1.IntroductionIn optical surface metrology it is of great importance that the absolute form is measured with high accuracy. The surface forms applied for optical systems are continuously increasing in curvature and complexity. In parallel, the demands on spatial resolution and accuracy are constantly growing. High-accuracy form measurements with high lateral resolution are therefore required. Conventional measurement devices such as coordinate measurement machines (CMMs)1, 2, 3, 4 can satisfy these requirements only partially, and recently, multiple distance sensor systems have been proposed as a promising approach to meet these demands.5, 6, 7 Although “distance” implies a fixed reference, sensors of form measuring setups with arbitrary offset, e.g., Fizeau interferometers, will also be termed distance sensors in the following. For absolute high-accuracy measurements, the elimination of errors is of first priority. Recent theoretical work7 has shown that multiple distance sensor systems are basically limited, which means that certain systematic errors cannot be eliminated in principle. For a small sensor head, the cumulation even of small systematic sensor errors leads to large overall errors.7 As a consequence, a small parabolic or spherical error can sum up to significant form errors, for example, in interferometers used for stitching interferometry. In Ref. 8 it has been recently shown theoretically that additional angular measurements of the sensor system’s head can basically solve this problem: the mathematical analysis reveals that for a properly designed multiple distance sensor system supplemented by additional angular information, the elimination of both scanning stage errors as well as systematic sensor offset errors is possible. In this paper, the experimental realization of the new method is presented. The multiple distance sensor system is realized by an interferometer and angular information is provided by an autocollimator. The system is called a traceable multiple sensor (TMS) system because the measurands of the system can be traced back to the SI units of angle and length with high accuracy. 2.Mathematical Background of the TMS PrincipleIn the TMS model,8 an array of distance sensors with constant spacing is scanned along the surface under test by a scanning stage (see Fig. 1). The analysis of the model accounts for piston and tilt errors, and , of the scanning stage at each position of the sensor system (see Fig. 2). In addition, systematic offset errors of the sensor elements are taken into account. The systematic offset errors are assumed to be constant while the sensor system is moved along the specimen. Angular information on the tilt of the distance sensor head is provided by an autocollimator, which is rigidly fixed to the optical table on which the specimen is placed (see Fig. 1). It was shown in Ref. 8 that the surface profile can be reconstructed exactly up to a straight line in the presence of scanning stage piston and tilt errors and systematic offset errors of the distance sensor elements provided that the distance measurements and the angular measurements are noise-free. In the presence of measurement noise, a least-squares solution is applied and the uncertainty of the reconstructed surface profile with respect to measurement noise can be derived. It turns out that the TMS principle yields a stable least-squares solution that reaches high accuracy. 3.Experimental Demonstrator SetupThe multiple distance sensor chosen for the demonstrator setup (see Fig. 3) is a commercial compact Twyman-Green interferometer with an aperture of for measuring specular optical surfaces.9 The pixels of the image detector array within the interferometer represent single distance sensors. In the experimental realization, the lateral sensor distance is given by the effective lateral distance of the sensor elements on the specimen’s surface. This is given by the image sensor’s pixel distance taking into account the magnification of the interferometer’s imaging system. Due to the restrictions in computer capabilities, not all pixels of the interferometer’s imaging array were used. Instead, a virtual sensor system consisting of 16 single sensors was constructed from the array as follows. First, a surface was constructed from the interferometer’s imaging array by appropriate surface modeling.10 Then, the obtained surface model was evaluated at the chosen 16 appropriate positions, and the resulting values assigned to the 16 virtual sensors. The sensor head is scanned along the surface under test by a low-cost linear stage. The specimen can have an extension of up to , limited only by the scan range of the scanning stage. The additional angular measurement is carried out by a calibrated autocollimator,11 which faces a small mirror that is mechanically fixed to the multiple distance sensor. The angle between specimen and autocollimator has to be highly stable, which is achieved by a rigid mechanical setup. The measurement is controlled by a standard computer while the data evaluation is performed by an additional fast computer with extended memory capacity. The TMS model (Ref. 8) implies that the lateral sensor distance is an integer multiple of the scanning step. For the 16 virtual distance sensors formed in this setup, the lateral sensor distance and the scanning step are , which determines the limit of the lateral resolution of the measurement. For different demands on the measurement, for example, a higher lateral resolution, the magnification of the interferometer’s imaging system, the number of evaluated pixels, and the scanning step can be designed appropriately. Basically, the lateral resolution can be reduced down to a few micrometers for the TMS method using an interferometer as the multiple distance sensor. To further improve the lateral resolution, other sensors could be used, for example, atomic force microscopes. 4.Test MeasurementsTo test the functionality of the instrument, exemplary surfaces were measured. The reduction of aperture, with respect to full-field interferometers capturing the whole surface in parallel, increases the range of measurable surface slope and curvature. Optical surfaces with peak-to-valley heights of at a specimen length of have been measured with this setup. Figure 4 shows a typical result for an aspherical structure on a flat substrate, and Fig. 5 shows the associated uncertainty with respect to the noise of the measurements. It has to be kept in mind that TMS inherently accounts for scanning stage tilt and piston errors as well as systematic distance sensor offset errors. This means that in contrast to stitching techniques that use distance sensors only, even the parabolic or spherical contributions of the surface form are determined with high accuracy. To further validate the measurement results of the TMS test setup, comparisons with other measurement principles have been started and will be published soon. The first results confirm the validity of the TMS setup at the uncertainty level of those instruments. 5.Future TasksThe uncertainty shown in Fig. 5 accounts only for the influence of the noise of the distance and angular measurements assuming correctness of the applied TMS model. Future work will address a more comprehensive uncertainty analysis, taking into account further influences, e.g., possible errors in the lateral positioning of the distance sensor system. First results have already been published.10, 12, 13 In particular, the following sources of uncertainty need to be quantitatively assessed:
6.ConclusionFor the measurement of the form of optical surfaces the new TMS system has been presented. It uses coupled multiple distance sensors, which are scanned along the surface under test. By using a small sensor head a high lateral resolution can be achieved. In addition to the multiple distance sensors, TMS utilizes an autocollimator measuring the tilt of the sensor head, which makes it possible to eliminate systematic errors of the distance sensors. In this way, the TMS system allows high-accuracy measurements of flat and curved surfaces with high lateral resolution. A demonstrator setup was shown and test measurements were presented confirming the high potential of the TMS method. AcknowledgmentThe authors would like to thank Ralf D. Geckeler and Joachim Gerhardt for fruitful discussions and helpful data analysis. ReferencesS. R. Clark and
J. E. Greivenkamp,
“Optical reference profilometry,”
Opt. Eng., 40 2845
–2851
(2001). https://doi.org/10.1117/1.1419192 0091-3286 Google Scholar
H. Shiozawa and
Y. Fukutomi,
“Development of an ultra-precision 3DCMM with the repeatability of nanometer order,”
JSPE Publ. Ser., 3 360
–365
(1999). Google Scholar
M. Negishi,
K. Watanabe,
K. Matsusita,
T. Kasahara, and
K. Hosaka,
“A high-precision coordinate measurement machine for aspherical optics,”
JSPE Publ. Ser., 3 354
–359
(1999). Google Scholar
H. Tsutsumi,
K. Yoshizumi, and
H. Takeuchi,
“Ultrahighly accurate 3D profilometer,”
Proc. SPIE, 5638 387
–394
(2005). 0277-786X Google Scholar
W. Gao and
S. Kiyono,
“High accuracy profile measurement of a machined surface by the combined method,”
Measurement, 19 55
–64
(1996). https://doi.org/10.1016/S0263-2241(96)00066-8 0263-2241 Google Scholar
W. Gao and
S. Kiyono,
“On-machine profile measurement of machined surface using the combined three-point method,”
JSME Int. J., Ser. C, 40 253
–259
(1997). 1340-8062 Google Scholar
I. Weingärtner and
C. Elster,
“System of four distance sensors for high-accuracy measurement of topography,”
Precis. Eng., 28 164
–170
(2004). https://doi.org/10.1016/j.precisioneng.2003.10.001 0141-6359 Google Scholar
C. Elster,
I. Weingärtner, and
M. Schulz,
“Coupled distance sensor systems for high-accuracy topography measurement: accounting for scanning stage and systematic sensor errors,”
Precis. Eng., 30 32
–38
(2006). 0141-6359 Google Scholar
Interferometer, FISBA Optik AG, CH-9016 St. Gallen and FISBA Optik GmbH, Berlin, http://www.fisba.ch/ Google Scholar
J. Gerhardt,
R. D. Geckeler,
M. Schulz, and
C. Elster,
“High-accuracy profile form measurement by a scanning system consisting of an angle sensor and coupled distance sensors,”
Proc. SPIE, 5856 385
–392
(2005). 0277-786X Google Scholar
ELCOMAT 3000, Moeller-Wedel Optical GmbH, http://www.moeller-wedel-optical.com/El-Autocolimators/E_Elcomat3000.htm/ Google Scholar
M. Schulz,
J. Gerhardt,
R. D. Geckeler, and
C. Elster,
“Traceable multiple sensor system for absolute form measurement,”
Proc. SPIE, 5878 84
–91
(2005). 0277-786X Google Scholar
M. Schulz and
R. D. Geckeler,
“Scanning form measurement for curved surfaces,”
Proc. SPIE, 5921 29
–39
(2005). 0277-786X Google Scholar
|