A single multimode optical fiber is used to excite and collect tissue autofluorescence as well as the fluorescence of an IV-injected fluorescent tumor marker. Measurements of the relative fluorescence intensity of a tumor marker as a function of the time after IV injection permit measurement of the kinetics of this substance in tumor, normal tissue, and skin. The authors believe that these are the first measurements of this kind in patients. Furthermore, the autofluorescence spectrum generated at several excitation wavelengths in different tissues is compared, for instance in the oesophagus, the bronchi, and the tongue. The measuring system is based on an optical multichannel analyzer which measures the fluorescence excited by monochromatic radiation from a spectrally filtered Xe lamp. A correlation between the observed pharmacokinetics and tumor properties like the degree of vascularization is of fundamental importance for each selected tumor marker. Also, the results of these measurements are used for the optical detection of tumors.
|