13 September 2018 Combination of analytical and experimental optical clearing of rodent specimen for detecting beta-carotene: phantom study
Author Affiliations +
Abstract
Recently, compression optical clearing (OC) was applied to detect dermal carotenoid using reflection spectroscopy. To enhance the precision and accuracy of reflection spectroscopy to better detect the spectral absorption of beta-carotene inside biological phantom, here, we simultaneously use compression and immersion OC using dimethyl sulfoxide. In addition, we analytically extract the absorption coefficient of beta-carotene using diffuse reflectance spectroscopy (as an analytical OC). Our results show that the presented analytical OC can be applied alone as a noninvasive method to measure cutaneous chromophores at deep tissues. Finally, we also improve the ability of the analytical clearing method mediated with experimental OC. Our result demonstrates that the combination of analytical and experimental clearing methods enhance the ability of diffuse reflection spectroscopy for extracting the absorption coefficient of beta-carotene as one of the chromospheres inside biological phantom.

1.

Introduction

Determining cutaneous carotenoid plays an important role in human health and nutrition studies.14 High-pressure liquid chromatography (HPLC) is the main in vitro method for determining the amount of beta-carotene. In addition to the high cost, some amount of beta-carotene will be lost after biopsy of tissue, and it will not give a precise value.3,5

Due to the limitation of HPLC, optical methods, such as Raman spectroscopy and reflection spectroscopy, have been utilized to determine the amount of beta-carotene in human and animal skin. The accuracy of these optical methods requires evaluation. In 2010, Ermakov and Gellermann6 investigated the correlation of in vivo skin tissue carotenoid obtained by resonance Raman spectroscopy with HPLC results. Darvin et al.7 measured the content of beta-carotene in the inner palm by in vivo resonance Raman spectroscopy.

The absorption spectrum of beta-carotene is masked by the blood and melanin absorption spectra. Strong scattering of skin complicates additional beta-carotene registration due to broadening and decreasing of absorption peaks. Therefore, one can apply optical clearing (OC) to enhance determination of beta-carotene spectra. Ermakov and Gellermann8 applied the compression mediated reflection spectroscopy to determine the amount of this choromophore. The mechanical compression results in a decrease in the value of the absorption coefficient of skin due to the decrease of total blood content at the location of compression.911 Some additional decrease of light scattering is also expected due to a lower blood volume fraction (as red blood cells are strong scatterers) and an increase in packing density of dermal collagen fibers.9,11 Therefore, the absorption spectrum of beta-carotene can be detectable. Nakhaeva et al.12 illustrated that the pressure exerted on the skin can weaken the effect of blood by reducing the depth of the dip in the green region of the spectrum that indicated the presence of blood in the tissue. Ansari et al.13 also reported that the mechanical compression altered the value of the absorption coefficient. These changes in optical properties can affect the reconstructed concentration of desired chromophore. In addition, the amount of transported blood depends on the value of mechanical pressure and the stiffness of tissue that affects the accuracy of compression-mediated reflectance spectroscopy. Therefore, it is important to verify the accuracy of compression-mediated reflection spectroscopy to determine cutaneous beta-carotene using other OC methods, such as immersion OC.

Polyatomic alcohol (glycerol, polypropylene glycol) and organic solvents [dimethyl sulfoxide (DMSO) and benzyl ether, benzyl alcohol, tetrahydrofuran] have been applied for immersion OC.1419 Mao et al.20 illustrated that some hydroxyl-terminated organic components’ agents (1-butanol, 1,4-butanediol, 1,3-propanediol, PEG200, PEG400, and glycerol) induce significant OC effects of porcine skin. Their results show that the OC effect of skin caused by these alcohols is related to the number of hydroxyl groups.

As Tuchin et al. proposed, the mechanism of immersion OC can be explained by dehydration of tissue constituents and the replacement of interstitial water with an agent that has a higher refractive index (n).14,21 Based on experimental results presented by Rylander et al.,17 application of glycerol causes a 60% decrease in the reduced scattering coefficient in the context of in vitro rodent skin. Zhu et al.22 illustrated that a combination of PEG-400 and thiazone (as immersion agents) can reduce the scattering coefficient (at a wavelength of 630 nm) to 60% of the initial value after 40 min of topical treatment of this mixture on rat skin in vitro.

Beside immersion OC, an analytical method also can be applied for measurement of chromophore inside tissues. For example, Zonios and Dimou23 applied an analytical form of diffuse reflectance from semi-infinite phantom to determine the absorption coefficient of phantom containing polystyrene beads and hemoglobin from bovine blood. Välisuo et al.24 applied a linear model based on the differential Beer–Lambert law to efficiently map the concentrations of blood and melanin in the skin. They calculated an optimal linear model around an operating point, taking into account the spectra of the selected wavelengths, the concentrations of these chromophores in the selected operating point, and the available wavelengths.

In the analytical OC method, a linear summation of different absorption spectra of key tissue constituents (blood, melanin, beta-carotene, and water) is considered to model the measured diffused reflectance with optimum values of chromophore concentration. It seems that immersion and analytical OC can be applied to enhance the capability of compression-mediated reflection spectroscopy to better reconstruct the concentration of chromophores inside tissues.

Motivation for this paper is threefold: first, we investigate the ability of both compression and immersion OC, as experimental OC, to improve the measurement of beta-carotene absorption. Second, we use an analytical OC method based on diffuse reflectance spectrum to detect the absorption spectrum of beta-carotene from experimental spectra. Finally, we show a new way to reconstruct effectively the absorption coefficient of beta-carotene obtained from diffuse reflection spectra based on both analytical and experimental clearing of samples. Hence, the originality of the current study is applying a combination of analytical and experimental OC methods to better measure the concentration of beta-carotene inside tissue phantom.

2.

Materials and Methods

2.1.

Experiment Set Up

As shown in Fig. 1, the experiment setup consists of a tungsten halogen lamp with 20-W power (HL2000; Ocean Optics Inc., Dunedin, Florida) and spectrometer (USB200, Ocean Optics) with a spectral range of 350 to 1100 nm and an optical fiber probe (R200-REF, Ocean Optics, Dunedin). This probe consists of a bundle of seven optical fibers—six illumination fibers around one read fiber—each of which is 200  μm in diameter (NA=0.22, length=2  m) and a distance of source–detector separation of 200  μm. A 3×0.25 stainless steel ferrule houses the fiber bundle. We also use a force gauge to measure the pressure.

Fig. 1

The scheme of experiment setup.

JBO_23_9_095002_f001.png

2.2.

Phantom Preparation

We prepared a two-layer phantom (Fig. 2). The underside layer consisted of Intralipid® 20% (3.3 ml), distilled water (13 ml), agar (0.7 g), and beta carotene [97.0% (UV) Sigma-Aldrich] in different concentrations of 0.2, 0.3, 0.4, and 0.6  μM. The upper layer is a mouse ear (2-month old, 20 g) with an average thickness of 250  μm (one-half of its full thickness is composed of two layers: the epidermis and the dermis, and below the skin layers, cartilage forms the structural support for the mouse ear).25 Eight ear specimens were obtained from Shahid Beheshti Medical Science University and stored at 4°C until experiments were performed (15 min). All animals participating in this experiment were cared for according to the ethic committee of Shahid Beheshti University, Iran.

Fig. 2

A schematic of phantom. Underside layer consists of Intralipid, agar- and beta-carotene, and ear of mouse applied as an upper layer.

JBO_23_9_095002_f002.png

2.3.

Optical Clearing Methods

Here, we used two different OC methods to detect absorption spectra of beta-carotene from reflectance spectra. First, the method of experimental OC (compression-immersion OC) is explained, and then the method of analytical method is stated.

2.3.1.

Experimental method: compression-immersion OC

Here, we utilized DMSO (99%, Sigma-Aldrich) as a chemical OC agent. A solution of DMSO 50% was prepared with ethanol as the solvent. Ear specimens were immersed in DMSO solution for 15 min. Then, the mechanical compression was applied by fiber probe on the specimen, and the reflectance spectra were measured, as depicted in Fig. 1.

2.3.2.

Analytical method

As mentioned in Ref. 23, the analytical form of diffuse reflectance R(λ) from semi-infinite phantom to determine optical properties can be stated as follows:

Eq. (1)

R(λ)=μs(λ)k1+k2μa(λ),
where k1 and k2 are the parameters depended on the probe geometry and refractive indices of the studied sample and were determined to be 0.025  mm1 and 0.075, respectively. μs(λ) is the wavelength-dependent reduced scattering coefficient, and μa(λ) is the wavelength-dependent absorption coefficient of tissue-like phantom. Assume R0 indicates the diffuse reflectance for μa(λ)=0, so μs(λ)=k1R0. This diffuse reflectance R0 can be estimated from the wavelength region of 600 to 900 nm using the scattering power law, as mentioned in Ref. 26 (see Fig. 3). Finally, the absorption coefficient is obtained as follows:

Eq. (2)

μa(λ)=k1k2.[R0R(λ)1].

Fig. 3

Estimation of R0 (dotted line) for measured results (solid line) using scattering power law. For the range λ>600  nm, the spectrum graph (solid line) can be fitted by the scattering power law aλb (dotted line), where a=26.65 and b=0.63 (with 95% confidence bounds).

JBO_23_9_095002_f003.png

In the visible spectrum, blood and melanin are the main chromophores that affect the absorption coefficient of our phantom, while variation of water and lipid does not have significant effects in this spectrum interval. Hence, the absorption coefficient of phantom can be expressed as follows:27

Eq. (3)

μa(λ)phantom=Mμa(λ)mel+Bμa(λ)Beta+Blμa(λ)blood,
where

Eq. (4)

μa(λ)blood=αBlμa(λ)HBO2+(1α)Blμa(λ)HB,
where M, B, and Bl denote melanin, beta-carotene, and blood volume fractions, respectively, and α is the degree of oxygenation of hemoglobin. The absorption spectrum of the mentioned chromophores is selected from the results presented by the “natural phenomena simulation group in University of Waterloo” and data presented by Jacques in 2013.28,29

3.

Results

3.1.

Experimental Clearing

Figures 4(a) and 4(b) demonstrate the effects of experimental clearing methods on diffuse reflectance and absorbance [the absorbance A(λ) is calculated by A(λ)=logR(λ)]. First, we apply compression OC and study the variation of diffuse reflectance (absorbance). The results including error bars show that the removal of blood dip between 500 and 570 nm depends on the amount of mechanical compression. This is because the transport of blood is a function of the amount of compression and Young’s module of tissue. One can see in Fig. 4(a) that, after the application of pressure and when blood is removed, the reflection increases due to less blood absorption. Indeed, scattering is also changed a bit in this range (scattering by RBC), but, in this wavelength range, blood absorption prevails.

Fig. 4

The influence of compression and immersion OC on the spectrum of reflectance and absorbance. The black line shows the (a) reflectance and (b) absorbance before applying experimental clearing. The red line depicts the variations of reflectance (absorption) due to compression OC (300  N/m2). The blue line presents the effects of the combination of immersion and compression OC on (a) reflectance and (b) absorbance. Eight ears are used while doing the experiment for investigating the repeatability of this method. Arrows point out absorption peak of beta-carotene.

JBO_23_9_095002_f004.png

Then, we simultaneously studied the influence of both compression and a chemical clearing agent on tissue phantoms. Blue lines including error bars in Figs. 4(a) and 4(b) reveal a noticeable change in absorbance due to utilizing DMSO, as the chemical OC agent, in OC of a mouse ear. Immersion of ears in DMSO leads to a decrease in optical scattering of them, so one can see an increase in optical depth (diffusion of DMSO in phantom causes a reduction in the scattering coefficient by refractive index matching).30 Results depict that the combined OC (before 600 nm) can enhance the absorption spectra, whereas these two methods have similar effects in an optical window (a wavelength longer than 600 nm).

As shown in Fig. 4 and during application of OC, beta-carotene absorption spectrum appears in its specified wavelength. We define an indicator AFactor=AB, as shown in Fig. 5(a), to clarify the effect of OC on absorbance spectrum (A is the absorbance at λ=490  nm and B is the absorbance at λ=505  nm). Figure 5(b) depicts the amount of AFactor for different values of beta-carotene concentration. The results show a linear relation between the concentration of beta-carotene and AFactor.

Fig. 5

(a) Illustration of AFactor obtained from absorbance spectra of optically cleared phantoms, where A is the absorbance at the wavelength 490 nm and B is the absorbance at the wavelength 505 nm. The black solid line depicts the spectrum before applying compression OC, and the red dotted line illustrates the absorbance spectra of optically cleared phantom after applying compression. (b) Relation between AFactor and concentration of beta-carotene for different phantoms. The black line shows the influence of compression OC, and the red line presents the effects of the combination of compression and immersion OC on AFactor.

JBO_23_9_095002_f005.png

3.2.

Analytical Clearing

Figure 6 compares the modeled absorption coefficient spectra and actual absorption coefficient obtained from diffuse reflectance spectra, as explained in Sec. 2.3.2. As shown in Fig. 6, these two curves are correlated with a high correlation coefficient of R2=0.94, so the fitting parameters indicating the volume fraction of chromophores can be obtained from modeled spectra.

Fig. 6

Modeling measured absorption coefficient (black solid line) using diffuse reflectance presented by the red dotted line.

JBO_23_9_095002_f006.png

As shown in Fig. 7, analytical OC could extract and quantify the absorption coefficient of each chromophore within tissue-like optical phantom without experimental clearing methods for detection of beta-carotene.

Fig. 7

Extracted absorption coefficient based on the optimum volume fraction of main chromophores.

JBO_23_9_095002_f007.png

4.

Discussion

Here, two methods for clearing of tissue-like phantom to extract the absorption of beta-carotene are studied.

4.1.

Experimental Clearing (Compression-Immersion OC)

First, we used mechanical compression to weaken the effect of high blood absorption and obtain the amount of beta-carotene in our phantoms. Ermakov and Gellermann8 investigated the correlation between results obtained by compression mediated reflection spectroscopy and Raman spectroscopy. They illustrated the feasibility of this OC to extract the absorbance spectrum of beta-carotene. Because the blood distribution inside biological tissue depends on the amount of mechanical pressure and the stiffness of tissue, the accuracy of compression mediated reflection spectroscopy can be a key limitation for detecting chromophore at deep depths.

Hence, we used DMSO as a chemical clearing agent during compression-immersion clearing to obtain beta-carotene absorption because DMSO is a well-known skin penetration enhancer and its permeability rate (176±42  g/m2h) is 10 times greater than that of water.17,31 In addition, DMSO induces a reversible conformational change of protein when it substitutes for water.17 By contrast, for example, glycerol does not have a high enough penetration rate since it can hardly penetrate into dermis to achieve OC by topical application because of the barrier of stratum corneum.19 As one can see in Fig. 4(b), applying DMSO as the OC agent improves the detection of the absorption spectrum of beta-carotene.

In this experiment, we use a phantom consisting of a mouse ear and beta-carotene at a depth of 0.2  mm from the surface of phantom. (In fact, we apply a bilirubin-free phantom having a low amount of melanin that influences the results.) Using DMSO and compression helps to decrease the influence of blood and mismatched refractive index to better detect beta-carotene absorption spectra. The efficacy of our method to measure the beta-carotene content in human skin depends on the scattering coefficient of skin. The light scattering in skin depends on different elements, such as melanin and other tissue structures. The presence of melanin (n=1.7) on the background of low refractive index of cell cytoplasm and interstitial fluid (n=1.34) causes high scattering in the epidermis layer of the skin.32 However, very white skin having a low melanin content has a high scattering coefficient.3335 Hence, in the range of 400 to 700 nm, high absorption and scattering limits the ability of experimental and analytical OC to efficiently detect the desired chromophores.

4.2.

Analytical Clearing (Extracting the Absorption Coefficient of Beta-Carotene)

Johansson36 measured the reflectance spectra of a thin cerebral phantom to estimate its absorption coefficient. Using linear regression, the absorption spectra were fitted to extract the main chromophore spectrum inside cerebral phantom. We applied this idea to perform analytical OC. That means, the blood and melanin spectra were digitally removed from the measured absorption spectra, then the absorption spectra of the desired chromophore should be apparent. The spectral elimination of dominant chromophores enhanced the capability of this method to detect the low value of concentration of chromophore at deep tissues without applying any experimental clearing. For example, Fig. 8 shows the spectral absorption coefficient for different values of concentration of beta-carotene inside the phantom. In general, a combination of all three considered modalities should be helpful for improving capabilities of the OC approach because of the reduction of tissue scattering and blood absorption impact on analytical reconstruction of chromophore concentration with the masked spectrum. To prove this claim, we also combine analytical clearing with experimental clearing methods. As shown in Fig. 9, the value of the extracted absorption coefficient of beta carotene is improved. Figure 10 also compares the value of the extracted absorption coefficient at a wavelength of 490 nm for these two methods. It is clearly specified that the estimated value of the absorption coefficient using this hybrid method (the combination of analytical and experimental clearing) is larger than experimental or analytical clearing alone. However, compression and immersion OC should be applied when analytical OC is not enough to get reliable data.

Fig. 8

Extracted spectral absorption coefficient of different values of concentration of beta-carotene using analytical OC.

JBO_23_9_095002_f008.png

Fig. 9

Extracted spectral absorption coefficient of different values of concentration of beta-carotene using combination of analytical and experimental OC.

JBO_23_9_095002_f009.png

Fig. 10

Relation between extracted absorption coefficient of beta-carotene at a wavelength of 490 nm with a concentration of beta-carotene inside phantoms for analytical clearing method and combination of analytical and experimental clearing method.

JBO_23_9_095002_f010.png

5.

Conclusion

This study aimed to improve compression OC using the analytical OC to measure beta-carotene inside the phantom. First, we utilize compression-immersion OC to decrease both the scattering coefficient of specimen and the effect of blood spectra. Our results show that this method enhances the ability of reflectance spectroscopy to predict the amount of beta-carotene. While the accuracy of this method depends on the value of pressure and the stiffness of the biological tissue, we also apply analytical OC to detect beta-carotene. Results depict that the analytical OC can measure the absorption coefficient of beta-carotene. Indeed, the analytical OC does not require any chemical agent or compression that might change or destroy the structure of tissues (as a noninvasive method). We finally apply the experimental clearing method besides analytical clearing to improve analytical capabilities of the OC approach for the extraction absorption coefficient of beta-carotene. The results show that analytical OC can be applied independently and in combination with compression or immersion, or both, when it is needed. The presented OC techniques can be applied for determining chromophores, such as melanin and lipid, in deep tissues that have good abilities for in vivo oncology and nutrition research, especially for spectral imaging of tumors.

Disclosures

The authors have no relevant financial interests in this article and no potential conflicts of interest to disclose.

Acknowledgments

E. A. G. and V. V. T. was supported by RFBR under Grant No. 17-02-00358 and the RF Governmental under Grant No. 14.Z50.31.0044.

References

1. 

M. C. Meinke et al., “Comparison of different cutaneous carotenoid sensors and influence of age, skin type, and kinetic changes subsequent to intake of a vegetable extract,” J. Biomed. Opt., 21 (10), 107002 (2016). https://doi.org/10.1117/1.JBO.21.10.107002 JBOPFO 1083-3668 Google Scholar

2. 

M. E. Darvin et al., “Optical methods for noninvasive determination of carotenoids in human and animal skin,” J. Biomed. Opt., 18 (6), 061230 (2013). https://doi.org/10.1117/1.JBO.18.6.061230 JBOPFO 1083-3668 Google Scholar

3. 

A. J. Speek, C. R. Temalilwa and J. Schrijver, “Determination of β-carotene content and vitamin A activity of vegetables by high-performance liquid chromatography and spectrophotometry,” Food Chem., 19 (1), 65 –74 (1986). https://doi.org/10.1016/0308-8146(86)90128-7 Google Scholar

4. 

J. T. Kumpulainen and J. T. Salonen, Natural Antioxidants and Anticarcinogens in Nutrition, Health and Disease, RSC, Cambridge (1999). Google Scholar

5. 

L. A. Chandler and S. J. Schwartz, “HPLC separation of cis- trans carotene isomers in fresh and processed fruits and vegetables,” J. Food Sci., 52 (3), 669 –672 (1987). https://doi.org/10.1111/jfds.1987.52.issue-3 JFDSAZ 0022-1147 Google Scholar

6. 

I. V. Ermakov and W. Gellermann, “Validation model for Raman based skin carotenoid detection,” Arch. Biochem. Biophys., 504 (1), 40 –49 (2010). https://doi.org/10.1016/j.abb.2010.07.023 Google Scholar

7. 

E. M. Darvin et al., “Comparison of two methods for noninvasive determination of carotenoids in human and animal skin: Raman spectroscopy versus reflection spectroscopy,” J. Biophotonics, 5 (7), 550 –558 (2012). https://doi.org/10.1002/jbio.201100080 Google Scholar

8. 

I. V. Ermakov and W. Gellermann, “Dermal carotenoid measurements via pressure mediated reflection spectroscopy,” J. Biophotonics, 5 (7), 559 –570 (2012). https://doi.org/10.1002/jbio.201100122 Google Scholar

9. 

E. A. Genina et al., “Optical clearing of biological tissues: prospects of application in medical diagnostics and phototherapy,” J. Biomed. Photonics Eng., 1 (1), 22 –58 (2015). https://doi.org/10.18287/JBPE-2015-1-1-22 Google Scholar

10. 

C. R. Idelson et al., “Effect of mechanical optical clearing on near‐infrared spectroscopy,” Lasers Surg. Med., 47 (6), 495 –502 (2015). https://doi.org/10.1002/lsm.v47.6 LSMEDI 0196-8092 Google Scholar

11. 

V. V. Tuchin, Optical Clearing of Tissues and Blood, SPIE Press, Bellingham, Washington (2006). Google Scholar

12. 

I. A. Nakhaeva et al., “The effect of external mechanical compression on in vivo water content in human skin,” Opt. Spectrosc., 118 (5), 834 –840 (2015). https://doi.org/10.1134/S0030400X15050124 OPSUA3 0030-400X Google Scholar

13. 

M. A. Ansari et al., “Study of the effect of mechanical pressure on determination of position and size of tumor in biological phantoms,” Appl. Opt., 52 (12), 2739 –2749 (2013). https://doi.org/10.1364/AO.52.002739 APOPAI 0003-6935 Google Scholar

14. 

V. V. Tuchin et al., “Light propagation in tissues with controlled optical properties,” J. Biomed. Opt., 2 (4), 401 –418 (1997). https://doi.org/10.1117/12.281502 JBOPFO 1083-3668 Google Scholar

15. 

L. Chen et al., “An effective balanced optical clearing method for intact biomedical imaging,” Sci. Rep., 7 (1), 12218 (2017). https://doi.org/10.1038/s41598-017-12484-3 SRCEC3 2045-2322 Google Scholar

16. 

J. Seo et al., “Clearing and labeling techniques for large-scale biological tissues,” Mol. Cells, 39 (6), 439 –446 (2016). https://doi.org/10.14348/molcells.2016.0088 Google Scholar

17. 

C. G. Rylander et al., “Dehydration mechanism of optical clearing in tissue,” J. Biomed. Opt., 11 (4), 041117 (2006). https://doi.org/10.1117/1.2343208 JBOPFO 1083-3668 Google Scholar

18. 

T. Yu et al., “Quantitative analysis of dehydration in porcine skin for assessing mechanism of optical clearing,” J. Biomed. Opt., 16 (9), 095002 (2011). https://doi.org/10.1117/1.3621515 JBOPFO 1083-3668 Google Scholar

19. 

X. Wen et al., “In vivo skin optical clearing by glycerol solutions: mechanism,” J. Biophotonics, 3 (1–2), 44 –52 (2010). https://doi.org/10.1002/jbio.200910080 Google Scholar

20. 

Z. Mao et al., “Influence of alcohols on the optical clearing effect of skin in vitro,” J. Biomed. Opt., 13 (2), 021104 (2008). https://doi.org/10.1117/1.2892684 JBOPFO 1083-3668 Google Scholar

21. 

D. Zhu et al., “Recent progress in tissue optical clearing,” Laser Photonics Rev., 7 (5), 732 –757 (2013). https://doi.org/10.1002/lpor.2013.7.issue-5 Google Scholar

22. 

D. Zhu et al., “Imaging dermal blood flow through the intact rat skin with an optical clearing method,” J. Biomed. Opt., 15 (2), 026008 (2010). https://doi.org/10.1117/1.3369739 JBOPFO 1083-3668 Google Scholar

23. 

G. Zonios and A. Dimou, “Modeling diffuse reflectance from semi-infinite turbid media: application to the study of skin optical properties,” Opt. Exp., 14 (19), 8661 –8674 (2006). https://doi.org/10.1364/OE.14.008661 OPEXFF 1094-4087 Google Scholar

24. 

P. Välisuo et al., “New closed-form approximation for skin chromophore mapping,” J. Biomed. Opt., 16 (4), 046012 (2011). https://doi.org/10.1117/1.3562976 JBOPFO 1083-3668 Google Scholar

25. 

P. So et al., “Two-photon deep tissue ex vivo imaging of mouse dermal and subcutaneous structures,” Opt. Express, 3 (9), 339 –350 (1998). https://doi.org/10.1364/OE.3.000339 OPEXFF 1094-4087 Google Scholar

26. 

S. Andree, C. Reble and J. Helfmann, “Spectral in vivo signature of carotenoids in visible light diffuse reflectance from skin in comparison to ex vivo absorption spectra,” Photonics Lasers Med., 2 (4), 323 –335 (2013). https://doi.org/10.1515/plm-2013-0032 Google Scholar

27. 

W. F. Cheong, S. A. Prahl and A. J. Welch, “A review of the optical properties of biological tissues,” IEEE. J. Quantum. Electron., 26 (12), 2166 –2185 (1990). https://doi.org/10.1109/3.64354 Google Scholar

29. 

S. L. Jacques, “Optical properties of biological tissues: a review,” Phys. Med. Biol., 58 (11), R37 (2013). https://doi.org/10.1088/0031-9155/58/11/R37 PHMBA7 0031-9155 Google Scholar

30. 

A. K. Bui et al., “Revisiting optical clearing with dimethyl sulfoxide (DMSO),” Lasers Surg. Med., 41 (2), 142 –148 (2009). https://doi.org/10.1002/lsm.v41:2 LSMEDI 0196-8092 Google Scholar

31. 

C. Ursin et al., “Permeability of commercial solvents through living human skin,” Am. Ind. Hyg. Assoc. J., 56 (7), 651 –660 (1995). https://doi.org/10.1080/15428119591016665 AIHAAP 0002-8894 Google Scholar

32. 

N. Sudheendran et al., “Assessment of tissue optical clearing as a function of glucose concentration using optical coherence tomography,” J. Innov. Opt. Health Sci., 3 (03), 169 –176 (2010). https://doi.org/10.1142/S1793545810001039 Google Scholar

33. 

D. Yudovsky and L. Pilon, “Retrieving skin properties from in vivo spectral reflectance measurements,” J. Biophotonics, 4 (5), 305 –314 (2011). https://doi.org/10.1002/jbio.201000069 JBOPFO 1083-3668 Google Scholar

34. 

G. Zonios et al., “Skin melanin, hemoglobin, and light scattering properties can be quantitatively assessed in vivo using diffuse reflectance spectroscopy,” J. Invest. Dermatol., 117 (6), 1452 –1457 (2001). https://doi.org/10.1046/j.0022-202x.2001.01577.x JIDEAE 0022-202X Google Scholar

35. 

R. R. Anderson and J. A. Parrish, “The optics of human skin,” J. Invest. Dermatol., 77 (1), 13 –19 (1981). https://doi.org/10.1111/1523-1747.ep12479191 JIDEAE 0022-202X Google Scholar

36. 

J. D. Johansson, “Spectroscopic method for determination of the absorption coefficient in brain tissue,” J. Biomed. Opt., 15 (5), 057005 (2010). https://doi.org/10.1117/1.3495719 JBOPFO 1083-3668 Google Scholar

Biography

Shadi Masoumi received her BS degree in atomic physics from Semnan University and her MS degree in photonics from Shahid Beheshti University, in 2018. She wrote her master thesis on tissue optical clearing methods in combination with diffuse reflection spectroscopy. Currently, she is a research assistant at laser and plasma research institute at Shahid Beheshti University, Iran. Her research interests include biomedical optics and biophotonics.

Mohammad Ali Ansari is assistant professor of photonics at Laser and Plasma Research Institute of Shahid Beheshti University, and head of Optical Bioimaging Lab. He obtained his PhD in photonics from Shahid Beheshti Univertity, Tehran, Iran, at 2010. His research areas includes fNIRS, diffuse optical tomography and photoacoustic imaging.

Ezeddin Mohajerani is professor of photonics at Laser and Plasma Research Institute of Shahid Beheshti University, and head of POMP Lab. He obtained his PhD in physics from University of Reading at UK in 1992. His research interest includes biophotonics, biospectroscopy, organic nanophotonics, organic nonlinear optics, smart organic materials, integrated optics, and organic optoelectronics. One of the active fields in his lab is studying noninvasive optical diagnosis and therapy of skin diseases and agriculture products based on spectroscopy. He has authored more than 150 articles, more than 10 in the field of tissue optics, optical diagnosis and therapy in Persian and English.

Elina A. Genina is a professor in the Departments of Optics and Biophotonics at Saratov State University. She is a coauthor of more than 250 peer-reviewed publications, nine book chapters, and nine patents. Her research interests include biomedical optics, laser medicine, and development of methods for control of tissue optical properties. She is a scientific secretary of International Symposium on Optics and Biophotonics (Saratov Fall Meeting). She has 2302 citations and h-index 22 (WoS, September 4, 2018).

Valery V. Tuchin is a professor and head of optics and biophotonics at Saratov State University (National Research University of Russia) and several other universities. His research interests include tissue optics, laser medicine, tissue optical clearing, and nanobiophotonics. He is a fellow of SPIE and OSA, has been awarded Honored Science Worker of the Russia, honored professor of Saratov University, SPIE educator award, FiDiPro (Finland), Chime Bell Prize of Hubei Province (China), and Joseph W. Goodman Book Writing Award (OSA/SPIE). He has 20607 citations and h-index 66 (Google Scholar, September 4, 2018).

© 2018 Society of Photo-Optical Instrumentation Engineers (SPIE) 1083-3668/2018/$25.00 © 2018 SPIE
Shadi Masoumi, Mohammad Ali Ansari, Ezeddin Mohajerani, Elina A. Genina, and Valery V. Tuchin "Combination of analytical and experimental optical clearing of rodent specimen for detecting beta-carotene: phantom study," Journal of Biomedical Optics 23(9), 095002 (13 September 2018). https://doi.org/10.1117/1.JBO.23.9.095002
Received: 30 June 2018; Accepted: 24 August 2018; Published: 13 September 2018
Lens.org Logo
CITATIONS
Cited by 9 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Absorption

Reflectance spectroscopy

Blood

Tissues

Chromophores

Optical clearing

Scattering

RELATED CONTENT


Back to Top