26 June 1992 High-resolution confocal transmission microscope: Part I - system design
Carol J. Cogswell, John W. O'Byrne
Author Affiliations +
Proceedings Volume 1660, Biomedical Image Processing and Three-Dimensional Microscopy; (1992) https://doi.org/10.1117/12.59581
Event: SPIE/IS&T 1992 Symposium on Electronic Imaging: Science and Technology, 1992, San Jose, CA, United States
Abstract
We are investigating using a transmitted light configuration in our experimental confocal microscope, in order to observe highly transparent biological preparations which typically yield little or no signal in reflection. Because refractive index and thickness variations in the specimen often deflect the focused image spot off-axis in a confocal transmission configuration, we are exploring using a CCD array detector in the transmitted beam path instead of a single-pinhole detector. Our method for detecting the brightest pixel values from the CCD diffraction image is shown to give a similar two-point response to that of a confocal fixed-pinhole transmission system. We also discuss the differences between reflection and transmission confocal modes and analyze the relative imaging properties of each.
© (1992) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Carol J. Cogswell and John W. O'Byrne "High-resolution confocal transmission microscope: Part I - system design", Proc. SPIE 1660, Biomedical Image Processing and Three-Dimensional Microscopy, (26 June 1992); https://doi.org/10.1117/12.59581
Lens.org Logo
CITATIONS
Cited by 9 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Confocal microscopy

Charge-coupled devices

Microscopes

Sensors

CCD image sensors

Image transmission

Reflection

Back to Top