19 May 2021 Re-absorption and nonradiative energy transfer in vibronic laser gain media
Jonathan W. Evans, Thomas R. Harris, Eric J. Turner, Martin M. Kimani, J. Matthew Mann, Ronald W. Stites, Gary Cook, Kenneth L. Schepler
Author Affiliations +
Abstract

Fe:ZnSe and other transition metals have broad upper and lower energy manifolds that give rise to broad absorption and fluorescence bands in the mid-infrared spectral region. Energy transfer between Fe2  +   ions via re-absorption of fluorescence (due to the spectral overlap of these bands) and phonon-assisted energy transfer (due to the lattice dynamics of ZnSe) becomes more probable with increasing Fe2  +   concentration. Here, we treat these processes as additional terms in the rate equations that govern the spontaneous decay of the Fe2  +   ion. This treatment gives insight into anomalous behavior seen in the thermal lifetime of the Fe:ZnSe system. We also apply the model to other transition metal-doped II–VI materials.

1.

Introduction

In the past decade, Fe:ZnSe has developed into an important mid-infrared (IR) laser gain medium with rapid and substantial power scaling in both pulsed and continuous-wave regimes.19 Several studies have reported the temperature-dependent fluorescent lifetime of this material from 10 to 300 K. However, the data produced by such studies differ substantially,1,10,11 and all studies report results that do not agree with the standard Mott–Seitz model12 of temperature-dependent fluorescent decay. In this work, which we initially presented in the conference proceedings of SPIE LASE,13 we develop a quantitative model that explains and resolves these discrepancies.

Vibronic laser gain media such as Fe:ZnSe and other transition metals have broad upper and lower energy manifolds that give rise to broad absorption and fluorescence bands. Energy transfer between active ions via re-absorption of fluorescence (due to the spectral overlap of these bands) and phonon-assisted energy transfer (due to the lattice dynamics of the crystal host) becomes more probable with increasing impurity concentration. We report a detailed model of the population dynamics of such systems, which includes contributions from fluorescent re-absorption, nonradiative energy transfer (NRET), and cascaded multi-phonon decay effects.

Spectroscopic measurements of the lifetime of several transition metal systems (including Fe:ZnSe) are presented and analyzed using this model. The model is fit to this lifetime data with excellent confidence intervals. The model shows qualitative agreement with expected trends, such as an increase in re-absorption and NRET with increasing ion concentration. The model provides additional insight for the design of mid-IR lasers based on Fe:ZnSe and other vibronic systems including systems based on Cr2+ and Co2+ ions.

2.

Background

Many authors have measured the temperature-dependent fluorescent lifetime of Fe:ZnSe. Adams et al.14 and Song et al.15 have measured and reported the fluorescence lifetime of Fe:ZnSe with Fe2+ impurity concentrations >1019  ions/cm3. Their results show the lifetime of Fe:ZnSe to approach 105  μs near 120 K. Deloach et al.16 and Jelínková et al.17 report similar results, though their concentrations were not specified. The data produced by these studies are shown in Fig. 1.

Fig. 1

The measured values of the fluorescence lifetime of Fe2+ ions in ZnSe with respect to changes in temperature.

OE_60_5_056103_f001.png

Myoung et al.18 measured the maximum value of the fluorescence lifetime of Fe:ZnSe with much lower Fe2+ concentrations of 1017  ions/cm3 to be 60  μs near 100 K. In our recent work,11 we measured the maximum value of the fluorescence lifetime of Fe:ZnSe with intermediate Fe2+ concentrations of 1018  ions/cm3 to be almost 70  μs at 100 K. These data are also included in Fig. 1.

It has been suggested that the reduction in the fluorescence lifetime observed by Myoung and in our work is attributable to the fact that the samples under investigation were produced with the post-growth thermal diffusion (PGTD) technique. However, the samples used by Song et al.15 were also produced in this manner, so that technique is apparently not the cause of the reduction in lifetime. We concluded in our recent work11 that the differences in the observed fluorescence lifetime from dataset to dataset are due to concentration effects. In that work, we developed a model of the population dynamics of such systems that includes contributions from effects that are known to be concentration dependent, namely, the re-absorption of fluorescence, NRET, and cascaded multi-phonon decay effects. In this work, we apply this model to additional lifetime data for Fe:ZnSe at different concentrations and Fe:CdMnTe, Cr:ZnS, Cr:ZnSe, Cr:CdSe, Cr:CdMnTe, Co:ZnS, Co:ZnSe, Co:ZnTe, and Co:CdTe.

3.

Theory

In this section, we will develop a description of the fluorescent decay from the upper state manifold in which the low-energy population experiences NRET between impurity ions and the higher-energy population experiences significant re-absorption (with subsequent re-emission) of fluorescent radiation. NRET will shorten the observed lifetime from its intrinsic value and re-absorption will lengthen it (Henderson and Imbusch19 provide a broad overview of the electronic dynamics of vibronic systems at high impurity concentrations. They note that phonon-assisted NRET between active ions occurs with greater probability with decreasing temperature and increasing impurity concentration. They also note that re-absorption of fluorescent radiation occurs with greater probability with increasing spectral overlap and with increasing ion concentration. The spectral overlap of the absorption and emission bands of Fe:ZnSe increases with increasing temperature due to the broadening of those bands as more levels become thermally accessible.). The relative probabilities of these processes are mediated by the occupation statistics of the sublevels of the manifold, which are assumed to be in thermal equilibrium.

We note that, in Fe:ZnSe, the energy difference between energy levels in the T25 manifold is much greater than kbT, and thus thermal statistics cannot be neglected. Furthermore, Fe2+ ion concentration appears to play a significant role in determining the observed fluorescence lifetime of Fe:ZnSe. Consequently, we have developed a model that describes the temperature dependence of the fluorescence lifetime in terms of thermally activated and concentration-dependent decay processes.

3.1.

Rate Equation Modeling of Energy Transfer Mechanisms

3.1.1.

Re-absorption

Fe:ZnSe and other transition metal media operate as quasi-four level laser systems. However, for the sake of simplicity, we will use a two level approximation to the upper state manifold and a single level approximation to the lower state manifold. We have introduced an additional upper state level to describe some recycling of the upper manifold population due to re-absorption of fluorescent emission and subsequent re-emission. Figure 2 shows the level scheme used in the following discussion. We assume some initial population of the meta-level and ignore the initial population of the ground level.

Fig. 2

An energy level diagram with levels (1) and (2) representing a single level with re-absorption.

OE_60_5_056103_f002.png

In this model, the two upper levels (1) and (2) together represent a single meta-level and each experiences spontaneous relaxation with the same spontaneous decay rate A20=A10. With some probability Pr, a photon emitted from this meta-level is re-absorbed. This process is represented by A12 and A21. The probability of re-absorption (we will assume 100% re-emission of absorbed photons) is

Eq. (1)

Pr=A21A21+A20=A21τ,

Eq. (2)

=A12A12+A10=A12τ.

So, A21=Pr/τ where τ is the observed lifetime of the upper energy levels. It is worth noting that the re-absorption does not compete with spontaneous emission process, rather re-absorption of a photon occurs only after spontaneous emission of a photon. Thus, in this model of the movement of population due to re-absorption, the observed decay rate of the upper levels is 1/τ=A20+A21. Consequently, A20=1/τA21.

We can write the rate equations for this system using the usual conventions:

Eq. (3)

dN2dt=A20N2A21N2+A12N1dN1dt=A10N1+A21N2A12N1dN0dt=A20N2+A10N1.

The fluorescence signal will be proportional to I(t)=(1Pr)N1(t)+(1Pr)N2(t). Solving the system analytically, and substituting A12A21, A10A20, and A20=1/τA21, we can then substitute into the expression for I(t) to obtain

Eq. (4)

I(t)=et/τ((N1[0]+N2[0])(Pr1)ePrt/τ)=(N1[0]+N2[0])(1Pr)e(1Pr)t/τ,
where N1[0] and N2[0] are the initial population in levels (1) and (2), respectively. Figure 3 shows normalized traces of I(t) as Pr increases. It is worth noting that the apparent lifetime of the trace increases with increasing Pr.

Fig. 3

Normalized traces of I(t) with several values of the re-absorption probability Pr. The Pr=0 trace represents the spontaneous rate (In Figs. 3 and 5, we have used τ=50  μs for illustrative purposes.).

OE_60_5_056103_f003.png

3.1.2.

Non-radiative energy transfer

It is straightforward to modify the rate equations to recast level (2) as a single upper level within the upper manifold and change level (1) to represent a radiation trap from which all paths to level (0) are nonradiative. Figure 4 shows the level scheme used in the following discussion.

Fig. 4

Non-radiative energy transfer to a radiation trap.

OE_60_5_056103_f004.png

We modify the previous system of rate equations using A20=1/τ and A12=A10=0 to represent non-radiative trapping, which occurs with some probability Pn. The solution for the fluorescence intensity becomes:

Eq. (5)

I(t)=e(1+Pn)t/τ(N2[0]N2[0]Pn)=N2[0](1Pn)e(1+Pn)t/τ.

Figure 5 shows normalized traces of I(t) as Pn increases. It is worth noting that the apparent lifetime of the decay decreases with increasing Pn.

Fig. 5

Traces of I(t) with several values of the re-absorption probability Pn.

OE_60_5_056103_f005.png

3.1.3.

Combining re-absorption and NRET

Modeling the combined system requires the introduction of two additional energy levels as shown in Fig. 6. The system of rate equations becomes slightly more complicated but can still be solved analytically. The model takes into consideration the florescence contribution from two separate levels, one of which sees the re-absorption of fluorescence and the other of which experiences radiation trapping. In this configuration, we suppose them to be separated by some activation energy, which is a free parameter in a previous model. Figure 6 shows this system.

Fig. 6

An electronic level diagram with levels (3) and (4) representing a single level with re-absorption and level (1) representing a radiation trap. Levels (1)–(4) belong to the upper state manifold.

OE_60_5_056103_f006.png

The rate equations can be written as

Eq. (6)

dN4dt=A40N4A43N4+A34N3dN3dt=A30N3A34N3+A43N4dN2dt=A20N2A21N2dN1dt=A21N2dN0dt=A40N4+A30N3+A20N2.

The fluorescence signal is

Eq. (7)

I(t)(1Pr)(N4[t]+N3[t])+(1Pn)N2[t],
which gives

Eq. (8)

I(t)=e(2+Pn+Pr)t/τ·(N2[0]e(1+Pr)t/τ(Pn1)+(N3[0]+N4[0])e(1+Pn+2Pr)t/τ(Pr1))=N2[0](1Pn)e(1+Pn)t/τ+(N3[0]+N4[0])(1Pr)e(1Pr)t/τ.

Clearly, the solution to this system represents decay from two independent levels with different rates to the same ground level. So, we simplify the five-level to its three-level equivalent (see Fig. 7). In this system, level (2) experiences a different relaxation rate than level (1) because photons emitted spontaneously from level (2) are re-absorbed with probability Pr and some of the population of level (1) moves to the ground state non-radiatively with probability Pn.

Fig. 7

Two levels, which undergo different parasitic processes, in thermal equilibrium.

OE_60_5_056103_f007.png

Now, suppose that the system obeys Fermi–Dirac statistics and is initially at equilibrium. So,

Eq. (9)

N1[0]=Ntotal·κ1=Ntotal(1+G2G1exp(ΔEkbT))1,

Eq. (10)

N2[0]=Ntotal·κ2=Ntotal(1+G1G2exp(ΔEkbT))1,
where ΔE=E2E1 is the energy difference between levels (1) and (2).

However, since the decay rates are different, the system is not in equilibrium for t>0 unless Pr=Pn=0, which gives the uninteresting solution: I(t)=Ntotalet/τ. So, we force the system to maintain thermal equilibrium by adding the equilibrium term to the system of rate equations (Fig. 7). A term AEQ represents the movement of the population to maintain thermal equilibrium. It is worth noting that the re-thermalization of the excited state manifold should happen on time scales much faster than the spontaneous lifetimes of any of the levels.

The rate equations are straightforward:

Eq. (11)

dN2dt=A20N2(N2N1)AEQdN1dt=A10N1+(N2N1)AEQdN0dt=A20N2+A10N1.

We impose the condition:

Eq. (12)

N2[t]N1[t]=N2[0]N1[0],
and solve for AEQ to obtain:

Eq. (13)

AEQ=N2[t]N1[t]N12[t]N22[t]τ1τ2τ1τ2,
where τ1=1/A10, and τ2=1/A20.

Substituting this expression back into Eq. (11) and solving again provides a new expression for the decay of the upper manifold. Thus,

Eq. (14)

I(t)(1Pr)N2[t]+(1Pn)N1[t],
becomes

Eq. (15)

I(t)=(N2[0](1Pr)+N1[0](1Pn))exp(N2[0]tτ1+N1[0]tτ2N2[0]τ1τ2+N1[0]τ1τ2)=Kexp(N2[0]t(N2[0]+N1[0])τ2)exp(N1[0]t(N2[0]+N1[0])τ1)=Kexp(κ2t(κ2+κ1)τ2)exp(κ1t(κ2+κ1)τ1),
where K=N2[0](1Pr)+N1[0](1Pn). It is worth noting that

Eq. (16)

κ2+κ1=(1+G1G2exp(ΔEkbT))1+(1+G2G1exp(ΔEkbT))1=1

So,

Eq. (17)

I(t)=Kexp(κ2tτ2)exp(κ1tτ1).

The product of exponentials is an exponential, so the form of the spontaneous decay of the manifold is clearly a single exponential with a lifetime τ, and

Eq. (18)

1τ=κ1τ1+κ2τ2.

Using W=1/τ, and substituting explicitly for κ2 and κ1, we obtain:

Eq. (19)

Wobs=W11+G2G1exp(ΔEkbT)+W21+G1G2exp(ΔEkbT).

3.2.

A Shorter Derivation

In this section, we offer a shorter derivation of Eq. (19) without the need for full rate equation analysis. Neglecting, for now, the energy distribution of its vibronic sublevels, the average relaxation rate of the upper manifold is simply the weighted sum of the relaxation rates of each of its electronic levels

Eq. (20)

NtotalWobs=n=1MNnWn,
where the total population is Ntotal=nNn and M is the total number of sublevels. For simplicity, we treat the manifold as consisting of two energy nondegenerate meta-levels. Level (1) has a population N1 and undergoes NRET with some probability P1. Level (2) has a population N2 and undergoes re-absorption with some probability P2 (we assume 100% re-emission).

It is useful to define a parameter that describes the partitioning of the manifold into these two meta-levels. The ratio of the degeneracies G1 and G2 of the meta-levels is

Eq. (21)

G2G1=n=m+1Mgnn=1mgn,
where M is the total number of sublevels in the manifold and gn is the degeneracy factor of the nth level. However, this partitioning of the manifold is arbitrary, so G2/G1 must be treated as a fit parameter.

We introduce thermal statistics by noting that the fractional population distribution is described as

Eq. (22)

N2N1=G2G1exp(ΔEkbT),
where ΔE is the separation of the two meta-levels. Substituting Eq. (22) into Eq. (20), we find

Eq. (23)

Wobs=W11+G2G1exp(ΔEkbT)+W21+G1G2exp(ΔEkbT)

3.3.

Multi-Phonon Cascade

The Mott–Seitz model describes the effect of temperature on the fluorescence lifetime of active ions decaying to the ground state by spontaneous emission in competition with a non-radiative process such as multi-phonon cascade (MPC).12,19,20 The measured lifetime τobs of the decay depends on the thermally dependent probability of the non-radiative decay process. The observed lifetime is

Eq. (24)

τobs=τrad·η=τrad·ArAr+Anrexp(E/kbT),
where τrad is the lifetime of the ion due to spontaneous radiative decay only, η is the fluorescence efficiency of the aggregate decay process, Ar is the probability of radiative decay, Anr is the probability of nonradiative decay, and the exponential term is the familiar Boltzmann statistical term with an activation energy E. This expression can be rewritten as

Eq. (25)

τobs=τrad1+Cexp(E/kbT),
where C describes the magnitude of nonradiative decay. We recast Eq. (25) in terms of related rates:

Eq. (26)

Wobs=Wrad(1+Cexp(E/kbT))=Wrad+Cexp(E/kbT)=Wrad+Wnr.

We use this functional form as the basis for the construction of this model. We will define the total decay rate Wtotal=Wobs+Wnr, where Wnr is a nonradiative decay term that we will introduce below and Wobs is the internal rate we derived in the previous sections:

Eq. (27)

Wobs=W11+G2G1exp(ΔEkbT)+W21+G1G2  exp(ΔEkbT).

NRET will increase the decay rate W1 relative to the spontaneous rate. In the absence of NRET, W1 will approach the radiative rate. Similarly, RET will artificially lengthen the observed fluorescence lifetime; this effect is equivalent to a decrease in the decay rate W2 relative to the spontaneous rate. In the absence of RET, W2 will approach the radiative rate. In the absence of both effects, Wrad will approach the radiative rate at all temperatures. Thus, we see that this model is merely a more general form of the Mott–Seitz model.

We now replace the nonradiative term Wnr in Eq. (26) with the Miyakawa–Dexter expression for nonradiative quenching (NRQ) due to MPC21 to obtain

Eq. (28)

Wtotal=Wobs+WMPC(p),
where

Eq. (29)

WMPC(p)=AM22πp(Sep)p2(1+1exp(hνeffkbT)1)p,
and where kb is the Boltzmann’s constant, e is the Euler’s constant, A is a coupling constant for the nonradiative term, M is an effective mass, S is the Huang–Rhys parameter, T is the temperature, and hνeff=E/p is the effective phonon energy required for p phonons to bridge the energy gap E between the energy levels involved in the optical transition.

Using this approach, we can construct a model for the observed fluorescence lifetime τobs=1/Wtotal that includes both energy transfer and MPC:

Eq. (30)

Wtotal=W11+G2G1exp(ΔEkbT)+W21+G1G2exp(ΔEkbT)+AM22πp(Sep)p2(1+1exp(hνeffkbT)1)p.

4.

Measuring the Fluorescence Lifetime

We performed measurements of the fluorescence lifetime of Fe2+ ions in chemical vapor deposition-grown ZnSe using a periodically poled lithium niobate (PPLN) optical parametric oscillator (OPO) pumped by a Spectra-Physics Nd:YVO4 laser. The pump laser operated at 1 kHz and the PPLN grating had a quasi-phasemathing period of 31  μm. Thus, at 100°C, the OPO emitted at 3050 nm. The full-width-half-maximum linewidth of the output was 8 nm. The average power was >20  mW and the average pulse length was 30  ns. Pulses from the OPO were focused into a sample of Fe:ZnSe, which was wrapped in indium foil and pressure mounted to a cold-finger within an evacuated cryostat. The decay of the optical output was recorded by focusing fluorescent emission through an Acton Spectra-Pro 750 monochromator onto a liquid-nitrogen cooled InSb detector.

The fluorescent decay signals were recorded with an oscilloscope and the lifetime τ of the impurity ions was determined as a least-sum-squared error fit of

Eq. (31)

I(t)=Aexp(tt0τ)+a0
to the measured decay trace of the collected fluorescence. Figure 8 shows a typical fit for an Fe:ZnSe sample.

Fig. 8

A numerical fit of Eq. (31) to an experimental decay trace from Fe:ZnSe.

OE_60_5_056103_f008.png

We performed measurements of the fluorescence lifetime of other transition metal gain media using the same pulsed excitation technique. Short laser pulses were focused into the sample of interest, which was wrapped in indium foil and pressure mounted to a cooling mount. Different pump lasers were used for the various media because no single pump laser was appropriate for all samples. The decay of the optical output was recorded by focusing fluorescent emission onto an appropriate photodiode and Eq. (31) was fit to the decay trace to determine the lifetime.

In the sections that follow, we will fit Eq. (30) to the datasets we have collected in this way as well as to datasets we have sampled from the relevant literature. Because we have access to the raw decay traces from our dataset, we are confident that these data are well fit by a single exponential decay model characterized by a single florescent lifetime τ. As we have shown, this single-exponential behavior is to be expected from a multiplicity of single-exponential processes that proceed simultaneously in a region that is (nearly) in thermal equilibrium. We do not have access to the decay traces for the literature data to which we have also fit our model. In some cases, the multi-variable fit has a lower degree of confidence than we obtained for fits to our dataset. This can be partially explained by three considerations: First, our data were taken using a great degree of computer control, which was not available to some earlier researchers. Second, the literate data were sampled graphically from published figures, which will introduce some degree of discretization error. Third, the energy transfer processes in the materials that other researchers have analyzed may not proceed in equilibrium, in which cases the observed decay may not be well characterized by a single exponential. Nevertheless, a single exponential decay model has been used in the lifetimes reported in all cases. Despite these sources of potential error, we show that Eq. (30) accommodates trends in temperature that are a general feature of such datasets.

5.

Analysis

We recast νeff in Eq. (30) in terms of p by substituting hνeff=Δ/p, where Δ=10Dq is the crystal field energy of the impurity ion and the value of Δ is obtained spectroscopically. So, we obtain a model of the form

Eq. (32)

Wtotal=W11+G2G1exp(ΔEkbT)+W21+G1G2exp(ΔEkbT)+AM22πp(Sep)p2(1+1exp(10Dq/pkbT)1)p,
which includes seven fit parameters: τ1=1/W1, τ2=1/W2, G2/G1, ΔE, A/M2, S, and p. The model can then be fit to the decay traces of various transition metal laser media.

5.1.

Fe2+-doped Materials

In our previous work,11 we fit Eq. (32) to the data we collected at that time and to two other measurements of the thermally resolved lifetime of Fe:ZnSe from the literature. The results are discussed in detail in that work; we summarize those results here briefly. Equation (32) fits the data very well and the confidence intervals are small relative to the fluorescence lifetime. Figures 9(a)9(c) shows the confidence intervals of the numerical fit are reasonable.

Fig. 9

A numerical fit of Eq. (30) to the lifetime data for Fe:ZnSe from the (a) Myoung et al.,18 (b) Evans et al.,11 and (c) Adams et al.14 Colored bands show the 90%, 95%, and 99% confidence intervals.

OE_60_5_056103_f009.png

The fit parameters of Eq. (32) are shown in Table 1. The parameters A/M2, S, and p, which are associated with nonradiative decay via MPC, are similar across all samples. The remaining parameters are associated with the radiative decay process and with energy transfer are seen to vary across samples. We hypothesized that concentration effects are responsible for the differences observed in these parameters between samples.

Table 1

The returned values of the fit parameters for our model of the temperature-dependent fluorescence lifetime of Fe2+ ions in ZnSe. It is worthy to note that A/M2, p, and S are similar across columns.

Myoung18Evans11Adams14
N (cm3)1.0×10178.6×10185×1019
τ1 (μs)54.5±0.651.8±1.034.5±3.6
τ2 (μs)80108245
G2/G14.84±0.255.64±1.87.5±0.9
ΔE (cm1)90.1±15.994.0±16.047.8±5.9
A/M2 (ms1)10.8±0.710.2±1.29.8±2.4
S5.465.475.44
p16.0±0.715.69±0.317.2±0.8

5.1.1.

Fe:ZnSe

In this work, we repeated the measurements of our previous work.11 We automated our data collection using LabVIEW and used a more stable laser (described in the experimental section) with a higher pulse repetition frequency for improved smoothing via averaging. The signal-to-noise ratio improvement achieved using this technique is evident in the confidence intervals of parameters of the fits. We examined Fe:ZnSe samples with various concentrations to eliminate experimental variables from the conclusion that concentration is the key variable affecting the lifetime of Fe:ZnSe. Each of the samples was purchased from IPG Photonics and the concentrations quoted by the vendor span two orders of magnitude. The fit of Eq. (32) to each of the samples is shown in Figs. 10(a)10(c).

Fig. 10

A numerical fit of Eq. (32) to the lifetime data for diffusion-doped Fe:ZnSe from IPG: (a) sample 1, (b) sample 2, and (c) sample 3.

OE_60_5_056103_f010.png

A notable trend with respect to concentration is observed. With increasing concentration, the curvature of lifetime curve with respect to temperature increases significantly with increasing Fe2+ concentration.

For comparison, a sample of Fe:ZnSe was grown solvothermally. Interestingly, the behavior of that material is consistent with the Mott–Seitz model (see Fig. 11). One might assume that a sample would approach that regime in the limit of low concentration, but the concentration of this sample, as inferred from optical transmission N=log(I/I0)/σL, is about the same as the sample interrogated to produce Fig. 10(b). Clearly the trends differ significantly, so concentration is not the only variable contributing to these observations.

The two samples are fabricated by different growth methods; the IPG crystal was made via PGTD and the in-house crystal by solvothermal growth. This is worth noting because previous work has shown that Fe:ZnSe lasers based on crystals grown by PGTD behave significantly different after treatment under hot isostatic press (HIP),22 which suggests that crystals grown via PGTD are highly defective. Such defects may increase the probability of inter-ionic energy transfer. Thus, it is possible that the observed differences in the thermal lifetime are due to energy transfer processes coupled to these defects.

Fig. 11

A numerical fit of Eq. (32) to the lifetime data for solvothermally grown Fe:ZnSe (N5.7×1018  cm3).

OE_60_5_056103_f011.png

Analysis of the data in this section shows that measured value of the lifetime below 260 K depends, in part, on the volume concentration of the Fe2+ dopant. Above 260 K, however, the effect of concentration is seen to be less significant. As shown in Fig. 12, the measured value of the lifetime from many samples of different ionic concentrations converge into relative agreement at temperatures >260  K. It is worth noting that the literature values were determined graphically, and some degree of sampling error is present in both Fig. 12 and Table 2.

Fig. 12

The measured values of the fluorescence lifetime of Fe2+ ions in ZnSe as reported in literature and as measured in this work (TW).

OE_60_5_056103_f012.png

Table 2

The measured values of the fluorescence lifetime of Fe2+ ions in ZnSe at room temperature as reported in literature and as measured in this work.

SourceT (K)N (cm−3)τ (ns)
Myoung et al.183011×1017326.8
Song et al.153001.5×1019377.1
Martinez23299444.1
Evans113058.6×1018581.7
Solvothermal (TW)3005.7×1018550.8
IPG 1 (TW)3008.45×1017541.6
IPG 2 (TW)3004.8×1018489.7
IPG 3 (TW)3001.5×1019515.0

The data from Fig. 12 near room temperature are shown in Table 2. Also included is a room temperature measurement of the fluorescent lifetime of Fe2+ ions in ZnSe nanoparticles reported by Martinez.23 The measured value of the fluorescent lifetime is approximately 500 ns across all concentrations. Values between 300 and 400 ns are reported at the highest and lowest values of concentration. No clear trend of the measured lifetime with respect to concentration can be discerned. This topic should be further investigated.

5.1.2.

Fe:CdMnTe

A sample of Fe:CdxMn1xTe was cut from a boule grown from melt by Brimrose Corporation using the Bridgman technique with an Fe2+ impurity concentration of 2.7×1018  ions/cm3 and an alloying fraction of approximately x=0.9. Mid-IR fluorescence was generated in the sample via excitation with 40-ns laser pulses centered at a wavelength of 3.45  μm from MgO:PPLN OPO (Λ=30.5  μm). The decay traces were recorded directly using an InSb detector and a digital oscilloscope using the same setup as the previous section. The temperature dependent fluorescence lifetimes and the fit of Eq. (32) to that data are shown in Fig. 13. The fit parameters are shown in Table 3.

Fig. 13

A numerical fit of Eq. (32) to the lifetime data for Fe:CdMnTe (N2.7×1018  cm3). It is worthy to note the incredibly tight confidence intervals of the fit.

OE_60_5_056103_f013.png

Table 3

The returned values of the fit parameters for our model of the temperature-dependent fluorescence lifetime of Fe2+ ions in ZnSe. It is worthy to note that p and S are similar across the Fe:ZnSe samples.

Fe:ZnSeFe:ZnSeFe:ZnSeFe:ZnSeFe:CdMnTe
IPG sample 1IPG sample 2IPG sample 3Solvothermal
N (cm3)8.5×10174.8×10181.5×10195.7×10182.7×1018
τ1 (μs)48.3±0.350.5±0.359.3±0.650.3±0.580.0±.2
τ2 (μs)54.1±5.758.3±7.512252195±12.9
G2/G16.544.304.444.949.0±1.6
ΔE (cm1)101±68.690.0±50.894.0±16.0110116±4.7
A/M2 (ms1)18.6±0.912.3±0.510.4±1.010.4±0.527.2±0.3
S5.355.445.505.495.84
p16.4±0.216.7±0.216.4±0.416.1±0.416.5±0.1

5.2.

Cr2+-Doped Materials

Thermally dependent lifetime datasets for Cr:ZnS, Cr:ZnSe, and Cr:ZnTe were obtained from DeLoach et al.16 and were fit using Eq. (32). Likewise, the thermally dependent lifetime datasets for Cr:ZnSe and Cr:CdSe were obtained from McKay24 and were fit to the same equation. The fit parameters for each material are shown in Table 4. The data and the corresponding fits are shown in Fig. 14 (Cr:ZnS), Fig. 15 (Cr:ZnTe), Figs. 16(a) and 16(b) (Cr:ZnSe), and Fig. 17 (Cr:CdSe).

Table 4

Fit parameters for the fluorescence lifetime several Cr2+-doped samples. The fit for Cr:ZnTe is not consistent with the other materials and the data does not fit our model well. The data may indicate strong concentration quenching effects.

Cr:ZnSCr:ZnSeCr:ZnTeCr:CdSe
DeLoach et al.16McKay24DeLoach et al.16McKay24
N (cm3)1.8×10191×10185×1018(26)×1018
τ1 (μs)6.81±0.075.21±0.072.63±0.145.37±0.088
τ2 (μs)8.7±1.312.5±0.68.76.00±0.08
G2/G14.083.110.664
ΔE (cm1)144±38188±14.0150150
A/M2 (μs1)1.006±0.00670.433±0.0240.0112.86±0.005
S5.14±0.075.11±.065.035.35
p27.3±0.629.5±1.715.825.0±0.7

Fig. 14

A numerical fit of Eq. (32) to the lifetime data for Cr:ZnS from DeLoach et al.16 (N1.8×1019  cm3) Colored bands show the 90%, 95%, and 99% confidence intervals.

OE_60_5_056103_f014.png

Fig. 15

A numerical fit of Eq. (32) to the lifetime data for Cr:ZnTe from DeLoach et al.16 (N5×1018  cm3) Colored bands show the 90%, 95%, and 99% confidence intervals.

OE_60_5_056103_f015.png

Fig. 16

A numerical fit of Eq. (32) to the lifetime data for Cr:ZnSe from (a) McKay24 and (b) Deloach.16 Colored bands show the 90%, 95%, and 99% confidence intervals.

OE_60_5_056103_f016.png

Fig. 17

A numerical fit of Eq. (32) to the lifetime data for Cr:CdSe from McKay24 (N1.75.9×1018  cm3). Colored bands show the 90%, 95%, and 99% confidence intervals.

OE_60_5_056103_f017.png

5.3.

Co2+-Doped Materials

Thermally dependent lifetime datasets for Co:ZnS, Co:ZnSe, and Co:ZnTe were obtained from DeLoach et al.16 and were fit using Eq. (32). The thermally dependent lifetime dataset for Co:CdTe collected for this work was also used to fit to the same equation. The fit parameters for each material are shown in Table 5. The data and the corresponding fits are shown in Fig. 18 (Co:ZnS), Fig. 19 (Co:ZnSe), Fig. 20 (Co:ZnTe), and Fig. 21 (Co:CdTe).

Table 5

Fit parameters for the fluorescence lifetime several Co2+-doped samples. The fits for Co:ZnS and Co:ZnTe did not require an additional radiative term. It is worth noting the dramatic difference in τ1 from host to host. The fit for Co:CdTe did not depend critically on the value of τ2 and the values of S and p returned do not seem consistent with the other materials shown.

Co:ZnSCo:ZnSeCo:ZnTeCo:CdTe
DeLoach et al.16DeLoach et al.16DeLoach et al.16This work
N (cm3)1.8×10193×10182.5×10193.5×1019
τ1 (μs)1180±6.8904588.8±10.0591±3.6
τ2 (μs)1318±65>1000
G2/G144
ΔE (cm1)202±19159.5±7.1
A/M2 (ms1)0.467±0.0401.29±0.140.33±0.0720.247±0.061
S4.984.244.688.61±.053
p19.3±0.522.2±0.621.9±1.832.7±0.73

Fig. 18

A numerical fit of Eq. (32) to the lifetime data for Co:ZnS from DeLoach et al.16 (N1.7×1019  cm3) Colored bands show the 90%, 95%, and 99% confidence intervals.

OE_60_5_056103_f018.png

Fig. 19

A numerical fit of Eq. (32) to the lifetime data for Co:ZnSe from DeLoach et al.16 (N3×1018  cm3) Colored bands show the 90% and 95% confidence intervals.

OE_60_5_056103_f019.png

Fig. 20

A numerical fit of Eq. (32) to the lifetime data for Co:ZnTe from DeLoach et al.16 (N2.5×1019  cm3) Colored bands show the 90%, 95%, and 99% confidence intervals.

OE_60_5_056103_f020.png

Fig. 21

A numerical fit of Eq. (32) to the lifetime data for Co:CdTe taken as a part of this work (N2.5×1019  cm3). Colored bands show the 90%, 95%, and 99% confidence intervals.

OE_60_5_056103_f021.png

6.

Tables

This section presents values returned from the fit of Eq. (32):

Wtotal=W11+G2G1exp(ΔEkbT)+W21+G1G2exp(ΔEkbT)+AM22πp(Sep)p2(1+1exp(10Dq/pkbT)1)p
to the data shown in the Analysis section. Values proceeded by a “∼” are approximate. Values in bold had to be held fixed for the fit to converge. Values reported without uncertainties are known to high degrees of confidence, so the uncertainties are omitted for formatting reasons.

7.

Conclusions

The model we have detailed in this work fits several measurements of the thermally resolved fluorescent lifetime for several II–VI crystals with different ionic dopants and with various impurity concentrations. The model offers a quantitative explanation for observations in transition metal laser gain media that have not been previously explained. Our analysis of the data for Fe:ZnSe shows conclusively that the Fe2+ concentration of the samples drives departure from the Mott–Seitz model for thermally dependent fluorescent lifetime. The model also provides strong evidence that NRQ via MPC is responsible for the temperate-activated reduction of the fluorescence lifetime in several classes of transition metal II–VI materials.

One interesting implication of this work—especially for Fe:ZnSe lasers, for which the lifetime of the sample can be double its radiative value—is that the energy storage lifetime of the gain medium can be increased, without decreasing the cross-section of the transition, as is the case in crystal field engineering techniques. This is possible because the observed increase in lifetime is due to re-absorption effects, not to a reduction in the spontaneous decay probability of the optical transition. Consequently, the saturation intensity is not affected.

Due to the large amount of available data, the trends in Fe:ZnSe are well understood, but future work will be required to understand the trends in other materials. Future work will focus on collecting more lifetime data for materials, which are promising candidates for use in mid-IR lasers.

Acknowledgments

We acknowledge and thank the Air Force Office of Scientific Research (AFOSR) and the Sensors Directorate of the Air Force Research Laboratory (AFRL) for funding this research. We thank Glen Perram of the Air Force Institute of Technology for technical discussions. Additionally, we thank Pamela Evans for proofreading this manuscript.

References

1. 

J. J. Adams et al., “Tunable laser action at 4.0 microns from Fe:ZnSe,” in Adv. Solid State Lasers, WD3 (1999). https://doi.org/10.1364/ASSL.1999.WD3 Google Scholar

2. 

V. Fedorov et al., “tunable solid-state lasers based on Fe2+-doped ZnSe crystals operating at low and room temperatures,” IEEE J. Quantum Electron., 42 907 –917 (2006). https://doi.org/10.1109/JQE.2006.880119 IEJQA7 0018-9197 Google Scholar

3. 

V. A. Akimov et al., “Efficient lasing in a Fe2+:ZnSe crystal at room temperature,” Quantum Electron., 36 (4), 299 (2006). https://doi.org/10.1070/QE2006v036n04ABEH013139 QUELEZ 1063-7818 Google Scholar

4. 

A. A. Voronov et al., “A continuous-wave Fe2+:ZnSe laser,” Quantum Electron., 38 (12), 1113 (2008). https://doi.org/10.1070/QE2008v038n12ABEH013837 QUELEZ 1063-7818 Google Scholar

5. 

M. E. Doroshenko et al., “Fe:ZnSe laser—comparison of active materials grown by two different methods,” Proc. SPIE, 7912 79122D (2011). https://doi.org/10.1117/12.876082 PSISDG 0277-786X Google Scholar

6. 

J. W. Evans, P. A. Berry and K. L. Schepler, “840 mW continuous-wave Fe:ZnSe laser operating at 4140 nm,” Opt. Lett., 37 5021 –5023 (2012). https://doi.org/10.1364/OL.37.005021 OPLEDP 0146-9592 Google Scholar

7. 

M. P. Frolov et al., “Study of a 2J pulsed Fe:ZnSe 4  μm laser,” Laser Phys. Lett., 10 (12), 125001 (2013). https://doi.org/10.1088/1612-2011/10/12/125001 1612-2011 Google Scholar

8. 

J. Evans, P. Berry and K. Schepler, “A passively Q-switched, CW-pumped Fe:ZnSe laser,” IEEE J. Quantum Electron., 50 204 –209 (2014). https://doi.org/10.1109/JQE.2014.2302233 QUELEZ 1063-7818 Google Scholar

9. 

D. V. Martyshkin et al., “High average power (35 W) pulsed Fe:ZnSe laser tunable over 3.84.2  μm,” in Conf. Lasers and Electro-Opt., SF1F.2 (2015). https://doi.org/10.1364/CLEO_SI.2015.SF1F.2 Google Scholar

10. 

N. Myoung et al., “Energy scaling of 4.3  μm room temperature Fe:ZnSe laser,” Opt. Lett., 36 94 –96 (2011). https://doi.org/10.1364/OL.36.000094 OPLEDP 0146-9592 Google Scholar

11. 

J. W. Evans et al., “Optical spectroscopy and modeling of Fe2+ ions in zinc selenide,” J. Luminesc., 188 541 –550 (2017). https://doi.org/10.1016/j.jlumin.2017.04.017 JLUMA8 0022-2313 Google Scholar

12. 

N. F. Mott and R. W. Gurney, Electronic Processes in Ionic Crystals, Dover, New York(1964). Google Scholar

13. 

J. W. Evans et al., “Re-absorption and nonradiative energy transfer in vibronic laser gain media,” Proc. SPIE, 10511 105110I (2018). https://doi.org/10.1117/12.2290822 PSISDG 0277-786X Google Scholar

14. 

J. J. Adams et al., “4.04.5-μm lasing of Fe:ZnSe Below 180 K, a new mid-infrared laser material,” Opt. Lett., 24 1720 –1722 (1999). https://doi.org/10.1364/OL.24.001720 OPLEDP 0146-9592 Google Scholar

15. 

Y. Song et al., “Mid-infrared light emission from a Fe2+:ZnSe polycrystal using quantum cascade laser pumping,” Appl. Phys. Lett., 105 (14), 141108 (2014). https://doi.org/10.1063/1.4897546 APPLAB 0003-6951 Google Scholar

16. 

L. DeLoach et al., “Transition metal-doped zinc chalcogenides: spectroscopy and laser demonstration of a new class of gain media,” IEE J. Quantum Electron., 32 885 –895 (1996). https://doi.org/10.1109/3.502365 Google Scholar

17. 

H. Jelínková et al., “Fe:ZnSe laser oscillation under cryogenic and room temperature,” Proc. SPIE, 8599 85990E (2013). https://doi.org/10.1117/12.2003840 PSISDG 0277-786X Google Scholar

18. 

N. Myoung et al., “Temperature and concentration quenching of mid–IR photoluminescence in iron doped ZnSe and ZnS Laser crystals,” J. Luminesc., 132 (3), 600 –606 (2012). https://doi.org/10.1016/j.jlumin.2011.10.009 JLUMA8 0022-2313 Google Scholar

19. 

B. Henderson and G. F. Imbusch, “Optical Spectroscopy of Inorganic Solids,” Clarendon Press, Oxford (1989). Google Scholar

20. 

V. Pagonis et al., “Modelling the thermal quenching mechanism in quartz based on time-resolved optically stimulated luminescence,” J. Luminesc., 130 (5), 902 –909 (2010). https://doi.org/10.1016/j.jlumin.2009.12.032 JLUMA8 0022-2313 Google Scholar

21. 

T. Miyakawa and D. L. Dexter, “Phonon sidebands, multiphonon relaxation of excited states, and phonon-assisted energy transfer between ions in solids,” Phys. Rev. B, 1 2961 –2969 (1970). https://doi.org/10.1103/PhysRevB.1.2961 Google Scholar

22. 

J. W. Evans, R. W. Stites and T. R. Harris, “Increasing the performance of an Fe:ZnSe laser using a hot isostatic press,” Opt. Mater. Express, 7 4296 –4303 (2017). https://doi.org/10.1364/OME.7.004296 Google Scholar

23. 

A. Martinez, “Spectroscopic characterization of iron-doped II–VI compounds for laser applications,” The University of Alabama at Birmingham, (2015). Google Scholar

24. 

J. B. McKay, Power scaling feasibility of chromium-doped II–VI laser sources and the demonstration of a chromium-doped zinc selenide face-cooled disk laser (2002). Google Scholar

Biography

Jonathan W. Evans is a research physicist at the Air Force Research Laboratory at Wright–Patterson Air Force Base, Ohio. He earned his BS degree in electrical engineering from Cedarville University in 2008. He earned his MS degree in electro-optics from the University of Dayton in 2010 and his PhD in optical sciences and engineering from Air Force Institute of Technology in 2014. His research interests include transition metal lasers, nonlinear parametric conversion, and spectroscopy of novel laser systems.

Thomas R. Harris has worked in the Optoelectronic Technology Branch of the Air Force Research Laboratory, Sensors Directorate at Wright–Patterson Air Force Base since 2015, first as an NRC Postdoctoral research associate and more recently as a contractor with Azimuth Corp. He earned his PhD in applied physics from the Air Force Institute of Technology in 2014. His interests include infrared spectroscopy and transition-metal and rare-earth doped laser host materials.

Eric J. Turner is a laser systems research engineer at KBR. He has experience in solid state laser materials research and development. He earned his BS degree in electrical engineering from Howard University and his MS degree in electro-optics from the University of Dayton.

Kenneth L. Schepler received his BS degree in physics from Michigan State University and MS and PhD degrees from the University of Michigan. He was a research physicist at the Air Force Research Laboratory (AFRL) at Wright–Patterson Air Force Base for over 32 years. He retired in 2014 and is now a research professor at CREOL. His interests include solid state lasers, laser spectroscopy, and nonlinear frequency conversion. He is a fellow of OSA and AFRL.

Biographies of the other authors are not available.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 Unported License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Jonathan W. Evans, Thomas R. Harris, Eric J. Turner, Martin M. Kimani, J. Matthew Mann, Ronald W. Stites, Gary Cook, and Kenneth L. Schepler "Re-absorption and nonradiative energy transfer in vibronic laser gain media," Optical Engineering 60(5), 056103 (19 May 2021). https://doi.org/10.1117/1.OE.60.5.056103
Received: 30 November 2020; Accepted: 21 April 2021; Published: 19 May 2021