5 January 2022 Perspective on the integration of optical sensing into orthopedic surgical devices
Author Affiliations +
Abstract

Significance: Orthopedic surgery currently comprises over 1.5 million cases annually in the United States alone and is growing rapidly with aging populations. Emerging optical sensing techniques promise fewer side effects with new, more effective approaches aimed at improving patient outcomes following orthopedic surgery.

Aim: The aim of this perspective paper is to outline potential applications where fiberoptic-based approaches can complement ongoing development of minimally invasive surgical procedures for use in orthopedic applications.

Approach: Several procedures involving orthopedic and spinal surgery, along with the clinical challenge associated with each, are considered. The current and potential applications of optical sensing within these procedures are discussed and future opportunities, challenges, and competing technologies are presented for each surgical application.

Results: Strong research efforts involving sensor miniaturization and integration of optics into existing surgical devices, including K-wires and cranial perforators, provided the impetus for this perspective analysis. These advances have made it possible to envision a next-generation set of devices that can be rigorously evaluated in controlled clinical trials to become routine tools for orthopedic surgery.

Conclusions: Integration of optical devices into surgical drills and burrs to discern bone/tissue interfaces could be used to reduce complication rates across a spectrum of orthopedic surgery procedures or to aid less-experienced surgeons in complex techniques, such as laminoplasty or osteotomy. These developments present both opportunities and challenges for the biomedical optics community.

1.

Introduction

In most developed countries, orthopedic surgery represents one of the most common surgical procedures.1 The number of patients and procedures will only increase with aging populations. Many surgical procedures have been developed successfully to treat common degenerative conditions, including total knee or hip replacement and spinal fusion, which are the most common hospital discharge procedures in the United States with about 1.5 million cases per year. Other procedures, such as partial and total shoulder replacement, partial hip replacement, kyphoplasty/vertebroplasty, and lumbar decompression, are also regularly performed by orthopedic surgeons as well as other specialists (neurosurgeons and interventional radiologists). Minimally invasive surgery (MIS) and percutaneous techniques have gradually been introduced to reduce complications and improve recovery times:24 examples in spinal surgery include pedicle screw placement, spinal fusion, vertebroplasty/kyphoplasty, and decompression. There has been more recent adoption in total hip arthroplasty, with renewed interest in anterior hip approaches.57 Other hip arthroplasty techniques have included bone resurfacing, which has been controversial with several large registries highlighting concerns over premature wear and metallosis.8,9

For minimally invasive approaches, adjunctive imaging is used frequently to plan and guide procedures, including endoscopy,10,11 x-ray fluoroscopy or CT scanning,1215 infrared navigation with visualization3,4,1618 and robotics with endoscopy.19,20 The overall goal is to provide information on the spatial position of surgical tools in either a two-dimensional plane or three-dimensional volume or direct visualization of the surgical area during the procedure to minimize surgical complications such as pedicle breach in percutaneous pedicle screw fixation. Intraoperative ultrasound is also being used increasingly via keyhole spinal surgery to guide indirect decompression of disc, bone and other pathologies that cause neurologic compression. In open surgery, imaging techniques such as fluoroscopy may be used in total hip revision to facilitate femoral cement removal, especially at the distal end where low visibility is accompanied by risk of shaft fracture and mutilation.21,22 Infrared navigation and instrument tracking have also been compared with endoscopic and open surgical approaches for other orthopedic procedures, such as total hip and knee replacement, but no significant improvements have been reported to date in the reduction of associated complications.23,24

Various unmet clinical needs in orthopedic surgery could potentially be addressed by optical spectroscopy or imaging, which come in many different forms depending on the light–tissue interaction being sensed. They include diffuse reflectance that depends on the light absorption and (elastic) scattering of the tissue, optical coherence tomography (OCT) that images tissue microstructures and fluorescence and Raman (inelastic scattering) that report molecular signatures. These techniques may be implemented at the working tip of surgical instruments to help guide the procedure and minimize risk of complications. Both morphological and physiological information can be obtained in real time and in a noninvasive manner.

The full potential of photonic sensing in orthopedic surgery has not been explored but over the past few years there have been some developments for specific procedures and to refine surgical workflow. Examples include: a cranial perforator based on diffuse reflectance spectroscopy (DFS) that stops automatically at the dura,25 an intramedullary nail system with a DFS sensor26 that prevents overdrilling, and a pedicle screw insertion device with DFS guidance2730 that facilitates spinal fixation. Although the first of these devices is used by neurosurgeons, the technology is relevant to orthopedic surgery, since similarly the objective is to avoid perforation through bone into critical adjacent normal tissues structures. While these examples illustrate the unmet clinical needs and the potential of optical techniques to provide greater control and reduced complications, work to date has been largely limited to preclinical or cadaveric studies.

There have also been only a few developments in the use of optical sensing for general orthopedic and spinal procedures, such as lumbar decompression, total hip or knee replacement, and shoulder replacement/fixation. These procedures utilize different surgical techniques/approaches into which optical guidance could be integrated, as summarized in Table 1. There are only limited data on the relevant optical properties of different layers of bone (periosteum, cortical bone, and endosteum). A recent study focused on percutaneous measurements of the optical absorption and reduced scattering coefficients of whole bone in the radius and distal tibia using a photon time-of-flight technique.31 This provided only average values of tissue, not specific to tissue layers, which may not be sufficiently reliable or accurate for optical sensing in critical locations (e.g., pedicle screw placement into cortical bone). This and other studies have shown strongly wavelength-dependent absorption and reduced scattering coefficients of bone in the range 0.1 to 0.5  cm1 and 4 to 12  cm1, respectively. Sekar et al. demonstrated a maximum CW penetration depth of between 7.5 and 15  mm at 785 nm with the lowest values measured at the trochanter due to strong scattering.31 In addition, the authors noted that at the penetration depths measured with their system in CW mode, they were objectively able to measure cortical bone at every position transcutaneously.31

Table 1

Overview of different surgical instrument methods potentially utilizing optical guidance.

ProcedureSurgical instrument methods
Total hip arthroplastyDrilling, shaving, shaping
Total knee arthroplastyDrilling, shaving, shaping
Shoulder fixationDrilling
Lumbar decompressionShaving, shaping
Pedicle screw placementDrilling, screw insertion

With this brief background, the objective here is to explore the potential roles of optical sensing in specific orthopedic and spinal surgery applications, particularly to reduce complications by increasing surgical precision and to enable extension of the surgical field with greater control and improved safety. Moreover, we aim to provide insights to the optical sciences/engineering community as a stimulus for technology innovation, with particular attention to the inherent challenges in integrating optical sensing within delicate surgical instruments and to clinical applications that would benefit from the improved precision.

2.

Optics-Enabled Orthopedic Technologies

2.1.

Orthopedic Applications and Challenges

2.1.1.

Total hip arthroplasty

The first use-case of a drill incorporating optical sensing is for total hip arthroplasty, for which specific drilling locations are well prescribed to avoid damage to the neurovasculature.3234 In general, the posterior-superior and posterior-inferior pelvic zones are considered “safe,” depending on the screw size. In general, screws <20  mm long should be used in the posterior–inferior direction to avoid the sciatic nerve, inferior gluteal nerve and blood vessels, as well as the internal pudendal nerve and vessels. However, there are some cases where either the iliac fossa has been damaged or the prescribed safe zones are too weak to anchor the acetabular implant. In these cases, the surgeon needs to drill in a direction where the pelvic bones are much thinner than the iliac fossa, so that there are vessels and nerves that could be damaged if the bone is pierced. However, these alternative approaches of drilling into the anterior–superior and anterior–inferior direction risk damaging the external iliac artery and vein (in the anterior–superior direction) or the obturator nerve, artery, and vein (in the anterior–inferior direction).34

Currently, the bone is drilled using haptic and/or auditory feedback to determine if bone/tissue boundary is being approached. In addition, clinical experience together with preoperative imaging inform the best path, and the drill and screw lengths should be chosen for whichever acetabulum quadrant was selected for optimal fixation. Other competing techniques such as navigation, intraoperative imaging, and robotics have been demonstrated, although their added benefit has not been established in total hip arthroplasty as compared with spinal approaches (laminectomy, pedicle screw placement) and knee arthroplasty: further randomized control studies are ongoing.5,20,23,24 In this complementary hybrid approach, development of an optics-enabled drill would act as an added safety device to warn of upcoming danger through feedback to the surgeon. This would not necessarily require automatic shut-off capabilities such as have been implemented in some orthopedic and cranial drills.25 We suggest that an optical sensor within the drill tip would report on whether the drill is encountering blood vessels and/or nerves and also sense the bone/tissue interface to warn of potential periosteum breach. This would increase confidence in utilizing the higher-risk quadrants without increasing complications. An example of how this drill could work is shown in Fig. 1.

Fig. 1

Concept of a “smart” optical drill for total hip arthroplasty.

JBO_27_1_010601_f001.png

There are, however, several challenges to this approach. Despite previous demonstrations of photonic devices integrated into solid drill bits such as cranial perforators,21 the bits used routinely in hip arthroplasty are flexible. They have a hollow core, which is well suited for fiberoptic placement, but large-core multimode glass fibers that are commonly used for such sensing have limited bend radius. More flexible plastic fibers might be an option but may have limited lifetime in the harsh drilling environment.

2.1.2.

Clavicle fracture fixation

In clavicle fracture fixation, rare (<1% of cases) but serious and life-threatening complications due to drilling into the subclavian neurovascular bundle have been reported.35 As in total hip arthroplasty, the procedure follows prescribed safe drilling angles and zones,36 despite which over-drilling still occurs, leading to possible laceration of the subclavian artery, damage to the brachial plexus or damage to the subclavian vein (not always due to the over-drilling itself but subsequent piercing by screws). This last complication may not be obvious during the procedure itself but leads to significant morbidity and need for revision surgery. Symptoms may not present for upwards of a decade, at which time claudication and critical limb ischemia may occur.35 In addition, other complications such as pseudoaneurysm, subclavian arteriovenous fistulas, air embolism, and severe postoperative radicular pain can occur that may not present for months or years.35

Most of the cases reporting damage to the associated vessels and nerves are in medial clavicle fixations, so that optical sensing could be used in these critical cases on the medial aspect of the clavicle where injury can occur. The most common cause of injury is overdrilling or taking an oblique approach through the clavicle instead of the prescribed superior–inferior direction. This can be exacerbated by portions of the subclavian vein adhering to the inferior border of the clavicle. It would then be of value if optical sensing could provide both feedback and a failsafe mechanism upon detecting imminent inferior breaching to stop the drill. This approach has been demonstrated with other optically enabled surgical devices.25,26 In addition to challenges similar to those in total hip arthroplasty, the screws and drill depth for medial clavicle fixation are relatively short, typically <18  mm in adults.31 One technical challenge, which is found across all the applications, is selecting the optimum optical properties for differentiating bone from surrounding soft tissues, especially where there may be vessel or nerve adhesions to the bone. Distinguishing between mixed fascia/muscle, blood, and nerves may be needed.

2.1.3.

Pedicle screw placement

Spinal surgery presents several scenarios where optical sensing could be incorporated, one being pedicle screw placement, in which drilling (as is commonly performed with percutaneous approaches) is technically challenging and complications from incorrect drilling have significant morbidity.13,37,38 The challenge in this application is in maintaining the correct trajectory within pedicular cancellous bone and avoiding excessive medial/lateral trajectory or, less commonly, anterior/superior/inferior trajectory that may result in cortical breaching. Deviating from the prescribed surgical path will inherently carry risk into injuring critical structures such as nerves or causing an incidental durotomy with a resultant cerebrospinal fluid (CSF) leak.38 Deviation from the planned trajectory may not be detected until after the screws have been inserted. A number of imaging modalities, including intraoperative fluoroscopy, CT and infrared navigation with tool tracking, have been reported for this application, with recent studies also involving robotics.2,4,15,19 The aim is to aid the surgeon in correct placement as well as getting the maximum purchase from the pedicle screws to minimize the risk of revision surgery, which is complex, invasive, and has higher complication rates and poorer outcomes, especially where screws have to be removed and repositioned. Optical sensing could play a complementary role in these procedures. A number of methods and approaches have been developed using optical sensing for percutaneous pedicle screw fixation, including integrating optical fibers within Kirschner K-wires, (a sharpened steel pin utilized for skeletal traction of the pedicle screw), to measure the diffuse reflectance, and photoacoustic imaging probes.2326 Optical sensing could be used locally in conjunction with systems such as navigation and global CT guidance to provide a second check so the prescribed drill path is maintained in the cancellous bone. One option is to integrate sensing onto the navigation drill guide rather than into the drill bit itself and then coregister the navigational and optical signals. A second option would be to have the sensor at the tip of the drill to distinguish cancellous from cortical bone, alerting the surgeon to deviation from the cancellous bone trajectory.

Unlike some of the other orthopedic applications, a common overdrilling problem in the spine relates not to the forward-facing boundary but rather to medial and lateral breaches. This presents a three-dimensional navigational challenge to maintain the appropriate trajectory. The reasons for this difference compared to total hip arthroplasty or clavicle fixation are due to the nature of screw placement. In the case of pedicle screw placement, the screw has a fixed prescribed length with the goal of insertion into the vertebral body. For an anterior breach, the screw would have to go through the entire vertebral body, where 2-mm deviations can lead to medial/lateral breach (the most common breach orientations), or there is a possibility of an in/out lateral breach where the screw leaves the pedicle and the tip reinserts into the vertebral body. Superior/inferior breaches also occur albeit at a lower frequency with inferior breaches considered as serious as medial breaches given the close course of nerve roots along the medial and inferior pedicle borders.39

2.1.4.

Lumbar decompression

Degenerative spinal stenosis is a common condition, with reported symptomatic cases of 9% in adults within a Japanese population and increasing with age.40 The typical intervention is removal of the vertebral ligaments and laminae to relieve pressure on the spinal canal and nerve roots as they exit the spine. In surgically managed cases, there are different approaches for decompression, based on clinical judgment and the degree of stenosis, but can include either laminectomy or laminotomy. These procedures can be performed through open or minimally invasive techniques. Especially for the latter, there is a significant surgical learning curve and there are complications even for the fully trained surgeon.10,4143 One of the most common complications is dural laceration, i.e., incidental cutting of the dura by the burr or other device (e.g., Kerrison or Cloward rongeurs—used to cut and pull away pieces of the lamina). This may or may not cause CSF leakage, with laceration rates reported in 15% to 20% of cases.44,45

In the minimally invasive approach, the operative field is viewed by endoscopy or microscopy, while open procedures provide direct vision. Robotic approaches have not been widely adopted to date.46,47 We envision optical guidance to prevent dural laceration, either by informing the surgeon when the dura is being approached or by stopping the device if imminent breach is probable by differentiating between bone and dura/CSF. This would also aid less experienced surgeons, shortening procedure times, and increasing confidence. An example of how this might work is presented in Fig. 2.

Fig. 2

Concept of an optics-enabled burr for laminectomy.

JBO_27_1_010601_f002.png

There are multiple technical challenges in this clinical application. First, the procedure itself is distinct from the other drilling procedures as the burr is used to shave back and forth to remove bone. Hence, the drill tip is not fully in contact with different pieces of bone, causing optical signal variations due to bone heterogeneity. Second, the geometry and size of the burr, which is spherical with diameter as small as 4 to 6 mm, make it more difficult to integrate fiberoptics. In the case of the hip, clavicle, and pedicle screws, the devices are large enough to carry the required number of fibers, but this is not the case for this procedure. It may then be necessary to add an accessory probe to provide illumination or detection separate from the drill bit itself, for example, using a wide-field imaging modality interfaced with the existing endoscopic system.

2.2.

Integration of Optical Devices into Orthopedic Tools

In the above applications, an optical device and algorithms would use native optical properties of bone and surrounding tissues to aid the surgeon in avoiding overdrilling through bone or into critical structures, either by providing feedback or by automatically stopping the drill. Similar approaches have been reported in neurosurgery where, for example, a cranial perforator system was developed to sense the approaching dura and stop the drill prior to piercing the skull.25 Optical fiber-integrated pedicle screws have also been reported utilizing DFS.27 Integration of optics into a K-wire highlights a similar approach.29 The integration of optics within an orthopedic surgical drill that has been utilized preclinically is shown in Fig. 3. These first advances have shown the promise of optical sensing as a low-cost, noninvasive method to improve the accuracy and safety of orthopedic procedures. Table 2 highlights some examples reporting use of such technologies.

Fig. 3

Implementation of two-wavelength DRS in a cranial perforator surgical drill bit to prevent plunging events during craniotomy. (a) Prototype of Stryker CD 4 drill with integrated DRS, (b) schematic of the integrated illumination and detection systems, (c) demonstration of successful stop of the drill in ex vivo (sheep) cranium using optical detection only, leaving 0.5-mm thick bone shelf just before the dura, (d) flow chart of 530- and 850-nm reflectance data acquisition and processing to detect the brain-bone boundary, applying threshold and slope algorithms. (Adapted from Ref. 25.)

JBO_27_1_010601_f003.png

Table 2

Reported approaches of optical integration into orthopedic procedures.

ProcedureClinical challengeTypical guidanceOptical techniqueVerification methodDetected structureRef.
Cranial perforationPenetrating into brainMechanically clutched burrDiffuse reflectanceEx vivo animalBone/brain interface25
Intramedullary nailingBreach and soft tissue damageX-rayDiffuse reflectanceEx vivo animalBone/muscle interface26
Pedicle screw placementLateral and medial breachCTDiffuse reflectance, photo-acousticsEx vivo humanCortical/cancellous bone2728.29.30

The use of native optical properties of tissues necessitates either a standard set of optical properties across a wide spectral range for the tissues of interest or a method of determining patient-specific tissue properties prior to drilling. Databases of optical properties have been generated for various tissue types to use as input to Monte Carlo simulations for photodynamic therapy (PDT).48 Similar to an approach used during PDT for prostate cancer, in which light penetration is measured to continually update optical absorption and scattering properties for dynamic dosimetry,49 real-time measurements could be employed to guide the surgeon as the drill moves into different tissue types. It may also be possible to employ a similar approach based on preoperative imaging, where patient-specific bone thickness could be used with varying muscle/fascia, blood, nerves/spinal cord, or CSF “background” tissues as a training set. However, this is nontrivial because of the large inter- and intrapatient variations in optical properties, as reported, for example, for bone31,5055 and CSF.5557 Hence, it may be difficult to achieve the accuracy needed without an intraoperative method of measuring the optical properties in individual patients. An alternate approach would be to collect a large data set to train an artificial intelligence (AI) algorithm that would subsequently be used for guidance.

Although this discussion has focused on the use of DFS and corresponding optical properties for detection of tissue boundaries or critical structures, other optical modalities could be integrated as alternative or complementary modalities. Raman spectroscopy, for example, has been used preclinically to detect osteoarthritic changes in human cartilage58 and transcutaneous in vivo detection of disordered bone.59 Raman, or possibly coherent anti-Stokes Raman (CARS) spectroscopy, could identify intrinsic biochemical “signatures” such as collagen between layers of bone, elastin/blood in vessels or myelin/lipids within nerves. Tissue autofluorescence or the use of exogenous fluorophores could be added for tissue discrimination, as increasingly used in guiding tumor resection. 6062 This would avoid the added technical complexity and costs associated with nonlinear techniques such as CARS or, to detect collagen, second harmonic generation. (A caveat in using collagen detection for guidance is that it is present in multiple tissue layers and at boundaries between bone/periosteum or bone/cartilage/dura in vertebrae.)

In general, regardless of the optical technique and depending on the procedure, detection of the following structures will be needed: bone (cortical, trabecular, osteoporotic), soft tissues (muscle, subcutaneous tissue, dura), connective tissues (ligaments, cartilage), and neurovascular tissues (spinal cord, nerves, blood vessels, CSF). Even with some canonical training sets of optical properties, additional perioperative or intraoperatively measurements will likely be needed, using techniques such as those listed in Table 3. However, some of these are either time-consuming or only acquire superficial optical properties which limits their utility.

Table 3

Reported methods to measure in vivo optical absorption and elastic scattering properties of human tissues.

MethodTissue typeWavelength (nm)Ref.
Spatially resolved reflectance measurementsEsophageal (normal, benign, malignant)63063
Brain and bladder420–450, 532, 63564
Esophageal wall63065
Skin and underlying tissues400–105066
Common nevi, dysplastic nevi, and malignant melanoma skin lesions483–91767
Prostate73268
Time-resolved reflectance measurementsSkin, subcutaneous fat and muscle83069
Bone760, 600–120031, 70
Spatial frequency domain imagingSkin450–800, 950–160071
Ovarian tissue73072
Low-coherence enhanced backscattering spectroscopyDuodenal mucosaTunable xenon lamp73

In addition, any optically enabled tool would need an accuracy at least equivalent to existing clinical devices, typically sub-mm for most of the above applications. For total hip arthroplasty and clavicle fracture fixation, the required accuracy would be 0.5  mm. Although 0.5 mm would exceed many of the leading navigation systems and beyond what current systems may manage we propose this value as a benchmark for these specific indications where navigation is not routinely used and where surgeons want to be able detect specific nerves or blood vessels. A larger uncertainty would be clinically acceptable for applications in the spine, specifically for pedicle screw placement (as reflected in Table 4), although for lumbar decompression there is currently no competing technology that reduces dura laceration rate. State-of-the-art navigation systems for pedicle screw placement have reported deviations of <1  mm in screw location compared to postoperative CT imaging in the lateral and axial directions:16,74 even if the accuracy of optical sensing only matched that of navigation and intraoperative CT, this should be beneficial in a number of ways. First, reduction in radiation dose within the OR would benefit both patients and operators. Second, optical sensing could allow for real-time imaging and feedback, which current CT navigation systems cannot achieve, and should entail lower equipment costs that enable wider dissemination. However, recent studies involving optical sensing in neurosurgery, such as the cranial perforator, have suggested that sub-mm optical accuracy is attainable as the group could reliably stop the cranial perforator within 0.5 mm from the surface of the dura routinely, albeit in an ex vivo setting.25

Table 4

Drilling parameters and competing technologies.

ProcedureDrill typeSpeed (rpm)Feed/shaving rate (mm/s)Accuracy required (mm)Estimated forward sensing required (mm)Competing approaches
Auto stopSurgeon cues
Hip arthroplastyOrthopedic drill with flexible drill bitUp to 75k4 to 50.512Safe zonesa, navigation, robotics
Shoulder fixationOrthopedic drill with flexible drill bitUp to 75k4 to 50.2 to 0.50.5 to 11.5 to 2Drill guides, safe zones
Pedicle screwHigh speed drill with drill guideUp to 90k4 to 50.5 to 1.01 to 1.52 to 3Navigation, CT guidance, fluoroscopic guidance
LaminectomyHigh speed drill with flex-burr in some casesUp to 90k1 to 20.1 to 0.30.25 to 0.751 to 1.5Endoscopic + microscopic imaging, robotic (open decompressionb)

aSafe zones are prescribed areas of tissue where procedures can be performed with minimal risk due to the absence of neurovascular structures.

bOpen decompression refers to procedure via a full skin incision where pressure on the spinal cord is relieved by removing the entire posterior portion of the vertebrae (the lamina).

With these various factors in mind, we suggest the following requirements for optical sensing in orthopedic surgical devices.

  • Field of view: 1.5 to 2× the drill tip/burr diameter;

  • Depth of view: either 2× or 3× the depth accuracy required (for autostop or surgical cues, respectively);

  • Spatial resolution: 0.1 mm by 0.1 mm (to differentiate neurovasculature from surrounding tissue);

  • Sensitivity: 90% to 95% (comparable to that of CT for pedicle screw placement);

  • Specificity: 70% to 80% (variable, depending on procedure and surgical experience).

A further consideration is the optical signal sampling rate required in each procedure, as summarized in Table 4, taking into account also the signal integration and analysis time and the time to stop the drill. We have provided estimates of the forward-sensing distance required, depending on whether autostop is implemented or the device is passive and simply alerts the surgeon. For the hip and shoulder procedures, sub-mm accuracy is required but the total drill time is only 5  s so the high rotational speed and axial feed rate may generate various forms of noise (vibrational, thermal from friction, and mechanical from the fiberoptic coupling) that may degrade the accuracy. A further consideration is that some procedures would require constant contact between the optical fiber and the tissue to maximize the signal.

The final aspect of integrating optical sensing into orthopedic surgical tools is the interaction with the surgeon. In general, providing visual and/or audio cues is likely safer and would be more easily adopted than autostop, at least initially. For example, visual cues could comprise green/red LEDs to indicate when the bit tip is still in bone or approaching a boundary, respectively. Alternatively, a single LED could be set to blink only when a boundary is being approached. Audio cues such as beeping are interesting in that many surgeons already use audio feedback to determine when they have moved into different types of bone, based on the changing pitch of the sound made by the drill bit moving toward a boundary. Haptic feedback is another option that has been trialed in surgical robotics,7577 although vibration of the tool is a challenge.

These various approaches could be integrated with other-modality devices and/or within robotics devices for MIS. An example is integrating optics within a pedicle screw that already utilizes neuronavigation or CT imaging. Communication between the modalities could be done wirelessly or using data cables coming off the drill (see Fig. 3). Semiautomated OCT imaging has been already tested preclinically in an ophthalmic surgery robotic system.78 A long-term goal would be optics integration with AI-enabled robotics for autonomous or semiautonomous surgery, analogous to CyberKnife radiation systems79 or the RAVENII robot for brain tumor ablation.80

A final aspect of the integration of optics into surgical devices is the possibility of changing the surgical approach for many of these procedures through opening avenues that were previously deemed too risky. For example, if critical neurovascular structures could be reliably detected, it would markedly alter current procedures such as laminoplasty or enable a direct anterior approach in total hip arthroplasty. Likewise, new approaches would be enabled in cranial procedures, transsphenoidal surgery, and shaving/oncological procedures.

3.

Conclusions

Orthopedic and spinal procedures are extremely common and their use increases with population ageing. Here, we have considered a number of such applications to illustrate how integrating optical sensing into the surgical tools could reduce risks and complication rates, allow surgeons to more confidently approach anatomical sites that are inherently less safe (e.g., as in hip arthroplasty), shorten operation times, and even enable novel surgical procedures. We envision that the integration of optics within orthopedic tools would use primarily the native optical properties of bone and surrounding soft tissues to provide near real-time feedback on upcoming structures (Fig. 1) or signaling, by visual and/or auditory means, when only a thin layer of bone remains (Fig. 2). This will complement other techniques such as navigation to ensure that the prescribed surgical plan is achieved with maximum safety and efficacy.

The need for such additional guidance will only expand with evolving minimally invasive and percutaneous procedures, so there are significant opportunities for further development of optical devices integrated into orthopedic surgical tools.

Disclosures

The authors declare that they have no competing interests.

Acknowledgments

We acknowledge Dr. Walter Messina for help with the figures and with input on future implementation of optics within surgical devices, Dr. Alexander Gregor for discussion surrounding potential applications within general and orthopedic surgery, and Dr. Sanathana Konugolu Venkata Sekar for discussions around optical properties of bone and clinical techniques. This work was supported by Science Foundation Ireland (SFI), Grant No. SFI/15/RP/2828

References

1. 

M. J. Hall et al., “Ambulatory surgery data from hospitals and ambulatory surgery centers: United States, 2010,” Natl. Health Stat. Rep., 102 1 –15 (2017). Google Scholar

2. 

N. S. Dahdaleh, Z. A. Smith and P. W. Hitchon, “Percutaneous pedicle screw fixation for thoracolumbar fractures,” Neurosurg. Clin. N. Am., 25 (2), 337 –346 (2014). https://doi.org/10.1016/j.nec.2013.12.011 Google Scholar

3. 

T. T. Kim et al., “Minimally invasive spinal surgery with intraoperative image-guided navigation,” Biomed. Res. Int., 2016 5716235 (2016). https://doi.org/10.1155/2016/5716235 Google Scholar

4. 

I. D. Siasios et al., “Percutaneous screw placement in the lumbar spine with a modified guidance technique based on 3D CT navigation system,” J. Spine Surg., 3 (4), 657 –665 (2017). https://doi.org/10.21037/jss.2017.12.05 Google Scholar

5. 

F. Migliorini et al., “Total hip arthroplasty: minimally invasive surgery or not? Meta-analysis of clinical trials,” Int. Orthop., 43 (7), 1573 –1582 (2019). https://doi.org/10.1007/s00264-018-4124-3 Google Scholar

6. 

G. Meermans et al., “The direct anterior approach in total hip arthroplasty: a systematic review of the literature,” Bone Joint J., 99-B (6), 732 –740 (2017). https://doi.org/10.1302/0301-620X.99B6.38053 Google Scholar

7. 

B. H. Brismar et al., “Early gain in pain reduction and hip function, but more complications following the direct anterior minimally invasive approach for total hip arthroplasty: a randomized trial of 100 patients with 5 years of follow up,” Acta Orthop., 89 (5), 484 –489 (2018). https://doi.org/10.1080/17453674.2018.1504505 Google Scholar

8. 

D. A. Marshall et al., “Hip resurfacing versus total hip arthroplasty: a systematic review comparing standardized outcomes,” Clin. Orthop. Relat. Res., 472 (7), 2217 –2230 (2014). https://doi.org/10.1007/s11999-014-3556-3 CORTBR 0009-921X Google Scholar

9. 

S. L. Ras Sørensen et al., “Revision rates for metal-on-metal hip resurfacing and metal-on-metal total hip arthroplasty: a systematic review,” Hip Int., 26 (6), 515 –521 (2016). https://doi.org/10.5301/hipint.5000444 Google Scholar

10. 

C.-W. Lee, K.-J. Yoon and S.-W. Kim, “Percutaneous endoscopic decompression in lumbar canal and lateral recess stenosis: the surgical learning curve,” Neurospine, 16 (1), 63 –71 (2019). https://doi.org/10.14245/ns.1938048.024 Google Scholar

11. 

R. G. Chiu et al., “Endoscopic versus open laminectomy for lumbar spinal stenosis: an international, multi-institutional analysis of outcomes and adverse events,” Global Spine J., 10 (6), 720 –728 (2020). https://doi.org/10.1177/2192568219872157 Google Scholar

12. 

M. S. Austin et al., “Image-guided spine surgery: a cadaver study comparing conventional open laminoforaminotomy and two image-guided techniques for pedicle screw placement in posterolateral fusion and nonfusion models,” Spine, 27 (22), 2503 –2508 (2002). https://doi.org/10.1097/00007632-200211150-00015 SPINDD 0362-2436 Google Scholar

13. 

V. Sarwahi et al., “Pedicle screw safety: how much anterior breach is safe? A cadaveric and CT-based study,” Spine, 42 (22), E1305 –e1310 (2017). https://doi.org/10.1097/BRS.0000000000002153 SPINDD 0362-2436 Google Scholar

14. 

V. Cordemans et al., “Pedicle screw insertion accuracy in terms of breach and reposition using a new intraoperative cone beam computed tomography imaging technique and evaluation of the factors associated with these parameters of accuracy: a series of 695 screws,” Eur. Spine J., 26 (11), 2917 –2926 (2017). https://doi.org/10.1007/s00586-017-5195-3 Google Scholar

15. 

G. Nayar et al., “Pedicle screw placement accuracy using ultra-low radiation imaging with image enhancement versus conventional fluoroscopy in minimally invasive transforaminal lumbar interbody fusion: an internally randomized controlled trial,” J. Neurosurg. Spine, 28 (2), 186 –193 (2018). https://doi.org/10.3171/2017.5.SPINE17123 Google Scholar

16. 

R. Jakubovic et al., “High Speed, high density intraoperative 3D optical topographical imaging with efficient registration to MRI and CT for craniospinal surgical navigation,” Sci. Rep., 8 (1), 14894 –14894 (2018). https://doi.org/10.1038/s41598-018-32424-z SRCEC3 2045-2322 Google Scholar

17. 

C. W. Kim et al., “Use of navigation-assisted fluoroscopy to decrease radiation exposure during minimally invasive spine surgery,” Spine J., 8 (4), 584 –590 (2008). https://doi.org/10.1016/j.spinee.2006.12.012 Google Scholar

18. 

C. J. Kleck et al., “One-step minimally invasive pedicle screw instrumentation using O-arm and stealth navigation,” Clin. Spine Surg., 31 (5), 197 –202 (2018). https://doi.org/10.1097/BSD.0000000000000616 Google Scholar

19. 

J. D. van Dijk et al., “Clinical pedicle screw accuracy and deviation from planning in robot-guided spine surgery: robot-guided pedicle screw accuracy,” Spine, 40 (17), E986 –E991 (2015). https://doi.org/10.1097/BRS.0000000000000960 SPINDD 0362-2436 Google Scholar

20. 

P. Subramanian et al., “A review of the evolution of robotic-assisted total hip arthroplasty,” Hip Int, 29 (3), 232 –238 (2019). https://doi.org/10.1177/1120700019828286 Google Scholar

21. 

J. M. Laffosse, “Removal of well-fixed fixed femoral stems,” Orthop. Traumatol. Surg. Res., 102 (1, Suppl.), S177 –S187 (2016). https://doi.org/10.1016/j.otsr.2015.06.029 Google Scholar

22. 

P. Megas et al., “Removal of well-fixed components in femoral revision arthroplasty with controlled segmentation of the proximal femur,” J. Orthop. Surg. Res., 9 (1), 137 (2014). https://doi.org/10.1186/s13018-014-0137-9 Google Scholar

23. 

K. Deep, S. Shankar and A. Mahendra, “Computer assisted navigation in total knee and hip arthroplasty,” SICOT J., 3 50 (2017). https://doi.org/10.1051/sicotj/2017034 Google Scholar

24. 

T. C. Kelley and M. L. Swank, “Role of navigation in total hip arthroplasty,” J. Bone Joint Surg. Am., 91 (1), 153 –158 (2009). https://doi.org/10.2106/JBJS.H.01463 Google Scholar

25. 

K. Grygoryev et al., “Cranial perforation using an optically-enhanced surgical drill,” IEEE Trans. Biomed. Eng., 67 (12), 3474 –3482 (2020). https://doi.org/10.1109/TBME.2020.2987952 IEBEAX 0018-9294 Google Scholar

26. 

M. Duperron et al., “Diffuse reflectance spectroscopy-enhanced drill for bone boundary detection,” Biomed. Opt. Express, 10 (2), 961 –977 (2019). https://doi.org/10.1364/BOE.10.000961 BOEICL 2156-7085 Google Scholar

27. 

G. Burström et al., “Diffuse reflectance spectroscopy accurately identifies the pre-cortical zone to avoid impending pedicle screw breach in spinal fixation surgery,” Biomed. Opt. Express, 10 (11), 5905 –5920 (2019). https://doi.org/10.1364/BOE.10.005905 BOEICL 2156-7085 Google Scholar

28. 

J. Shubert and M. A. L. Bell, “Photoacoustic imaging of a human vertebra: implications for guiding spinal fusion surgeries,” Phys. Med. Biol., 63 (14), 144001 (2018). https://doi.org/10.1088/1361-6560/aacdd3 PHMBA7 0031-9155 Google Scholar

29. 

A. Swamy et al., “Diffuse reflectance spectroscopy, a potential optical sensing technology for the detection of cortical breaches during spinal screw placement,” J. Biomed. Opt., 24 (1), 017002 (2019). https://doi.org/10.1117/1.JBO.24.1.017002 JBOPFO 1083-3668 Google Scholar

30. 

A. Swamy et al., “Diffuse reflectance spectroscopy for breach detection during pedicle screw placement: a first in vivo investigation in a porcine model,” Biomed. Eng. Online, 19 47 (2020). https://doi.org/10.1186/s12938-020-00791-2 Google Scholar

31. 

S. K. V. Sekar et al., “In vivo, non-invasive characterization of human bone by hybrid broadband (600–1200 nm) diffuse optical and correlation spectroscopies,” PLoS One, 11 (12), e0168426 (2016). https://doi.org/10.1371/journal.pone.0168426 POLNCL 1932-6203 Google Scholar

32. 

S. K. Hwang, “Vascular injury during total hip arthroplasty: the anatomy of the acetabulum,” Int. Orthop., 18 (1), 29 –31 (1994). https://doi.org/10.1007/BF00180175 Google Scholar

33. 

J. S. Kirkpatrick et al., “The relationship of the intrapelvic vasculature to the acetabulum. Implications in screw-fixation acetabular components,” Clin. Orthop. Relat. Res., 258 183 –190 (1990). CORTBR 0009-921X Google Scholar

34. 

Y. Kawasaki et al., “Location of intrapelvic vessels around the acetabulum assessed by three-dimensional computed tomographic angiography: prevention of vascular-related complications in total hip arthroplasty,” J. Orthopaedic Sci., 17 (4), 397 –406 (2012). https://doi.org/10.1007/s00776-012-0227-7 Google Scholar

35. 

H. D. Clitherow and G. I. Bain, “Major neurovascular complications of clavicle fracture surgery,” Shoulder Elbow, 7 (1), 3 –12 (2015). https://doi.org/10.1177/1758573214546058 Google Scholar

36. 

D. Qin et al., “Safe drilling angles and depths for plate-screw fixation of the clavicle: avoidance of inadvertent iatrogenic subclavian neurovascular bundle injury,” J. Trauma, 69 (1), 162 –168 (2010). https://doi.org/10.1097/TA.0b013e3181bbd617 Google Scholar

37. 

R. J. Mobbs and D. A. Raley, “Complications with K-wire insertion for percutaneous pedicle screws,” Clin. Spine Surg., 27 (7), 390 –394 (2014). https://doi.org/10.1097/BSD.0b013e3182999380 Google Scholar

38. 

D. A. Raley and R. J. Mobbs, “Retrospective computed tomography scan analysis of percutaneously inserted pedicle screws for posterior transpedicular stabilization of the thoracic and lumbar spine: accuracy and complication rates,” Spine, 37 (12), 1092 –1100 (2012). https://doi.org/10.1097/BRS.0b013e31823c80d8 SPINDD 0362-2436 Google Scholar

39. 

Z. A. Smith et al., “Incidence of lumbar spine pedicle breach after percutaneous screw fixation: a radiographic evaluation of 601 screws in 151 patients,” Clin. Spine Surg., 27 (7), 358 –363 (2014). https://doi.org/10.1097/BSD.0b013e31826226cb Google Scholar

40. 

Y. Ishimoto et al., “Prevalence of symptomatic lumbar spinal stenosis and its association with physical performance in a population-based cohort in Japan: the Wakayama Spine Study,” Osteoarthritis Cartilage, 20 (10), 1103 –1108 (2012). https://doi.org/10.1016/j.joca.2012.06.018 Google Scholar

41. 

K. Nomura and M. Yoshida, “Assessment of the learning curve for microendoscopic decompression surgery for lumbar spinal canal stenosis through an analysis of 480 cases involving a single surgeon,” Global Spine J., 7 (1), 54 –58 (2017). https://doi.org/10.1055/s-0036-1583943 Google Scholar

42. 

S. M. Park et al., “Learning curve for lumbar decompressive laminectomy in biportal endoscopic spinal surgery using the cumulative summation test for learning curve,” World Neurosurg., 122 e1007 –e1013 (2019). https://doi.org/10.1016/j.wneu.2018.10.197 Google Scholar

43. 

J. A. Sclafani and C. W. Kim, “Complications associated with the initial learning curve of minimally invasive spine surgery: a systematic review,” Clin. Orthop. Relat. Res., 472 (6), 1711 –1717 (2014). https://doi.org/10.1007/s11999-014-3495-z CORTBR 0009-921X Google Scholar

44. 

C. Herren et al., “Incidental durotomy in decompression for lumbar spinal stenosis: incidence, risk factors and effect on outcomes in the Spine Tango registry,” Eur. Spine J., 26 (10), 2483 –2495 (2017). https://doi.org/10.1007/s00586-017-5197-1 Google Scholar

45. 

M. H. Khan et al., “Postoperative management protocol for incidental dural tears during degenerative lumbar spine surgery: a review of 3,183 consecutive degenerative lumbar cases,” Spine, 31 (22), 2609 –2613 (2006). https://doi.org/10.1097/01.brs.0000241066.55849.41 SPINDD 0362-2436 Google Scholar

46. 

B. N. Staub and S. S. Sadrameli, “The use of robotics in minimally invasive spine surgery,” J. Spine Surg., 5 (Suppl. 1), S31 –S40 (2019). https://doi.org/10.21037/jss.2019.04.16 Google Scholar

47. 

C. C. Zygourakis et al., “Technique: open lumbar decompression and fusion with the Excelsius GPS robot,” Neurosurg. Focus, 45 (VideoSuppl), V6 (2018). https://doi.org/10.3171/2018.7.FocusVid.18123 Google Scholar

48. 

L. Lilge et al., “Minimal required PDT light dosimetry for nonmuscle invasive bladder cancer,” J. Biomed. Opt., 25 (6), 068001 (2020). https://doi.org/10.1117/1.JBO.25.6.068001 JBOPFO 1083-3668 Google Scholar

49. 

J. Swartling et al., “System for interstitial photodynamic therapy with online dosimetry: first clinical experiences of prostate cancer,” J. Biomed. Opt., 15 (5), 058003 (2010). https://doi.org/10.1117/1.3495720 JBOPFO 1083-3668 Google Scholar

50. 

A. Bashkatov et al., “Optical properties of human cranial bone in the spectral range from 800 to 2000 nm,” Proc. SPIE, 6163 616310 (2006). https://doi.org/10.1117/12.697305 PSISDG 0277-786X Google Scholar

51. 

M. Firbank et al., “Measurement of the optical properties of the skull in the wavelength range 650-950 nm,” Phys. Med. Biol., 38 (4), 503 –510 (1993). https://doi.org/10.1088/0031-9155/38/4/002 PHMBA7 0031-9155 Google Scholar

52. 

S. Haleh, G. Hirac and P. Frédéric, “Optical properties of mice skull bone in the 455- to 705-nm range,” J. Biomed. Opt., 22 (1), 010503 (2017). https://doi.org/10.1117/1.JBO.22.1.010503 JBOPFO 1083-3668 Google Scholar

53. 

A. Pifferi et al., “Optical biopsy of bone tissue: a step toward the diagnosis of bone pathologies,” J. Biomed. Opt., 9 (3), 474 –480 (2004). https://doi.org/10.1117/1.1691029 JBOPFO 1083-3668 Google Scholar

54. 

I. Steinberg et al., “Quantitative study of optical and mechanical bone status using multispectral photoacoustics,” J. Biophotonics, 9 (9), 924 –933 (2016). https://doi.org/10.1002/jbio.201500206 Google Scholar

55. 

G. E. Strangman, Q. Zhang and Z. Li, “Scalp and skull influence on near infrared photon propagation in the Colin27 brain template,” NeuroImage, 85 (1), 136 –149 (2014). https://doi.org/10.1016/j.neuroimage.2013.04.090 NEIMEF 1053-8119 Google Scholar

56. 

A. Custo et al., “Effective scattering coefficient of the cerebral spinal fluid in adult head models for diffuse optical imaging,” Appl. Opt., 45 (19), 4747 –4755 (2006). https://doi.org/10.1364/AO.45.004747 APOPAI 0003-6935 Google Scholar

57. 

J. Steinbrink et al., “Determining changes in NIR absorption using a layered model of the human head,” Phys. Med. Biol., 46 (3), 879 –896 (2001). https://doi.org/10.1088/0031-9155/46/3/320 PHMBA7 0031-9155 Google Scholar

58. 

R. Kumar et al., “Optical investigation of osteoarthritic human cartilage (ICRS grade) by confocal Raman spectroscopy: a pilot study,” Anal. Bioanal.Chem., 407 (26), 8067 –8077 (2015). https://doi.org/10.1007/s00216-015-8979-5 ABCNBP 1618-2642 Google Scholar

59. 

J. R. Peterson et al., “Early detection of burn induced heterotopic ossification using transcutaneous Raman spectroscopy,” Bone, 54 (1), 28 –34 (2013). https://doi.org/10.1016/j.bone.2013.01.002 Google Scholar

60. 

F. Poulon et al., “Optical properties, spectral, and lifetime measurements of central nervous system tumors in humans,” Sci. Rep., 7 (1), 13995 (2017). https://doi.org/10.1038/s41598-017-14381-1 SRCEC3 2045-2322 Google Scholar

61. 

M. Sibai et al., “Quantitative spatial frequency fluorescence imaging in the sub-diffusive domain for image-guided glioma resection,” Biomed. Opt. Express, 6 (12), 4923 –4933 (2015). https://doi.org/10.1364/BOE.6.004923 BOEICL 2156-7085 Google Scholar

62. 

P. V. Butte et al., “Intraoperative delineation of primary brain tumors using time-resolved fluorescence spectroscopy,” J. Biomed. Opt., 15 (2), 027008 (2010). https://doi.org/10.1117/1.3374049 JBOPFO 1083-3668 Google Scholar

63. 

P. Bargo et al., “In vivo determination of optical properties of normal and tumor tissue with white light reflectance and an empirical light transport model during endoscopy,” J. Biomed. Opt., 10 (3), 034018 (2005). https://doi.org/10.1117/1.1921907 JBOPFO 1083-3668 Google Scholar

64. 

T. J. Beck et al., “Clinical determination of tissue optical properties in vivo by spatially resolved reflectance measurements,” Proc. SPIE, 5138 96 –105 (2003). https://doi.org/10.1117/12.508664 PSISDG 0277-786X Google Scholar

65. 

R. Bays et al., “Clinical determination of tissue optical properties by endoscopic spatially resolved reflectometry,” Appl. Opt., 35 (10), 1756 –1766 (1996). https://doi.org/10.1364/AO.35.001756 APOPAI 0003-6935 Google Scholar

66. 

R. M. P. Doornbos et al., “The determination of in vivo human tissue optical properties and absolute chromophore concentrations using spatially resolved steady-state diffuse reflectance spectroscopy,” Phys. Med. Biol., 44 (4), 967 –981 (1999). https://doi.org/10.1088/0031-9155/44/4/012 PHMBA7 0031-9155 Google Scholar

67. 

G. Zonios et al., “In vivo optical properties of melanocytic skin lesions: common nevi, dysplastic nevi and malignant melanoma,” Photochem. Photobiol., 86 (1), 236 –240 (2010). https://doi.org/10.1111/j.1751-1097.2009.00630.x PHCBAP 0031-8655 Google Scholar

68. 

K. K.-H. Wang and T. C. Zhu, “Reconstruction of in-vivo optical properties for human prostate using interstitial diffuse optical tomography,” Opt. Express, 17 (14), 11665 –11672 (2009). https://doi.org/10.1364/OE.17.011665 OPEXFF 1094-4087 Google Scholar

69. 

A. Kienle and T. Glanzmann, “In vivo determination of the optical properties of muscle with time-resolved reflectance using a layered model,” Phys. Med. Biol., 44 (11), 2689 –2702 (1999). https://doi.org/10.1088/0031-9155/44/11/301 PHMBA7 0031-9155 Google Scholar

70. 

M. S. Patterson, B. Chance and B. C. Wilson, “Time resolved reflectance and transmittance for the non-invasive measurement of tissue optical properties,” Appl. Opt., 28 (12), 2331 –2336 (1989). https://doi.org/10.1364/AO.28.002331 APOPAI 0003-6935 Google Scholar

71. 

T. Kono et al., “Optical characteristics of human skin with hyperpigmentation caused by fluorinated pyrimidine anticancer agent,” Biomed. Opt. Express, 10 (8), 3747 –3759 (2019). https://doi.org/10.1364/BOE.10.003747 BOEICL 2156-7085 Google Scholar

72. 

S. Nandy et al., “Characterizing optical properties and spatial heterogeneity of human ovarian tissue using spatial frequency domain imaging,” J. Biomed. Opt., 21 (10), 101402 (2016). https://doi.org/10.1117/1.JBO.21.10.101402 JBOPFO 1083-3668 Google Scholar

73. 

N. N. Mutyal et al., “A fiber optic probe design to measure depth- limited optical properties in-vivo with low-coherence enhanced backscattering (LEBS) spectroscopy,” Opt. Express, 20 (18), 19643 –19657 (2012). https://doi.org/10.1364/OE.20.019643 OPEXFF 1094-4087 Google Scholar

74. 

S. Linsler et al., “The use of intraoperative computed tomography navigation in pituitary surgery promises a better intraoperative orientation in special cases,” J. Neurosci. Rural Pract., 7 (4), 598 –602 (2016). https://doi.org/10.4103/0976-3147.186977 Google Scholar

75. 

J. K. Koehn and K. J. Kuchenbecker, “Surgeons and non-surgeons prefer haptic feedback of instrument vibrations during robotic surgery,” Surg. Endosc., 29 (10), 2970 –2983 (2015). https://doi.org/10.1007/s00464-014-4030-8 Google Scholar

76. 

A. Abiri et al., “Multi-modal haptic feedback for grip force reduction in robotic surgery,” Sci. Rep., 9 (1), 5016 (2019). https://doi.org/10.1038/s41598-019-40821-1 SRCEC3 2045-2322 Google Scholar

77. 

K. Bark et al., “In vivo validation of a system for haptic feedback of tool vibrations in robotic surgery,” Surg. Endosc., 27 (2), 656 –664 (2013). https://doi.org/10.1007/s00464-012-2452-8 Google Scholar

78. 

C.-W. Chen et al., “Semiautomated optical coherence tomography-guided robotic surgery for porcine lens removal,” J. Cataract Refractive Surg., 45 (11), 1665 –1669 (2019). https://doi.org/10.1016/j.jcrs.2019.06.020 Google Scholar

79. 

W. Kilby et al., “The CyberKnife Robotic Radiosurgery System in 2010,” Technol. Cancer Res. Treatment, 9 (5), 433 –452 (2010). https://doi.org/10.1177/153303461000900502 Google Scholar

80. 

D. Hu et al., “Semi-autonomous simulated brain tumor ablation with RAVENII surgical robot using behavior tree,” in IEEE Int. Conf. Rob. and Autom., 3868 –3875 (2015). https://doi.org/10.1109/ICRA.2015.7139738 Google Scholar

Biography

Carl Fisher is a researcher in Biophotonics@Tyndall at the Tyndall National Institute in Cork, IE. He received his PhD in medical biophysics from the University of Toronto in 2016 and received his B.Sc (Hons) degree in biochemistry from McGill University in 2009. He is currently pursuing his degree in medicine at the University College Cork. His research interests include integration of optics in surgical devices and use of biophotonics for therapeutic applications.

James Harty is a professor of orthopaedic surgery at the University College Cork. He has special interest in adult lower limb reconstruction. His extensive research interests include revision hip and knee arthroplasty as well as biomechanics of fracture healing.

Albert Yee is the Holland Bone and Joint Program Chief and Marvin Tile Chair, Division Chief of Orthopaedics at Sunnybrook Health Sciences Centre. He is a professor (University of Toronto) and co-director of the Department of Surgery Spine Program. His research focuses on translational spine studies utilizing pre-clinical surgical models to evaluate minimally invasive therapeutics. This work has led to first in human clinical trials and FDA approval and commercialization of new spine technology.

Celina L. Li is a graduate student in Biophotonics@Tyndall, IPIC, Tyndall National Institute, Ireland. She received her double BSc degrees in biology and physics from the University of British Columbia, Vancouver, Canada, in 2017, and her MSc degree in medical biophysics from the University of Toronto, Toronto, Canada, in 2020. Her research interest includes diffuse reflectance spectroscopy, machine learning, and translation research.

Katarzyna Komolibus received her MSc degree in electronics and telecommunication, specialized in optoelectronics and optical fiber technology, from Wroclaw University of Technology in 2011 and her PhD in applied physics from Cork Institute of Technology, where she worked on carrier dynamics in III–V semiconductor nanostructures in 2016. Since 2017, she has been working as a postdoctoral researcher at Biophotonics@Tyndall. Her research interests include integration of optics into surgical instruments and multimodal spectroscopy techniques for diagnostic and therapeutic applications.

Konstantin Grygoryev graduated from UCC with Hons. BSc degree in neuroscience and received his PhD in area of MEMS. During his research career, he has worked on microneedle biosensors and their integration into ECG and has developed a fabrication process for polymer microneedle electrodes, still used today. He has also worked on developing an electrochemical bacterial sensor for integration in milk processing plants. In 2015, he joined the photonics group and has since worked on integration of photonics-based surgical guidance methods into existing and new surgical tools. His main expertise is designing, building, programming and testing of miniaturized, custom electro-optical sensing systems.

Huihui Lu received her PhD in chemistry from the National University of Ireland, Cork. Her prior experience include medical device industry of in-vitro diagnostics devices development and point-of-care device instrumentation, integrating silicon photonic platform to enable future portable and wearable devices and multimodality optical techniques for surgical guidance. Her research interests include biosensor and assay solutions, silicon photonic integration for on-chip sensing, and multimodality spectroscopy techniques for point-of-care diagnostic devices and healthcare applications.

Ray Burke has a background in physics and electronics. He has more than 25 years R&D experience in ICT and Medtech, working for large multinational corporations (Boston Scientific) as a senior design engineer and a research manager. He is PI on three large EU projects POSITION, LAA Start, and MedPhab and leads the research for these in-vivo medical devices research programs. He leads the engagements with Medtech, clinicians, and currently supervises PhD and MD students.

Brian C. Wilson is a professor/senior scientist at the University of Toronto/Princess Margaret Cancer Centre, where he directs a translational research program in photonics and nanotechnologies applied mainly to cancer diagnostics, interventional guidance, and phototherapies. He is a fellow of SPIE and OSA and holds the Britton Chance and Michaels S. Feld Awards in biomedical photonics, respectively, as well as the Robert L. Noble Award in translational cancer research from the Canadian Cancer Society.

Stefan Andersson-Engels received his MSc degree in physics in 1985 and his PhD in 1990 from Lund University. He was a postdoc at McMasters University in Canada 1990 to 1991 and became a full professor at Lund University in 1999. In 2016, he was recruited as the head of Biophotonics@Tyndall National Institute and is a deputy director of the IPIC Centre. His research interests include tissue optics and applications of light in biomedical diagnostics and treatments.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 Unported License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Carl Fisher, James Harty, Albert J. M. Yee, Celina L. Li, Katarzyna Komolibus, Konstantin Grygoryev, Huihui Lu, Ray Burke, Brian C. Wilson, and Stefan Andersson-Engels "Perspective on the integration of optical sensing into orthopedic surgical devices," Journal of Biomedical Optics 27(1), 010601 (5 January 2022). https://doi.org/10.1117/1.JBO.27.1.010601
Received: 5 August 2021; Accepted: 23 November 2021; Published: 5 January 2022