Paper
23 February 2006 Highly accurate calibration of optical radiation at infrared spectrum using thermopile TS-76
Lei Zhang, Xiao-bing Zheng, Ji Wang, Ping Xie, Feng Chen, Shuang Li, Wei Zhang
Author Affiliations +
Abstract
Infrared radiometric calibration is of critical importance for information quantification of satellite remote sensing of environment at infrared spectrum. In this paper, thin-film thermopile sensors (IPHT-Jena, type TS-76) are used as transfer detectors in the spectral range from 514nm to 1550nm. The relative uncertainty in the spectral responsivity of the thin-film thermopiles was about 1.5%. Further work will be focused on calibrating TS-76 with a monochromator to extend the spectral range from 1800nm to 3000nm and reducing uncertainty budget of TS-76.
© (2006) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Lei Zhang, Xiao-bing Zheng, Ji Wang, Ping Xie, Feng Chen, Shuang Li, and Wei Zhang "Highly accurate calibration of optical radiation at infrared spectrum using thermopile TS-76", Proc. SPIE 6150, 2nd International Symposium on Advanced Optical Manufacturing and Testing Technologies: Optical Test and Measurement Technology and Equipment, 615041 (23 February 2006); https://doi.org/10.1117/12.676519