Paper
17 September 2012 Updating the Chandra HETGS efficiencies using in-orbit observations
Author Affiliations +
Abstract
The efficiencies of the gratings in the High Energy Transmission Grating Spectrometer (HETGS) were updated using in-light observations of bright continuum sources. The procedure first involved verifying that fluxes obtained from the +1 and -1 orders match, which checks that the contaminant model and the CCD quantum efficiencies agree. Then the fluxes derived using the high energy gratings (HEGs) were compared to those derived from the medium energy gratings (MEGs). The flux ratio was fit to a low order polynomial, which was allocated to the MEGs above 1 keV or the HEGs below 1 keV. The resultant efficiencies were tested by examining fits to blazar spectra.
© (2012) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Herman L Marshall "Updating the Chandra HETGS efficiencies using in-orbit observations", Proc. SPIE 8443, Space Telescopes and Instrumentation 2012: Ultraviolet to Gamma Ray, 844348 (17 September 2012); https://doi.org/10.1117/12.927209
Lens.org Logo
CITATIONS
Cited by 2 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Magnetoencephalography

Quantum efficiency

Sensors

Calibration

Data modeling

Charge-coupled devices

Spectroscopy

RELATED CONTENT

X-ray CCD calibration for the AXAF CCD Imaging Spectrometer
Proceedings of SPIE (October 31 1996)
The AXAF CCD Imaging Spectrometer
Proceedings of SPIE (July 14 1986)
Absolute effective areas of the HETGS
Proceedings of SPIE (November 19 1998)