18 August 2021 Binary photoacoustic tomography for improved vasculature imaging
Author Affiliations +
Abstract

Significance: The proposed binary tomography approach was able to recover the vasculature structures accurately, which could potentially enable the utilization of binary tomography algorithm in scenarios such as therapy monitoring and hemorrhage detection in different organs.

Aim: Photoacoustic tomography (PAT) involves reconstruction of vascular networks having direct implications in cancer research, cardiovascular studies, and neuroimaging. Various methods have been proposed for recovering vascular networks in photoacoustic imaging; however, most methods are two-step (image reconstruction and image segmentation) in nature. We propose a binary PAT approach wherein direct reconstruction of vascular network from the acquired photoacoustic sinogram data is plausible.

Approach: Binary tomography approach relies on solving a dual-optimization problem to reconstruct images with every pixel resulting in a binary outcome (i.e., either background or the absorber). Further, the binary tomography approach was compared against backprojection, Tikhonov regularization, and sparse recovery-based schemes.

Results: Numerical simulations, physical phantom experiment, and in-vivo rat brain vasculature data were used to compare the performance of different algorithms. The results indicate that the binary tomography approach improved the vasculature recovery by 10% using in-silico data with respect to the Dice similarity coefficient against the other reconstruction methods.

Conclusion: The proposed algorithm demonstrates superior vasculature recovery with limited data both visually and based on quantitative image metrics.

1.

Introduction

Photoacoustic tomography (PAT) is an imaging technique that enables us to visualize vasculature networks at high resolution at depths much greater than optical microscopy.1,2 PAT can also resolve tissue chromophores such as oxyhemoglobin and deoxyhemoglobin, which allows estimation of tissue oxygentation levels.1,3 Accurate estimation of tissue oxygentation is very important in the fields of neuroimaging, cancer diagnostics, and critical care medicine.4,5 PAT involves collection of broadband acoustic waves generated by thermoelastic expansion of tissue upon absorption of light (illuminated at near-infrared wavelengths).6,7 PAT image reconstruction involves recovery of the acoustic source distribution from the detected broadband acoustic waves, which can be performed using analytical or model-based methods.8 Analytical methods are computationally less expensive compared to model-based image reconstruction.8,9 In contrary, model-based schemes are known to generate accurate images compared to analytical schemes, specifically in scenarios where irregular acquisition geometry is involved or limited data is acquired.9,10 Even with the model-based schemes, PAT inversion becomes difficult when the number of ultrasound detector positions is limited, and the situation gets aggravated due to the presence of noise in the data. More recently deep learning-based image postprocessing schemes have been developed for artifact removal and for limited data situations in PAT/microscopy;1113 however, these schemes were trained on simulated data and generating large experimental data is very difficult.

Segmentation approaches were proposed to delineate skin lining in mesoscopic photoacoustic images.14 Active contours-based image segmentation was used to obtain the PAT region of interest (ROI), the obtained ROI was then used to perform fluence correction to make PAT more quantitative.15,16 Different computational techniques have been reported in the literature to obtain vascular network in photoacoustic imaging. Vesselness filter has been proposed to enable accurate recovery of vasculature in PAT;1719 however, questions have been raised on the accuracy of using the vesselness filter (due to generation of artificial vasculature).20 Probabilistic approach involving different steps such as smoothing and filtering the data, clustering, vessel-segmentation using clusters, and morphological filling was developed for segmenting vessels in photoacoustic images.21 More recently convolution neural network-based approaches were also developed for jointly performing the image reconstruction and segmentation.22 Further, deep learning-based approach was proposed for segmenting animal boundary using multi-modal photoacoustic and ultrasound data.23 Further, deep learning-based method has also been developed for animal brain imaging.24

Notably segmenting vasculature is very important in the context of diagnosis and surgical guidance, specifically for cardiovascular diseases.25,26 In this work, we developed a binary tomography approach wherein we assume the reconstruction distribution contains only two unknowns namely the background and the absorbers. The major contributions of this work include: (a) application of binary tomographic methods for photoacoustic tomographic imaging involving spherical wave propagation as opposed to straight line approximation applied in x-ray computed tomography; (b) showing the utility of binary tomographic methods for improving photoacoustic tomographic imaging of vasculature; and (c) establishing the binary tomographic method superiority in photoacoustic tomographic imaging for providing approximate priors to improve model-based image reconstruction.

Binary tomography problem is solved based on a dual optimization approach with an auxiliary variable posed as a constrained optimization problem. We compared the binary tomography algorithm against standard backprojection approach and model-based reconstruction schemes such as Tikhonov regularization and sparsity-based regularization.27 The performance of these different algorithms was validated with in-silico data, physical phantom, and in-vivo brain vasculature imaging data.

2.

Methods

2.1.

Acoustic Forward Problem

The acoustic forward model in PAT can be written as

Eq. (1)

Ax=b,
where A is the system matrix constructed as explained in Refs. 9 and 28, b indicates the acoustic time series data measured at the detector locations, and x is the initial value of pressure rise at each pixel of the photoacoustic image. Note that the size of the matrix A is NM×NP (NM indicates the times series data at different transducer locations and NP indicates the total number of pixels).

Backprojection is the simplest approach to reconstruct the photoacoustic image from the measured acoustic data and is mathematically given as xbp=ATb.29 The reconstructed photoacoustic image quality is often very less using the backprojection approach (xbp) especially when limited data is available.

2.2.

Model-Based Reconstructions

2.2.1.

L2-Norm-based (Tikhonov) regularization

The Tikhonov regularization method has been used by many researchers in the context of photoacoustic imaging. Tikhonov method generates smoother solutions, and the characteristics of the solution are governed by the choice of the regularization parameter. The Tikhonov regularization approach minimizes the following objective function:30

Eq. (2)

ΩTikh=minx(Axb22+λx22),
where the regularization parameter is indicated by λ, which trades-off between the residue of the linear equations and L2-norm of (x). Smoother photoacoustic images could be obtained using higher regularization parameter value. In contrary, a smaller value of λ will increase the noise/high-frequency information in the reconstructed photoacoustic image. Minimization of the objective function (ΩTikh) can be performed using normal equations, which results in:9,30

Eq. (3)

xTikh=(ATA+λI)1ATb.

Further, since the system matrix is having large dimension (i.e., 51,200×40,000), a least squares QR-based dimensionality reduction was used to perform this inversion.9,31 The regularization parameters used in the Tikhonov regularization approach was chosen heuristically to result in the best possible figure of merit.

2.2.2.

L1-Norm-based (sparsity) regularization

Photoacoustic imaging is widely used for cardiovascular imaging due to its ability to image vasculature networks, hence it is reasonable to assume that PA images are sparse in nature. Therefore, it is expected that using non-smooth regularizers, such as the L1-norm-based approach could reconstruct accurate photoacoustic images. Many numerical optimization techniques were proposed for performing sparse recovery-based reconstruction.9,32,33 The L1-norm-based objective function is given by3335

Eq. (4)

Ω1=minx(Axb22+λx1),
where .1 represents the L1-norm. A majorization minimization framework was utilized to solve the L1-norm-based minimization in PAT, as explained in Refs. 32 and 35. The reconstruction parameters used in the majorization minimization approach were chosen heuristically to result in the best possible figure of merit.

2.3.

Binary Tomography

Binary tomography tries to reconstruct imaging volumes having binary distribution, i.e., x(u0,0,u1) with few number of detection positions.36 Performing reconstruction with limited data is particularly essential in the context of PAT, which has direct implication in terms of data-acquisition time or system cost. Binary tomography will try to directly estimate the segmented output from the acquired photoacoustic data. The binary tomography problem is defined as36

Eq. (5)

x*=argminx[infΦ(12Axb2)]subjected to  x=sgn(Φ),
where sgn(.) denotes element wise signum function [with sgn(0)=0] and Φ is the auxiliary variable. As can be seen, the above minimization problem is trying to estimate the solution that will have binary distribution based on the sign of the auxiliary variable. The Lagrangian for the above problem is given as36

Eq. (6)

L(x,Φ,v)=12Axb2+vT(xsgn(Φ)).

The variable v is the Lagrangian multiplier associated with equality constraint, i.e., x=sgn(Φ). The above optimization can be solved using a dual optimization approach:

Eq. (7)

v*=argmaxv[g(v)],
where g(v) is the dual function, which is given as

Eq. (8)

g(v)=infΦ,x[L(x,Φ,v)].

The dual problem in the case of binary tomography (with u0<u1) comes out to be

Eq. (9)

v*=argminvRN[12vATb(ATA)+p(v)],
where p(v)=i(|u0|max(vi,0)+|u1|max(vi,0)) is an asymmetric one-norm. The primal solution is given as x*=u0+(u1u0)H(v*), with H(.) being a heavy-side function. This dual problem is solved using proximal gradient descent. The proximal operator for the binary tomography problem (the above given asymmetric one-norm function) is given by asymmetric soft-thresholding operation, which is defined as

Eq. (10)

Sa<b(t)=t|b|  t|b|Sa<b(t)=t+|a|  t|a|Sa<b(t)=0|a|<t<|b|.

It is important to note that the binary tomography approach is trying to assign a class to the solution based on the sign of the auxiliary variable, hence a direct segmented output will be obtained from the acquired acoustic time series data. This will in essence avoid a two-step approach of reconstructing the image and then performing image segmentation.

2.4.

Figure of Merit

The proposed scheme was evaluated using a Dice similarity coefficient, which is used to measure the similarity between two sets of data. Let xtrue be the ground truth and xrecon be the reconstructed image and then the Dice similarity coefficient is computed as

Eq. (11)

DC=2|xtruexrecon||xtrue|+|xrecon|,
where |.| indicates the number of elements (cardinality) in the image.

2.5.

Numerical Simulations and Experiments

The different methods explained here were assessed with two numerical phantoms namely the Derenzo phantom (to understand the ability to recover objects of different sizes at different locations in the imaging domain) and a blood vessel phantom (as the main aim of PAT is to recover vasculature structures). The photoacoustic pressure intensity was considered as 1 kPa and speed of sound was considered to be homogeneous and constant at 1500  m/s. Note that numerical simulations are used to evaluate the different algorithms as it is difficult to compare the capabilities of different algorithms with realistic in-vivo cases due to unknown ground truth.

A two-dimensional imaging region of size 20.1×20.1  mm (corresponding to 402×402  pixels) was considered, and this imaging region housed the Derenzo phantom and the numerical blood vessel phantom. A larger computational grid having of size 50.1×50.1  mm (corresponding to 1002×1002  pixels) was considered for propagating the acoustic wave fields (forward model) and detecting the pressure signals by transducer. A total of 60 or 80 detectors (transducer positions) were placed on the circle of radius 22 mm. The detected simulated data were then added with additive white Gaussian noise to result in an SNR of 30, 40, and 60 dB in the in-silico data. Further, the transducer was modeled to have a center frequency of 2.25 MHz (with bandwidth of 70%), resulting in maximum detectable frequency as 3.0375 MHz. A total of about 512 samples were acquired with a sampling period of 50 ns (simulating a typical data acquisition (DAQ) sampling rate used in experiments). The in-silico data were then used to reconstruct the photoacoustic image using the different algorithms explained in Secs. 2.12.3. Note that the forward model was run on a finer grid, i.e., 20.1×20.1  mm (corresponding to 402×402  pixels), whereas the reconstruction was performed on a coarser domain, i.e., 20.1×20.1  mm (corresponding to 201×201  pixels) to avoid inverse crime. The dimension of the model matrix in the case of 60 detectors was 30,720×40,401 (corresponding to 40,401: 201×201 reconstruction grid and 30,720: 60 detector position each acquiring 512 time samples), and for the case of 80 detectors is 40,960×40,401 (corresponding to 40,401: 201×201 reconstruction grid and 40,960: 80 detector position each acquiring 512 time samples). The speed of sound in all these cases was considered to be constant at 1500  m/s.

Next, these different algorithms were evaluated using experimental photoacoustic data. Initially, a horse-hair phantom was built. This phantom was arranged in a triangular shape, with side-length and diameter of hair measured to be 10 and 0.15 mm, respectively, these three hairs were glued to the pipette tips adhered on acrylic slab. Photoacoustic signals from the horse-hair phantom data were acquired using the experimental setup shown in Fig. 2 of Ref. 37. The phantom was illuminated with a Q-switched Nd:YAG laser operating at 532 nm having a pulse width of 5 ns with 10 Hz repetition rate. The nanosecond laser pulses were delivered on the sample using Four right-angle uncoated prisms (PS911, Thorlabs) and one uncoated plano-concave lens (LC1715, Thorlabs). Note that the light fluence at the sample was about 9  mJ/cm2. The laser and the DAQ systems were synchronized using a sync signal from the laser. An ultrasound transducer having 13 mm diameter active area and about 70% nominal bandwidth (with center frequency of 2.25 MHz) was rotated around the sample (over 360 deg) for recording the photoacoustic signals. The generated PA signals due to photoacoustic effect were first amplified and then filtered using a pulse amplifier (Olympus-NDT, 5072PR). The amplified signals were then recorded using a DAQ card (GaGe, compuscope 4227) with a sampling frequency of 25 MHz. The experimental data were reconstructed using a model matrix with a size of 51,200×40,000 (corresponding to 40,000: 200×200 reconstruction grid and 51,200: 100 detector position each acquiring 512 time samples). In-vivo rat brain imaging data were collected using the same experimental setup.37 These in-vivo data were used to compare the reconstruction performance of different algorithms. The reconstructions presented in this work were implemented on a system with Intel i9-9920X processor with 3.50 GHz, 288 cores, and 128 GB memory.

3.

Results and Discussions

Figure 1 shows the reconstructed images obtained using the different algorithms with the Derenzo phantom and 80 detector position. Figures 1(a)1(c) indicate the reconstruction results corresponding to the backprojection approach with 30, 40, and 60 dB signal-to-noise ratio in data (SNRd). The reconstruction results corresponding to the L2-norm-based reconstruction for different SNRd (30, 40, and 60 dB) are shown in Figs. 1(d)1(f). As expected, the reconstructions obtained using the L2-norm approach are more accurate compared to the backprojection algorithm, particularly in reducing the streak artifacts. The L1-norm-based reconstruction outputs are shown in Figs. 1(g)1(i) for different SNRd values, and it can be seen that the L1-norm-based reconstruction shows improved resolution characteristics compared to L2-norm-based reconstruction. Figures 1(j)1(l) show the reconstruction corresponding to the binary tomography approach with different levels of noise in the data, as can be seen the reconstruction outputs using the binary tomography approach is able to localize the objects of different sizes with reduced blur. Further, the contrast using the proposed approach is found to be higher compared to other reconstruction schemes.

Fig. 1

Reconstruction results using Derenzo phantom and 80 detector positions using (a)–(c) backprojection approach and SNRd=30, 40, and 60 dB, respectively; (d)–(f) Tikhonov regularization approach and SNRd=30, 40, and 60 dB, respectively; (g)–(i) L1-norm approach and SNRd=30, 40, and 60 dB, respectively; and (j)–(l) binary tomography approach and SNRd=30, 40, and 60 dB, respectively.

JBO_26_8_086004_f001.png

Next numerical experiments were performed by reducing the number of detector positions to 60. Figure 2 shows the reconstructed images obtained using with the Derenzo phantom and 60 detector position. Figures 2(a)2(c) show the reconstruction results obtained using the backprojection algorithm with SNRd as 30, 40, and 60 dB, respectively. In comparison with 80 detector locations, the smaller absorbers were not resolved accurately while using only 60 detector location [as shown with red arrows in Figs. 2(a)2(c) and green arrows in Fig. 1(a)]. However, these smaller absorbers were accurately resolved using other methods with 60 detector positions (as can be seen in Fig. 2). The reconstruction results corresponding to the Tikhonov regularization approach for different SNRd (30, 40, and 60 dB) is shown in Figs. 2(d)2(f). The results indicate that the Tikhonov approach was successful in resolving smaller absorbers compared to backprojection approach. The L1-norm regularization-based reconstruction images are shown in Figs. 2(g)2(i) for different SNRd values. As can be seen with red arrows in Figs. 2(g)2(i), there are few locations in the imaging domain with large negative values resembling artifacts.16,38 Figures 2(j)2(l) show the reconstruction results corresponding to the binary tomography approach with different levels of noise in the data, as can be seen the reconstruction outputs using the binary tomography approach has reduced blur and is able to localize the objects of different sizes. Further, the obtained contrast using the proposed approach seems to better compare to other reconstruction schemes. However, it can be seen that spurious artifacts do arise using the proposed approach, and the larger circular absorbers seem to have donut shape [as can be seen with red arrows in Fig. 2(l)]. The line profile along the dashed red line shown in Fig. 1(j) for SNRd=30  dB and having 80 and 60 detectors is given in Fig. 3. Figure 3 indicates that the proposed method is able to accurately recover the bigger circles compared to other reconstruction methods. Note that the line-profile with binary tomography and other reconstructions is having smaller size compared to the ground truth. This might be due to lesser number of detector positions, limited bandwidth of the transducer, and the choice of threshold being considered during reconstruction.

Fig. 2

Similar effort as Fig. 1, with 60 acoustic detector positions.

JBO_26_8_086004_f002.png

Fig. 3

The line plot comparison for the different algorithms (L2-norm, L1-norm, and binary tomography cases) with SNRd=30  dB and having 80 and 60 detectors.

JBO_26_8_086004_f003.png

Another numerical experiment using blood vasculature structure was taken up next. Figure 4 shows the results pertaining to blood vasculature phantom and 80 detector locations. Figures 4(a)4(c) show the vasculature distribution using the backprojection approach at different SNRd values. Figures 4(d)4(f) indicate the reconstruction distribution obtained using L2-norm regularization with SNRd values being 30, 40, and 60 dB, respectively. The L1-norm regularization-based reconstruction at SNRd=30, 40, and 60 dB is shown in Figs. 4(g)4(i), respectively. As in earlier cases, the L2-norm-based approach seems to have lesser artifacts compared to backprojection approach; L1-norm regularization seems to have better resolution characteristics compared to L2-norm regularization, which is in accordance with earlier works.33,35 Note that L1-norm reconstruction produces reconstructions with improved resolution compared to L2-norm reconstruction due to the presence of soft-thresholding operation during L1-norm minimization. Finally, Figs. 4(j)4(l) show the reconstruction results corresponding to the binary tomography algorithm. Overall, Fig. 4 indicates that the binary tomography approach is able to recover the vasculature distribution. However, the binary tomography approach seems to present few discontinuities while recovering the vasculature like distribution [shown by green arrows in Fig. 4(k)]. Finally, Figs. 14 collectively indicate that binary tomography approach holds potential while recovering structures with binary contrast.

Fig. 4

Same effort as Fig. 1 with 80 detectors and blood vessel phantom.

JBO_26_8_086004_f004.png

The Dice similarity scores were also computed for the different in-silico cases demonstrated in Figs. 14, and the same has been shown in Table 1. The Dice similarity scores indicate that the proposed binary tomography approach seems to generate the most accurate representation to the ground truth. For the cases of backprojection, Tikhonov regularization, and L1-based regularization, the reconstructed output was first made non-negative by thresholding with 0, and then segmentation was performed by choosing the mean value as the segmentation threshold.

Table 1

Dice similarity scores for the different methods shown in Figs. 1–4. The best results are shown in bold.

Phantom (data noise)BackprojectionTikhonov regularizationL1-based regularizationBinary tomography
Fig. 1 (30 dB)0.85750.90020.95530.9752
Fig. 1 (40 dB)0.85790.90130.95760.9753
Fig. 1 (60 dB)0.85830.90120.95900.9756
Fig. 2 (30 dB)0.85580.88590.95600.9740
Fig. 2 (40 dB)0.85590.88630.96000.9743
Fig. 2 (60 dB)0.85620.88630.96090.9742
Fig. 4 (30 dB)0.84850.89030.94010.9697
Fig. 4 (40 dB)0.84830.89320.94640.9704
Fig. 4 (60 dB)0.84870.89290.94650.9705

Next, the binary tomography approach was evaluated using experimental data. Initially, horse-hair physical phantom data were reconstructed using the backprojection, L2-norm regularization, L1-norm regularization, and binary tomography approaches. The results corresponding to the same is shown in Fig. 5. Similar to numerical simulations, the L2-norm regularization scheme [Fig. 5(b)] is found to reconstruct images devoid of artifacts compared to backprojection scheme [shown by red arrow in Fig. 5(a)]. Similarly, the L1-norm regularization-based reconstruction seems to be having increased resolution at cost of amplified noise [shown by red arrow in Fig. 5(c)]. The binary tomography approach is able to generate accurate reconstruction of the horse-hair phantom; however, a discontinuity is observed on one of the hair [same is the case with other methods too; shown by red arrows in Figs. 5(b)5(d)]. Finally, the different algorithms explained in Sec. 2 were used to reconstruct the in-vivo rat brain vasculature distribution. The reconstruction of the vasculature in the rat brain using backprojection approach is shown in Fig. 6(a). The L2-norm and L1-norm-based regularized reconstruction is shown in Figs 6(b) and 6(c), respectively. Figure 6 indicates that L2-norm and L1-norm-based regularized reconstruction seems to outperform the backprojection type reconstruction. Finally, Fig. 6(d) shows the reconstruction results pertaining to the binary tomographic approach. It is evident from Fig. 6 that binary tomography is able to recover larger vasculature; however, the noise and spurious structures seem to appear using the binary tomography approach [double-edge-like structure shown by red arrows in Fig. 6(d)]. Note that it is difficult to compare the reconstructed outputs for experimental data cases using quantitative metrics due to the unavailability of ground truth. Further, metrics such as signal-to-noise ratio and peak-signal-to-noise ratio might heavily bias the comparison toward the binary tomography approach [as background is suppressed (see Figs. 5 and 6)], therefore, we have avoided reporting the quantitative metrics.

Fig. 5

Reconstruction results corresponding to horse hair phantom experiment using (a) backprojection, (b) Tikhonov regularization; (c) L1-norm regularization, and (d) binary tomography.

JBO_26_8_086004_f005.png

Fig. 6

Reconstruction results corresponding to rat brain imaging experiment using (a) backprojection, (b) Tikhonov regularization, (c) L1-norm regularization, and (d) binary tomography.

JBO_26_8_086004_f006.png

To demonstrate the visual comparison of different methods, the segmented results obtained using the backprojection, Tikhonov regularization, and L1-norm regularization are shown in Figs. 7(a)7(c), respectively. The reconstructions were performed with 80 detector positions and SNRd=30  dB. Note that the image segmentation was performed using a k-means clustering algorithm having two classes. The images shown here are the image pairs compared to the ground truth, wherein the green color indicates the false positive outcomes and pink color indicates the false negative outcomes. The results corresponding to the binary tomography approach are shown in Fig. 7(d). Note that there are very few artifacts observed using the binary tomography approach compared to the other standard reconstruction schemes.

Fig. 7

Segmentation comparison with the ground truth using (a) backprojection, (b) Tikhonov regularization, (c) L1-norm regularization, and (d) binary tomography. Here, green color indicates the false positives, and pink color indicates the false negatives obtained using the different algorithms.

JBO_26_8_086004_f007.png

The results indicate that the proposed binary tomography approach is able to directly reconstruct and delineate the absorbers from the experimental projections; this avoids the two-step approaches (of performing image segmentation after reconstruction). The results also show the appearance of artifacts with the binary tomography approach. However, the binary tomography scheme could be used to obtain the approximate distribution, which can be used as initial seed points for advanced segmentation approaches. Note that the reconstructions obtained using the binary tomography scheme provide only two classes (background and the absorber), making this algorithm to be useful in identifying the vasculature networks. Application like quantification of venous sinus distention for the diagnosis of intracranial hypotension relies more on accurate identification of the size of vasculature than the actual signal strength.39 However, in the case of multiscale vasculature, i.e., big arteries/veins to smaller capillaries, this problem can be posed as multiple classes and the binary tomography scheme could fail. In situations of multiple class problems, we could use a discrete tomographic reconstruction approach, which will be taken up as future work. PAT involves reconstruction of the initial pressure rise distribution, which directly relates to the absorbed energy density; however, the emphasis of binary tomography is to recover the vasculature structures accurately, thereby enabling its utility in scenarios such as therapy monitoring and hemorrhage detection in different organs.

The reconstruction times for the Tikhonov, L1-norm, and binary tomography reconstructions are 75.12, 1968.83, and 15.68 s, respectively. Further, the L1-norm-based reconstruction tends to be more noisy, as the L1-norm is not defined close to zero and the derivative of the L1-norm will be an approximation thereby amplifying the noise. The regularization parameters were chosen to result in the best possible reconstruction using the Tikhonov, L1-norm, and binary tomography methods. The regularization parameter does have an influence on the image quality, and the same was modeled and removed using a two-step approach as explained in Refs. 9 and 40. However, in this work, we have chosen the value optimally to result in the best possible figure of merit and avoid biasing the comparison. Notably the quantitative results obtained (in terms of Dice similarity scores) using the binary tomography approach is the best among the different methods with an advantage of being single step scheme. The proposed approach directly works with the acquired photoacoustic data (being robust at different noise levels as found in simulations) and hence does not require large training data to perform image segmentation using learning methods.22 Also recent works have focused on performing accurate reconstruction methods with limited angle photoacoustic data,41 evaluation of binary tomography reconstruction with limited angle data will be taken up as future work. Importantly, our work has focused on two-dimensional image reconstruction of vascular networks; extending the same to three-dimensional geometry should be straight forward as the proposed approach relies on optimization framework (extending the same will be similar to other three-dimensional inversion schemes33,42). Finally, the proposed approach can also be extended to image vasculature networks at mesoscopic depths wherein photoacoustic imaging demonstrated many clinical and preclinical applications.43

4.

Conclusion

The proposed binary PAT method is capable of providing accurate reconstruction of vascular networks directly from the photoacoustic sinogram data. Binary tomography approach relies on solving a dual optimization problem, wherein the primal solution is defined based on a heavy-side function. Further, the dual optimization problem was solved using proximal gradient descent approach with proximal operator being asymmetric soft thresholding. Notably, the reconstructed solution will have a binary distribution based on the sign of the auxiliary variable introduced during the optimization. The binary tomography algorithm was compared with backprojection, Tikhonov regularization, and L1-norm-based regularization approach with in-silico, physical phantom, and in-vivo brain vasculature data. The results showed that the proposed approach was able to produce results that were closer to the ground truth and the same was reflective from the Dice similarity coefficient. Binary tomography approach could be potentially useful for many applications in the context of photoacoustic microscopy, mesoscopy, and tomography.

Disclosures

The authors declare no conflicts of interest.

Acknowledgments

Jaya Prakash would like to acknowledge the IISc Startup Grant, DBT-IYBA Program, and the Infosys Young Investigator Award. Phaneendra K. Yalavarthy would like to acknowledge the DST-ICPS cluster funding (No. T-851) for the Data Science Program. Part of this work was supported through WIPRO-GE Collaborative Laboratory of Artificial Intelligence in Healthcare and Medical Imaging.

Code, Data, and Materials Availability

The associate code and data that support the findings of this study are available from the corresponding author upon reasonable request.

References

1. 

L. V. Wang and S. Hu, “Photoacoustic tomography: in vivo imaging from organelles to organs,” Science, 335 (6075), 1458 –1462 (2012). https://doi.org/10.1126/science.1216210 SCIEAS 0036-8075 Google Scholar

2. 

L. V. Wang and J. Yao, “A practical guide to photoacoustic tomography in the life sciences,” Nat. Methods, 13 (8), 627 –638 (2016). https://doi.org/10.1038/nmeth.3925 1548-7091 Google Scholar

3. 

A. Taruttis and V. Ntziachristos, “Advances in real-time multispectral optoacoustic imaging and its applications,” Nat. Photonics, 9 (4), 219 –227 (2015). https://doi.org/10.1038/nphoton.2015.29 NPAHBY 1749-4885 Google Scholar

4. 

Q. Fu et al., “Photoacoustic imaging: contrast agents and their biomedical applications,” Adv. Mater., 31 (6), 1805875 (2019). https://doi.org/10.1002/adma.201805875 ADVMEW 0935-9648 Google Scholar

5. 

K. S. Valluru, K. E. Wilson and J. K. Willmann, “Photoacoustic imaging in oncology: translational preclinical and early clinical experience,” Radiology, 280 (2), 332 –349 (2016). https://doi.org/10.1148/radiol.16151414 RADLAX 0033-8419 Google Scholar

6. 

L. V. Wang, “Tutorial on photoacoustic microscopy and computed tomography,” IEEE J. Sel. Top. Quantum Electron., 14 (1), 171 –179 (2008). https://doi.org/10.1109/JSTQE.2007.913398 IJSQEN 1077-260X Google Scholar

7. 

L. V. Wang and H.-I. Wu, Biomedical Optics: Principles and Imaging, John Wiley & Sons, Hoboken, New Jersey (2012). Google Scholar

8. 

A. Rosenthal, V. Ntziachristos and D. Razansky, “Acoustic inversion in optoacoustic tomography: a review,” Curr. Med. Imaging, 9 (4), 318 –336 (2013). https://doi.org/10.2174/15734056113096660006 Google Scholar

9. 

J. Prakash et al., “Basis pursuit deconvolution for improving model-based reconstructed images in photoacoustic tomography,” Biomed. Opt. Express, 5 (5), 1363 –77 (2014). https://doi.org/10.1364/BOE.5.001363 BOEICL 2156-7085 Google Scholar

10. 

P. Burgholzer et al., “Exact and approximative imaging methods for photoacoustic tomography using an arbitrary detection surface,” Phys. Rev. E, 75 046706 (2007). https://doi.org/10.1103/PhysRevE.75.046706 Google Scholar

11. 

T. Vu et al., “A generative adversarial network for artifact removal in photoacoustic computed tomography with a linear-array transducer,” Exp. Biol. Med., 245 (7), 597 –605 (2020). https://doi.org/10.1177/1535370220914285 EXBMAA 0071-3384 Google Scholar

12. 

T. Vu et al., “Deep image prior for undersampling high-speed photo acoustic microscopy,” Photoacoustics, 22 100266 (2021). https://doi.org/10.1016/j.pacs.2021.100266 Google Scholar

13. 

R. Manwar et al., “Photoacoustic signal enhancement: Towards utilization of low energy laser diodes in real-time photoacoustic imaging,” Sensors, 18 (10), (2018). https://doi.org/10.3390/s18103498 SNSRES 0746-9462 Google Scholar

14. 

S. Nitkunanantharajah et al., “Skin surface detection in 3D optoacoustic mesoscopy based on dynamic programming,” IEEE Trans. Med. Imaging, 39 (2), 458 –467 (2020). https://doi.org/10.1109/TMI.2019.2928393 ITMID4 0278-0062 Google Scholar

15. 

S. Mandal, X. L. D. Ben and D. Razansky, “Visual quality enhancement in optoacoustic tomography using active contour segmentation priors,” IEEE Trans. Med. Imaging, 35 (10), 2209 –2217 (2016). https://doi.org/10.1109/TMI.2016.2553156 Google Scholar

16. 

J. Prakash et al., “Maximum entropy based non-negative optoacoustic tomographic image reconstruction,” IEEE Trans. Biomed. Eng., 66 2604 –2616 (2019). https://doi.org/10.1109/TBME.2019.2892842 IEBEAX 0018-9294 Google Scholar

17. 

T. Oruganti, J. G. Laufer and B. E. Treeby, “Vessel filtering of photoacoustic images,” 319 –328 (2013). https://doi.org/10.1117/12.2005988 Google Scholar

18. 

P. Zhang et al., “In vivo superresolution photoacoustic computed tomography by localization of single dyed droplets,” Light: Sci. Appl., 8 (1), 36 (2019). https://doi.org/10.1038/s41377-019-0147-9 Google Scholar

19. 

L. Lin et al., “Single-breath-hold photoacoustic computed tomography of the breast,” Nat. Commun., 9 (1), 2352 (2018). https://doi.org/10.1038/s41467-018-04576-z NCAOBW 2041-1723 Google Scholar

20. 

A. Longo et al., “Assessment of hessian-based Frangi vesselness filter in optoacoustic imaging,” Photoacoustics, 20 100200 (2020). https://doi.org/10.1016/j.pacs.2020.100200 Google Scholar

21. 

P. Raumonen and T. Tarvainen, “Segmentation of vessel structures from photoacoustic images with reliability assessment,” Biomed. Opt. Express, 9 2887 –2904 (2018). https://doi.org/10.1364/BOE.9.002887 BOEICL 2156-7085 Google Scholar

22. 

Y. E. Boink, S. Manohar and C. Brune, “A partially-learned algorithm for joint photo-acoustic reconstruction and segmentation,” IEEE Trans. Med. Imaging, 39 (1), 129 –139 (2020). https://doi.org/10.1109/TMI.2019.2922026 ITMID4 0278-0062 Google Scholar

23. 

B. Lafci et al., “Deep learning for automatic segmentation of hybrid optoacoustic ultrasound (OPUS) images,” IEEE Trans. Ultrasonics, Ferroelectr. Freq. Control, 68 (3), 688 –696 (2021). https://doi.org/10.1109/TUFFC.2020.3022324 Google Scholar

24. 

R. Manwar et al., “Deep learning protocol for improved photoacoustic brain imaging,” J. Biophotonics, 13 (10), e202000212 (2020). https://doi.org/10.1002/jbio.202000212 Google Scholar

25. 

K. M. Kempski et al., “In vivo photoacoustic imaging of major blood vessels in the pancreas and liver during surgery,” J. Biomed. Opt., 24 (12), 121905 (2019). https://doi.org/10.1117/1.JBO.24.12.121905 JBOPFO 1083-3668 Google Scholar

26. 

M. A. Lediju Bell, “Photoacoustic imaging for surgical guidance: Principles, applications, and outlook,” J. Appl. Phys., 128 (6), 060904 (2020). https://doi.org/10.1063/5.0018190 JAPIAU 0021-8979 Google Scholar

27. 

J. Prakash et al., “Fractional regularization to improve photoacoustic tomographic image reconstruction,” IEEE Trans. Med. Imaging, 38 1935 –1947 (2019). https://doi.org/10.1109/TMI.2018.2889314 ITMID4 0278-0062 Google Scholar

28. 

C. Shaw et al., “Least squares QR-based decomposition provides an efficient way of computing optimal regularization parameter in photoacoustic tomography,” J. Biomed. Opt., 18 (8), 080501 (2013). https://doi.org/10.1117/1.JBO.18.8.080501 JBOPFO 1083-3668 Google Scholar

29. 

G. L. Zeng, Medical Image Reconstruction: A Conceptual Tutorial, Springer-Verlang, Berlin, Heidelberg (2010). Google Scholar

30. 

G. H. Golub, P. C. Hansen and D. P. O’Leary, “Tikhonov regularization and total least squares,” SIAM J. Matrix Anal. Appl., 21 (1), 185 –194 (1999). https://doi.org/10.1137/S0895479897326432 Google Scholar

31. 

C. C. Paige and M. A. Saunders, “LSQR: an algorithm for sparse linear equations and sparse least squares,,” ACM Trans. Math. Software, 8 43 –71 (1982). https://doi.org/10.1145/355984.355989 ACMSCU 0098-3500 Google Scholar

32. 

J. Prakash et al., “Sparse recovery methods hold promise for diffuse optical tomographic image reconstruction,” IEEE J. Sel. Top. Quantum Electron., 20 (2), (2014). https://doi.org/10.1109/JSTQE.2013.2278218 IJSQEN 1077-260X Google Scholar

33. 

Y. Han et al., “Three-dimensional optoacoustic reconstruction using fast sparse representation,” Opt. Lett., 42 (5), 979 (2017). https://doi.org/10.1364/OL.42.000979 OPLEDP 0146-9592 Google Scholar

34. 

M. M. Betcke et al., “Acoustic wave field reconstruction from compressed measurements with application in photoacoustic tomography,” IEEE Trans. Comput. Imaging, 3 (4), 710 –721 (2017). https://doi.org/10.1109/TCI.2017.2706029 Google Scholar

35. 

A. Majumdar and R. K. Ward, “On the choice of compressed sensing priors and sparsifying transforms for MR image reconstruction: an experimental study,” Signal Process.: Image Commun., 27 (9), 1035 –1048 (2012). SPICEF 0923-5965 Google Scholar

36. 

A. Kadu and T. van Leeuwen, “A convex formulation for binary tomography,” IEEE Trans. Comput. Imaging, 6 1 –11 (2020). https://doi.org/10.1109/TCI.2019.2898333 Google Scholar

37. 

N. Awasthi et al., “Image-guided filtering for improving photoacoustic tomographic image reconstruction,” J. Biomed. Opt., 23 (9), 091413 (2018). https://doi.org/10.1117/1.JBO.23.9.091413 JBOPFO 1083-3668 Google Scholar

38. 

L. Ding et al., “Efficient non-negative constrained model-based inversion in optoacoustic tomography,” Phys. Med. Biol., 60 6733 –6750 (2015). https://doi.org/10.1088/0031-9155/60/17/6733 PHMBA7 0031-9155 Google Scholar

39. 

P. Rajendran et al., “In vivo detection of venous sinus distension due to intracranial hypotension in small animal using pulsed-laser-diode photoacoustic tomography,” J. Biophotonics, 13 (6), e201960162 (2020). https://doi.org/10.1002/jbio.201960162 Google Scholar

40. 

J. Prakash et al., “Model-resolution-based basis pursuit deconvolution improves diffuse optical tomographic imaging,” IEEE Trans. Med. Imaging, 33 (4), (2014). https://doi.org/10.1109/TMI.2013.2297691 ITMID4 0278-0062 Google Scholar

41. 

K. B. Chowdhury et al., “A synthetic total impulse response characterization method for correction of hand-held optoacoustic images,” IEEE Trans. Med. Imaging, 39 (10), 3218 –3230 (2020). https://doi.org/10.1109/TMI.2020.2989236 Google Scholar

42. 

R. Nuster, P. Slezak and G. Paltauf, “High resolution three-dimensional photoacoustic tomography with CCD-camera based ultrasound detection,” Biomed. Opt. Express, 5 2635 –2647 (2014). https://doi.org/10.1364/BOE.5.002635 BOEICL 2156-7085 Google Scholar

43. 

P. K. Upputuri and M. Pramanik, “Recent advances toward preclinical and clinical translation of photoacoustic tomography: a review,” J. Biomed. Opt., 22 (4), 041006 (2016). https://doi.org/10.1117/1.JBO.22.4.041006 JBOPFO 1083-3668 Google Scholar

Biographies of the authors are not available.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 Unported License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Jaya Prakash, Sandeep Kumar Kalva, Manojit Pramanik, and Phaneendra K. Yalavarthy "Binary photoacoustic tomography for improved vasculature imaging," Journal of Biomedical Optics 26(8), 086004 (18 August 2021). https://doi.org/10.1117/1.JBO.26.8.086004
Received: 19 April 2021; Accepted: 29 June 2021; Published: 18 August 2021