying parafilm, which remains even after washing. The intensity recorded in each coregistered pixel in the Fluobeam® and endoscope images was compared to determine if there was correlation between the images and hence whether the data acquired with our endoscope faithfully recapitulates that acquired with the gold standard Fluobeam® system. Intensity scatter plots of these data clearly show a direct relationship between fluorescence intensity in the Fluobeam® and endoscope images for 4 of the 5 EMRs [Fig. 8(c)]. Spearman’s rank correlation coefficient reveals a moderate but significant correlation between the Fluobeam® and endoscope images in these 4 of 5 EMRs (). The fluorescence signal in central 80% of the EMRs within Fluobeam® images, given as , was found to be , , , , AU for EMRs A to E, respectively. The lack of significant correlation in EMR D is therefore likely due to the relatively uniform fluorescence observed in both Fluobeam® and endoscope images. Using EMR B, which had two large regions of ND and dysplastic tissue (see Sec. 2.5.2), we then coregistered the NIR endoscopy data to the histology grid [Fig. 9(a)]. The expected negative binding relationship between dysplastic and ND tissue was confirmed [Fig. 9(b)]. These results provide a preliminary indication that WGA-IR800CW fluorescence imaging with our endoscope would be capable of distinguishing between disease pathologies in esophageal tissue. 4.DiscussionThe ability to detect areas of dysplasia in patients with Barrett’s esophagus is a significant clinical challenge. The current standard of care, including WLE with random and targeted biopsies, suffers significant limitations due to the poor contrast available for dysplastic tissue in regions of Barrett’s. If it were possible to accurately detect these lesions, they could be resected or ablated, preventing the progression of disease to EAC, for which prognosis is bleak with 5-year survival rates reported as low as 20%.1 To address this clinical challenge, we combined a NIR fluorescently labeled wheat germ agglutinin (WGA-IR800CW) that shows differential binding to normal, Barrett’s, dysplastic, and cancerous tissue36 with a bimodal endoscope capable of acquisition and coregistration of white light reflectance and NIR fluorescence signals in real time. We show here that our system is able to detect WGA-IR800CW at concentrations as low as 110 nM at the shortest working distances with a field of view of 63 deg and an image resolution of . When applied to image WGA-IR800CW in a realistic tissue scenario using ex vivo mouse stomachs, we were able to discriminate gastric- from squamous-type tissue. We then demonstrated, in a pilot study, the performance of the system as applied to endoscopic mucosal resection specimens from a patient with Barrett’s esophagus and compared our data to a gold standard system for detecting NIR fluorescence. These results showed an encouraging correlation between fluorescence signal intensities recorded with the two systems on a per-pixel basis, relating directly to histopathological outcome.45 While these results are promising, some instrumentation challenges remain for clinical translation. In this work, we were imaging ex vivo specimens on a flat surface. Optimal sensitivity in this geometry was found for the shortest WDs in the range found in vivo as expected. Clinical endoscopy may encompass WDs up to several centimeters within a single image. If we wish to perform quantitative fluorescence imaging in vivo, we will need to correct for the variable working distance within a given image. This may be achieved by constructing a pixel-wise ratiometric image of a fluorescence image and a coregistered white light reflectance image, which has shown to perform well in fluorescence phantoms.35 In addition, a number of developments could be included to ensure that we achieve acceptable SNR at a wider range of WDs. We could employ a more tailored light source, such that more spectral power is deposited across the excitation band of the IRDye800CW, or increase the overall illumination power of the system given that we remain significantly below the power used in current commercial endoscopes. Improved sensitivity would also allow us to improve our FOV to approach the 140 deg FOVs that are common in commercial endoscopes. We will also need to overcome coregistration challenges to relate our fluorescence data to the “ground truth” histopathological analysis in future studies of ex vivo specimens. Inaccuracies in the histology grid registration may arise due to: unavoidable deformations and artifacts in the processing of EMRs; and the assignment of a single majority pathological grade to a large 1 mm area, which may contain mixed pathologies. The manual coregistration of the grid to the fluorescence image then adds a further challenge. Application in a clinical setting in vivo also requires an optimization of the dye concentration to be used with our system. The sensitivity assays performed here used solutions of the dye in well plates. An important next step will be to calibrate how the sprayed concentration of WGA-IR800CW relates to the final bound concentration observed on the tissue following application and washing with a spray catheter. A quantitative assessment of the applied concentration of dye and specific/nonspecific binding against a known concentration of sialic acid residues will allow us determine the optimum spraying concentration for use in future clinical trials. Our approach has several advantages over other molecular imaging methods being explored for NIR endoscopy. First, lectins are cheap, stable at low pH, heat stable,47 and have a known, validated molecular target that has been shown to be down regulated in the progression to EAC.36 Furthermore, these lectins can be topically applied with a short incubation time, minimizing the length of the procedure compared to, for example, intravenous administration of antibodies, which requires a long clearance time. Second, the use of a wide-field bimodal fluorescence endoscope allows this approach to image the entire esophagus for improved endoscopic surveillance. We therefore believe that our approach will facilitate rapid translation into future clinical trials. 5.ConclusionsWe have developed a bimodal white light reflectance and NIR fluorescence endoscope that provides wide-field surveillance for dysplastic tissue in patients with Barrett’s esophagus using the molecular imaging contrast agent WGA-IR800CW. Our endoscopy system could distinguish between squamous and gastric tissue types in mouse stomachs ex vivo and compared favorably to gold standard measurements when applied to ex vivo endoscopic mucosal resections from patients. Based on these data, we can move forward with clinical translation of this device, performing further optimization of the instrument hardware and image processing software based on our findings, prior to implementation in clinical trials. AcknowledgmentsThis work was funded by the EPSRC-CRUK Cancer Imaging Centre in Cambridge and Manchester (Grant no. C197/A16465), CRUK (C14303/A17197) and the EU FP7 agreement FP7-PEOPLE-2013-CIG-630729. Additional funds were provided by the University of Cambridge MRC Confidence in Concept Award. SEB is supported by a CRUK Career Establishment Award (Grant no. C47594/A16267). ReferencesC. Hur et al.,
“Trends in esophageal adenocarcinoma incidence and mortality,”
Cancer, 119
(6), 1149
–1158
(2013). http://dx.doi.org/10.1002/cncr.27834 CANCAR 0008-543X Google Scholar
R. Anaparthy and P. Sharma,
“Progression of Barrett oesophagus: role of endoscopic and histological predictors,”
Nat. Rev. Gastroenterol. Hepatol., 11
(9), 525
–534
(2014). http://dx.doi.org/10.1038/nrgastro.2014.69 Google Scholar
T. K. Desai et al.,
“The incidence of oesophageal adenocarcinoma in non-dysplastic Barrett’s oesophagus: a meta-analysis,”
Gut, 61
(7), 970
–976
(2012). http://dx.doi.org/10.1136/gutjnl-2011-300730 GUTTAK 0017-5749 Google Scholar
L. C. Duits et al.,
“Barrett’s oesophagus patients with low-grade dysplasia can be accurately risk-stratified after histological review by an expert pathology panel,”
Gut, 64
(5), 700
–706
(2015). http://dx.doi.org/10.1136/gutjnl-2014-307278 GUTTAK 0017-5749 Google Scholar
N. J. Shaheen et al.,
“Radiofrequency ablation in Barrett’s esophagus with dysplasia,”
N. Engl. J. Med., 360
(22), 2277
–2288
(2009). http://dx.doi.org/10.1056/NEJMoa0808145 NEJMAG 0028-4793 Google Scholar
K. N. Phoa et al.,
“Multimodality endoscopic eradication for neoplastic Barrett oesophagus: results of an European multicentre study (EURO-II),”
Gut, 65
(4), 555
–562
(2015). http://dx.doi.org/10.1136/gutjnl-2015-309298 GUTTAK 0017-5749 Google Scholar
R. C. Fitzgerald et al.,
“British Society of Gastroenterology guidelines on the diagnosis and management of Barrett’s oesophagus,”
Gut, 63
(1), 7
–42
(2014). http://dx.doi.org/10.1136/gutjnl-2013-305372 GUTTAK 0017-5749 Google Scholar
D. S. Levine et al.,
“Safety of a systematic endoscopic biopsy protocol in patients with Barrett’s esophagus,”
Am. J. Gastroenterol., 95
(5), 1152
–1157
(2000). http://dx.doi.org/10.1111/j.1572-0241.2000.02002.x Google Scholar
S. J. Spechler et al.,
“American Gastroenterological Association technical review on the management of Barrett’s esophagus,”
Gastroenterology, 140
(3), e18
–e52
(2011). http://dx.doi.org/10.1053/j.gastro.2011.01.031 GASTAB 0016-5085 Google Scholar
M. B. Sturm and T. D. Wang,
“Emerging optical methods for surveillance of Barrett’s oesophagus,”
Gut, 64
(11), 1816
–1823
(2015). http://dx.doi.org/10.1136/gutjnl-2013-306706 GUTTAK 0017-5749 Google Scholar
M. I. Canto et al.,
“In vivo endomicroscopy improves detection of Barrett’s esophagus-related neoplasia: a multicenter international randomized controlled trial (with video),”
Gastrointest. Endoscopy, 79
(2), 211
–221
(2014). http://dx.doi.org/10.1016/j.gie.2013.09.020 Google Scholar
J. Song et al.,
“Meta-analysis of the effects of endoscopy with narrow band imaging in detecting dysplasia in Barrett’s esophagus,”
Dis. Esophagus, 28
(6), 560
–566
(2014). http://dx.doi.org/10.1111/dote.12222 Google Scholar
P. Sharma et al.,
“Standard endoscopy with random biopsies versus narrow band imaging targeted biopsies in Barrett’s oesophagus: a prospective, international, randomised controlled trial,”
Gut, 62
(1), 15
–21
(2013). http://dx.doi.org/10.1136/gutjnl-2011-300962 GUTTAK 0017-5749 Google Scholar
H. C. Wolfsen et al.,
“Prospective, controlled tandem endoscopy study of narrow band imaging for dysplasia detection in Barrett’s esophagus,”
Gastroenterology, 135
(1), 24
–31
(2008). http://dx.doi.org/10.1053/j.gastro.2008.03.019 GASTAB 0016-5085 Google Scholar
C. Jayasekera et al.,
“Added value of narrow band imaging and confocal laser endomicroscopy in detecting Barrett’s esophagus neoplasia,”
Endoscopy, 44
(12), 1089
–1095
(2012). http://dx.doi.org/10.1055/s-0032-1325734 ENDCAM Google Scholar
M. Singh et al.,
“Observer agreement in the assessment of narrowband imaging system surface patterns in Barrett’s esophagus: a multicenter study,”
Endoscopy, 43
(9), 745
–751
(2011). http://dx.doi.org/10.1055/s-0030-1256631 ENDCAM Google Scholar
R. Singh et al.,
“Narrow-band imaging with magnification in Barrett’s esophagus: validation of a simplified grading system of mucosal morphology patterns against histology,”
Endoscopy, 40
(6), 457
–463
(2008). http://dx.doi.org/10.1055/s-2007-995741 ENDCAM Google Scholar
D. F. Boerwinkel et al.,
“Third-generation autofluorescence endoscopy for the detection of early neoplasia in Barrett’s esophagus: a pilot study,”
Dis. Esophagus, 27
(3), 276
–284
(2014). http://dx.doi.org/10.1111/dote.12094 Google Scholar
L. M. Wong Kee Song et al.,
“Autofluorescence imaging,”
Gastrointest. Endoscopy, 73
(4), 647
–650
(2011). http://dx.doi.org/10.1016/j.gie.2010.11.006 Google Scholar
M. Giacchino et al.,
“Clinical utility and interobserver agreement of autofluorescence imaging and magnification narrow-band imaging for the evaluation of Barrett’s esophagus: a prospective tandem study,”
Gastrointest. Endoscopy, 77
(5), 711
–718
(2013). http://dx.doi.org/10.1016/j.gie.2013.01.029 Google Scholar
D. F. Boerwinkel et al.,
“Effects of autofluorescence imaging on detection and treatment of early neoplasia in patients with Barrett’s esophagus,”
Clin. Gastroenterol. Hepatol., 12
(5), 774
–781
(2014). http://dx.doi.org/10.1016/j.cgh.2013.10.013 Google Scholar
W. L. Curvers et al.,
“Endoscopic tri-modal imaging is more effective than standard endoscopy in identifying early-stage neoplasia in Barrett’s esophagus,”
Gastroenterology, 139
(4), 1106
–1114
(2010). http://dx.doi.org/10.1053/j.gastro.2010.06.045 GASTAB 0016-5085 Google Scholar
W. L. Curvers et al.,
“Endoscopic trimodal imaging versus standard video endoscopy for detection of early Barrett’s neoplasia: a multicenter, randomized, crossover study in general practice,”
Gastrointest. Endoscopy, 73
(2), 195
–203
(2011). http://dx.doi.org/10.1016/j.gie.2010.10.014 Google Scholar
J. D. Horwhat et al.,
“A randomized comparison of methylene blue-directed biopsy versus conventional four-quadrant biopsy for the detection of intestinal metaplasia and dysplasia in patients with long-segment Barrett’s esophagus,”
Am. J. Gastroenterol., 103
(3), 546
–554
(2008). http://dx.doi.org/10.1111/j.1572-0241.2007.01601.x Google Scholar
S. Ngamruengphong, V. K. Sharma and A. Das,
“Diagnostic yield of methylene blue chromoendoscopy for detecting specialized intestinal metaplasia and dysplasia in Barrett’s esophagus: a meta-analysis,”
Gastrointest. Endoscopy, 69
(6), 1021
–1028
(2009). http://dx.doi.org/10.1016/j.gie.2008.06.056 Google Scholar
S. Tholoor et al.,
“Acetic acid chromoendoscopy in Barrett’s esophagus surveillance is superior to the standardized random biopsy protocol: results from a large cohort study (with video),”
Gastrointest. Endoscopy, 80
(3), 417
–424
(2014). http://dx.doi.org/10.1016/j.gie.2014.01.041 Google Scholar
G. Longcroft-Wheaton et al.,
“Acetic acid spray is an effective tool for the endoscopic detection of neoplasia in patients with Barrett’s esophagus,”
Clin. Gastroenterol. Hepatol., 8
(10), 843
–847
(2010). http://dx.doi.org/10.1016/j.cgh.2010.06.016 Google Scholar
S. H. Yun et al.,
“Comprehensive volumetric optical microscopy in vivo,”
Nat. Med., 12
(12), 1429
–1433
(2006). http://dx.doi.org/10.1038/nm1450 1078-8956 Google Scholar
H. C. Wolfsen et al.,
“Safety and feasibility of volumetric laser endomicroscopy in patients with Barrett’s esophagus (with videos),”
Gastrointest. Endoscopy, 82
(4), 631
–640
(2015). http://dx.doi.org/10.1016/j.gie.2015.03.1968 Google Scholar
M. J. Gora et al.,
“Tethered capsule endomicroscopy enables less invasive imaging of gastrointestinal tract microstructure,”
Nat. Med., 19
(2), 238
–240
(2013). http://dx.doi.org/10.1038/nm.3052 1078-8956 Google Scholar
M. L. James and S. S. Gambhir,
“A molecular imaging primer: modalities, imaging agents, and applications,”
Physiol. Rev., 92
(2), 897
–965
(2012). http://dx.doi.org/10.1152/physrev.00049.2010 PHREA7 0031-9333 Google Scholar
P. Habibollahi et al.,
“Optical imaging with a cathepsin B activated probe for the enhanced detection of esophageal adenocarcinoma by dual channel fluorescent upper GI endoscopy,”
Theranostics, 2
(2), 227
–234
(2012). http://dx.doi.org/10.7150/thno.4088 Google Scholar
S. Realdon et al.,
“In vivo molecular imaging of HER2 expression in a rat model of Barrett’s esophagus adenocarcinoma,”
Dis. Esophagus, 28
(4), 394
–403
(2015). http://dx.doi.org/10.1111/dote.12210 Google Scholar
M. B. Sturm et al.,
“Targeted imaging of esophageal neoplasia with a fluorescently labeled peptide: first-in-human results,”
Sci. Transl. Med., 5
(184), 184ra61
(2013). http://dx.doi.org/10.1126/scitranslmed.3004733 STMCBQ 1946-6234 Google Scholar
B. P. Joshi et al.,
“Multimodal endoscope can quantify wide-field fluorescence detection of Barrett’s neoplasia,”
Endoscopy, 48
(2), A1
–A13
(2015). http://dx.doi.org/10.1055/s-0034-1392803 ENDCAM Google Scholar
E. L. Bird-Lieberman et al.,
“Molecular imaging using fluorescent lectins permits rapid endoscopic identification of dysplasia in Barrett’s esophagus,”
Nat. Med., 18
(2), 315
–321
(2012). http://dx.doi.org/10.1038/nm.2616 1078-8956 Google Scholar
C. S. von Holstein et al.,
“Detection of adenocarcinoma in Barrett’s oesophagus by means of laser induced fluorescence,”
Gut, 39
(5), 711
–716
(1996). http://dx.doi.org/10.1136/gut.39.5.711 GUTTAK 0017-5749 Google Scholar
J. J. Tjalma et al.,
“Molecular-guided endoscopy targeting vascular endothelial growth factor a for improved colorectal polyp detection,”
J. Nucl. Med., 57
(3), 480
–485
(2016). http://dx.doi.org/10.2967/jnumed.115.166975 JNMEAQ 0161-5505 Google Scholar
P. B. Garcia-Allende et al.,
“Towards clinically translatable NIR fluorescence molecular guidance for colonoscopy,”
Biomed. Opt. Express, 5
(1), 78
–92
(2014). http://dx.doi.org/10.1364/BOE.5.000078 BOEICL 2156-7085 Google Scholar
R. A. Sheth et al.,
“Pilot clinical trial of indocyanine green fluorescence-augmented colonoscopy in high risk patients,”
Gastroenterol. Res. Pract., 2016
(9), 6184842
(2016). http://dx.doi.org/10.1155/2016/6184842 Google Scholar
J. Glatz et al.,
“Near-infrared fluorescence cholangiopancreatoscopy: initial clinical feasibility results,”
Gastrointest. Endoscopy, 79
(4), 664
–668
(2014). http://dx.doi.org/10.1016/j.gie.2013.10.008 Google Scholar
K. Sato et al.,
“Near infrared photoimmunotherapy in the treatment of pleural disseminated NSCLC: preclinical experience,”
Theranostics, 5
(7), 698
–709
(2015). http://dx.doi.org/10.7150/thno.11559 Google Scholar
M. Elter, S. Rupp and C. Winter,
“Physically motivated reconstruction of fiberscopic images,”
in Proc. Int. Conf. Pattern Recognit.,
599
–602
(2006). http://dx.doi.org/10.1109/ICPR.2006.932 Google Scholar
, “Electron-multiplying (EM) gain,”
(2015) http://www.qimaging.com/resources/pdfs/emccd_technote.pdf ( December ). 2015). Google Scholar
M. di Pietro et al.,
“Detection of dysplasia in Barrett’s oesophagus using lectin-based near infra-red molecular imaging: an ex-vivo study on human tissue,”
in Proc. of the British Society of Gastroenterology Meeting,
(2016). Google Scholar
P. G. J. Barten, Contrast Sensitivity of the Human Eye and Its Effects on Image Quality, SPIE Press, Bellingham, Washington
(1999). Google Scholar
Y. Nagata and M. M. Burger,
“Wheat germ agglutinin. Isolation and crystallization,”
J. Biol. Chem., 247
(7), 2248
–2250
(1972). JBCHA3 0021-9258 Google Scholar
Biography |