Open Access
1 September 2005 Algorithms for automated oximetry along the retinal vascular tree from dual-wavelength fundus images
Author Affiliations +
Abstract
We present an automated method to perform accurate, rapid, and objective measurement of the blood oxygen saturation over each segment of the retinal vascular hierarchy from dual-wavelength fundus images. Its speed and automation (2 s per entire image versus 20 s per segment for manual methods) enables detailed level-by-level measurements over wider areas. An automated tracing algorithm is used to estimate vessel centerlines, thickness, directions, and locations of landmarks such as bifurcations and crossover points. The hierarchical structure of the vascular network is recovered from the trace fragments and landmarks by a novel algorithm. Optical densities (OD) are measured from vascular segments using the minimum reflected intensities inside and outside the vessel. The OD ratio (ODR=OD600/OD570) bears an inverse relationship to systemic HbO2 saturation (SO2). The sensitivity for detecting saturation change when breathing air versus pure oxygen was calculated from the measurements made on six subjects and was found to be 0.0226 ODR units, which is in good agreement with previous manual measurements by the dual-wavelength technique, indicating the validity of the automation. A fully automated system for retinal vessel oximetry would prove useful to achieve early assessments of risk for progression of disease conditions associated with oxygen utilization.

1.

Introduction

Oxygen utilization is altered in many disorders of the retina.1, 2, 3, 4, 5, 6 The present work is motivated by the need for developing automatic and noninvasive ways of measuring the percent blood O2 saturation from retinal images. Evaluations of blood oxygenation along vascular segments of individual artery and vein networks can show, respectively, the distributions of oxygen supply and extraction from local regions. Here we describe automated image analysis techniques based on dual-wavelength oximetry7 and present results from normal subjects of indices from retinal vessels that are proportional to O2 saturation. Additional information concerning the vessel diameter, which has previously been correlated with blood flow changes,8 is forthcoming in the analysis. Future applications for an automatic system could include large-scale screening for assessing risk for progression of the stages of diabetic retinopathy, postoperative monitoring or image-guided intervention with patients receiving laser treatments for proliferative retinal vessel disease, and application to change detection studies of the retinal blood oxygen saturation.

Existing methods of vessel oximetry acquire images at multiple wavelengths, where measurements are limited to a small region in the image.7, 9, 10, 11, 12 In the present method, digital images of continuous vessel segments of several millimeters length, containing artery and vein pairs, are acquired. The automated image analysis identifies serial segments of the vessels in two images, one sensitive to changes in the percent of HbO2 present in the blood, and the other a reference image that is not sensitive.7 Figure 1 shows two images acquired simultaneously at 570nm and 600nm . The 570-nm image, which is insensitive to HbO2 change, has higher contrast between the blood vessels and the background than does the image at 600nm , where differential contrast between the artery and vein comes from the higher and lower HbO2 content in each vessel. The apparent OD of the blood inside vessels is measured from reflectance values taken from inside and outside of the vessels. Algorithms that trace the blood vessels, reconstruct the vascular hierarchy, and measure the apparent OD of the blood contained in the retinal vessels are described below.

Fig. 1

Sample images from subject 3 at oxygen-insensitive and oxygen-sensitive wavelengths. (a) Insensitive image acquired at 570nm . (b) Sensitive image acquired at 600nm and registered to the 570-nm image. The 570-nm image shows higher contrast between the blood vessels and the background, and reflected light from the vein (upper vessel) and artery (lower vessel) are comparable at this wavelength. At 600nm , the vessel contrast is low and the artery appears lighter than the vein.

054013_1_036505jbo1.jpg

Many advances and refinements in technique for retinal oximetry have been made over the past four decades.7, 9, 10, 11, 12, 13, 14, 15, 16, 17, 18, 19, 20 Hickam 9, 10 were the first to report studies of retinal oxygenation in man by noninvasive measurements, using photographic methods to record vessel OD at oxygen-sensitive and insensitive wavelengths, and ratio analysis to obtain relative measures of oxygen saturation. Later, Laing 11 extended this method by showing that blood saturation varied linearly with the ratio over physiological and hypoxic ranges of saturation. Delori developed a three-wavelength oximetry method to obtain absolute oxygen saturation values from single vessel segments.12 Delori’s oximeter was the first to demonstrate altered blood saturation in optic atrophy.13 A method to perform retinal vessel oximetry by spectroscopic recording was introduced by Schweitzer 14 Relative measures of oxygen saturation along retinal vessels using a ratiometric measurement were reported by Beach, 7 using a digital imaging technique to simultaneously record oxygen-sensitive and insensitive images. Jensen15 and Critten 16 have employed the ratio technique successfully to measure oxygen saturation changes in retinal vessels. Recently Khoobehi 17 employed hyperspectral imaging to monitor the relative spatial changes in retinal oxygen saturation of the retina and optic disc tissue in the monkey. Uzumcu 18 obtained measurements at six wavelengths and estimated the parameters related to the oxygen saturation using an iterative algorithm. A scanning laser ophthalmoscope that images at four wavelengths was used by Smith 19, 20 They suggested that the choice of wavelengths is critical and used an optimized wavelength set. The analysis was done on a small selected region from the images. Studies of the relationship between oxygen utilization and retinal disorders are also extensively discussed in the literature.1, 2, 3, 4, 5, 6

The motivation for our image analyses was to enable higher levels of automation and precision using the dual-wavelength approach with digital imaging. We have also added the capability to automatically distinguish the oxygen saturation values from the different levels of the vessel hierarchy. Our vessel hierarchy algorithm automatically reconstructs the vascular structure and hence enables the comparison at the corresponding levels in the vascular hierarchy. This ability should be especially useful in venous networks for localization of areas of relative hypoxia.

Several methods have been proposed in the literature for detection of vascular features from retinal images.21, 22, 23, 24, 25, 26, 27, 28, 29, 30, 31, 32, 33, 34, 35, 36, 37, 38, 39, 40, 41, 42, 43, 44, 45, 46 These methods can be broadly classified into four categories.The first approach is based on adaptive filtering or segmentation.21, 22, 23, 24, 25, 26, 27, 28, 29, 30 In the second method, referred to as vessel tracking or tracing, local image properties are used to trace the vasculature recursively starting from an initial point.31, 32, 33, 43, 44, 45, 46 Some of these algorithms exploit a priori knowledge on the position of one or more reference points belonging to the vessels. In some of the methods commonly used in quantitative coronary analysis, the initial and end points of the vessel are entered manually.32 The disadvantage of these methods is that they are not fully automated. The third approach uses mathematical morphology to detect long tubular objects representing blood vessels.38, 39, 40, 41 The fourth approach uses classification methods to segment vessels.42 Chaudhuri assume a Gaussian cross-sectional profile for the vessels and detect them by convolving with a matched filter.21 Hoover use the matched filter response with an adaptive threshold probing.22 Gang detect vessels using amplitude modified second-order Gaussian filter.23 Other methods based on matched filters are described in Refs. 24, 25. Directional edge detectors were used by Li, 28 while local image characteristics were exploited for detecting the blood vessels in window-based methods described in Refs. 29, 30. A parameterized variation of the exploratory approach known as “snakes” is discussed in Refs. 34, 35. Aylward and Bullitt36 describe a method based on intensity ridge traversal to extract centerlines of tubular objects. Jiang and Mojon37 implement adaptive local threshold using multiple threshold verification and a curvilinear structure model based on angle, width, and contrast. Zana and Klein use mathematical morphology to detect vessel-like structures.38 Morphological processing is also described in Refs. 39, 40, 41 for finding blood vessels from retinal images. An unsupervised classification algorithm was used by Fontaine for segmenting the vasculature.42 In this paper we use the exploratory tracing algorithm developed by Can. 43 and further developed in Refs. 44, 45, 46. The algorithm will be described in detail in the next section.

2.

Fully-Automatic Retinal Vessel Tracing Algorithm

Previous research in this group has developed algorithms for rapid automated tracing and feature extraction from retinal images.43, 44, 45, 46 These algorithms were originally inspired by the need to construct computer vision tools for assisting ophthalmic procedures such as laser retinal surgery and perimetry. The tracing method is based on adaptive exploratory processing of the image, directly at the image intensity level, avoiding image-wide pixel-processing operations such as gray-scale filters, edge operators, morphological filters, etc. The algorithm is designed such that the computations are restricted to the most relevant and promising locality of pixels. In the present work we extend these algorithms to develop an automated system to measure the oxygen content in the blood from images of the retina obtained at two different wavelengths.

The tracing algorithm is based on a parallel edge model. Over short distances, the vessel segments can be assumed to be linear and having parallel edges oriented along a particular direction. These parallel edges can be detected by using a set of two-dimensional correlation kernels. The exploratory algorithm starts from seed points that are selected automatically using a two-step detection and verification stage from a uniformly spaced grid. Each seed is assigned a direction during the verification stage and the tracing algorithm proceeds by tracing in these initial directions from each valid seed point. The algorithm takes a step in this direction and then determines the direction to proceed at that point. Given a point pk with initial direction sk , the directional templates are used to estimate the next point pk+1 and direction sk+1 in a recursive fashion. Figure 2(a) shows how the algorithm takes a step in the current direction and then estimates an orientation at the new point. The exploratory algorithm terminates if it meets certain stopping conditions. Interested readers are referred to Refs. 43, 44 for details of the tracing algorithm. At each point in the vessel centerline, the algorithm estimates the width, boundary points and also the direction. From the centerline network, the branching and crossover points are detected as the intersection points of the centerline trace segments. These points are referred to as landmarks and help in determining the hierarchy of the vascular tree as described below. Each landmark has two or more vessel segments associated with it and also stores the angles that these segments make with the landmark. This is illustrated in Fig. 2(b).

Fig. 2

Recursive tracing algorithm and landmarks. (a) The vessel edges can be modeled to be parallel over small distances. The templates for two different orientations are shown in panel (a). Separate templates are used for the left and right edges. The direction of tracing at point pk is sk . From that point the algorithm takes a step in the direction sk . The new direction sk+1 is found at pk+1 by the application of the templates. (b) Landmarks obtained from the vessel centerlines. Each landmark has a set of traces associated with it and also the angle the vessel extends at it. This information is used by the vessel linking algorithm to reconstruct the vascular hierarchy.

054013_1_036505jbo2.jpg

3.

Reconstructing the Vascular Hierarchy

The detected trace segments from the exploratory tracing algorithm do not give any information about the vascular hierarchy. Also the same vessel might be traced in parts as several fragments. If we are able to automatically identify fragments that belong to the same vessel and also its branches, we would be able to make much more accurate measurements from a single connected vessel rather than making the measurements from the individual fragments. The information about the vascular hierarchy is also desirable for picking appropriate vessels to compare the oxygen saturation for detecting changes. The traces are therefore processed using a recursive algorithm that reconstructs the vascular structure.

The traces are first sorted according to the decreasing order of the product of their lengths and average width and put in a priority list. This is done to make sure that the vessel linking algorithm starts from segments that are significantly wide and long. Starting from the first vessel in the sorted list, the algorithm checks whether this segment has any landmarks associated with it. If there are no landmarks associated with this parent segment, it is given a unique name and removed from the list. On the other hand, if the parent segment has one or more landmarks associated with its end points, the segments that are associated with these landmarks are processed and named one at a time as follows. While selecting from multiple segments, associated through the same landmark, the segments that are closer in direction to the current segment are processed first followed by segments that make larger angles with the current segment. Also if there are multiple approximately collinear segments associated with the current segment, it is considered probable that the other segments are part of another vessel and are therefore, not named. When naming the segments, the following rules are followed.

  • 1. If the intensity of the current segment and the previous segment are significantly different (30%), then the current segment is not named.

  • 2. If the directions of the current segment and the previous segment are the same or if their widths are comparable, the current segment is given the same name as that of the previous segment.

  • 3. If 1 and 2 are not satisfied, the current segment is given a name that represents it as the child of the previous segment. This is done by appending -D- (to denote daughter) followed by a number, to the end of the name of the previous segment. It is also ensured that no two branches of the same vessel at the same level have the same name.

In order to avoid naming a single segment more than once, the named segments are removed from the list and eliminated from further consideration. Once all the connected segments under one parent segment have been processed, the algorithm checks the list and process the next segment. The algorithm proceeds until there are no more elements in the list to be processed. A pseudo code description of the above algorithm is provided in Fig. 3 , and Fig. 4 shows the original image, traced image, initial trace segments and their numbers and the reconstructed vessel structure. The vessel linking algorithm is limited by the accuracy of the initial tracing and landmark extraction and cannot recover from errors in the tracing. Another source of error is significant changes in the width or intensity of parts of the same vessel. For example, segment 10 in Fig. 4(c) is really a daughter of the vein marked as V. But in this case, the intensity of segment 10 was significantly higher from that of V, and hence the algorithm mistook it to be part of another vessel. Subsequent segments (64 and 15) are also named based on the name given for segment 10. Hence naming errors at the higher levels of the hierarchy are propagated to the lower levels.

Fig. 3

Pseudo code for the recursive vessel segment naming algorithm. Starting with the longest traced segment, find the segments connected to that segment and name them in a recursive fashion. This results in a hierarchical representation of the vascular network.

054013_1_036505jbo3.jpg

Fig. 4

(a) The original oxygen-insensitive image from subject 1 to be processed using the recursive vessel linking algorithm. Vein (top vessel) and artery (small vessel below the vein). (b) Traces and landmarks for the image shown in (a). The green points are the vessel centerlines and the red lines indicate the detected landmarks (bifurcations and crossovers). (c) Initial trace fragments and trace numbers for the traces shown in (b). Each fragment is colored differently for clarity. (d) Illustrating the hierarchical linking of vessel traces to reconstruct the vascular tree. The hierarchical names given to the fragments shown in (c) are shown in red along the original fragments. The starting number for the vessel names have been replaced with A and V for clarity in differentiating arteries and veins.

054013_1_036505jbo4.jpg

4.

Optical Density Measurements

Images of the eye are obtained at 570nm and 600nm . The tracing algorithm is applied to the image obtained at 570nm . This is because at this wavelength, there is better contrast between the vessels and the background resulting in better detection of the vasculature. The 600-nm image is registered to the 570-nm image using a similarity transformation. This is a rigid transformation that allows for translation, rotation, and scaling between two images. The registration parameters for our images were determined using the image registration feature in IPLab (Scanalytics).

The apparent OD can be calculated on the vessel segments by finding the minimum intensity inside the vessel and the average outside intensity. For each point pk in a vessel centerline, let rk and lk denote the right and left boundary points. In order to find the pixel with the minimum reflection at this particular cross section of the vessel, the pixels along the straight line joining rk and lk are examined. The minimum intensity inside the vessel at pk is given by:

Eq. 1

Iin(pk)=argmin{I[trk+(1t)lk]},
where, I(m) is the image intensity at point m and t is a parameter that varies from 0 to 1.

Extravascular light reflection at point pk is measured from points lying at one vessel diameter distance on either side of the vessel boundary. It is given by:

Eq. 2

Iout(pk)=12{I(rk+Duk)+I(lkDuk)},
where D is the diameter of the vessel at that point and uk is the unit vector in the direction perpendicular to sk . In order to minimize the effects of noise on the measurements of extravascular light reflection, we can substitute the values in the above equations with a local average. Equation 2 is modified to:

Eq. 3

Iout(pk)=12{Avgw[rk+Duk]+Avgw[lkDuk]},
where

Eq. 4

Avgw[p]=1(2w+1)2i=wwj=wwI(px+i,py+j).
Figures 5 and 6 show, respectively, the above notation as applied to the image of the vessel, and the sample points for calculating the minimum values inside vessels and average values outside vessels. Now we describe two methods to calculate the apparent OD for a segment.

Fig. 5

Illustration of the ODR calculation method. The blue dots on the vessel indicate the centerline point pk , the left boundary point, lk and the right boundary point, rk . The unit vector in the direction perpendicular to the vessel orientation is denoted by uk and D is the diameter of the vessel at pk . The two red dots on the outside of the vessel indicate the points at which the extra vascular reflection is measured and the black boxes represent the local neighborhood over which the average is calculated. The white dot represents the minimum intensity point inside the vessel.

054013_1_036505jbo5.jpg

Fig. 6

Illustration of the OD calculation method. (a) The centerlines in white and the pixels across each cross section used for finding the minimum intensity in blue (b) The minimum intensity pixel found across each cross section in white and the points that are used for calculating the extravascular reflectance in red.

054013_1_036505jbo6.jpg

Case 1. In this method, the apparent OD is calculated by averaging the inside intensities and the outside intensities and finding one OD for the segment. Let Iin,seg and Iout,seg be the average inside and outside values for trace segment i with M points:

Eq. 5

Iin,seg(i)=1Mk=1MIin(pk),

Eq. 6

Iout,seg(i)=1Mk=1MIout(pk).
Then the OD for segment i is calculated as:

Eq. 7

OD570,seg(i)=log10(Iout,segIin,seg).

Case 2. This variation of the method is aimed at dealing with any illumination gradient across the image. In this method, we calculate the OD for each centerline point in the trace segment and then average the ODs to obtain the overall OD for the segment. The OD at centerline point pk is given by:

Eq. 8

OD570(pk)=log10(Iout(pk)Iin(pk));
and the average OD for segment i with M points is calculated as:

Eq. 9

OD570,seg(i)=1Mk=1MOD(pk),

Eq. 10

=1Mk=1Mlog10(Iout(pk)Iin(pk)).
We also calculate the standard deviation of the OD values for each segment. Denoting the standard deviation as SD(i) , it is calculated as:

Eq. 11

SD570,seg(i)={1M1k=1M[OD570(pk)OD570,seg(i)]2}12.
Two lists are created one each for the OD570 and the SD570 values of the segments.

The 600-nm image that was registered to the 570-nm image is then processed as described above to get OD600 and the SD600 . The centerlines and boundary points used for this are the same as the ones obtained from tracing the 570-nm image. The ratio of the OD at 600nm to that at 570nm is computed as the ODR:

Eq. 12

ODR=OD600OD570.
The ODR measurement was found to be related to the oxygen saturation in the blood. This will be discussed in detail inSec. 6.

5.

Description of the Retinal Image Analysis

Experimental data were obtained from seven volunteer subjects and the recordings confirmed to an approved institutional review board protocol. Subjects’ eyes were dilated with 1% topical tropicamide and 2.5% phenylephrine. Recordings were made simultaneously from retinal artery vein pairs using a dual-wavelength camera that is described in Ref. 7. The present instrument uses an inexpensive 16-bit digital camera (Apogee U-2) and upgraded optics to improve wavelength selection and the pixel resolution is 9×9μm . Images containing larger retinal vessels and side-branch vessels were recorded during room air breathing and again while the subject breathed pure oxygen (10Lmin) from a mask that was attached over the nose and mouth. The same retinal areas were selected for low and high oxygen. Arterial saturation was read with a pulse oximeter using a finger probe to assess changes in arterial saturation. The retina was illuminated with the xenon flash from the fundus camera (Topcon TRC).

Recorded dual-wavelength retinal images were processed first by separating them into 570-nm and 600-nm images as shown in Fig. 1, and these separate images were registered. The centerline traces of vessels in the 570-nm image were found using the tracing algorithm. Results from the tracing were used as inputs to the vessel hierarchy algorithm. Figures 4(a) and 4(b) show, respectively, a 570-nm vessel image with individually identified vessel segments overlaid by centerline tracings and connecting landmarks that are used to link segments based on their hierarchical structure. The full complement of detected segments in the image was then resolved into serialized segments along the vascular tree as shown in Fig. 4(d). Vessel segment hierarchy is defined by us using a naming scheme where the large parent vessel is identified with a single number, daughter vessels having the same number and the designation D which follows, and multiple daughters or side branches that occur along the vessel moving distally being assigned higher numbers. This naming scheme is marked on the image of vessels to confirm validity.

The algorithm for vessel hierarchy has been validated by an ophthalmologist (one of the authors). For each image, the vessel names were superimposed on the segments and presented for validation. Major vessels and its branches were selected and the segments that were given a name inconsistent with the convention were marked out. The success rate of the vessel linking algorithm is defined as the number of trace segments correctly named by the algorithm to the total number of trace segments. For the images considered, the success rate was found to be 87%. It has to be noted that the success of the vessel linking algorithm is limited by accuracy of the vessel tracing algorithm. Specifically, out of the 13% of the errors made by the algorithm, 7% were caused due to the errors in finding the bifurcation and crossing points by the tracing algorithm. The other errors were caused due to the big differences in width and intensity of the branches of the same vessel, in which case the algorithm mistook the branch to be part of a different vessel.

Along centerline traces, lines perpendicular to the vessel orientation were followed to sample minimum-valued pixels that estimate the reflected light from the blood column as shown in Fig. 6(a). Along the same line, approximately one vessel diameter from the outer wall, the extravascular reflection was averaged from neighboring pixels. OD was determined either by averaging values from inside and outside the vessel and then applying Eq. 7 for Case 1, or by finding the OD for each pair of values along the length of the segment and then averaging these as in Eq. 9 for Case 2, in either case ending up with a particular OD for each segment. Case 2 should reduce the influence of light gradients on OD values since these are determined at points along vessels rather than from area-averaged values. Centerline traces from 570-nm images were applied to the registered 600-nm and ODs were obtained for 600-nm images. ODRs were then calculated as described above.

6.

Determination of Oxygen Change Sensitivity

Data from six subjects were used to determine the oximeter response resulting from changing the inspired gas from room air to pure oxygen. Table 1 compares ODRs from normal and high oxygen conditions in ten artery and eleven vein segments of the superior temporal vessel arcades. Systemic arterial oxygen saturation averaged 97% in the subjects during room air breathing and 100% during pure oxygen. ODs were determined from averaged intensities along the vessel segments (Case 1). These vessels began one disk diameter from the rim of the optic disk. The ODR of the arteries and veins decreases when breathing gas was switched from room air to pure oxygen. The mean change for the artery was 0.0226 ODR units. This change is very close to the expected 0.0236 ODR units for a blood saturation increase of three percentages, assuming an instrument sensitivity of 0.00787 ODR/change in percent saturation that was established with a similar methodology and vessel analysis using Case 1.7 The high degree of agreement between the two methods confirms that our automated analysis gives the same result as the previous vessel tracking technique which required user intervention to pick vessels and determine the size of each scanned segment before performing the analysis.

Table 1

Comparison of ODR (Case 1) at different SO2 levels for the selected arteries and veins. SO2 was read from a pulse oximeter with a finger probe. The ODR decreases as the systemic SO2 increases. The vessel naming convention is that A (V) represents a first level vessel, and A-D-* (V-D-*) represents a daughter vessel branching off from the main vessel. The first number in the name given by the vessel linking algorithm has been replaced with A and V for clarity in distinguishing arteries (A*) and veins (V*) . SO2 in room air=97% , in pure O2=100% .

SubjectVesselidentifierODRroom air Mean±SD room airODRpure O2 Mean±SD pure O2 Difference
Arteries
Subject 1A0.2450 0.1688±0.0379 0.2362 0.1462±0.0387 0.0226
Subject 2A0.21820.1810
Subject 3A0.15670.1423
A-D-10.19190.1412
Subject 4A0.13670.1218
Subject 5A0.14020.1204
A-D-10.16440.1542
A-D-20.14410.1397
Subject 6A0.13120.0950
A-D-10.15990.1354
Veins
Subject 1V0.5104 0.4065±0.0957 0.4184 0.3315±0.0758 0.0750
V-D-10.45930.4087
V-D-20.47590.3727
Subject 2V0.48740.3295
Subject 3V0.39870.3080
V-D-10.36510.3330
Subject 4V0.32350.3043
Subject 5V0.51730.4125
V-D-10.26370.2234
Subject 6V-10.27820.1622
V-20.23190.1774

Oxygen breathing caused a greater change in the ODRs of veins; the decrease observed for the veins was 3.31 times that of the artery, which corresponds to an increase in venous saturation of approximately 10%. The venous saturation changes should significantly exceed those in arteries since pure oxygen causes the tissue oxygen source to switch from hemoglobin to dissolved oxygen in plasma, with the result that there is less desaturation of the venous blood. Table 2 gives results when the vessel ODs were found by the alternate method of first determining OD at each location along the segments and then averaging these (Case 2). Case 2 had little effect on the outcome; ODRs of arteries and veins (room air) were 2.3% higher and 1.2%lower, respectively, than the corresponding values from Case 1. In arteries, there was no significant difference between the two methods in the response to breathing oxygen. In veins, Case 2 resulted in a 9% greater decrease in the ODR. Figure 7 shows vessels from two subjects for room air (left panels) and pure O2 (right panels), with the ODR calculated using Case 2 and vessel diameters marked on selected veins and arteries. Both the datasets show the reduction in the ODR, which corresponds to increased saturation, and in the reduced diameter, which resulted from oxygen-induced vasoconstriction. In the lower panels, the responses to O2 were greater; reduced vessel diameters are easily discerned from the images and both parameters changed by larger degrees. Table 3 gives the intermediate steps for the ODR calculation for the two subjects shown in Fig. 7. Although the table has many entries, all of the values are computed by the automatic algorithm, and the user only needs to analyze the results.

Fig. 7

Optical density ratios (ODR) and the measured vessel widths (pixels) in parentheses for selected arteries (A) and veins (V) from subject 5 (a,b) and subject 6 (c,d). (a,c) Room air breathing, (b,d) pure O2 breathing.

054013_1_036505jbo7.jpg

Table 2

Comparison of ODR (Case 2) at different SO2 levels for the selected arteries (A*) and veins (V*) . The vessel naming convention is the same as in Table 1. SO2 in room air=97% , in pure O2=100% .

SubjectVesselidentifierODRroom air Mean±SD room airODRpure O2 Mean±SD pure O2 Difference
Arteries
Subject 1A0.2321 0.1726±0.0340 0.2207 0.1500±0.0317 0.0226
Subject 2A0.22500.1860
Subject 3A0.15690.1461
A-D-10.18730.1504
Subject 4A0.13850.1317
Subject 5A0.15620.1205
A-D-10.18130.1545
A-D-20.15980.1405
Subject 6A0.13000.1155
A-D-10.15840.1343
Veins
Subject 1V0.5163 0.4018±0.0900 0.4194 0.3198±0.0931 0.0820
V-D-10.45830.4096
V-D-20.47630.3690
Subject 2V0.48450.3292
Subject 3V0.39750.3070
V-D-10.36840.3326
Subject 4V0.32330.3026
Subject 5V0.48810.4136
V-D-10.27380.2237
Subject 6V-10.28270.1569
V-20.23140.1725

Table 3

Intermediate ODR calculation (Case 2) for the segments shown in Fig. 7. The table shows the name of the segment, the index of the original trace segments that were linked together by the algorithm, the individual ODs at 570nm and 600nm , and the ODR. The standard deviation (SD) is calculated over the points in each segment. The vessel naming convention is the same as in Table 1. SO2 in room air=97% , in pure O2=100% .

ImageVesselidentifierNumber oforiginal tracefragments linkedto form thesegment 570nm 600nm ODR
ODSDODSD
Arteries
Subject 5(room air)A20.18630.04940.02910.00990.1562
A-D-120.17820.05680.03230.00830.1813
A-D-230.16460.01200.02630.00650.1598
Subject 5(high O2 )A30.21080.01520.02540.01470.1205
A-D-140.17220.02770.02660.01770.1545
A-D-220.17940.01580.0220.01910.1405
Subject 5(room air)A10.62920.09930.08180.01680.1300
A-D-110.57970.14730.09180.02550.1584
Subject 5(high O2 )A20.59390.05890.06860.02010.1155
A-D-130.45500.12760.06110.02950.1343
Veins
Subject 6(room air)V50.23030.08200.11240.00900.4881
V-D-110.09570.02010.02620.00680.2738
Subject 6(pure O2 )V90.20190.02250.08350.00890.4136
V-D-110.07780.02490.01740.00610.2237
Subject 6(room air)V-130.55400.06240.15660.01920.2827
V-210.71100.07960.16450.02170.2314
Subject 6(pure O2 )V-120.69320.13520.11960.01520.1725
V-220.71690.12440.11250.01250.1569

Repeatability of this method should be assessed from artery data since the saturation change is more consistent than in veins. For results from arteries in Table 1, the coefficient of variation (COV=SEMEAN) for saturation changes equals 0.21 (N=10) . This measure of variability in part reflects differences in the actual rise in retinal arterial saturation among the subjects that resulted from breathing pure oxygen.

The automated method also allows studying the changes in vessel diameters with oxygen saturation. The vessel diameters are estimated by the tracing algorithm at each trace point and the mean width for each segment is calculated. Table 4 shows the change in diameter of the blood vessels under room air and pure oxygen breathing for the six subjects. The mean decrease was 7.5% for the arteries and 9.93% for the veins for the increase of 3% in oxygen saturation.

Table 4

Comparison of the vessel diameters at different O2 levels. The percentage decrease in the diameter is calculated as (D1−D2)∕D1100* , where D1 and D2 are the measured diameters of the vessel at room air and pure O2 breathing, respectively, in pixels. The vessel naming convention is the same as in Table 1. SO2 in room air=97% , in pure O2=100% .

SubjectVesselidentifierDiameterroom air (D1) Diameterpure O2 (D2) Percentage decrease
Arteries
Subject 1A10.6010.431.60
Subject 2A8.498.92 5.06
Subject 3A14.1711.8316.51
A-D-110.4710.024.30
Subject 4A16.5312.2925.65
Subject 5A9.959.524.32
A-D-18.098.10 0.12
A-D-26.646.167.82
Subject 6A14.3113.158.11
A-D-110.188.9711.89
Mean decrease±SD for arteries 7.50±8.83
Veins
Subject 1V15.0012.9213.87
V-D-111.6410.1912.46
V-D-210.519.0913.51
Subject 2V11.449.9413.11
Subject 3V14.9912.7315.08
V-D-18.367.4311.12
Subject 4V8.6310.72 24.22
Subject 5V12.1811.317.14
V-D-14.324.173.47
Subject 6V-113.059.8124.83
V-213.6011.0318.90
Mean decrease±SD for veins 9.93±12.61

Plotted in Fig. 8 are values of the ODR versus distance along a vessel segment using Case 2, for air and pure O2 breathing. The expected reduction in the ODR with increased blood oxygen saturation is evident, while a gradient in saturation along the vessel is also observed for both cases. These data show how it would be possible to measure oxygen gradients along retinal vessels, or even discontinuous changes resulting from vessel occlusion.

Fig. 8

ODR plotted against position along a vein segment during room air (diamond symbols) and pure O2 (circles) breathing for subject 4. In this case, pure O2 increased the vein saturation (reduced ODR). The vessel showed a continuous positive gradient in saturation left to right (negative ODR gradient).

054013_1_036505jbo8.jpg

Vessel traces have been applied to larger images of the retina that were taken with the full frame of the CCD camera using the 55-deg field-of-view setting on the fundus camera. Figure 9 shows images of temporal vessels with trace points included. In Fig. 9(a), the 570-nm recording wavelength produces high contrast between both artery and vein against the fundus background. These traces followed the larger vessels and the numerous smaller side branches faithfully. In Fig. 9(b) the 600-nm recording, which produces lower vessel contrast, is also faithfully traced. In this, the only case of a thinned retina that we have worked with, our method correctly identified retinal vessels in both the oxygen-insensitive and oxygen-sensitive images. Most of the small traces seen in Fig. 9(b) between the large artery and the vein (darker vessel) do connect with one of the large vessels, hence these are likely to be smaller branching arteries. Even though the myopic eye in Fig. 9 was traced faithfully, there is a possibility that the algorithm might mistake choroidal pigment bands in myopic subjects as vessels. Hence, more tests with this kind of subjects are necessary to know whether the algorithm will always succeed where there are choroidal pigment bands.

Fig. 9

(a) Illustration over larger areas of the retina with traces of vessel interior (white points) and exterior (red points). Oxygen-insensitive image showing vessels arch over the macula (dark region at bottom) from subject 7. The nerve fiber layer is visible near vessels. (b) Oxygen-sensitive image (600nm) from subject 7, where vessels arch below the macula. Retinal thinning as a condition of myopia produces intensity variations (light areas). The traces were obtained from the corresponding 570-nm image and are superimposed on the registered 600-nm image. Based on these traces, the minimum intensity points inside the vessels (indicated in white) were then determined by looking at the intensities in the 600-nm image.

054013_1_036505jbo9.jpg

6.1.

Implementation and Execution Time

The exploratory tracing algorithm was implemented in C++ and the vessel hierarchy reconstruction and OD measurement algorithms were implemented in Matlab. After the images were registered, the automatic algorithms took around 2s on an average for each subject for the analysis on a PC with Pentium IV Processor and 1GB of RAM. This can be compared to 20s for just a single vessel segment with 50 points included needed for the earlier analysis.7 Also, the automatic method produces ODR measurements over all the vessels in the field of the image, whereas the earlier analysis took almost an order of magnitude more time for a single segment.

7.

Conclusions

The main contribution of this work is the adaptation of automated tracing algorithms for oxygen measurement from dual wavelength images. Automation for identification and analysis of retinal vessels is important for development of clinically useful technology for diagnosis. The steps taken thus far show that oxygen changes are forthcoming from segments of prominent vessels, including arteries and veins in dual-wavelength images. The vessel tracing and linking algorithms make it possible to obtain OD and oxygen change measurements from larger areas than were previously described in the literature. The vessel linking algorithm starts from a main vessel segment and names its side branches and bifurcations hierarchically. The algorithm was found to perform well in all the subjects under study, and we believe that this is the right start for solving complex problems associated with vessel hierarchies. Also our vessel tracing algorithm is able to find the vessel diameters with sufficient precision to detect oxygen vasoconstriction. This is significant since the vessel diameter is an important parameter in the analysis of retinal blood flow. Automated OD measurements have also been successful in smaller (<50μm) veins that branch from the main vessels. Although there is room for improvement, the result show it is feasible for automatic image analysis to indicate areas of relative hypoxia based on distributions of venous blood saturation. One problem that remains is to successfully determine O2 saturation values in the smaller artery segments with automation. The very low vessel OD, strong central reflection and variation in the extravascular reflection at 600nm make this determination difficult.

The choice of measurement wavelengths in the described method are such that the retinal vessels can be easily traced by measuring density against the fundus background, and sensitivity to saturation change is high. Retinal vessels become more visible against the fundus background at wavelengths where hemoglobin light absorption is strongest, between 540 and 580nm where our reference wavelength is set. The sensing wavelength, which is longer than 580nm , provides less vessel contrast but gives the highest ratio of light absorption between saturated and desaturated blood. At wavelengths longer than 610nm , the retinal arteries become difficult to detect. Turbid media causes significant signal loss by scattering as measuring wavelength decreases toward blue wavelengths. Thus the present choice of wavelengths should perform reasonably well in patients with moderately turbid media, although high turbidity limits our method as it does color fundus imaging.

Three wavelength methods have been reported that provide absolute calibration of blood oxygen saturation from retinal vessels.12 By comparison, the two wavelength method described here provides relative saturation indices using the optical density ratio (ODR). The absolute sensitivity of the ODR to saturation changes was previously determined.7 Thus, the method can be used to determine saturation changes in single vessels, or between two vessels. The artery-vein saturation difference (A-V difference) is an important parameter for oxygen utilization. [Venous saturation can be estimated from the A-V difference and the sensitivity of the ODR if the arterial saturation is known, such as during the breathing of pure oxygen.] Venous saturation during normoxia can then be assessed from the decrease in the ODR relative to hyperoxia. Simultaneous imagery of the fundus at two wavelengths can be easily acquired using image splitting techniques; this is significantly more difficult to achieve for higher numbers of wavelength.

Results from the automated analysis were essentially similar to earlier findings with dual-wavelength imaging that employed an automated analysis that had to be initiated manually.7 An area for improvement is in the registration of the 570-nm and 600-nm images. Highly accurate feature based registration algorithms have been developed by our group for retinal images.47, 48, 49 These algorithms depend on reliable feature extraction from the images. In the case of dual wavelength images, the central reflex for arteries in the 600-nm image is so strong that the two sides of the arteries get traced separately as two segments. Although minimum values we seek are along either of these traces, it would be better if the centerline of the vessel was traced as one segment. Since in the 570-nm image, this same vessel gets traced as a single segment through the centerline, the registration algorithm would tend to associate one or the other of the traces on each side of center in the 600-nm image to the centerline from the 570-nm image. This could result in wrong correspondences during registration, causing convergence to a wrong solution. Further research is planned to make the feature extraction better in retinal images with “hollow” vessels, which would allow the registration to be done automatically. Our results indicate that any light gradients present with the current method of fundus camera imaging should not significantly influence the calculation of vessel ODRs, which are the measure here for relative oxygen saturation.

Future work will also concentrate on refining the methods to enable analysis of smaller vessels, particularly the more distal networks of veins that drain blood from local areas of the retina. These vessels carry blood away from the retinal areas that could be affected by altered metabolic activity. This information would allow identification of local hypoxia that is associated with disorders of the retina, and in conjunction with automated techniques, allow for screening of disorders in clinical rather than research environments. The ODR values, while not absolute measures of saturation, are highly sensitive to saturation differences that would be present between areas of normal and low oxygenation. Thus, a diagnostic image that automatically reveals areas of relative hypoxia on the fundus image is one of the important goals of this work. For the other goal of investigating oxygen utilization in the normal and diseased retina, the present stage of development can lead to an image-based measurement of saturations and vessel diameters across the retina. We have demonstrated the feasibility of the methods on normal retina in this paper. Future measurements in patients with retinal disorders will be performed to assess clinical potential for this method.

Acknowledgments

The Rensselaer component of this research was supported in part by the National Science Foundation Experimental Partnerships grant EIA-0000417, the Center for Subsurface Sensing and Imaging Systems, under the Engineering Research Centers Program of the National Science Foundation (Award Number EEC-9986821). We thank our colleagues Kenneth Fritzsche and Charles V. Stewart of the RPI Retina team for use of the retinal vessel tracing core. The team at the Institute for technology Development and LSU acknowledges support from the NIH: EY14776 (Photon Industries, Inc.), EY014872, EY02377, and an unrestricted grant to LSU from Research to Prevent Blindness. Luis Martinez (ITD) also assisted with data collection in these experiments.

References

1. 

F. C. Delori, J. Sebag, G. T. Feke, and J. J. Weiter, “Oxygen saturation of retinal vessels in optic atrophy,” Invest. Ophthalmol. Visual Sci., 27 221 (1986). 0146-0404 Google Scholar

2. 

J. S. Tiedeman, S. E. Kirk, S. Srinivas, and J. M. Beach, “Retinal oxygen consumption during hyperglycemia in patients with diabetes without retinopathy,” Ophthalmologica, 105 31 –36 (1998). 0030-3755 Google Scholar

3. 

R. D. Braun and R. A. Linsenmeier, “Retinal oxygen tension and the electroretinogram during arterial occlusion in the cat,” Invest. Ophthalmol. Visual Sci., 36 (3), 523 –541 (1995). 0146-0404 Google Scholar

4. 

E. Stefansson, M. B. Landers, and M. L. Wolbarsht, “Oxygenation and vasodilatation in relation to diabetic and other proliferative retinopathies,” Ophthalmic Surg., 14 209 –226 (1983). 0022-023X Google Scholar

5. 

E. Stefansson, R. Machemer, E. de Juan Jr., B. W. McCuen II, J. Peterson, “Retinal oxygenation and laser treatment in patients with diabetic retinopathy,” Am. J. Ophthalmol., 113 (1), 36 –38 (1992). 0002-9394 Google Scholar

6. 

J. M. Beach, J. S. Tiedeman, M. F. Hopkins, and Y. S. Sabharwal, “Multi-spectral fundus imaging for early detection of diabetic retinopathy,” Proc. SPIE, 3603 114 –121 (1999). 0277-786X Google Scholar

7. 

J. M. Beach, K. J. Schwenzer, S. Srinivas, D. Kim, and J. S. Tiedeman, “Oximetry of retinal vessels by dual-wavelength imaging: calibration and influence of pigmentation,” J. Appl. Physiol., 86 748 –758 (1999). 8750-7587 Google Scholar

8. 

B. Khoobehi, G. A. Peyman, and K. D. Vo, “Relationship between blood velocity and retinal vessel diameter,” Invest. Ophthalmol. Visual Sci., 33 (4), 804 (1992). 0146-0404 Google Scholar

9. 

J. B. Hickam, R. Frayser, and J. C. Ross, “A study of retinal venous blood oxygen saturation in human subjects by photographic means,” Circulation, 27 375 –383 (1963). 0009-7322 Google Scholar

10. 

J. B. Hickam and R. Frayser, “Studies of the retinal circulation in man: observations on vessel diameter, arteriovenous oxygen difference, and mean circulation time,” Circulation, 33 (1966). 0009-7322 Google Scholar

11. 

R. A. Laing, A. J. Cohen, and E. Friedman, “Photographic measurements of retinal blood oxygen saturation: falling saturation rabbit experiments,” Invest. Ophthalmol. Visual Sci., 14 606 –610 (1975). 0146-0404 Google Scholar

12. 

F. C. Delori, “Noninvasive technique for oximetry of blood in retinal vessels,” Appl. Opt., 27 1113 –1125 (1988). 0003-6935 Google Scholar

13. 

J. Sebag, F. C. Delori, G. T. Feke, and J. J. Weiter, “Effects of optic atrophy on retinal blood flow and oxygen saturation in humans,” Arch. Ophthalmol. (Chicago), 107 (2), 222 –226 (1989). 0003-9950 Google Scholar

14. 

D. Schweitzer, M. Hammer, J. Kraft, E. Thamm, E. Konigsdorffer, and J. Strobe, “In vivo measurement of the oxygen saturation of retinal vessels in healthy volunteers,” IEEE Trans. Biomed. Eng., 46 (12), 1454 –1465 (1999). https://doi.org/10.1109/10.804573 0018-9294 Google Scholar

15. 

P. K. Jensen, “Non-invasive retinal oximetry in normal human subjects,” Computer Assisted Fundus Image Analysis, 28 –30 (2003) Google Scholar

16. 

M. Crittin, H. Schmidt, and C. E. Riva, “Hemoglobin oxygen saturation (So2) in the human ocular fundus measured by reflectance oximetry: preliminary data in retinal veins,” Spectrochim. Acta, Part A, 219 (4), 289 –291 (2002). 0584-8539 Google Scholar

17. 

B. Khoobehi, J. M. Beach, and H. Kawano, “Hyperspectral imaging for measurement of oxygen saturation in the optic nerve head,” Invest. Ophthalmol. Visual Sci., 45 (5), 1464 –1472 (2004). 0146-0404 Google Scholar

18. 

M. Uzumcu, F. M. Vos, A. M. Vossepoel, and G. L. van der Heijde, “Theoretical analysis of a spectrophotometric technique for measuring oxygen saturation in retinal vessels,” Proc. ASCI 2000, 6th Annual Conf. Adv. School Comp. Imag., 117 –121 (2000) Google Scholar

19. 

A. Lompado, M. H. Smith, K. R. Denninghoff, and L. W. Hillman, “Multispectral confocal scanning laser ophthalmoscope for retinal vessel oximetry,” Spectral Imaging: Instrumentation, Applications, and AnalysisProc. SPIE, 3920 67 –73 (2000) (0277-786X) Google Scholar

20. 

M. H. Smith, K. R. Denninghoff, L. W. Hillman, and R. A. Chipman, “Oxygen saturation measurements of blood in retinal vessels during blood loss,” J. Biomed. Opt., 3 296 –303 (1998). https://doi.org/10.1117/1.429856 1083-3668 Google Scholar

21. 

S. Chaudhuri, S. Chatterjee, N. Katz, M. Nelson, and M. Goldbaum, “Detection of blood vessels in retinal images using two-dimensional matched filters,” IEEE Trans. Med. Imaging, 8 263 –269 (1989). https://doi.org/10.1109/42.34715 0278-0062 Google Scholar

22. 

A. Hoover, V. Kouznetsova, and M. Goldbaum, “Locating blood vessels in retinal images by piecewise threshold probing of a matched filter response,” IEEE Trans. Med. Imaging, 19 (3), 203 –210 (2000). https://doi.org/10.1109/42.845178 0278-0062 Google Scholar

23. 

L. Gang, O. Chutatape, and S. M. Krishnan, “Detection and measurement of retinal vessels in fundus images using amplitude modified second-order Gaussian filter,” IEEE Trans. Biomed. Eng., 49 (2), 168 –172 (2002). https://doi.org/10.1109/10.979356 0018-9294 Google Scholar

24. 

L. Zhou, M. S. Rzeszotarski, L. J. Singerman, and J. M. Chokreff, “The detection and quantification of retinopathy using digital angiograms,” IEEE Trans. Med. Imaging, 13 (4), 619 –626 (1994). 0278-0062 Google Scholar

25. 

F. Miles and A. Nuttall, “Matched filter estimation of serial blood vessel diameters from video images,” IEEE Trans. Med. Imaging, 12 (2), 147 –152 (1993). https://doi.org/10.1109/42.232243 0278-0062 Google Scholar

26. 

L. Zhou, M. S. Rzeszotarski, L. J. Singerman, and J. M. Chokreff, “The detection and quantification of retinopathy using digital angiograms,” IEEE Trans. Med. Imaging, 13 (4), 619 –626 (1994). 0278-0062 Google Scholar

27. 

F. Miles and A. Nuttall, “Matched filter estimation of serial blood vessel diameters from video images,” IEEE Trans. Med. Imaging, 12 (2), 147 –152 (1993). https://doi.org/10.1109/42.232243 0278-0062 Google Scholar

28. 

H. Li and O. Chutatape, “Fundus image features extraction,” Proc. 22nd Ann. Inter. Conf. IEEE Eng. Med. Biol. Soc., 4 3071 –3073 (2000). Google Scholar

29. 

T. Pappas and J. S. Lim, “A new method for estimation of coronary artery dimensions in angiograms,” IEEE Trans. Acoust., Speech, Signal Process., 36 (9), 1501 –1513 (1988). https://doi.org/10.1109/29.90378 0096-3518 Google Scholar

30. 

A. Pinz, S. Bernogger, P. Datlinger, and A. Kruger, “Mapping the human retina,” IEEE Trans. Med. Imaging, 17 (4), 606 –619 (1998). https://doi.org/10.1109/42.730405 0278-0062 Google Scholar

31. 

R. Polli and G. Valli, “An algorithm for real-time vessel enhancement and detection,” Comput. Methods Programs Biomed., 52 1 –22 (1997). https://doi.org/10.1016/S0169-2607(96)01773-7 0169-2607 Google Scholar

32. 

Y. Sun, “Automated identification of vessel contours in coronary arteriograms by an adaptive tracking algorithm,” IEEE Trans. Med. Imaging, 8 (1), 78 –88 (1989). https://doi.org/10.1109/42.20365 0278-0062 Google Scholar

33. 

L. Gagnon, M. Lalonde, M. Beaulieu, and M. C. Boucher, “Procedure to detect anatomical structures in optical fundus images,” Proc. SPIE, 4322 1218 –1225 (2001). 0277-786X Google Scholar

34. 

L. H. Staib and J. S. Duncan, “Boundary finding with parametrically deformable models,” IEEE Trans. Pattern Anal. Mach. Intell., 14 (11), 1061 –1075 (1992). https://doi.org/10.1109/34.166621 0162-8828 Google Scholar

35. 

A. Klein, F. Lee, and A. A. Amini, “Quantitative coronary angiography with deformable spline models,” IEEE Trans. Med. Imaging, 16 (5), 468 –482 (1997). https://doi.org/10.1109/42.640737 0278-0062 Google Scholar

36. 

S. R. Aylward and E. Bullitt, “Initialization, noise, singularities, and scale in height ridge traversal for tubular object centerline extraction,” IEEE Trans. Med. Imaging, 21 (2), (2002). 0278-0062 Google Scholar

37. 

X. Jiang and D. Mojon, “Adaptive local thresholding by verification-based multithreshold probing with application to vessel detection in retinal images,” IEEE Trans. Pattern Anal. Mach. Intell., 25 (1), 131 –137 (2003). 0162-8828 Google Scholar

38. 

F. Zana and J. C. Klein, “Segmentation of vessel-like patterns using mathematical morphology and curvature evaluation,” IEEE Trans. Image Process., 10 (7), 1010 –1019 (2001). https://doi.org/10.1109/83.931095 1057-7149 Google Scholar

39. 

G. E. Øien and P. Osnes, “Diabetic retinopathy: automatic detection of early symptoms from retinal images,” Proc. Norwegian Signal Proc. Symp., (1995) Google Scholar

40. 

A. Osareh, M. Mirmehdi, B. Thomas, and R. Markham, “Classification and localization of diabetic-related eye disease,” Proc. 7th Euro. Conf. Computer Vision, 502 –516 (2002) Google Scholar

41. 

T. Walter, J. C. Klein, P. Massin, and A. Erginay, “A contribution of image processing to the diagnosis of diabetic retinopathy detection of exudates in color fundus images of the human retina,” IEEE Trans. Med. Imaging, 21 (10), 1236 –1243 (2002). https://doi.org/10.1109/TMI.2002.806290 0278-0062 Google Scholar

42. 

M. Fontaine, L. Macaire, J. Postaire, M. Valette, and P. Labalette, “Fundus images segmentation by unsupervised classification,” Proc. Vision Interface, (1999) Google Scholar

43. 

A. Can, H. Shen, J. N. Turner, H. L. Tanenbaum, and B. Roysam, “Rapid automated tracing and feature extraction from live high-resolution retinal fundus images using direct exploratory algorithms,” IEEE Trans. Inf. Technol. Biomed., 3 (2), 125 –138 (1999). 1089-7771 Google Scholar

44. 

K. Fritzsche, A. Can, H. Shen, C. Tsai, J. Turner, H. L. Tanenbuam, C. V. Stewart, and B. Roysam, “Automated model based segmentation, tracing and analysis of retinal vasculature from digital fundus images,” Angiography and Plaque Imaging: Advanced Segmentation Techniques, (2002) Google Scholar

45. 

G. Lin, C. V. Stewart, B. Roysam, K. Fritzsche, and H. L. Tanenbaum, “Predictive scheduling algorithms for real-time feature extraction and spatial referencing: application to retinal image sequences,” IEEE Trans. Biomed. Eng., 51 (1), 115 –125 (2002). 0018-9294 Google Scholar

46. 

H. Shen, B. Roysam, C. V. Stewart, J. N. Turner, and H. L. Tanenbaum, “Optimal scheduling of tracing computations for real-time vascular landmark extraction from retinal fundus images,” IEEE Trans. Inf. Technol. Biomed., 5 (1), 77 –91 (2001). 1089-7771 Google Scholar

47. 

A. Can, C. V. Stewart, B. Roysam, and H. L. Tanenbaum, “A feature-based, robust, hierarchical algorithm for registering pairs of images of the curved human retina,” IEEE Trans. Pattern Anal. Mach. Intell., 24 (3), 347 –364 (2002). https://doi.org/10.1109/34.990136 0162-8828 Google Scholar

48. 

A. Can, C. V. Stewart, B. Roysam, and H. L. Tanenbaum, “A feature-based algorithm for joint, linear estimation of high-order image-to-mosaic transformations: mosaicing the curved human retina,” IEEE Trans. Pattern Anal. Mach. Intell., 24 (3), 412 –419 (2002). https://doi.org/10.1109/34.990145 0162-8828 Google Scholar

49. 

C. V. Stewart, C. L. Tsai, and B. Roysam, “The dual-bootstrap iterative closest point algorithm with application to retinal image registration,” IEEE Trans. Med. Imaging, 22 (11), 1379 –1394 (2003). https://doi.org/10.1109/TMI.2003.819276 0278-0062 Google Scholar
©(2005) Society of Photo-Optical Instrumentation Engineers (SPIE)
Harihar Narasimha-Iyer, James M. Beach, Bahram Khoobehi, Jinfeng Ning, Hiroyuki Kawano, and Badrinath Roysam "Algorithms for automated oximetry along the retinal vascular tree from dual-wavelength fundus images," Journal of Biomedical Optics 10(5), 054013 (1 September 2005). https://doi.org/10.1117/1.2113187
Published: 1 September 2005
Lens.org Logo
CITATIONS
Cited by 21 scholarly publications and 4 patents.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Image segmentation

Oxygen

Arteries

Veins

Oximetry

Image processing algorithms and systems

Blood

Back to Top