Paper
7 March 2016 Development of 3D holographic endoscope
Meriç Özcan, Duygu Önal Tayyar
Author Affiliations +
Abstract
Here we present the development of a 3D holographic endoscope with an interferometer built around a commercial rigid endoscope. We consider recording the holograms with coherent and incoherent light separately without compromising the white light imaging capacity of the endoscope. In coherent light based recording, reference wave required for the hologram is obtained in two different ways. First, as in the classical holography, splitting the laser beam before the object illumination, and secondly creating the reference beam from the object beam itself. This second method does not require path-length matching between the object wave and the reference wave, and it allows the usage of short coherence length light sources. For incoherent light based holographic recordings various interferometric configurations are considered. Experimental results on both illumination conditions are presented.
© (2016) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Meriç Özcan and Duygu Önal Tayyar "Development of 3D holographic endoscope", Proc. SPIE 9771, Practical Holography XXX: Materials and Applications, 97710A (7 March 2016); https://doi.org/10.1117/12.2211569
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Endoscopes

Holography

Interferometers

Holograms

Holographic interferometry

3D image reconstruction

Light sources

Back to Top